Home > Hitachi > Spectrophotometer > Hitachi F 2500 Manual

Hitachi F 2500 Manual

    Download as PDF Print this page Share this page

    Have a look at the manual Hitachi F 2500 Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 913 Hitachi manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    							1 - 42 
    1.11  Photometer Menu Commands  
     
    The following commands are available in the Spectrophotometer 
    menu.  
    • Set Wavelength   
    • Zero Adjust   
    • Auto Pre-scan   
    • Shutter Open/Close   
     
    (1) Set Wavelength  
     
    This command sets the excitation/emission wavelength to a 
    specified wavelength.   
    Click the Set Wavelength command or button, and the 
    dialog box shown below will come up.   
     
     
     
     
    (2) Auto Pre-scan  
     
    This command automatically searches for an optimum 
    excitation/emission wavelength for the sample under 
    measurement.    Other peaks than the fluorescence peak 
    (such as peaks of excitation scattered light and Raman 
    spectrum) are ignored.    On completion of automatic pre-
    scanning, the optimum excitation/emission wavelength is 
    indicated on the monitor window (in the current wavelength 
    indication field detailed in 1.6).     
     
    (3) Zero Adjust  
     
    This command automatically closes the shutter and 
    measures a zero point in detection.    When the zero point 
    is changed, e.g., after the photomultiplier voltage is 
    changed, automatic zero-point adjustment is carried out.   
     
    (4) Shutter Open/Close  
     
    This command is available for opening/closing the 
    excitation beam shutter.    The current state is indicated on 
    the right side of the monitor window.     
    1.11  Photometer Menu Commands  
    						
    							1 - 43 
    1.12  Utility Menu Commands  
     
    The Utility menu contains the following commands: 
    • Instrumental Response   
    • Analog Output Gain 
    • EX Power Monitor   
    • Sens. Check   
    • Lamp Usage   
    • Lamp Off (only with Model F-2500)   
     
    (1) Instrumental Response  
     
    When you use this function.    You need the Spectral 
    correction accessories kit (P/N 251-0126).       
     
    (2)  Analog Output Gain   
     
    Use this command for setting a data value corresponding to 
    1 V of analog output.    For details, refer to 8.4.   
     
    (3)  EX Power Monitor   
     
    Use this command for checking output of the monitor 
    detector.    When the EX Power Monitor is turned on 
    (clicked), monitor value data is displayed regardless of 
    measurement condition.   
     
    (4) Sens. Check  
     
    Use this command for checking sensitivity.    You can check 
    sensitivity/drift with this command.    For details, refer to the 
    maintenance edition.   
     
    (5) Lamp Usage  
     
    Use this command for checking/resetting the usage period 
    of the xenon lamp.    A cumulative turn-on time of the xenon 
    lamp is indicated.    After lamp replacement, reset this 
    indication.    For details, refer to the maintenance edition.   
     
    (6)  Lamp Off (only with Model F-2500)   
     
    Use this command for turning off the xenon lamp.   
    For turning on the xenon lamp again, restarting the 
    spectrophotometer is necessary.     
    1.12  
    						
    							1 - 44 
    1.13    Termination of FL Solutions Program   
     
    After completion of measurement, shut down the instrument in 
    the following procedure.   
     
    (1)  From the File (F) menu, select the Exit (X) command.   
     
     
     
    Fig. 1-42 
     
     
    Select Yes, and the FL Solutions program will then be 
    terminated.   
     
    (2)  Turn off the power switch of the spectrophotometer main 
    unit.  
     
    (3)  Click the Start button of the Windows 98 and select the 
    Shut Down (U) item.    In the Shut Down Windows dialog 
    box, select “Shut down” and click OK.  
     
    (4)  Turn off power to the personal computer and the CRT 
    display monitor.   
    (At step (3), the personal computer may be turned off in 
    some settings.)   
     
    (5)  Turn off power to the printer.   
      
    1.13    Termination of FL Solutions Program  
    						
    							2 - 1 
    2.  WAVELENGTH SCAN  
     
     
    2.1  Flow of Operation  
     
    This section explains the condition setting and operating method 
    for wavelength scan.     
    Measurement is made in the following procedure.   
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
    Fig. 2-1    Operational Flow (wavelength scan)   
    End 
    Start 
    Method setting 
    Click  button.  
     Scan range 
     Data mode 
     Scan speed     
     
    Sample name/comment input
    Click  button.  
    Pre-scan 
    Measurement 
    Data processing 
     Spectrum tracing 
     Peak detection, etc.   
    Print  (See page 2-2.) 
    (See page 2-15.) 
     
    (See page 2-16.)   
     Click 
     button.  
    (See page 2-16.)   
     Click 
     button.  
    (Refer to Section 6.) 
    Click 
     button.  
    (See page 2-19.)     
     
     
     
    2.1  
    						
    							2 - 2 
    2.2    Creating an Analysis Method   
     
    Set the analytical conditions (method) prior to measurement.   
    From the Edit menu, select the Method command or click the 
            button.    The dialog box shown in Fig. 2-2 will appear.   
     
     
     
    Fig. 2-2  Analysis Method 
     
     
    The following tabs are provided on the Analysis Method window 
    for the wavelength scan mode.    These tabs are explained on 
    the subsequent pages.     
     General 
     Instrument 
     Monitor 
     Processing 
     Report  
    2.2    Creating an Analysis Method  
    						
    							2 - 3 
    (1) General  
     
    The parameters shown in Fig. 2-3 are provided under the 
    General tab.   
     
