Presets and System setup Vivid7/Vivid7 PRO Users Manual397 FC092326-03 Imaging • Global: enables the user to configure display-related settings. • Application: enables configuration of the probe and application specific settings. • Application menu: enables configuration of the Measurement menu. The Global setup sheet Figure 12-3: The Global setup sheet
Presets and System setup 398Vivid7/Vivid7 PRO Users Manual FC092326-03 Cineloop store Crop images Doppler Parameter Description Cineloop store: • Time before/after heart cycle: sets the total storage time span of the cineloop in ECG mode. • Time span (no ECG): sets the total storage time span of the cineloop with no ECG. • Preview loop before store: when selected enable review of cineloops before storage. Parameter Description Crop images: : In the Analysis screen, removes top and bottom of the image when more than two images have been selected. Parameter Description Doppler: • Show KHz scale: when selected, displays the KHz scale on the left side of the Doppler spectrum (see page 112).
Presets and System setup Vivid7/Vivid7 PRO Users Manual399 FC092326-03 Patient Info Scan Info Parameter Description Patient Info: • Title bar Line 1 & 2: selects from the pop-up menu the patient information to display on the scanning screens Title bar (see page 52). • Anonymous patient: when checked, no patient information is displayed on the scanning screens Title bar. Parameter Description Scan Info: • : displays scan information on the video record.
Presets and System setup 400Vivid7/Vivid7 PRO Users Manual FC092326-03 Application The Application category enables the configuration of probe/ application specific settings (presets). The application-specific settings can be stored and used as default presets with this probe. Figure 12-4: The Application setup sheet (example)
Presets and System setup Vivid7/Vivid7 PRO Users Manual401 FC092326-03 The Probe/application configuration parame- ters Parameter Description Image Store settings: • Single frame (live store): : Store cineloop. : Store single frame image only. • Number of heart cycles: Select the number of heart cycles to store (Single frame must be unchecked). Auto freeze: • Freeze 2D image in Doppler: the last 2D or color flow image is displayed when entering in Doppler mode. • Auto freeze after: sets the time after which the system enters in freeze when not in use. Footswitch functionality: Configures the footswitch pedal for the selected application. Select the operation to perform for each pedal from the associated Pedal pop-up menu. Templates and Packages: Defines the default stress protocol associated to the application. Select the default Protocol to be associated to the selected application from the pop-up menu. Auto invert on steer: In Color flow, the color bar is inverted when steering the color flow sector angle.
Presets and System setup 402Vivid7/Vivid7 PRO Users Manual FC092326-03 Create a new Application The application created is probe dependant. Select the desired probe before configuring a new application. 1. Press APPLICATION on the Control panel. 2. Trackball to Preset... in the Application pop-up menu. The Application setup sheet (see Figure 12-4) is displayed. 3.Adjust the parameters as desired (see page 401). 4. Press New. A Dialogue window is displayed. 5.Enter a name for the new application. 6. Press OK To edit an application 1. Press APPLICATION and select the application to edit. 2. Press APPLICATION again. 3. Trackball to Preset... in the Application pop-up menu. The Application setup sheet (see Figure 12-4) is displayed. 4. Change the parameters as desired (see page 401). 5. Press Save to store the changes. Applicable only on user-defined applications. Create new application: Press New to create a new Application. A dialogue window is displayed where the operator is asked to give a name to the new application. Remove current application: Press Delete to remove the current application. Factory Application settings cannot be deleted. Save image/appl. settings Press Save to store the changes applied to the current setting. Not applicable on factory application settings. Parameter Description
Presets and System setup Vivid7/Vivid7 PRO Users Manual403 FC092326-03 Deletion of an Application 1. Press APPLICATION and select the application to delete. 2. Press APPLICATION again. 3. Trackball to Preset... in the Application pop-up menu. The Application setup sheet (see Figure 12-4) is displayed. 4. Press Delete to remove the selected application.
Presets and System setup 404Vivid7/Vivid7 PRO Users Manual FC092326-03 Application menu The Application menu category enables rearrangement of the the Application menu to best suit the users requirements. The Application menu is a two-levels pop-up menu. The first level called Application, displays the most frequently used applications in any desired order. The second level called More... displays the less frequently used applications. Figure 12-5: The Application menu setup sheet (example) 1. First menu level 2. Second menu level3. Moving tools
Presets and System setup Vivid7/Vivid7 PRO Users Manual405 FC092326-03 Configuration of the Application menu The Application menu can be configured by moving the applications up and down inside the pop-up menu and from one level to the other. To move an application inside one level 1. Trackball to the application to move. 2. Press SELECT. 3. Press . The application is moved one step up. Press Default to get factory setting.4. Press . The application is moved one step down. To move an application from one level to the other 1. Trackball to the application to move. 2. Press SELECT. 3. Press as many times as necessary: • if the application to move is in the More menu • if the application to move is in the Applications menu till the application has moved to the other menu.
Presets and System setup 406Vivid7/Vivid7 PRO Users Manual FC092326-03 Measure The Measure setup category enables the organization of the Factory default Measurement menu or the creation of user- defined Measurement menus to best suit the users requirements. Figure 12-6: The Measurement menu setup sheet 1. Configuration window (see next pages for details)2. The measurement menu (displays updated configuration)