Cardiac Measurement and Analysis Vivid7/Vivid7 PRO Users Manual227 FC092326-03 Post-measurement assignment labels Each type of measurement, within each mode, can be associated with a set of pre-defined parameter labels. Parameter labels can be assigned to the highlighted measurement by the user. To assign a parameter label to a measure- ment: 1. Trackball to the actual measurement in the Measurement result table (see Figure 6-5). The selection of a measurement with- out pre-defined la- bels will not display the Parameter la- bel menu.2. Press SELECT. A Parameter label menu is displayed. 3. Trackball through the Parameter label menu to highlight the required label. 4. Press SELECT to assign the highlighted parameter label to the measurement. The assigned measurements may be reviewed in the Worksheet (see page 230). Up to five assigned measurements with the same label can be stored in the patient archive. CAUTIONOnly assigned measurements will be saved. Measurements without assignment will be lost when scanning is resumed.
Cardiac Measurement and Analysis 228Vivid7/Vivid7 PRO Users Manual FC092326-03 Figure 6-5: Measurement assignment 1. Parameter Label menu 2. Selected label Assignment 3. Assigned measurement
Cardiac Measurement and Analysis Vivid7/Vivid7 PRO Users Manual229 FC092326-03 Controlling the Measurement result table display The display of the Measurement result table can be minimized and moved to prevent the table obscuring parts of the ultrasound image. Minimizing the Measurement result table 1. Trackball to the symbol on the heading of the Measurement result table (see Figure 6-6). 2. Press SELECT. Repeat step 1 to en- large the Measure- ment result table.The Measurement result table is minimized to the heading bar. Moving the Measurement result table Alternative: Rotate the assignable RESULT WINDOW to move the Measure- ment result table from corner to cor- ner on the Acquisi- tion window. 1. Trackball to the symbol on the heading of the Measurement result table (see Figure 6-6). 2. Press SELECT to grab the table. 3. Trackball the Measurement result table to a new position. 4. Press SELECT to anchor the table. Figure 6-6: Measurement result table display tools 1. Minimize/maximize table 2. Move table
Cardiac Measurement and Analysis 230Vivid7/Vivid7 PRO Users Manual FC092326-03 Worksheet The worksheet function enables the user to review, edit, delete or print data independently of a report. All measurements and calculations taken during the examination can be viewed at any time using the worksheet. Over view Figure 6-7: The Worksheet screen Using Worksheet 1. Press WORKSHEET on the control panel and select the 1. Measurement type 2. Measurement parameter 3. Value: Averaging, Max, Min or Last4. Measured / calculated values 5. Value type 6. Measurement type selection
Cardiac Measurement and Analysis Vivid7/Vivid7 PRO Users Manual231 FC092326-03 measurement type (see Figure 6-7). To scroll through pages 1. Select PAGE DOWN or PAGE UP. To select the type of value 1. Trackball to the relevant cell in the Method column. 2. Press SELECT. A pop-up menu is displayed showing the different options available (Figure 6-8). Figure 6-8: The Calculation method options list. 3. Trackball to the required option. 4. Press SELECT. The value is updated accordingly. Excluding or including measurements One or more measurement values from a set of measurements for a parameter can be excluded when doing average calculation. To exclude a measurement 1. Trackball to the measurement value to exclude. 2. Press UPDATE MENU. The Worksheet menu is displayed. When excluded the measurement dis- play turns grey.3. Trackball to Exclude Value. 4. Press SELECT. To include a measurement 1. Trackball to the measurement value to include. 2. Press UPDATE MENU. The Worksheet menu is displayed. 1. Average of the measurements taken 2. Maximum measurement 3. Minimum measurement. 4. Last measurement that was taken
Cardiac Measurement and Analysis 232Vivid7/Vivid7 PRO Users Manual FC092326-03 3. Trackball to Include Value. 4. Press SELECT. Manually changing a value Individual measured values can be manually changed using the alphanumeric keyboard. To manually change a value 1. Trackball to the value that is to be changed. 2. Press SELECT. An asterisk indi- cates that the value has been manually altered. The calcula- tion type is changed to Edit.3. Use the alphanumeric keyboard to enter the required value. To restore automatic calculation 1. Trackball to the relevant cell in the Method column. 2. Press SELECT. A pop-up menu is displayed showing the different calculation options available (Figure 6-8). 3. Press SELECT. The value is re-calculated according the method selected. Deleting measurement parameter 1. Trackball to the measurement parameter to delete. 2. Press UPDATE MENU. The Worksheet menu is displayed. 3. Trackball to Delete Value. 4. Press SELECT.
