Home > Hitachi > Spectrophotometer > Hitachi F 2500 Manual

Hitachi F 2500 Manual

    Download as PDF Print this page Share this page

    Have a look at the manual Hitachi F 2500 Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 913 Hitachi manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    							2 - 18 
    2.5    Conducting a Wavelength Scan   
     
    (1)  There are two methods of starting the scan.     
     
    (a)  Select the Measure command from the 
    Spectrophotometer menu.   
     
    (b)  Click the     (measurement) button on the 
    measurement toolbar.   
    For halting the measurement midway, click the   
    button on the toolbar.   
     
    (2)  Upon clicking the          button, the window shown in   
    Fig. 2-10 appears when using a sample table.    Now click 
    Yes.  
     
     
     
    Fig. 2-10 
     
     
    (3)  The dialog box as in Fig. 2-11 will open.    Set each sample 
    and click OK.  
     
     
     
    Fig. 2-11 
     
    2.5    Conducting a Wavelength Scan  
    						
    							2 - 19 
    (4)  When automatic saving is set, data will be saved after 
    completion of measurement.    Upon opening its file, a data 
    processing window will appear as in Fig. 2-12.     
    Unless automatic saving is set, select the Save As 
    command from the File menu and save data in file.   
     
     
     
    Fig. 2-12    Data Processing Window (wavelength scan) 
     
     
    (a) Spectrum  
    The spectrum of a measured sample is traced.   
     
    (b) Peak table  
    Peak wavelength, peak height and valley wavelength 
    are displayed.   
    (a) Spectrum 
    (b) Peak table 2.5  
    						
    							2 - 20 
    2.6  Data Printout  
     
    Printout can be provided even when the function for automatic 
    printout after measurement is not used.    The following two 
    procedures are available.   
     
    (1)  Select the Print command from the File menu.    Or click the   
              print button on the toolbar.    The field active on the 
    data processing window (e.g., (a) in Fig. 2-12) will be 
    printed.    What is displayed by clicking the Print Preview 
    command in the File menu will be printed.     
     
     
     
    Fig. 2-13 
     
     
    2.6  Data Printout  
    						
    							2 - 21 
    (2)  Select the Report command in the Data menu.    Or click 
    the     button on the toolbar.  The Print Preview window 
    will appear, so click Print.    Select the items to be printed 
    at the Report tab under Properties.   
     
     
     
    Fig. 2-14 
     
     
    NOTE:  If characters overflow from the print frame in 
    repeat measurement or in overlaying, then change 
    the printing orientation from vertical to horizontal.   
     
    2.6  
    						
    							3 - 1 
    3.  TIME SCAN 
     
     
    3.1  Flow of Operation  
     
    This section explains the parameter setting and operating 
    method for time scan.     
    Measurement is performed in a flow shown below. 
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
    Fig. 3-1    Operational Flow (time scan) 
    End 
    Start 
    Method setting 
    Click  button.  
     Measuring 
    wavelength 
     Scan time 
     
    Sample name/comment input
    Click  button.  
    Measurement 
    Data processing 
     Rate calculation 
     Peak detection 
     Spectrum tracing 
     Spectral calculation 
    Print  (See page 3-2.) 
    (See page 3-13.) 
     
    (See page 3-15.)   
     Click 
     button.  
    (Refer to Section 6.) 
    Click 
     button.  
    (See page 3-17.)     
     
     
    3.1  
    						
    							3 - 2 
    3.2    Creating an Analysis Method   
     
    Set the analytical conditions (method) prior to measurement.   
    From the Edit menu, select the Method command or click the       
    button.    The dialog box shown in Fig. 3-2 will appear.   
     
     
     
    Fig. 3-2  Analysis Method 
     
     
    The following tabs are provided on the Analysis Method window 
    for the time scan mode.    These tabs are explained on the 
    subsequent pages.   
    • General 
    • Instrument 
    • Monitor 
    • Processing 
    • Report  
     
    3.2    Creating an Analysis Method  
    						
    							3 - 3 
    (1) General  
     
    The parameters shown in Fig. 3-3 are provided under the 
    General tab.   
     
