    							Cardiac Measurement and Analysis
    Vivid7/Vivid7 PRO Users Manual217
    FC092326-03
    2D Area measurements
    1. Generate the 2D image.
    2. Press 
    FREEZE to stop the cineloop.
    Alternative: Press 
    CALIPER and AREA 
    assignable.
    3. Press MEASURE on the Control Panel.
    4. Select Area (trace) in the Measurement Menu (see 
    Figure 6-2).
    5. Trackball the cursor to the start point of the measurement.
    See the Status bar 
    to get the next step 
    to perform.6. Press SELECT to anchor the start point of the measurement.
    7. Trace the area (planetary) with the Trackball.
    The measurement 
    display color on the 
    2D image changes 
    from green to red af-
    ter completion of 
    the measurement.The current area and circumference values are displayed 
    in the Measurement result table and are instantaneously 
    updated when moving the cursor.
    8. Press 
    SELECT to complete the measurement.
    The measurement result is displayed in the Measurement 
    result table.
    9. To assign a label to the measurement, see page 227.
    The measurements 
    displayed on the 2D 
    image and the cor-
    responding results 
    are numbered.10. Repeat steps 5 through 8 to make additional area 
    measurements.
    2D area measurement ratio
    1. Generate the 2D image.
    2. Press 
    FREEZEto stop the cineloop.
    Alternative: Press 
    CALIPER, MORE as-
    signable and 
    AREA 
    RATIO.
    3. Press MEASURE on the Control Panel.
    4. Select Area ratio in the Measurement Menu (see Figure 6-
    2).
    5. Perform two area measurements as described in steps 5 
    through 8 in the above section.
    See the Status bar 
    to get the next step 
    to perform.The measurement results including the ratio (%) of the two 
    measured areas are displayed in the Measurement result 
    table.
    Editing 2D Area measurements
    1. Trackball the cursor to the anchor point of the area 
    measurement to modify.
    2. Press 
    SELECT twice (Double-click) to select the anchor 
    point.
    The selected marker turns green and is unanchored.
    3. With the Trackball, reposition the marker.
    4. Press the 
    SELECT to anchor. 
    						
    							Cardiac Measurement and Analysis
    218Vivid7/Vivid7 PRO Users Manual
    FC092326-03
    2D Volume measurements
    The measurements described in this section enable volume 
    measurement in a defined zone. The measurements tool 
    generates results by two methods:
    For measurement 
    formulae, refer to 
    the Reference 
    Manual.•  Method of Disk (displayed as Vmod in the Measurement 
    result table), known as Simpsons method.
    •  Area/Length method (displayed as Va -l in the 
    Measurement result table).
    To perform a volume measurement:
    1. Generate the 2D image.
    2. Press 
    FREEZE to stop the cineloop.
    Alternative: Press 
    CALIPER and 
    VOLUME assignable.
    3. Press MEASURE on the Control Panel.
    4. Select Vo l u m e in the Measurement Menu (see Figure 6-2).
    5. Trackball the cursor to the start point where a volume is to 
    be measured.
    See the Status bar 
    to get the next step 
    to perform.6. Press SELECT to anchor the start point of the measurement.
    7. Trackball the cursor to draw the length.
    8. Press 
    SELECT to anchor the second point.
    9. Drag the cursor with the Trackball to outline the area of 
    interest.
    The measurement 
    display color on the 
    2D image changes 
    from green to red af-
    ter completion of 
    the measurement.The current area, circumference and Area/Length Volume 
    (Va-l) values are displayed in the Measurement result table 
    (see Figure 6-2) and are instantaneously updated when 
    moving the cursor.
    10. Press 
    SELECT to complete the measurement.
    The measurement results including Vmod (Simpson) are 
    displayed in the Measurement result table (see Figure 6-
    2).
    The measurements 
    displayed on the 2D 
    image and the cor-
    responding results 
    are numbered.11. To assign a label to the measurement, see page 227.
    12. Repeat steps 5 through 10 to make additional volume 
    measurements.
    2D Depth measurements
    The measurements described in this section enable depth 
    measurement from the probe to a selected point.
    To perform a depth measurement:
    1. Generate the 2D image.
    2. Press 
    FREEZE to stop the cineloop. 
    						
