GE Vivid 7 User Manual
Have a look at the manual GE Vivid 7 User Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.
Getting started Vivid7/Vivid7 PRO Users Manual17 FC092326-03 S Video OutNTSC/PAL S- VideoVCR or video monitorSignal level 1Vp-p SVGA OutRGB high resolution videoComputer monitor USBUniversal serial busPrinter Ethernet10/100 Base-TX Ethernet IEEE 8023Network device Socket Signal type Device type Note
Getting started 18Vivid7/Vivid7 PRO Users Manual FC092326-03 Switching On/Off To switch on the unit: 1. Switch on the circuit breaker on the rear of the unit (see Figure 1-4). 2. Press (on/off button) on the top left of the control panel (see Figure 1-4). After initialization the default scanning screen (2D mode) is displayed, the active probe being the one connected to the left most connector socket. Switching off the unit When the Vivid 7/Vivid 7 PRO is switched off, the system performs an automatic shutdown sequence. The unit can be switched off into two states. • Standby mode: most of the system is powered down, but a certain portion of the unit remains energized. The standby mode allows a shorter reboot time when the system is used on a daily basis or moved from one place to another. • Full shutdown: the entire system is shut down. It is recommended to perform a full shutdown at least once a week. Full shutdown After switching off the system, wait at least ten seconds be- fore turning it on again.1. Press (on/off button) on the top left of the control panel. The Exit dialogue window is displayed. Figure 1-3: The Exit dialogue window
Getting started Vivid7/Vivid7 PRO Users Manual19 FC092326-03 In case of total lock- up of the system, hold the on/off but- ton down a few sec- onds to turn the system off.2. Select Shutdown. The shutdown process takes a few seconds and is completed when the control panel illumination is turned off. To switch off before moving the unit, follow the additional steps below: 1. Set the circuit breaker to OFF. 2. Remove the plug from the mains power socket. 3. Secure the unit power cable around the cable storage hooks on the rear of the unit. Figure 1-4: The Circuit breaker and On/Off button Standby mode 1. Press (on/off button) on the top left of the control panel. The Exit dialogue window is displayed (Figure 1-3). 2. Select Standby. The system enters in standby mode. If the unit is to be moved while in standby, follow the additional steps below: 1. Set the circuit breaker to OFF. 2. Remove the plug from the mains power socket. F 1F1
Getting started 20Vivid7/Vivid7 PRO Users Manual FC092326-03 At the new location: 1. Secure the power plug in the wall outlet. 2. Switch on the circuit breaker on the rear of the unit (Figure 1-4). 3. Press (on/off button) (Figure 1-4). When turning on a system in standby it takes a few seconds before it responds. Do not push the on/off button again during this period (A second push will initiate a full shutdown). If the power cable is removed from the wall outlet the system will remain in Standby mode for approximately 15 min using an internal battery. If the system is unplugged for a longer time period a full shutdown is automatically performed. The system can be used regardless of the battery level. The battery is automatically charged when the system is plugged to the wall outlet and the circuit breaker is switched on. It takes about 10 hours to charge a flat battery. If the system is left in the OFF condition (Circuit breaker in the OFF position and system unplugged from the wall out let) for an extended period of time (3 to 5 days or more), the system may not boot up or may beep when turned ON. Should this occur, the system needs to recharge the UPS batteries. This could take 15 minutes to as long as 10 hours, depending on the battery age, system input voltage and system temperature. To eliminate draining the UPS batteries, the system should remain plugged into the wall outlet with the circuit breaker in the ON position.
Getting started Vivid7/Vivid7 PRO Users Manual21 FC092326-03 Moving and transporting the unit Wheels The wheels of the unit are controlled by the pedals situated between the front wheels of the unit (see Figure 1-5). Examine the wheels frequently for defects to avoid breaking or jamming. Figure 1-5: The Vivid 7/Vivid 7 PRO ultrasound unit pedalsWheel Characteristics Front Swivel, swivel lock and full lock Rear Swivel, but do not lock 1. Swivel lock pedal (Free position) 2. Full lock pedal (Free position) 3. Front wheel 12 3
Getting started 22Vivid7/Vivid7 PRO Users Manual FC092326-03 Wheel operation To engage the pedal in full lock To release the brake To engage swivel lock Pedal Action Press on pedal no. 2 Pedal Action Press on pedal no.1 Pedal Action Press on pedal no.1 12 12
Getting started Vivid7/Vivid7 PRO Users Manual23 FC092326-03 To release swivel lock Moving the unit To prepare the unit to be moved 1. If not locked, push the keyboard console to its park position. 2. Turn the system off, including the circuit breaker (see page 18), and remove the plug from the wall. 3. Disconnect all cables linking the unit to any off-board peripheral devices and network. Note the marks on each cable to recon- nect them later.4. Secure the unit’s power cable. 5. Place all probes in the probe holder. Ensure that the probe cables do not protrude from the unit or interfere with the wheels. The park port may also be used for a probe connection al- though it is not ac- tive.6. Ensure that no loose items are left on the unit 7. Unlock the brake (see page 22). To ensure safety while moving the unit 1. Ensure that the keyboard console is in locked position. 2. Proceed cautiously when crossing door or elevator thresholds. Grasp the front handle grips or the back handle Pedal Action Press on pedal no. 2 WARNINGDo not move the unit if the keyboard console is in free position. Ensure that the hands of the patient are away from the console arm when moving the keyboard console.
Getting started 24Vivid7/Vivid7 PRO Users Manual FC092326-03 bar and push or pull. Do not attempt to move the unit using cables or probe connectors. Take extra care while moving the unit on inclines. 3. Ensure that the unit does not strike the walls or door frames. 4. Ensure the pathway is clear. 5. Move the unit slowly and carefully. 6. Use two or more persons to move the unit over long distances or on inclines. Transpor ting the unit Take extra care when transporting the unit by vehicle. In addition to the moving precautions listed on page 23, follow the procedure described below. 1. If not locked, push the keyboard console to its park position. 2. Disconnect all probes and secure them in their boxes. 3. Ensure that the transporting vehicle is appropriate for the unit’s weight. The recommended load capacity is a minimum of 190 kg (419 lbs). 4. Park the vehicle on a level surface for loading and unloading. 5. Secure the unit while it is on the lift, to prevent rolling. Do not attempt to hold it in place by hand. Cushion the unit and strap the lower part so that it does not break loose. 6. Ensure that the unit is secured inside the vehicle. Secure it with straps to prevent movement while in transit. 7. Drive cautiously to prevent vibration damage. Reinstalling at a new location 1. When the unit is in place at a new location, lock the wheel CAUTIONAvoid ramps that are steeper than 10 degrees. WARNINGDo not move/lift the unit if the keyboard console is in free position.
Getting started Vivid7/Vivid7 PRO Users Manual25 FC092326-03 brakes (see page 22). 2. Follow the installation procedure described on page 12. Unit acclimation time Following transport the unit may be very cold or hot. Allow the unit to acclimate before being switched on. Acclimation will take one hour for each 2.5 ºC increment when the unit’s temperature is below 10 ºC or above 40 ºC. ºC02.557.510354042.5 ºF32 36.5 41 45.5 50 95 104 108.5 Hours43210023 ºC45 47.5 50 52.5 55 57.5 60 ºF113 117.5 122 126.5 131 135.5 140 Hours45678910
Getting started 26Vivid7/Vivid7 PRO Users Manual FC092326-03 System description System over view Figure 1-6: the Vivid 7/Vivid 7 PRO ultrasound system 1 2 3 4 5 6 78 9 10 11 12 13