    							Quantitative Analysis
    Vivid7/Vivid7 PRO Users Manual237
    FC092326-03
    Quantitative Analysis window
    Over view
     Figure 7-1: The Quantitative analysis window (here with TVI data)
    The color cineloop window
    1. Color cineloop window
    2. Tissue cineloop window
    3. Analysis window
    4. Sample Area5. Time at cursor position and velocity at cursor 
    position
    6. Trackball Assignments
    7. Sample area tools
    7 
    						
    							Quantitative Analysis
    238Vivid7/Vivid7 PRO Users Manual
    FC092326-03
    Displays TVI, Tissue Tracking, Strain, Strain 
    rate or Angio color-coded data.
    Sample area (1):
    Indicates sampling position of the velocity 
    (TVI), displacement (Tissue Tracking), percent 
    deformation (Strain), deformation rate (Strain 
    rate) or intensity (Contrast) trace. The sample 
    area is color-coded: the first sample area is 
    yellow, the second green...etc.
    a) Shown only when a sample area is 
    selected (pointed at).
    b) With Contrast data only.
    c) Shown only when pointing at an 
    anchored sample area.The cineloop windows system menu
    This menu is entered by pressing 
    UPDATE 
    MENU when the QA trackball cursor is within 
    one of the Cineloop windows.
    • Delete all Sample areas: removes all traces 
    at once.
    • Disable frame
    b: the current frame is 
    excluded from the cineloop display.
    • Set Sample area Shape
    a: enables resizing 
    of a selected sample area by setting height, 
    width and tilt angle. The trackball marker 
    must be pointed at an anchored sample area.
    • Label Sample area...
    a: set a descriptive 
    name to the sample area. The label is useful 
    for identification of the sample area when 
    exporting data. 
    • Delete anchor
    c: remove anchoring from a 
    dynamic sample area (see also page 244 
    and page 245).
    • Delete Sample area
    a: removes selected 
    sample area from the Cineloop window and 
    belonging trace in the Analysis window. The 
    trackball marker must be pointed at an 
    anchored sample area.
    • Cancel: exits the System menu.
    1 
    						
    							Quantitative Analysis
    Vivid7/Vivid7 PRO Users Manual239
    FC092326-03
    The Tissue cineloop window
    Displays 2D data
    Sample area (1):
    Indicates sampling position of the velocity 
    (TVI), displacement (Tissue Tracking), percent 
    deformation (Strain), deformation rate (Strain 
    rate) or intensity (Contrast) trace. The sample 
    area is color-coded: the first sample area is 
    yellow, the second green...etc.
    Sample area tools:
    •  : creates a sample area based on 
    freehand drawing.
    •  : creates a sample area with a pre-
    defined circular/elliptic shape (configurable, 
    see page 250)
    1 
    						
    							Quantitative Analysis
    240Vivid7/Vivid7 PRO Users Manual
    FC092326-03
    The analysis window
    TVI:
    Displays velocity trace
    1. Y axis: velocity scale (cm/s)
    2. X axis: Time (s)
    3. ECG
    4. Time at cursor position
    5. Velocity at cursor position
    6. Velocity at frame marker position 
    (color coded)
    Tissue Tracking:
    Displays tissue displacement trace
    1. Y axis: displacement scale (mm)
    2. X axis: time (s)
    3. ECG with Tracking start and Tracking 
    end markers
    4. Time at cursor position
    5. Displacement at cursor position
    6. Displacement at frame marker 
    position (color coded)
    Strain rate:
    Displays Strain rate trace (rate of deformation 
    (s
    -1))
    1. Y axis: s-1
    2. X axis: time (s)
    3. ECG
    4. Time at cursor position
    5. Strain rate at cursor position
    6. Strain rate at frame marker position
    1
    2
    3
    45
    -2.3
    6 
    						
    							Quantitative Analysis
    Vivid7/Vivid7 PRO Users Manual241
    FC092326-03
    Strain:
    Displays Strain trace (extent of tissue 
    deformation (%))
    1. Y axis: percent displacement
    2. X axis: time (s)
    3. ECG with Strain start and Strain end 
    markers
    4. Time at cursor position
    5. % deformation at cursor position
    6. % deformation at frame marker 
    position (color coded)
    Contrast:
    Displays time-intensity curve
    1. Y axis: Intensity scale (logarithmic) 
    (dB) or linear acoustic units (AU).
    2. X axis: Time (s) or dT (s), elapsed 
    time from previous frame.
    3. ECG: displays ECG trace, the current 
    frame marker and the start and stop 
    markers for the cineloop.
    4. Time at cursor position
    5. Intensity (dB or AU) at cursor position
    6. Intensity (dB or AU) at frame marker 
    position (color coded)
    1
    2
    3
    45
    -2.3
    6 
    						
    							Quantitative Analysis
    242Vivid7/Vivid7 PRO Users Manual
    FC092326-03
    a) With contrast data only.
    b) Shown only in zoom mode.The analysis window system menu:
    This menu is entered by pressing 
    UPDATE 
    MENU when the QA cursor is within one of the 
    analysis window.
    • Delete all Sample areas: removes all traces 
    at once.
    • Analysis signal: toggles trace display 
    between velocity, displacement, strain rate, 
    stain or greyscale intensity curves.
    • Drift compensation: compensates drifting 
    of strain or Tissue Tracking curves by either 
    resetting the curve to zero at the tracking 
    start point (cycle resetting) or by linear 
    compensation throughout the cycle (linear 
    compensation)
    • Horizontal scale: set horizontal unit as time 
    (s) or time interval (dt) between frames.
    • Vertical auto-scaling: selects between full 
    unit range or a range according to the 
    maximum and minimum values of the 
    displayed trace(s).
    • Vertical unit
    a: toggles between logarithmic 
    (dB) and linear acoustical units (AU).
    • Line style: selects between solid line only or 
    solid line with square markers at each data 
    point.
    • Smoothing: smooths the trace displayed by 
    applying a filter over a defined time window. 
    Both the filter type and time window are user-
    selectable. The type of filter available is 
    depending on the analysis signal displayed.
    • Export traces: saves trace data in ASCII 
    format, readable in spreadsheet programs. If 
    present, trace data for physiological traces 
    are also exported.
    • Curve fitting
    a: toggles between Wash-in, 
    Wash-out and off.
    • Unzoom
    b: restores full analysis window 
    display when in zoom mode.
    • Cancel: exits the System menu. 
    						
