    							Cardiac Measurement and Analysis
    Vivid7/Vivid7 PRO Users Manual227
    FC092326-03
    Post-measurement assignment labels
    Each type of measurement, within each mode, can be 
    associated with a set of pre-defined parameter labels. 
    Parameter labels can be assigned to the highlighted 
    measurement by the user.
    To assign a parameter label to a measure-
    ment:
    1. Trackball to the actual measurement in the Measurement 
    result table (see Figure 6-5).
    The selection of a 
    measurement with-
    out pre-defined la-
    bels will not display 
    the Parameter la-
    bel menu.2. Press SELECT.
    A Parameter label menu is displayed.
    3. Trackball through the Parameter label menu to highlight the 
    required label.
    4. Press 
    SELECT to assign the highlighted parameter label to 
    the measurement.
    The assigned measurements may be reviewed in the 
    Worksheet (see page 230). Up to five assigned measurements 
    with the same label can be stored in the patient archive.
    CAUTIONOnly assigned measurements will be saved. Measurements 
    without assignment will be lost when scanning is resumed. 
    						
    							Cardiac Measurement and Analysis
    228Vivid7/Vivid7 PRO Users Manual
    FC092326-03
     Figure 6-5: Measurement assignment
    1. Parameter Label menu
    2. Selected label
    Assignment
    3. Assigned 
    measurement 
    						
    							Cardiac Measurement and Analysis
    Vivid7/Vivid7 PRO Users Manual229
    FC092326-03
    Controlling the Measurement result table 
    display
    The display of the Measurement result table can be minimized 
    and moved to prevent the table obscuring parts of the 
    ultrasound image.
    Minimizing the Measurement result 
    table
    1. Trackball to the symbol   on the heading of the 
    Measurement result table (see Figure 6-6).
    2. Press 
    SELECT.
    Repeat step 1 to en-
    large the Measure-
    ment result table.The Measurement result table is minimized to the heading 
    bar.
    Moving the Measurement result table
    Alternative: Rotate 
    the assignable  
    RESULT WINDOW to 
    move the Measure-
    ment result table 
    from corner to cor-
    ner on the Acquisi-
    tion window.
    1. Trackball to the symbol   on the heading of the 
    Measurement result table (see Figure 6-6).
    2. Press 
    SELECT to grab the table.
    3. Trackball the Measurement result table to a new position.
    4. Press 
    SELECT to anchor the table.
     Figure 6-6: Measurement result table display tools
    1. Minimize/maximize 
    table
    2. Move table 
    						
    							Cardiac Measurement and Analysis
    230Vivid7/Vivid7 PRO Users Manual
    FC092326-03
    Worksheet
    The worksheet function enables the user to review, edit, delete 
    or print data independently of a report. All measurements and 
    calculations taken during the examination can be viewed at any 
    time using the worksheet.
    Over view
     Figure 6-7: The Worksheet screen
    Using Worksheet
    1. Press WORKSHEET on the control panel and select the 
    1. Measurement type
    2. Measurement parameter 
    3. Value: Averaging, Max, Min or Last4. Measured / calculated values
    5. Value type
    6. Measurement type selection 
    						
    							Cardiac Measurement and Analysis
    Vivid7/Vivid7 PRO Users Manual231
    FC092326-03
    measurement type (see Figure 6-7).
    To scroll through pages
    1. Select PAGE DOWN or PAGE UP.
    To select the type of value
    1. Trackball to the relevant cell in the Method column.
    2. Press 
    SELECT.
    A pop-up menu is displayed showing the different options 
    available (Figure 6-8).
     Figure 6-8: The Calculation method options list.
    3. Trackball to the required option.
    4. Press 
    SELECT.
    The value is updated accordingly.
    Excluding or including measurements
    One or more measurement values from a set of measurements 
    for a parameter can be excluded when doing average 
    calculation.
    To exclude a measurement
    1. Trackball to the measurement value to exclude.
    2. Press 
    UPDATE MENU.
    The Worksheet menu is displayed.
    When excluded the 
    measurement dis-
    play turns grey.3. Trackball to Exclude Value.
    4. Press 
    SELECT.
    To include a measurement
    1. Trackball to the measurement value to include.
    2. Press 
    UPDATE MENU.
    The Worksheet menu is displayed.
    1. Average of the 
    measurements taken
    2. Maximum measurement
    3. Minimum measurement.
    4. Last measurement that was 
    taken 
    						
