Hitachi F 2500 Manual
Have a look at the manual Hitachi F 2500 Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 913 Hitachi manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.
1 - 42 1.11 Photometer Menu Commands The following commands are available in the Spectrophotometer menu. • Set Wavelength • Zero Adjust • Auto Pre-scan • Shutter Open/Close (1) Set Wavelength This command sets the excitation/emission wavelength to a specified wavelength. Click the Set Wavelength command or button, and the dialog box shown below will come up. (2) Auto Pre-scan This command automatically searches for an optimum excitation/emission wavelength for the sample under measurement. Other peaks than the fluorescence peak (such as peaks of excitation scattered light and Raman spectrum) are ignored. On completion of automatic pre- scanning, the optimum excitation/emission wavelength is indicated on the monitor window (in the current wavelength indication field detailed in 1.6). (3) Zero Adjust This command automatically closes the shutter and measures a zero point in detection. When the zero point is changed, e.g., after the photomultiplier voltage is changed, automatic zero-point adjustment is carried out. (4) Shutter Open/Close This command is available for opening/closing the excitation beam shutter. The current state is indicated on the right side of the monitor window. 1.11 Photometer Menu Commands
1 - 43 1.12 Utility Menu Commands The Utility menu contains the following commands: • Instrumental Response • Analog Output Gain • EX Power Monitor • Sens. Check • Lamp Usage • Lamp Off (only with Model F-2500) (1) Instrumental Response When you use this function. You need the Spectral correction accessories kit (P/N 251-0126). (2) Analog Output Gain Use this command for setting a data value corresponding to 1 V of analog output. For details, refer to 8.4. (3) EX Power Monitor Use this command for checking output of the monitor detector. When the EX Power Monitor is turned on (clicked), monitor value data is displayed regardless of measurement condition. (4) Sens. Check Use this command for checking sensitivity. You can check sensitivity/drift with this command. For details, refer to the maintenance edition. (5) Lamp Usage Use this command for checking/resetting the usage period of the xenon lamp. A cumulative turn-on time of the xenon lamp is indicated. After lamp replacement, reset this indication. For details, refer to the maintenance edition. (6) Lamp Off (only with Model F-2500) Use this command for turning off the xenon lamp. For turning on the xenon lamp again, restarting the spectrophotometer is necessary. 1.12
1 - 44 1.13 Termination of FL Solutions Program After completion of measurement, shut down the instrument in the following procedure. (1) From the File (F) menu, select the Exit (X) command. Fig. 1-42 Select Yes, and the FL Solutions program will then be terminated. (2) Turn off the power switch of the spectrophotometer main unit. (3) Click the Start button of the Windows 98 and select the Shut Down (U) item. In the Shut Down Windows dialog box, select “Shut down” and click OK. (4) Turn off power to the personal computer and the CRT display monitor. (At step (3), the personal computer may be turned off in some settings.) (5) Turn off power to the printer. 1.13 Termination of FL Solutions Program
2 - 1 2. WAVELENGTH SCAN 2.1 Flow of Operation This section explains the condition setting and operating method for wavelength scan. Measurement is made in the following procedure. Fig. 2-1 Operational Flow (wavelength scan) End Start Method setting Click button. Scan range Data mode Scan speed Sample name/comment input Click button. Pre-scan Measurement Data processing Spectrum tracing Peak detection, etc. Print (See page 2-2.) (See page 2-15.) (See page 2-16.) Click button. (See page 2-16.) Click button. (Refer to Section 6.) Click button. (See page 2-19.) 2.1
2 - 2 2.2 Creating an Analysis Method Set the analytical conditions (method) prior to measurement. From the Edit menu, select the Method command or click the button. The dialog box shown in Fig. 2-2 will appear. Fig. 2-2 Analysis Method The following tabs are provided on the Analysis Method window for the wavelength scan mode. These tabs are explained on the subsequent pages. General Instrument Monitor Processing Report 2.2 Creating an Analysis Method