    							Cardiac Measurement and Analysis
    Vivid7/Vivid7 PRO Users Manual219
    FC092326-03
    3. Press MEASURE on the Control Panel.
    See the Status bar 
    to get the next step 
    to perform.4. Press the assignable POINT to select the depth 
    measurement function.
    5. Trackball the cursor to the position to measure.
    The current distance from the probe is displayed in the 
    Measurement result table and is instantaneously updated 
    when moving the cursor.
    The measurements 
    displayed on the 2D 
    image and the cor-
    responding results 
    are numbered.6. Press SELECT to anchor the point.
    The depth value (cm) is displayed in the Measurement 
    result table. 
    						
    							Cardiac Measurement and Analysis
    220Vivid7/Vivid7 PRO Users Manual
    FC092326-03
    M-Mode Measurements
    In M-Mode, the user can perform distance and time 
    measurements. This measurement package has also the 
    following pre-defined measurement studies: 
    • LA/Ao
    • LV
    • RV
    M-Mode Length measurements
    1. Generate the M-Mode image.
    2. Press 
    FREEZE to stop the cineloop.
    Alternative: Press 
    CALIPER on the con-
    trol panel and press 
    CALIPER assignable.
    3. Press MEASURE on the Control Panel.
    4. Select caliper in the Measurement Menu.
    5. Trackball the cursor to the start point of the measurement.
    6. Press 
    SELECT to anchor the start point of the measurement.
    See the Status bar 
    to get the next step 
    to perform.7. Trackball the cursor to the measurement end point.
    The current distance value is displayed in the 
    Measurement result table and is instantaneously updated 
    when moving the cursor.
    8. Press 
    SELECT to anchor the end point of the measurement.
    The measurement 
    display color on the 
    M-Mode changes 
    from green to red af-
    ter completion of 
    the measurement.The measurement result is displayed in the Measurement 
    result table.
    9. To assign a label to the measurement, see page 227.
    10. Repeat steps 5 through 8 to make additional length 
    measurements.
    The measurements 
    displayed on the M-
    Mode image and the 
    corresponding re-
    sults are numbered.Editing M-Mode Length measurements
    1.  Trackball the cursor to one of the anchor points of the 
    measurement to modify.
    2. Press 
    SELECT twice (double-click).
    The selected marker turns green and is unanchored.
    3. With the Trackball, reposition the marker to a new position.
    4. Press 
    SELECT.
    Ao/LA study
    1. Generate the M-Mode image.
    2. Press 
    FREEZE to stop the cineloop. 
    						
    							Cardiac Measurement and Analysis
    Vivid7/Vivid7 PRO Users Manual221
    FC092326-03
    Alternative: Press 
    CALIPER on the con-
    trol panel and press 
    the 
    AO/LA assign-
    able.
    3. Press MEASURE on the Control Panel.
    4. Select Ao/LA in the Measurement Menu.
    5. Trackball the cursor along the time axis to the required point 
    to start measurement of Aorta root diameter.
    See the Status bar 
    to get the next step 
    to perform.6. Press SELECT.
    The starting point for the measurement is anchored.
    7. Trackball to the end point of the measurement.
    The current value is 
    updated while mov-
    ing the cursor.8. Press SELECT.
    The measurement end point is anchored and the value is 
    displayed in the Measurement result table.
    A new free-moving cursor is displayed on the image, ready 
    for the next measurement.
    9. Repeat steps 5, through 8 to measure Left Atrium.
    The LA value is displayed in the Measurement result table. 
    The Ao/LA ratio is displayed in the Measurement result 
    table.
    LV study
    The LV study consists of measurements in fixed-time mode in 
    both systole and diastole of:
    •  Interventricular septum thickness (IVS)
    •  Left ventricular internal dimension (LVID)
    •  Left ventricular posterior wall thickness (LVPW)
    The following parameters are also calculated:
    •  EDV (End diastole volume)
    •  ESV (End systole volume)
    •  SV (Stroke volume)
    • EF (Ejection Fraction)
    To perform LV study
    1. Generate the M-Mode image.
    2. Press 
    FREEZE to stop the cineloop.
    Alternative: Press 
    CALIPER on the con-
    trol panel and press 
    the 
    LV STUDY as-
    signable.
    3. Press MEASURE on the Control Panel.
    4. Select LV study in the Measurement Menu.
    5. Trackball the cursor along the time axis to the required point 
    to start measurement of IVSd.
    6. Press 
    SELECT.
    The starting point for the measurement is anchored.
    7. Trackball to the end point of the measurement. 
    						