    							Cardiac Measurement and Analysis
    232Vivid7/Vivid7 PRO Users Manual
    FC092326-03
    3. Trackball to Include Value.
    4. Press 
    SELECT.
    Manually changing a value
    Individual measured values can be manually changed using 
    the alphanumeric keyboard.
    To manually change a value
    1. Trackball to the value that is to be changed.
    2. Press 
    SELECT.
    An asterisk indi-
    cates that the value 
    has been manually 
    altered. The calcula-
    tion type is changed 
    to Edit.3. Use the alphanumeric keyboard to enter the required value.
    To restore automatic calculation
    1. Trackball to the relevant cell in the Method column.
    2. Press 
    SELECT.
    A pop-up menu is displayed showing the different 
    calculation options available (Figure 6-8).
    3. Press 
    SELECT.
    The value is re-calculated according the method selected.
    Deleting measurement parameter
    1. Trackball to the measurement parameter to delete.
    2. Press 
    UPDATE MENU.
    The Worksheet menu is displayed.
    3. Trackball to Delete Value.
    4. Press 
    SELECT. 
    						
    							Quantitative Analysis
    Vivid7/Vivid7 PRO Users Manual233
    FC092326-03
    Chapter 7 
    Quantitative Analysis
    This chapter includes the following information:
    • Introduction  ................................................................................... ..  235
    • Accessing the Quantitative analysis package  ........................... ..  236
    • Quantitative Analysis window  ..................................................... ..  237• Overview  .................................................................................. 237
    • Generation of a trace  .................................................................... ..  244 • About the sample area  ............................................................. 244
    • To generate a trace  .................................................................. 244
    • Manual tracking of the sample area (dynamic anchored sample 
    area) ...........................................................................................  245
    • Zooming in the Analysis window  .............................................. 246
    • Deletion of a trace  ......................................................................... ..  247
    • To delete all traces  ................................................................... 247
    • To delete one specific trace  ..................................................... 247
    • Frame disabling  ............................................................................ ..  248
    • Disabling frames from the frame marker  .................................. 248
    • Disabling frames in the cineloop windows  ............................... 248
    • Re-enabling all frames  ............................................................. 249
    • Optimizing sample area ................................................................ ..  250 • Reshaping a sample area  ........................................................ 250
    • Labelling a sample area  ........................................................... 251
    • Optimizing the trace display  ........................................................ ..  252 • Optimizing the Y-axis  ............................................................... 252
    • Trace smoothing  ...................................................................... 253
    • Switching modes or traces  .......................................................... ..  255 • To switch mode  ........................................................................ 255
    • To switch trace  ......................................................................... 255
    • Curve fitting analysis .................................................................... ..  256 
    						
    							Quantitative Analysis
    234Vivid7/Vivid7 PRO Users Manual
    FC092326-03
    • Anatomical M-Mode ...................................................................... ...  265
    • Introduction  ............................................................................... 265
    • Using Anatomical M-Mode ........................................................ 265 
    						
    							Quantitative Analysis
    Vivid7/Vivid7 PRO Users Manual235
    FC092326-03
    Introduction
    The quantitative analysis software package is designed for 
    analysis of TVI, Tissue Tracking, Strain, Strain rate and 
    Contrast related raw data.
    The main features of these options are:
    For TVI:
    •  Multiple Time -motion trace display from selected points in 
    the myocardium.
    •  Arbitrary Curved anatomical M-Mode
    For Tissue Tracking:
    •  Multiple tissue displacement trace display from selected 
    segments in the myocardium.
    •  Arbitrary Curved anatomical M-Mode
    For Strain rate:
    •  Multiple Strain rate (Rate of deformation (s-1)) trace display 
    from selected segments in the myocardium.
    •  Arbitrary Curved anatomical M-Mode
    For Str ain:
    •  Multiple Strain (extend of tissue deformation (%)) trace 
    display from selected segments in the myocardium.
    •  Arbitrary Curved anatomical M-Mode
    For Contrast:
    •  Time-Intensity analysis from multiple region of interest
    • Curve fitting
    •  Arbitrary Curved anatomical M-Mode
    If not otherwise specified, the topics described in this chapter 
    are applicable for both TVI and contrast data. 
    						
    							Quantitative Analysis
    236Vivid7/Vivid7 PRO Users Manual
    FC092326-03
    Accessing the Quantitative analysis package
    In replay mode:
    1. Open an examination and recall a TVI or contrast loop.
    2. Press the assignable 
    Q ANALYSIS.
    The Quantitative Analysis screen is displayed (see 
    Figure 7-1).
    In live
    1. Press FREEZE.
    Note: if in 2D mode outside a contrast application, press 
    ALT and MORE assignable.
    2. Press the assignable 
    Q ANALYSIS.
    The Quantitative Analysis screen is displayed (see 
    Figure 7-1). 
    						