Home > GE > Ultrasound System > GE Vivid 7 User Manual

GE Vivid 7 User Manual

    Download as PDF Print this page Share this page

    Have a look at the manual GE Vivid 7 User Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    Page
    of 536
    							Presets and System setup
    Vivid7/Vivid7 PRO Users Manual407
    FC092326-03
    Parameter Description
    Add measure:
    Create or select from the pop-up list a 
    measurement to be added to a folder.
    Add folder:
    Enables the user to create its own folder 
    with the desired measurements. The 
    folder is displayed the Measurement 
    menu.
    M&A Categories:
    Enables selection of the measurement 
    categories to display in the Measurement 
    menu. Only checked items will be 
    displayed.
    • Create Copy: Enables copy of a 
    selected measurement category 
    (selection is done by selecting the 
    category name). 
    • Delete: enables deletion of user-defined 
    measurement categories.
    • Factory Default: restores factory 
    display.
    2D, MM and Dop. radio buttons:
    Enables the display of mode related 
    Measurement menu in the configuration 
    window. 
    						
    							Presets and System setup
    408Vivid7/Vivid7 PRO Users Manual
    FC092326-03
    Configuration tools:
    Deletes selected entry (folder 
    or measurement) in the 
    Measurement menu. The 
    factory entries cannot be 
    deleted.
    Moves selected measurement 
    or folder up or down inside the 
    Measurement menu.
    Folder:
    Displayed when a folder is selected in the 
    Measurement Menu.
    Shows the entire contents of a selected 
    folder.
    •  : the items is displayed in the 
    Measurement menu.
    •  : The item is hidden from the 
    Measurement menu.
    Measurement:
    Displayed when a measurement is 
    selected in the Measurement Menu.
    Shows all the parameters related to the 
    selected measurement.
    •  : the items is displayed in the 
    Measurement menu.
    •  : The item is hidden from the 
    Measurement menu.
    Only checked parameters will be 
    displayed in the Measurement result 
    window, the worksheet and the report.
    Auto sequence:
    Prompts the next measurement in the 
    folder. Parameter Description 
    						
    							Presets and System setup
    Vivid7/Vivid7 PRO Users Manual409
    FC092326-03
    Configuration of the Measurement 
    menu
    Display of the Measurement categories
    1. Press M&A categories in the Configuration window.
    The M&A categories are displayed in a pop-up window 
    (see page 407).
    2. Check the categories to be displayed.
    Uncheck the categories to hide.
    To copy a Measurement categor y
    1. Press M&A categories in the Configuration window.
    The M&A categories are displayed in a pop-up window 
    (see page 407).
    2. Move the trackball marker over the M&A category name.
    3. Press 
    SELECT to highlight the category.
    4. Press Create copy.
    A copy of the selected measurement category is displayed 
    in the Measurement menu.
    Factory Measure-
    ment categories 
    cannot be renamed.To rename the Measurement category:
    1. Select the Measurement category in the Measurement 
    menu.
    2.Enter a new name in the Measurement field.
    Selection of a Measurement categor y
    1. Trackball to the Measurement menu heading.
    2. Press 
    SELECT.
    The measurement categories are displayed in a sub-
    menu.
    3. Trackball to the measurement category of interest.
    4. Press 
    SELECT.
    The measurement category is displayed.
    Moving an item in the Measurement menu
    1. Trackball to the entry to move into the Measurement menu.
    2. Press 
    SELECT.
    3. Press   or   to move the selection up or down inside 
    the Measurement menu. 
    						
    							Presets and System setup
    410Vivid7/Vivid7 PRO Users Manual
    FC092326-03
    Deleting an item in the Measurement menu
    Only user created 
    items can be delet-
    ed.1. Trackball to the entry to delete in the Measurement menu.
    2. Press 
    SELECT.
    3. Press   to delete the item.
    Display/hide a folder or a measurement in the 
    Measurement menu
    The Measurement menu (Folders and Measurements) can be 
    configured to display only the entries (folders and 
    measurements) of interest.
    To hide a folder or a measurement:
    1.Uncheck the actual folder or measurement in the Folder or 
    Measurement field in the Configuration window.
    To display a hidden folder or measurement:
    1.Check the actual folder or measurement in the Folder or 
    Measurement field in the Configuration window.
    Creating a user-defined folder
    1. If the folder is to be inside another folder, select the actual 
    folder in the Measurement menu.
    2. Press Add folder.
    The Measurement menu is updated.
    3. Select the new folder and Enter the folder name in the 
    Name text field.
    Adding a measurement to a folder
    The user can either add a pre-defined measurement or create 
    a new measurement with user-defined parameters to a folder.
    1. Select the folder in the Measurement menu.
    2. Press Add Measure.
    The Add measurement dialogue window is displayed 
    (Figure 12-7). 
    						
