    							Scanning Modes
    Vivid7/Vivid7 PRO Users Manual97
    FC092326-03
    Using 2D
    Refer to page 400 
    about creating pre-
    sets.The 2D-Mode is the systems default mode.
    1. Press 
    2D on the control panel to access 2D mode.
    Check the Displays 
    brightness and con-
    trast setting before 
    adjusting the unit 
    imaging controls2. Optimize the image by adjusting the image controls 
    described in the previous section.
    If necessary use preset for optimum performance with 
    minimum adjustment.
    Optimizing 2D
    The following controls can be adjusted to optimize the 2D Mode 
    display:
    • Use the Gain and TGC controls to optimize the overall 
    image.
    • Use the Depth control to adjust the range to be imaged.
    • Use the Focus control to center the focal point(s) around 
    the region of interest.
    • Use the Frequency (move to higher frequencies) or the 
    Frame rate control (move to lower frame rate) to increase 
    resolution in image
    • Use the Frequency (move to lower frequency) to increase 
    penetration.
    • Use the Reject control to reduce noise in the image.
    • Use the DDP control to optimize imaging in the blood flow 
    regions and make a cleaner, less noisy image.
    WARNINGAlways use the minimum power required to obtain acceptable 
    images in accordance with applicable guidelines and policies. 
    						
    							Scanning Modes
    98Vivid7/Vivid7 PRO Users Manual
    FC092326-03
    M-Mode
    M-Mode over view
     Figure 3-3: The cardiac M-Mode screen (top/bottom)
    1. Time motion cursor 
    conventional M-
    Mode
    2. Time motion cursor 
    curved anatomical 
    M-Mode
    3. Time motion cursor 
    anatomical M-Mode
    4. Depth scale
    5. Focus marker
    6. Time scale
    7. Status window:
    8. Soft menu 
    						
    							Scanning Modes
    Vivid7/Vivid7 PRO Users Manual99
    FC092326-03
     Figure 3-4: The M-Mode controls on the front panel
    This unit has three types of M-Mode: 
    •  Conventional M-Mode (MM): displays a distance/time plot 
    of a cursor line in the axial plane of the 2D-image.
    •  Anatomical M-Mode (AMM): displays a distance/time plot 
    from a cursor line, which is independent from the axial 
    plane. AMM is available in greyscale, color, TVI, Tissue 
    Tracking, Strain rate and Strain modes.
    •  Curved Anatomical M-Mode (CAMM): displays a distance/
    time plot from a free-drawn cursor line. CAMM is available 
    in greyscale, color, TVI, Tissue Tracking, Strain rate and 
    1. Assignable keys:
    • Horizontal sweep R
    • Frequency/Resolution
    • Focus
    • Contour
    • Up/Down
    • Color maps
    2. Zoom
    3. Img Size
    4. Soft menu
    • Compress R
    • Reject R
    • Dynamic Range
    • Power
    5. Freeze
    6. M-Mode
    7. Access to AMM & 
    CAMM
    8. Gain R
    Controls marked with R are 
    also available in freeze and 
    cine replay
    1
    2
    3
    4
    5
    6
    8
    7 
    						
    							Scanning Modes
    100Vivid7/Vivid7 PRO Users Manual
    FC092326-03
    Strain modes.
    M-Mode and 2D Mode display areas can be side by side or top/
    bottom. Conventional M-Mode can be combined with Color 
    (see ’Color M-Mode overview’ on page 105).
    M-Mode controls
    M-Mode assignable controls
    Horizontal sweep
    Adjusts the horizontal refresh rate of the M-Mode area of the 
    display. Horizontal sweep does not change the acquisition 
    resolution, so that user can change the horizontal sweep in 
    replay (with no loss of quality).
    On some low fre-
    quencies, the sys-
    tem switches 
    automatically to 
    second-harmonic 
    mode. The word 
    “Octave” appears 
    in the status win-
    dow.Frequency/Resolution
    Enables the adjustment of the probes operating frequency. 
    Rotate the knob clockwise to increase the frequency. The 
    selected frequency is displayed in the status window. For some 
    probes/applications the lowest frequency settings will be 
    Octave imaging settings.
    Focus
    Changes the location of the focal point(s). A triangular focus 
    marker indicates the depth of the focal point.
    Contour
    Controls image processing related to the extent of edge 
    enhancement applied to an image.
    Up/Down
    Flips the M-Mode display 180 degrees.
    Color maps
    Displays a color map menu to optimize the greyscale 
    presentation. The menu enables an option from a list of non-
    linear gray-curves or different colorized curves to be selected. 
    						
    							Scanning Modes
    Vivid7/Vivid7 PRO Users Manual101
    FC092326-03
    M-Mode Soft menu controls
    Compress
    Controls the amount of contrast in the image. An index number 
    is displayed in the status window to indicate the relative level of 
    compression.
    Reject
    Adjust reject level. When this control is increased, low-level 
    echoes are rejected and appear darker in the image. An index 
    number is displayed in the status window to indicate the 
    relative level of rejection.
    Power
    When power is re-
    duced, it reduces 
    the signal-to-noise 
    ratio, so that the im-
    age, spectrum or 
    color scan may be-
    come noisier.Controls the amount of acoustic power applied in all modes. 
    When power is set to maximum, it is equal to or less than the 
    maximum acoustic power permitted by the FDA. The Thermal 
    Index (TI) and the Mechanical Index (MI) are displayed on the 
    screen.
    Dynamic Range
    Enables control of the dynamic range or contrast of the image. 
    When dynamic range is set to High, the image is softer and 
    more low-level data is visible. 
    						
