Home > Hitachi > Spectrophotometer > Hitachi F 2500 Manual

Hitachi F 2500 Manual

    Download as PDF Print this page Share this page

    Have a look at the manual Hitachi F 2500 Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 913 Hitachi manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    							8 - 4 
    For remeasuring any of the standards after their measurements, 
    click its data as shown in Fig. 8-6.    Then click the Remeasure 
    button.  
     
     
     
     
    Fig. 8-6 
     
     
    A message box will open as in Fig. 8-7, so press the sipping 
    lever.  
     
     
     
    Fig. 8-7 
     
     
    When a standard is remeasured, the calibration curve is 
    recalculated and sample data is also recalculated based on the 
    curve and then displayed.    To saved the recalculated data, click 
    the End button or the          (save) button on the toolbar.     
     
    A message box is displayed in both standard and sample 
    measurement sessions.    So operate according to the messages.   
    When the message box appears, press the sipping lever of the 
    sample sipper.   
     
    NOTE:  When the sample sipper is used, blank measurement 
    cannot be made.  8.1.2  When Using a 
    Sample Table 
    8.1    Use of Sample Sipper  
    						
    							8 - 5 
    Select the Measure command from the Spectrophotometer menu 
    or click the          (measurement) button on the toolbar, and the 
    window in Fig. 8-8 will open.   
     
     
     
    Fig. 8-8 
     
     
    Press the sipping lever and measurement starts.     
    For measurement of the 2nd and subsequent standards, a 
    message box as in Fig. 8-9 will open.    For remeasurement or 
    for skipping measurement of the next standard, select the 
    relevant item with the mouse (put a bullet in its radio button) and 
    then click Yes.  When clicking No, the current measurement 
    sequence will be terminated.     
    For restarting measurement, click the     button.  
    Measurement then starts from the 1st standard.   
     
     
     
    Fig. 8-9 
     
     
    8.1.2  
    						
    							8 - 6 
    On completion of standard measurement, the window in Fig.   
    8-10 will open.   
     
     
     
    Fig. 8-10 
     
     
    By clicking OK with the mouse, a calibration curve is drawn and 
    then samples can be measured.     
    By clicking Cancel, the measurement is terminated, and the 
    standards’ data is not saved.     
    For remeasurement of the final standard, select “Repeat the last 
    measurement” and click OK.    Set the final standard in place and 
    press the sipping lever.    When repeat measurement has been 
    set, measurement is started from the 1st one (Repl.No.1) for that 
    standard. 
     
    In the case of sample measurement also, a message box as in 
    Fig. 8-11 will appear.    Press the sipping lever, and 
    measurement of the 1st sample starts.     
     
     
     
    Fig. 8-11 
     
    8.1    Use of Sample Sipper  
    						
    							8 - 7 
    For measurement of the 2nd sample, a message box as in   
    Fig. 8-12 opens.   
     
     
     
    Fig. 8-12 
     
     
    Press the sipping lever for measuring the 2nd sample.     
    For skipping this sample (and measuring the 3rd sample), select 
    “Skip this sample?” and click the Yes button with the mouse.   
    For remeasurement, select “Repeat the last measurement?” and 
    click Yes button with the mouse.    When repeat measurement 
    has been set, measurement is started from the 1st one 
    (Repl.No.1) for the 1st sample.   
    In case of interrupt measurement, select “Insert a new sample 
    into the sequence” and click the Yes button with the mouse.     
    A message box as in Fig. 8-13 opens, so click the OK button.   
     
     
     
    Fig. 8-13 
     
     
    8.1.2  
    						
    							8 - 8 
    For remeasurement after a series of measurements, click the 
    data to be remeasured as in Fig. 8-14.    Then click the 
    Remeasure button.   
     
     
    Fig. 8-14 
     
     
    A message box will open as in Fig. 8-15.    Press the sipping 
    lever, and measurement starts.    When remeasuring a standard, 
    the calibration curve is recalculated and sample data is also 
    recalculated based on the curve and then displayed.    To save 
    the recalculated data, click the     (Save) button. 
     
     
    Fig. 8-15 
     
     
    (1)  When not Using Sample Table   
     
    Set your analysis method, and when the monitor window 
    appears, press the sipping lever.    Measurement starts.     
     
    (2)  When Using Sample Table   
     
    Select the Measure command from the Spectrophotometer 
    menu or click the     (measurement) button on the toolbar.  
    A message box opens.    Follow the messages given.    8.1.3 In Wavelength 
    Scan or Time 
    Scan Mode   
    8.1    Use of Sample Sipper  
    						
    							8 - 9 
    8.2    Use of AS-3000 Intelligent Autosampler   
     
    Set the rack parameters, injection volume and other items with 
    the AS-3000 intelligent autosampler.    For details, refer to the 
    instruction manual attached to the AS-3000.     
     
    Select the Measure command from the Spectrophotometer menu 
    or click the     (measurement) button on the toolbar.   
    Then, press the START key on the AS-3000 intelligent 
    autosampler.  Measurement will start.  
     
    NOTE:  In measurement without using a sample table, 
    remeasurement, interrupt measurement and blank 
    measurement cannot be done. 
     
    NOTE:  In measurement when using a sample table, 
    remeasurement, interrupt measurement and blank 
    measurement cannot be done.    But remeasurement of 
    standards is possible.    The standard to be 
    remeasured should be set in the initial test tube 
    position.  
     
