Quantitative Analysis Vivid7/Vivid7 PRO Users Manual237 FC092326-03 Quantitative Analysis window Over view Figure 7-1: The Quantitative analysis window (here with TVI data) The color cineloop window 1. Color cineloop window 2. Tissue cineloop window 3. Analysis window 4. Sample Area5. Time at cursor position and velocity at cursor position 6. Trackball Assignments 7. Sample area tools 7
Quantitative Analysis 238Vivid7/Vivid7 PRO Users Manual FC092326-03 Displays TVI, Tissue Tracking, Strain, Strain rate or Angio color-coded data. Sample area (1): Indicates sampling position of the velocity (TVI), displacement (Tissue Tracking), percent deformation (Strain), deformation rate (Strain rate) or intensity (Contrast) trace. The sample area is color-coded: the first sample area is yellow, the second green...etc. a) Shown only when a sample area is selected (pointed at). b) With Contrast data only. c) Shown only when pointing at an anchored sample area.The cineloop windows system menu This menu is entered by pressing UPDATE MENU when the QA trackball cursor is within one of the Cineloop windows. • Delete all Sample areas: removes all traces at once. • Disable frame b: the current frame is excluded from the cineloop display. • Set Sample area Shape a: enables resizing of a selected sample area by setting height, width and tilt angle. The trackball marker must be pointed at an anchored sample area. • Label Sample area... a: set a descriptive name to the sample area. The label is useful for identification of the sample area when exporting data. • Delete anchor c: remove anchoring from a dynamic sample area (see also page 244 and page 245). • Delete Sample area a: removes selected sample area from the Cineloop window and belonging trace in the Analysis window. The trackball marker must be pointed at an anchored sample area. • Cancel: exits the System menu. 1
Quantitative Analysis Vivid7/Vivid7 PRO Users Manual239 FC092326-03 The Tissue cineloop window Displays 2D data Sample area (1): Indicates sampling position of the velocity (TVI), displacement (Tissue Tracking), percent deformation (Strain), deformation rate (Strain rate) or intensity (Contrast) trace. The sample area is color-coded: the first sample area is yellow, the second green...etc. Sample area tools: • : creates a sample area based on freehand drawing. • : creates a sample area with a pre- defined circular/elliptic shape (configurable, see page 250) 1
Quantitative Analysis 240Vivid7/Vivid7 PRO Users Manual FC092326-03 The analysis window TVI: Displays velocity trace 1. Y axis: velocity scale (cm/s) 2. X axis: Time (s) 3. ECG 4. Time at cursor position 5. Velocity at cursor position 6. Velocity at frame marker position (color coded) Tissue Tracking: Displays tissue displacement trace 1. Y axis: displacement scale (mm) 2. X axis: time (s) 3. ECG with Tracking start and Tracking end markers 4. Time at cursor position 5. Displacement at cursor position 6. Displacement at frame marker position (color coded) Strain rate: Displays Strain rate trace (rate of deformation (s -1)) 1. Y axis: s-1 2. X axis: time (s) 3. ECG 4. Time at cursor position 5. Strain rate at cursor position 6. Strain rate at frame marker position 1 2 3 45 -2.3 6
Quantitative Analysis Vivid7/Vivid7 PRO Users Manual241 FC092326-03 Strain: Displays Strain trace (extent of tissue deformation (%)) 1. Y axis: percent displacement 2. X axis: time (s) 3. ECG with Strain start and Strain end markers 4. Time at cursor position 5. % deformation at cursor position 6. % deformation at frame marker position (color coded) Contrast: Displays time-intensity curve 1. Y axis: Intensity scale (logarithmic) (dB) or linear acoustic units (AU). 2. X axis: Time (s) or dT (s), elapsed time from previous frame. 3. ECG: displays ECG trace, the current frame marker and the start and stop markers for the cineloop. 4. Time at cursor position 5. Intensity (dB or AU) at cursor position 6. Intensity (dB or AU) at frame marker position (color coded) 1 2 3 45 -2.3 6
Quantitative Analysis 242Vivid7/Vivid7 PRO Users Manual FC092326-03 a) With contrast data only. b) Shown only in zoom mode.The analysis window system menu: This menu is entered by pressing UPDATE MENU when the QA cursor is within one of the analysis window. • Delete all Sample areas: removes all traces at once. • Analysis signal: toggles trace display between velocity, displacement, strain rate, stain or greyscale intensity curves. • Drift compensation: compensates drifting of strain or Tissue Tracking curves by either resetting the curve to zero at the tracking start point (cycle resetting) or by linear compensation throughout the cycle (linear compensation) • Horizontal scale: set horizontal unit as time (s) or time interval (dt) between frames. • Vertical auto-scaling: selects between full unit range or a range according to the maximum and minimum values of the displayed trace(s). • Vertical unit a: toggles between logarithmic (dB) and linear acoustical units (AU). • Line style: selects between solid line only or solid line with square markers at each data point. • Smoothing: smooths the trace displayed by applying a filter over a defined time window. Both the filter type and time window are user- selectable. The type of filter available is depending on the analysis signal displayed. • Export traces: saves trace data in ASCII format, readable in spreadsheet programs. If present, trace data for physiological traces are also exported. • Curve fitting a: toggles between Wash-in, Wash-out and off. • Unzoom b: restores full analysis window display when in zoom mode. • Cancel: exits the System menu.