    							Scanning Modes
    102Vivid7/Vivid7 PRO Users Manual
    FC092326-03
    Using M-Mode
    Conventional M-Mode
    1. To access M-Mode from any other scan mode, press MM on 
    the control panel.
    2. Use the trackball to position the cursor over the required 
    area of the image.
    Gain, Frequency, 
    Focus, Dynamic 
    Range and Com-
    pression affect also 
    the 2D image.3. Adjust horizontal sweep, Gain, Frequency, Focus, Dynamic 
    Range, Compression and Contour to optimize the display if 
    necessary.
    4. Press 
    FREEZE to stop imaging.
    Anatomical M-Mode
    Anatomical M-
    Mode can also be 
    used with previous-
    ly acquired digitally 
    stored 2D images.1. From the 2D Live or M-Mode view, press ALT. 
    The alternative modes are displayed on the assignable.
    2. Press 
    AMM
    OR
    1. From the 2D Live, press 
    FREEZE.
    2. Press 
    MM to access the Anatomical M-Mode.
    The Trackball as-
    signable Pos (Posi-
    tion) is activated.3. Use the trackball to position the cursor over the required 
    area of the image.
    The Trackball as-
    signable Angle is 
    activated.4. Press TRACKBALL to allow free rotation of the solid full-
    length cursor line throughout the 2D image.
    5. Rotate the solid cursor line to the desired direction.
    The Trackball as-
    signable Pos is acti-
    vated.6. Press TRACKBALL twice and reposition the intersection 
    point to the desired position along the cursor line.
    7. Repeat steps 4. and 5. to change the angle of the solid 
    cursor line if necessary.
    The M-Mode area of 
    the display updates 
    as the M-Mode sig-
    nal is constructed.8. Press TRACKBALL to activate scrolling control on the 
    trackball.
    9. Use the trackball to scroll through the data acquired at that 
    location. The M-Mode display will vary accordingly.
    Cur ved Anatomical M-Mode
    Curved Anatomi-
    cal M-Mode can al-
    so be used with 
    previously ac-
    quired digitally 
    stored 2D images.1. From the 2D or M-Mode view press ALT. 
    The alternative modes appear on the assignable display.
    2. Press 
    CURVED AMM. 
    						
    							Scanning Modes
    Vivid7/Vivid7 PRO Users Manual103
    FC092326-03
    3. Use the trackball to position the starting point of the time 
    motion curve.
    The time motion 
    curve can be edited 
    by following the 
    curve back to the de-
    sired point and re-
    drawn as desired. 
    Following the curve 
    back to the starting 
    point will delete the 
    time motion curve.4. Press SELECT to anchor the starting point of the time motion 
    curve.
    5. Use the trackball to position the second point of the time 
    motion curve.
    6. Press 
    SELECT to anchor the second point of the time motion 
    curve.
    7. Repeat step 5. and 6. up to seven times to draw a complete 
    time motion curve.
    Optimizing M-Mode
    Refer to page 400 
    about creating pre-
    sets.The use of preset gives optimum performance with minimum 
    adjustment. If necessary, the following controls can be adjusted 
    to further optimize the M-Mode display:
    • Adjust Horizontal sweep to optimize the display resolution.
    • Adjust Gain and TGC controls to adjust the range to be 
    imaged.
    Except for Con-
    tour, all the con-
    trols listed in the 
    optimizing M-
    Mode section will 
    also affect the 2D 
    image.• Use the Frequency (move to higher frequencies) or the 
    Frame rate control (move to lower frame rate) to increase 
    resolution in image.
    • Use the Frequency (move to lower frequency) to 
    increase penetration.
    • Adjust Focus to move the focal point(s) around the region 
    of interest in the M-Mode display.
    • Adjust Dynamic range to optimize the useful range of 
    incoming echoes to the available greyscale.
    • Adjust Compress and Contour to further optimize the 
    display.
    • Adjust Reject to reduce noise while taking care not to 
    eliminate significant low-level diagnostic information. 
    						
    							Scanning Modes
    104Vivid7/Vivid7 PRO Users Manual
    FC092326-03
    Color Mode 
    Color 2D Mode over view
     Figure 3-5: The Color Mode screen
    1. Probe orientation 
    marker
    2. Color bar
    3. Color sector marker
    4. Status window:
    5. Soft menu
    Controls marked with R 
    are also available in 
    freeze and cine replay. 
    						
    							Scanning Modes
    Vivid7/Vivid7 PRO Users Manual105
    FC092326-03
    Color M-Mode over view
     Figure 3-6: The Color M-Mode screen (top/bottom display)
    1. Time motion cursors
    2. Color bar
    3. Focus marker
    4. Flow sector marker
    5. Time scale
    6. Status window:
    7. Soft menu
    Controls marked with R 
    are also available in 
    freeze and cine replay. 
    						
    							Scanning Modes
    106Vivid7/Vivid7 PRO Users Manual
    FC092326-03
     Figure 3-7: The Color Mode controls on the front panel
    1. Assignable keys:
    • Horizontal sweep 
    (Color M-Mode only) R
    • Scale
    • LVR
    • Invert R
    • Variance R
    • Dual focus
    • Color maps
    • Angio (Color 2D only)
    • Cineloop (in Freeze, 
    Color 2D mode only)
    2. Zoom
    • Display zoom
    • HR zoom
    3. Img Size (Color M-
    Mode)
    4. Soft menu
    • Sample Volume
    • Tissue Priority R
    • DDP (Color 2D only) R
    • Frequency
    • Radial avg.
    • Lateral avg.
    • Power
    5. Freeze
    6. Color
    7. Access to AMM & 
    CAMM
    8. Gain R
    Controls marked with R are 
    also available in freeze and 
    cine replay
    6
    1
    2
    5
    4
    3
    8
    7 
    						