     
     
    Fig. 2-3  General Tab 
     
     
    (a) Measurement (mode)  
    Select a measurement mode from among the four 
    listed below.   
     Wavelength scan  
     Time scan  
     Photometry  
      3-D Scan (displayed when the ‘3-D Scan program’ 
    (option) is installed in combination with Model   
    F-2500)  
     
    Here, select “Wavelength scan.”    For Photometry, 
    Time scan and 3-D Scan, refer to Sections 3, 4 and 5.   
     
    (b) Operator  
    Enter an operator name.     
     
    (c) Instrument  
    The model of the connected instrument is indicated.   
     
    (d)  Use sample table (Set measurement sample) 
    Apply a check mark here when a sample table is used.     
    2.2  
    						
    							2 - 4 
    (e) Sampling  
     Standard  
    When sample is exchanged manually.   
     
    (f) Comments  
    Enter a description or notes on measuring conditions.   
     
    (g) Load  
    Click this button to read out saved analytical conditions.   
    The file opening window appears upon clicking, so 
    select an analytical conditions file.   
     
    (h) Save  
    By clicking this button after setting the parameters for 
    all five tabs, the analytical conditions are saved.   
    Clicking this button overwrites parameters on the 
    existing ones.   
     
    (i) Save As  
    Click this button for renaming the set conditions for 
    saving.    The “Save As” window will appear, so enter a 
    file name.   
     
    (2) Instrument  
     
    When you click the Instrument tab, the window in Fig. 2-4 
    appears.  
     
     
     
    Fig. 2-4  Instrument Tab 
    2.2    Creating an Analysis Method  
    						
    							2 - 5 
    (a) Scan mode  
    Select one of the following scan modes.    In case of 
    Synchronous, excitation wavelength appears on the 
    abscissa of the graph.     
     Excitation  
     Emission  
     Synchronous  
     
    (b) Data mode  
    Select one of the following data modes.   
    The Luminescence mode is for measuring 
    luminescence in the biological/chemical field and only 
    the sample side signal is acquired without performing 
    ratio photometry.    The Phosphorescence mode is for 
    acquiring only phosphorescence after extinction of 
    excitation beam while turning a chopper on the 
    excitation side.   
     Fluorescence  
     Luminescence  
      Phosphorescence (with Model F-4500)   
     
    (c) EM WL  
    This is the start wavelength for wavelength scan on the 
    excitation side.    Input the shorter wavelength side of 
    measuring wavelength range.   
     
    Model Input Range   
    F-4500 200 to 900 nm or 200 to 890 nm in 
    Synchronous scan mode   
    F-2500 220 to 800 nm or 220 to 790 nm in 
    Synchronous scan mode   
     
     
    (d)  EX Start WL   
    This is the start wavelength for wavelength scan on the 
    excitation side.    Input the shorter wavelength side of 
    measuring wavelength range.   
     
    Model Input Range  
    F-4500  200 to 890 nm   
    F-2500  220 to 790 nm   
     
    2.2  
    						
    							2 - 6 
    (e)  EX End WL   
    This is the end wavelength for wavelength scan on the 
    excitation side.    Input the longer wavelength side of 
    measuring wavelength range.   
     
    Model Input Range  
    F-4500  210 to 900 nm   
    F-2500  230 to 800 nm   
     
     
    (f) EX WL  
    An excitation wavelength can be input.    It is the end 
    wavelength for wavelength scan on the excitation side.   
    Input the longer wavelength side of measuring 
    wavelength range.   
     
    Model Input Range  
    F-4500  200 to 900 nm   
    F-2500  220 to 800 nm   
     
     
    (g)  EM Start WL   
    This is the start wavelength for wavelength scan on the 
    emission side.    Input the shorter wavelength side of 
    measuring wavelength range.     
     
    Model Input Range  
    F-4500  200 to 890 nm   
    F-2500  220 to 790 nm   
     
     
    (h)  EM End WL   
    This is the end wavelength for wavelength scan on the 
    emission side.    Input the longer wavelength side of 
    measuring wavelength range.     
     
    Model Input Range  
    F-4500  210 to 900 nm   
    F-2500  230 to 800 nm   
     
    2.2    Creating an Analysis Method  
    						
    							2 - 7 
    (i) Scan speed  
    Set a wavelength scan speed.  (Unit:  nm/min)  
     
    F-4500 F-2500  
    30000 3000 
    12000 1500 
    2400 300 
    1200 60 
    300 15 
    60  
    15  
     
     
    For the Phosphorescence data mode, select one of the 
    following values.   
     240 nm/min  
     60 nm/min 
     15 nm/min 
     
    (j) Delay  
    After pressing the Measure button, measurement is 
    started following the delay time set here.    It is used for 
    temperature stabilization, etc.    In repeat measurement, 
    it is the time until the start of the first measurement.   
    The delay time is invalid for the second and 
    subsequent measurements.   
    Input range:    0 to 9999 s   
     
    (k) EX Slit  
    Select a slit width for the excitation side.    Unit:    nm   
     
    F-4500 F-2500  
    1.0 2.5 
    2.5 5.0 
    5.0 10.0 
    10.0 20.0 
    20.0  
     
     
    2.2  
    						
    All Hitachi manuals Comments (0)

    Related Manuals for Hitachi F 2500 Manual