Quantitative Analysis Vivid7/Vivid7 PRO Users Manual243 FC092326-03 The Trackball assignments QA: Pointing tool in Quantitative analysis mode. Scroll/Speed: • When the cineloop is stopped, enables scrolling through the cineloop. • When the cineloop is running, enables control of the cine replay speed.
Quantitative Analysis 244Vivid7/Vivid7 PRO Users Manual FC092326-03 Generation of a trace Up to eight traces can be generated. About the sample area The sample area can be in three different states: • Free sample area: freely moving sample area (QA cursor) before anchoring. The free sample area disappears when the QA cursor is moved over a static an- chored frame.• Static sample area: the free sample area is anchored by pressing SELECT on the Trackball area. • Dynamic anchored sample area: the sample area is anchored in two or more frames (see Manual tracking below). In these particular frames, the sample area is displayed with an anchor. The sample area moves smoothly between the anchored positions when playing/ scrolling the cineloop. To generate a trace Trace from a pre-defined sample area The shape of the pre-defined sample area is configurable (see page 250). 1. If the Trackball assignment is not on QA, press TRACKBALL until QA is highlighted. The trace and sam- ple area are color- coded. First gener- ated trace is yellow, second green...etc.2. If necessary, select the sample area Shape button . 3. Trackball to one of the Cineloop windows. The trackball cursor is changed to a sample area (white circle). A preview of the trace is displayed in the Analysis window. 4. Press SELECT to anchor the sample area. In this frame the sample are is marked with an anchor. If the cineloop has more than one heart cycle a sample area will also be anchored in the corresponding frame in the next heart cycles. The trace is updated accordingly in the Analysis window. The Strain cursor In Strain and Strain rate modes, the sample area displays a Strain cursor showing the segment along the beam direction that is used for Strain and Strain rate calculations. Make sure
Quantitative Analysis Vivid7/Vivid7 PRO Users Manual245 FC092326-03 that the Strain cursor is within the myocardium when anchoring the sample area. Trace from a freehand sample area 1. Select the Pencil button . 2. Trackball to one of the Cineloop windows. The trackball cursor is changed to a cross. 3. Press and hold down the SELECT button while drawing a sample area with the trackball. 4. Release the SELECT button. The sample area is automatically closed. The trace is updated accordingly in the Analysis window. Manual tracking of the sample area (dynamic anchored sample area) The sample area can be moved within the loop to ensure that data in the trace are generated from the same anatomical location during the cyclic motion of the heart. 1. Place a sample area over a region of interest. Note the anatomical location of the sample area. 2. Using the trackball, scroll to a new frame. 3. Press TRACKBALL until the QA trackball assignment is selected. 4. Trackball to the sample area. 5. Press SELECT. The sample area is unanchored. 6. Drag the sample area to the corresponding anatomical location in the new frame. In the original frame and this par- ticular frame the sample area is marked with an an- chor.When the sample area is anchored in more than one frame, linear interpolation is performed, so that the sample area is smoothly moved between the anchored positions in the selected frames when running the cineloop. 7. Press TRACKBALL until the Scroll trackball assignment is selected. 8. Using the Trackball, scroll through the cineloop and control that the sample area follows the moving anatomical structure. 9. Add anchored sample areas in several frames to obtain a more accurate displacement of the sample area.
Quantitative Analysis 246Vivid7/Vivid7 PRO Users Manual FC092326-03 To move a dynamic anchored sample area 1. Press TRACKBALL until the Scroll trackball assignment is selected. In these frames the sample area is marked with an an- chor.2. Using the trackball, browse through the cineloop to display one of the frames where the sample area was anchored. 3. Press TRACKBALL until the QA trackball assignment is selected. 4. Trackball to the sample area to move, in one of the Cineloop windows. 5. Press SELECT. The sample area is unanchored. 6. Drag the sample area to a new location. 7. Press SELECT to anchor the sample area to the new location. Zooming in the Analysis window 1. In the Analysis window, press and hold down the SELECT key while dragging the trackball cursor to define the zooming area. 2. Release the SELECT key. The selected area is displayed in the Analysis window. To unzoom 1. Press UPDATE MENU in the trackball area on the control panel. The System menu is displayed. 2. Trackball to Unzoom. 3. Press SELECT.