Quantitative Analysis Vivid7/Vivid7 PRO Users Manual233 FC092326-03 Chapter 7 Quantitative Analysis This chapter includes the following information: • Introduction ................................................................................... .. 235 • Accessing the Quantitative analysis package ........................... .. 236 • Quantitative Analysis window ..................................................... .. 237• Overview .................................................................................. 237 • Generation of a trace .................................................................... .. 244 • About the sample area ............................................................. 244 • To generate a trace .................................................................. 244 • Manual tracking of the sample area (dynamic anchored sample area) ........................................................................................... 245 • Zooming in the Analysis window .............................................. 246 • Deletion of a trace ......................................................................... .. 247 • To delete all traces ................................................................... 247 • To delete one specific trace ..................................................... 247 • Frame disabling ............................................................................ .. 248 • Disabling frames from the frame marker .................................. 248 • Disabling frames in the cineloop windows ............................... 248 • Re-enabling all frames ............................................................. 249 • Optimizing sample area ................................................................ .. 250 • Reshaping a sample area ........................................................ 250 • Labelling a sample area ........................................................... 251 • Optimizing the trace display ........................................................ .. 252 • Optimizing the Y-axis ............................................................... 252 • Trace smoothing ...................................................................... 253 • Switching modes or traces .......................................................... .. 255 • To switch mode ........................................................................ 255 • To switch trace ......................................................................... 255 • Curve fitting analysis .................................................................... .. 256
Quantitative Analysis 234Vivid7/Vivid7 PRO Users Manual FC092326-03 • Anatomical M-Mode ...................................................................... ... 265 • Introduction ............................................................................... 265 • Using Anatomical M-Mode ........................................................ 265
Quantitative Analysis Vivid7/Vivid7 PRO Users Manual235 FC092326-03 Introduction The quantitative analysis software package is designed for analysis of TVI, Tissue Tracking, Strain, Strain rate and Contrast related raw data. The main features of these options are: For TVI: • Multiple Time -motion trace display from selected points in the myocardium. • Arbitrary Curved anatomical M-Mode For Tissue Tracking: • Multiple tissue displacement trace display from selected segments in the myocardium. • Arbitrary Curved anatomical M-Mode For Strain rate: • Multiple Strain rate (Rate of deformation (s-1)) trace display from selected segments in the myocardium. • Arbitrary Curved anatomical M-Mode For Str ain: • Multiple Strain (extend of tissue deformation (%)) trace display from selected segments in the myocardium. • Arbitrary Curved anatomical M-Mode For Contrast: • Time-Intensity analysis from multiple region of interest • Curve fitting • Arbitrary Curved anatomical M-Mode If not otherwise specified, the topics described in this chapter are applicable for both TVI and contrast data.
Quantitative Analysis 236Vivid7/Vivid7 PRO Users Manual FC092326-03 Accessing the Quantitative analysis package In replay mode: 1. Open an examination and recall a TVI or contrast loop. 2. Press the assignable Q ANALYSIS. The Quantitative Analysis screen is displayed (see Figure 7-1). In live 1. Press FREEZE. Note: if in 2D mode outside a contrast application, press ALT and MORE assignable. 2. Press the assignable Q ANALYSIS. The Quantitative Analysis screen is displayed (see Figure 7-1).