2 - 3 (1) General The parameters shown in Fig. 2-3 are provided under the General tab. Fig. 2-3 General Tab (a) Measurement (mode) Select a measurement mode from among the four listed below. Wavelength scan Time scan Photometry 3-D Scan (displayed when the ‘3-D Scan program’ (option) is installed in combination with Model F-2500) Here, select “Wavelength scan.” For Photometry, Time scan and 3-D Scan, refer to Sections 3, 4 and 5. (b) Operator Enter an operator name. (c) Instrument The model of the connected instrument is indicated. (d) Use sample table (Set measurement sample) Apply a check mark here when a sample table is used. 2.2
2 - 4 (e) Sampling Standard When sample is exchanged manually. (f) Comments Enter a description or notes on measuring conditions. (g) Load Click this button to read out saved analytical conditions. The file opening window appears upon clicking, so select an analytical conditions file. (h) Save By clicking this button after setting the parameters for all five tabs, the analytical conditions are saved. Clicking this button overwrites parameters on the existing ones. (i) Save As Click this button for renaming the set conditions for saving. The “Save As” window will appear, so enter a file name. (2) Instrument When you click the Instrument tab, the window in Fig. 2-4 appears. Fig. 2-4 Instrument Tab 2.2 Creating an Analysis Method
2 - 5 (a) Scan mode Select one of the following scan modes. In case of Synchronous, excitation wavelength appears on the abscissa of the graph. Excitation Emission Synchronous (b) Data mode Select one of the following data modes. The Luminescence mode is for measuring luminescence in the biological/chemical field and only the sample side signal is acquired without performing ratio photometry. The Phosphorescence mode is for acquiring only phosphorescence after extinction of excitation beam while turning a chopper on the excitation side. Fluorescence Luminescence Phosphorescence (with Model F-4500) (c) EM WL This is the start wavelength for wavelength scan on the excitation side. Input the shorter wavelength side of measuring wavelength range. Model Input Range F-4500 200 to 900 nm or 200 to 890 nm in Synchronous scan mode F-2500 220 to 800 nm or 220 to 790 nm in Synchronous scan mode (d) EX Start WL This is the start wavelength for wavelength scan on the excitation side. Input the shorter wavelength side of measuring wavelength range. Model Input Range F-4500 200 to 890 nm F-2500 220 to 790 nm 2.2
2 - 6 (e) EX End WL This is the end wavelength for wavelength scan on the excitation side. Input the longer wavelength side of measuring wavelength range. Model Input Range F-4500 210 to 900 nm F-2500 230 to 800 nm (f) EX WL An excitation wavelength can be input. It is the end wavelength for wavelength scan on the excitation side. Input the longer wavelength side of measuring wavelength range. Model Input Range F-4500 200 to 900 nm F-2500 220 to 800 nm (g) EM Start WL This is the start wavelength for wavelength scan on the emission side. Input the shorter wavelength side of measuring wavelength range. Model Input Range F-4500 200 to 890 nm F-2500 220 to 790 nm (h) EM End WL This is the end wavelength for wavelength scan on the emission side. Input the longer wavelength side of measuring wavelength range. Model Input Range F-4500 210 to 900 nm F-2500 230 to 800 nm 2.2 Creating an Analysis Method
2 - 7 (i) Scan speed Set a wavelength scan speed. (Unit: nm/min) F-4500 F-2500 30000 3000 12000 1500 2400 300 1200 60 300 15 60 15 For the Phosphorescence data mode, select one of the following values. 240 nm/min 60 nm/min 15 nm/min (j) Delay After pressing the Measure button, measurement is started following the delay time set here. It is used for temperature stabilization, etc. In repeat measurement, it is the time until the start of the first measurement. The delay time is invalid for the second and subsequent measurements. Input range: 0 to 9999 s (k) EX Slit Select a slit width for the excitation side. Unit: nm F-4500 F-2500 1.0 2.5 2.5 5.0 5.0 10.0 10.0 20.0 20.0 2.2