    For measurement after completion of settings: 
     
    (1)  Select the Measure command from the Spectrophotometer 
    menu in the FL Solutions program or click the 
    (measurement) button on the toolbar.     
    The window in Fig. 8-16 opens.   
     
     
    Fig. 8-16    For Start from Calibration Curve Creation 
     
     
    (2) Press the START key on the AS-3000.   
     
    (3)  Measurement is started.    8.2.1  When Not Using   
    a Sample Table   
    8.2.2  When Using a 
    Sample Table   
     
    8.2.1  
    						
    							8 - 10 
    8.3  For Analog Output  
     
    For connection, use the analog output terminal.    Select the 
    Analog Output Gain command from the Utility menu.     
     
     
     
    Fig. 8-17 
     
     
    This is a setting of data value corresponding to analog output of 
    1 V.   
    Analog output value (V) becomes as follows.   
     
     Model F-4500 
    Low :  (Data value - photometric value lower limit)/1024   
    High :  (Data value - photometric value lower limit)/8192   
     
      Model F-2500   
    Low :  (Data value)/1024   
    High :  (Data  value)/8192   
     
    8.3  For Analog Output  
    						
    							 A -
    1
    APPENDIX A    DETAILS ON QUANTITATION 
     
     
    A.1 Foreword 
    In the Photometry mode of the F-2500 spectrophotometer, there 
    are four calibration types. 
     
      Linear working curve 
      Quadratic working curve 
      Cubic working curve 
      Multiple segment working curve 
     
    These are explained in more detail below. 
     
    A.2  Linear Working Curve (1st order) 
     
    Regression line is determined via least squares method from a 
    maximum of 20 data.    The calculation formula is as follows : 
    x = A1· y + A0 
    A1 = 
    ()
    yx
    nyx
    y
    nyii i i
    ii−
    ∑⋅
    ∑ ∑
    −
    ∑ ∑1
    1
    22, A0 = x
    nAy
    nii∑
    −⋅∑
    1 
    where,  x :    Concentration of each sample (input value) 
    y :    Data on each sample (measured value) 
    n :    Number of samples 
     
    A.3  Quadratic Working Curve (2nd order) 
     
    A quadratic curve is determined via least squares method from a 
    maximum of 20 data.    The calculation formula is as follows : 
    x = A2 · y
    2 + A1 · y + A0 
    A2 = 
    ()() ()()
    ()()()}
    Sy xSyy SyxSyy
    SyySy y Syy22
    22 22−
    −
    
     S(yy) = ()yy
    nii
    22∑−∑ 
    A1 = 
    ()()()()
    ()()()}
    SyxSy y Sy xSyy
    SyySy y Syy22 2 2
    22 22−
    −
    
     S(yx) = yxyx
    niiii∑−⋅
    ∑ ∑ 
    A0 = 
    x
    nAx
    nAy
    nii i∑
    −∑
    −∑
    122
     S(yy2) = yyy
    niii 32∑−⋅
    ∑ ∑ 
     S(y
    2x) = yxyx
    niiii 22∑−⋅
    ∑ ∑ 
     S(y
    2y2) = ()yy
    nii
    422∑−∑  
    						
    							 A -
    2
    where,  x :    Concentration of each sample (input value) 
    y :    Data on each sample (measured value) 
    n :    Number of samples 
     
    A.4  Cubic Working Curve (3rd order) 
     
    A cubic curve is determined via least squares method from a 
    maximum of 20 data.    The calculation formula is as follows : 
    X = A3 · Y
    3 + A2 · y2 + A1 · y + A0 
    where,  x :    Concentration of each sample (input value) 
    y :    Data on each sample (measured value) 
    n :    Number of samples 
     
    A.5  Multiple Segment Working Curve (Segmented) 
     
    The calibration curve is apt to bend when measuring turbid 
    samples or the like.    Use of a quantification program allows 
    correcting the calibration curve by using up to 20 standards.   
    Figure A-1 shows an example of curve correction.    For the part 
    which exceeds the measuring range of the standards, simply 
    extend the line as it is. 
     
     
    Fig. A-1    Correction of Curve Bending 
     
    STD5
    STD4
    STD1
    STD6 
    STD3
    STD2
    CONC
    Data  
    						
    							 A -
    3
     
     
    (1)  A correct calibration curve can be created only when the 
    measured value increases or decreases monotonically 
    versus the concentration value.      Especially when the 
    inclination is negative-going, be sure to measure a blank 
    having a data value larger than the other standards.     
    A curve not showing a monotonic increase will appear as in 
    Fig. A-2.    And when the data value of the blank is small, 
    regardless of a negative-going inclination, the curve will 
    appear as in Fig. A-3. 
     
     
    Fig. A-2    Example of Curve without Monotonic Increase 
    (unsuitable curve) 
     
     
    STD5
    CONC 
    Data
    STD4
    STD1
    STD2
    STD3
    STD6
    Data
    STD1
    CONC
    STD2
    STD3 
    STD4 
    STD5
    Fig. A-3  Example of Negative-going Curve where 
    Measured Value of Blank is Small   
    (unsuitable curve)  
    						
    All Hitachi manuals Comments (0)

    Related Manuals for Hitachi F 2500 Manual