    							Presets and System setup
    Vivid7/Vivid7 PRO Users Manual411
    FC092326-03
     Figure 12-7: The Add Measurement dialogue window
    To add an existing measurement with pre-defined 
    parameters
    1. Select Use Copy of.
    2. Select the measurement in the pop-up menu.
    3. Press OK.
    To create a measurement with user-defined parameters
    1. In the Add measurement dialogue window, select Blank.
    2. Press OK.
    3. In the Measurement field, enter the measurement name.
    4. Select the desired measurement type in the Tool pop-up 
    menu.
    The actual measurement parameters are displayed.
    5. Trackball to the first parameter in the Parameter column.
    6. Press 
    SELECT twice to activate the free text cursor.
    7.Enter a name for the parameter.
    Adding a user-defined function to a measure-
    ment
    1. Select the actual measurement in the Measurement menu.
    The parameters related to the measurement are listed in 
    Parameter list.
    2.Press twice on an empty line at the bottom of the 
    Parameter list.
    The free text cursor is activated.
    3.Enter a name for the parameter.
    4. Press on the Parameter line once again.
    The Edit Formula dialogue window is displayed. 
    						
    							Presets and System setup
    412Vivid7/Vivid7 PRO Users Manual
    FC092326-03
     Figure 12-8: The Edit formula dialogue window
    5. Create a formula by selecting parameters, operators and 
    functions.
    Unit: All results are calculated in SI units. If the unit is not 
    specified or not one in the table below, the displayed value 
    will be in SI units. All conversions to other units must 
    include a conversion factor in the formula. If the unit is one 
    in the table below, the conversion factor will be included in 
    the formula.
    6. Press Check to control the formula.
    7. Press OK to store the parameter.
    CAUTIONMake sure that the formula gives a SI unit that is compatible with 
    the conversion unit.
    CAUTIONGE Ultrasound do not take any responsibility for the correctness 
    of the user-defined functions.
    Calculation SI Alternative unit
    Time s
    Ratio %
    Frequency bpm
    Angle rad
    Distance m 
    						
    							Presets and System setup
    Vivid7/Vivid7 PRO Users Manual413
    FC092326-03
    Velocity m/s
    Acceleration
    m/s
    2
    Area
    m2
    Volume
    m3
    Volume flow
    m3/s
    Pressure mmHg
    Pressure/time mmHg/s
    Mass kg
    OtherCalculation SI Alternative unit 
    						
    							Presets and System setup
    414Vivid7/Vivid7 PRO Users Manual
    FC092326-03
    Repor t
    The Report configuration category is divided in three sheets:
    • Te m p l a t e s: enables the configuration of the Te m p l a t e  
    selection menu
    • Diagnostic codes: enables the creation of pre-defined text 
    inputs to be used in the Diagnosis information field in the 
    Examination list window (see Figure 8-10, page 286).
    • Comment texts: enables the creation of pre-defines text 
    inputs to be used in the Comment information field in the 
    Examination list window (see Figure 8-10, page 286). 
    						
    							Presets and System setup
    Vivid7/Vivid7 PRO Users Manual415
    FC092326-03
    The Report templates sheet
     Figure 12-9: The Report templates sheet
    1. Application selection pop-up menu
    2. List of factory and user-defined report templates 
    for the selected application
    3. List of report templates shown in the Templat e 
    selection menu for the selected application
    4. Add/remove selected template to/from the 
    Template selection menu.5. Move selected template up/down in the Tem plate 
    selection menu.
    6. Open selected template in the template 
    designer.
    7. Delete selected template (user-defined 
    templates only)
    8. Export template to MOD or CD-R 
    						
    							Presets and System setup
    416Vivid7/Vivid7 PRO Users Manual
    FC092326-03
    Configuration of the Template selection menu
    The configuration of the Template selection menu is done by:
    •  Selecting the application specific Template list (left field) 
    and the application specific Template selection menu (right 
    field).
    •  Inserting or removing report templates to/from the Te m p l a t e  
    selection menu.
    •  Sorting the report templates in the Template selection 
    menu.
    Procedure
    Selecting the application specific template list and 
    Template selection menu:
    1. Trackball to the Application selection pop-up menu over the 
    list of report templates (left field, see Figure 12-9).
    2. Press 
    SELECT.
    The pop-up menu is displayed.
    3. Trackball to the desired application.
    4. Press 
    SELECT.
    The list of factory and user-defined report templates (left 
    field, see Figure 12-9) is updated showing the available 
    templates for the selected application.
    5. Likewise select the application for the Template selection 
    menu (right field, see Figure 12-9).
    Inserting a template to the Template selection menu
    1. Trackball to the template to insert in the Template list (left 
    field).
    2. Press 
    SELECT.
    3. Press the Right arrow button .
    The selected template is inserted in the Template selection 
    menu (right field).
    Removing a template from the Template selection 
    menu
    1. Trackball to the template to remove in the Te m p l a t e  
    selection menu (right field).
    2. Press 
    SELECT.
    3. Press the Left arrow button .
    The selected template is removed from the Te m p l a t e  
    selection menu (right field). 
    						
    All GE manuals Comments (0)