     
     
    Fig. 3-3  General Tab 
     
     
    (a) Measurement (mode)  
    Select a measurement mode from among the four 
    listed below.   
    • Wavelength scan  
    • Time scan  
    • Photometry  
    •  3-D Scan (displayed when the ‘3-D Scan program’ 
    (option) is installed in combination with Model   
    F-2500)  
     
    Here, select “Time scan.”    For Wavelength scan, 
    Photometry and 3-D Scan, refer to Sections 2, 4 and 5.   
     
    (b) Operator  
    Enter an operator name.     
     
    (c) Instrument  
    The model of the connected instrument is indicated.   
     
    (d)  Use sample table (Set measurement sample) 
    Apply a check mark here when a sample table is used.     
     
    3.2  
    						
    							3 - 4 
    (e) Sampling  
    • Standard  
    When sample is exchanged manually.   
     
    (f) Comments  
    Enter a description or notes on measuring conditions.   
     
    (g) Load  
    Click this button to read out saved analytical conditions.   
    The file opening window appears upon clicking, so 
    select an analytical conditions file.   
     
    (h) Save  
    By clicking this button after setting the parameters for 
    all five tabs, the analytical conditions are saved.   
    Clicking this button overwrites parameters on the 
    existing ones.   
     
    (i) Save As  
    Click this button for renaming the set conditions for 
    saving.    The “Save As” window will appear, so enter a 
    file name.   
     
    (2) Instrument  
     
    This window appears when Time scan is specified for 
    Measurement under the General tab on the Analysis 
    Method window.    When you click the Instrument tab, the 
    window in Fig. 3-4 appears.   
     
     
     
    Fig. 3-4  Instrument Tab 
    3.2    Creating an Analysis Method  
    						
    							3 - 5 
    (a) Data mode  
    Select one of the following data modes.   
    The Luminescence mode is for measuring 
    luminescence in the biological/chemical field and only 
    the sample side signal is acquired without performing 
    ratio photometry.   
    If phosphorescence life is shorter than 20 ms, 
    “Phosphorescence short life” should be selected.     
    In this case, unit time and measuring time are 
    automatically set to ms and 20 ms.     
    • Fluorescence  
    • Luminescence  
    •  Phosphorescence (only with Model F-4500)   
    •  Phosphorescence Life Time (only with Model   
    F-4500)  
    •  Phosphorescence Life Time (Short) (only with 
    Model F-4500)   
     
    (b) EX WL 
    Input an excitation wavelength.   
     
    Model Input Range  
    F-4500  0, 200 to 900 nm 
    F-2500  0, 200 to 800 nm 
     
     
    (c) EM WL  
    Input an emission wavelength.   
     
    Model Input Range  
    F-4500  0, 200 to 900 nm 
    F-2500  0, 200 to 800 nm 
     
     
    (d)  Time unit (only with Model F-4500) 
    Set a time unit.   
    • s  
    • ms  
     
    3.2  
    						
    							3 - 6 
    (e) Scan time  
    Set a scan time.     
     
    Model Input Range  
    F-4500  10 to 3600 s   
    1000 to 9999 ms   
    F-2500  10 to 9000 s 
     
    According to the set scan time, data sampling interval 
    is set as listed below.   
     
    • Model F-4500   
     
    Measuring Time   
    (s) Sampling Interval   
    (ms) 
    ~ 3  1 
    ~ 7  2 
    ~ 18  5 
    ~ 36  10 
    ~ 72  20 
    ~ 180  50 
    ~ 360  100 
    ~ 720  200 
    ~ 1800  500 
    ~ 3600  1000 
     
    In the Phosphorescence mode, the minimum sampling 
    interval is 50 ms.    It is fixed at 40 µs in 
    phosphorescence short-life measurement.   
     
    • Model F-2500 
     
    Measuring Time   
    (s) Sampling Interval   
    (ms) 
    ~ 120  100 
    ~ 240  200 
    ~ 600  500 
    ~ 1200  1000 
    ~ 2400  2000 
    ~ 6000  5000 
    ~ 9000  10000 
     
     
    3.2    Creating an Analysis Method  
    						
    All Hitachi manuals Comments (0)