    							Quantitative Analysis
    Vivid7/Vivid7 PRO Users Manual243
    FC092326-03
    The Trackball assignments
    QA:
    Pointing tool in Quantitative analysis mode.
    Scroll/Speed:
    •  When the cineloop is stopped, enables 
    scrolling through the cineloop.
    •  When the cineloop is running, enables 
    control of the cine replay speed. 
    						
    							Quantitative Analysis
    244Vivid7/Vivid7 PRO Users Manual
    FC092326-03
    Generation of a trace
    Up to eight traces can be generated.
    About the sample area
    The sample area can be in three different states:
    • Free sample area: freely moving sample area (QA cursor) 
    before anchoring.
    The free sample area 
    disappears when the 
    QA cursor is moved 
    over a static an-
    chored frame.• Static sample area: the free sample area is anchored by 
    pressing 
    SELECT on the Trackball area.
    • Dynamic anchored sample area: the sample area is 
    anchored in two or more frames (see Manual tracking 
    below). In these particular frames, the sample area is 
    displayed with an anchor. The sample area moves 
    smoothly between the anchored positions when playing/
    scrolling the cineloop.
    To generate a trace
    Trace from a pre-defined sample area
    The shape of the pre-defined sample area is configurable (see 
    page 250).
    1. If the Trackball assignment is not on QA, press 
    TRACKBALL 
    until QA is highlighted.
    The trace and sam-
    ple area are color-
    coded. First gener-
    ated trace is yellow, 
    second green...etc.2. If necessary, select the sample area Shape button . 
    3. Trackball to one of the Cineloop windows.
    The trackball cursor is changed to a sample area (white 
    circle).
    A preview of the trace is displayed in the Analysis window.
    4. Press 
    SELECT to anchor the sample area.
    In this frame the sample are is marked with an anchor.
    If the cineloop has more than one heart cycle a sample 
    area will also be anchored in the corresponding frame in 
    the next heart cycles.
    The trace is updated accordingly in the Analysis window.
    The Strain cursor
    In Strain and Strain rate modes, the sample area displays a 
    Strain cursor showing the segment along the beam direction 
    that is used for Strain and Strain rate calculations. Make sure  
    						
    							Quantitative Analysis
    Vivid7/Vivid7 PRO Users Manual245
    FC092326-03
    that the Strain cursor is within the myocardium when anchoring 
    the sample area.
    Trace from a freehand sample area
    1. Select the Pencil button .
    2. Trackball to one of the Cineloop windows.
    The trackball cursor is changed to a cross.
    3. Press and hold down the 
    SELECT button while drawing a 
    sample area with the trackball.
    4. Release the 
    SELECT button.
    The sample area is automatically closed.
    The trace is updated accordingly in the Analysis window.
    Manual tracking of the sample area 
    (dynamic anchored sample area)
    The sample area can be moved within the loop to ensure that 
    data in the trace are generated from the same anatomical 
    location during the cyclic motion of the heart.
    1. Place a sample area over a region of interest.
    Note the anatomical location of the sample area.
    2. Using the trackball, scroll to a new frame.
    3. Press 
    TRACKBALL until the QA trackball assignment is 
    selected.
    4. Trackball to the sample area.
    5. Press 
    SELECT.
    The sample area is unanchored.
    6. Drag the sample area to the corresponding anatomical 
    location in the new frame.
    In the original 
    frame and this par-
    ticular frame the 
    sample area is 
    marked with an an-
    chor.When the sample area is anchored in more than one 
    frame, linear interpolation is performed, so that the sample 
    area is smoothly moved between the anchored positions in 
    the selected frames when running the cineloop.
    7. Press 
    TRACKBALL until the Scroll trackball assignment is 
    selected.
    8. Using the Trackball, scroll through the cineloop and control 
    that the sample area follows the moving anatomical 
    structure.
    9. Add anchored sample areas in several frames to obtain a 
    more accurate displacement of the sample area. 
    						
    							Quantitative Analysis
    246Vivid7/Vivid7 PRO Users Manual
    FC092326-03
    To move a dynamic anchored sample area
    1. Press TRACKBALL until the Scroll trackball assignment is 
    selected.
    In these frames the 
    sample area is 
    marked with an an-
    chor.2. Using the trackball, browse through the cineloop to display 
    one of the frames where the sample area was anchored.
    3. Press 
    TRACKBALL until the QA trackball assignment is 
    selected.
    4. Trackball to the sample area to move, in one of the 
    Cineloop windows.
    5. Press 
    SELECT.
    The sample area is unanchored.
    6. Drag the sample area to a new location.
    7. Press 
    SELECT to anchor the sample area to the new 
    location.
    Zooming in the Analysis window
    1. In the Analysis window, press and hold down the SELECT 
    key while dragging the trackball cursor to define the 
    zooming area.
    2. Release the 
    SELECT key.
    The selected area is displayed in the Analysis window.
    To unzoom
    1. Press UPDATE MENU in the trackball area on the control 
    panel.
    The System menu is displayed.
    2. Trackball to Unzoom.
    3. Press 
    SELECT. 
    						