    							Cardiac Measurement and Analysis
    222Vivid7/Vivid7 PRO Users Manual
    FC092326-03
    8. Press SELECT.
    The IVSd measurement end point is anchored and the 
    value is displayed in the Measurement result table.
    The end point of the IVSd is also the start point for the LVIDd.
    1. Trackball to the end point of the LVIDd measurement.
    2. Press 
    SELECT.
    The LVIDd measurement end point is anchored and the 
    value is displayed in the Measurement result table.
    The end point of the LVIDd is also the start point for the 
    LVPWd.
    1. Trackball to the end point of the LVPWd measurement.
    2. Press 
    SELECT.
    The LVPWd measurement end point is anchored and the 
    value is displayed in the Measurement result table.
    3. Repeat steps 5, through 2 to measure IVS, LVID and LVPW 
    in systole.
    RV study
    The RV study consists of measurement in fixed-time mode of 
    Right ventricular internal dimension (RVID) in both diastole and 
    systole.
    To perform RV study
    1. Generate the M-Mode image.
    2. Press 
    FREEZE to stop the cineloop.
    Alternative: Press 
    CALIPER on the con-
    trol panel and press 
    the 
    RV STUDY as-
    signable.
    3. Press MEASURE on the Control Panel.
    4. Select RV study in the Measurement Menu.
    5. Trackball the cursor along the time axis to the required point 
    to start measurement of RVIDd.
    6. Press 
    SELECT.
    The starting point for the measurement is anchored.
    The current value is 
    updated while mov-
    ing the cursor.7. Trackball to the end point of the measurement.
    8. Press 
    SELECT.
    The measurement end point is anchored and the RVIDs 
    measurement value is displayed in the Measurement 
    result table.
    A new free-moving cursor is displayed on the image, ready 
    for the next measurement.
    9. Repeat steps 5, through 8 to measure RVIDs. 
    						
    							Cardiac Measurement and Analysis
    Vivid7/Vivid7 PRO Users Manual223
    FC092326-03
    The RVIDs value is displayed in the Measurement result 
    table. 
    						
    							Cardiac Measurement and Analysis
    224Vivid7/Vivid7 PRO Users Manual
    FC092326-03
    Doppler Measurements
    The following measurements may be calculated on Doppler 
    mode spectra:
    For measurement 
    formulae, refer to 
    the Reference 
    Manual.•  Maximum (peak) and mean velocity
    •  Maximum and mean pressure gradient
    • Pressure half-time (PHT)
    •  Velocity time integral (VTI)
    •  Mitral valve area (MVA), derived from PHT
    Velocity and Pressure point measurements
    1. Generate the spectrum to be measured.
    2. Press 
    FREEZE to stop the cineloop.
    Alternative: Press 
    CALIPER on the con-
    trol panel and press 
    the 
    POINT CALIPER 
    assignable.
    3. Press MEASURE on the Control Panel.
    4. Select Point Caliper in the Measurement Menu.
    5. Trackball the cursor to the position to measure.
    The current velocity is displayed in the Measurement result 
    table and is instantaneously updated when moving the 
    cursor.
    The measurement 
    display on the spec-
    trum and the corre-
    sponding results 
    are numbered.6. Press SELECT to anchor the point.
    The velocity (m/s) and pressure (mmHg) values are 
    displayed in the Measurement result table.
    Velocity and Pressure caliper measurements
    1. Generate the spectrum to be measured.
    2. Press 
    FREEZE to stop the cineloop.
    Alternative: Press 
    CALIPER on the con-
    trol panel and press 
    the 
    CALIPER assign-
    able.
    3. Press MEASURE on the Control Panel.
    4. Select Caliper in the Measurement Menu.
    5. Trackball the cursor to the start point of the measurement.
    6. Press 
    SELECT to anchor the start point of the measurement.
    7. Trackball the cursor to the measurement end point.
    The current velocity and pressure values are displayed in 
    the Measurement result table and are instantaneously 
    updated when moving the cursor.
    The measurement 
    display color on the 
    spectrum changes 
    from green to red af-
    ter completion of 
    the measurement.8. Press SELECT to anchor the end point of the measurement.
    The following measurement results are displayed in the 
    Measurement result table:
    •  Velocity and pressure at anchor point positions
    •  Velocity (V3) and pressure (p3) differences between 
    anchor point position 
    						
    							Cardiac Measurement and Analysis
    Vivid7/Vivid7 PRO Users Manual225
    FC092326-03
    •  Time difference (dT) between anchored points position
    The measurement 
    display on the spec-
    trum and the corre-
    sponding result are 
    numbered.9. To assign a label to the measurement, see page 227.
    10. Repeat steps 5 through 8 to make additional 
    measurements.
    Manual Doppler trace measurements
    Adjust Compress 
    and reject controls 
    to optimize the 
    Doppler signal.1. Generate the spectrum to be measured.
    2. Press 
    FREEZE to stop the cineloop.
    3. Press 
    MEASURE on the Control Panel.
    Alternative: Press 
    CALIPER on the con-
    trol panel and press 
    the 
    MANUAL TRACE 
    assignable
    4. Select Manual trace in the Measurement Menu.
    A vertical green cursor is displayed on the spectrum.
    5. Trackball the cursor to the start point on the left side of the 
    trace.
    6. Press 
    SELECT to anchor the start point of the measurement.
    7. With the trackball, trace the Doppler envelope.
    The trace can be adjusted, while tracing, by moving the 
    cursor backward to erase portion of the trace (or the entire 
    trace) and then create the trace again.
    The measurement 
    display color on 
    spectrum changes 
    from green to red af-
    ter completion of 
    the measurement.8. Press SELECT to complete the trace.
    The following measurement results are displayed in the 
    Measurement result table:
    •  Maximum and mean Velocities
    •  Maximum and mean pressures
    • Env. Ti 
    •  Velocity time integral (VTI)
    9. Trackball the cursor to the start point of the next heart beat.
    10. Press the 
    SELECT to anchor the next heart beat starting 
    point.
    The heart rate (BPM) is displayed in the Measurement 
    result table.
    Automatic Doppler trace measurements
    Adjust Compress 
    and reject controls 
    to optimize the 
    Doppler signal.1. Generate the spectrum to be measured.
    2. Press 
    FREEZE to stop the cineloop.
    3. Press 
    MEASURE on the Control Panel.
    Alternative: Press 
    CALIPER on the con-
    trol panel and press 
    the 
    AUTO TRACE as-
    signable.
    4. Select Auto Trace in the Measurement Menu.
    A vertical green cursor is displayed on the spectrum.
    5. Trackball the cursor to the starting point.
    6. Press 
    SELECT to anchor the start point of the measurement. 
    						
    							Cardiac Measurement and Analysis
    226Vivid7/Vivid7 PRO Users Manual
    FC092326-03
    7. Trackball to the end trace position.
    8. Press 
    SELECT to anchor the end point of the trace.
    The trace is automatically generated and the following 
    measurements are displayed in the Measurement result 
    table:
    •  Maximum and mean Velocities
    •  Maximum and mean pressures
    • Env. Ti 
    •  Velocity time integral (VTI)
    9. Trackball the cursor to the next heart beat.
    10. Press 
    SELECT to anchor the next heart beat starting point.
    The heart rate (BPM) is displayed in the Measurement 
    result table.
    MV E/A ratio
    Adjust Compress 
    and reject controls 
    to optimize the 
    Doppler signal.1. Generate the spectrum to be measured.
    2. Press 
    FREEZE to stop the cineloop.
    3. Press 
    MEASURE on the Control Panel.
    Alternative: Press 
    CALIPER on the con-
    trol panel and press 
    the 
    MV E/A RATIO as-
    signable.
    4. Select MV E/A ratio in the Measurement Menu.
    5. Trackball the cursor to the peak of the E wave.
    6. Press 
    SELECT to anchor the point.
    7.Drag cursor to baseline to mark dT.
    8. Press 
    SELECT on the trackball area to anchor the second 
    point.
    9. Trackball the cursor to the peak of A wave.
    10. Press 
    SELECT to anchor the point.
    the velocity at peak for E and A waves and the calculated 
    E/A ratio are displayed in the Measurement result table. 
    						