Hitachi F 2500 Manual
Have a look at the manual Hitachi F 2500 Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 913 Hitachi manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.
8 - 4 For remeasuring any of the standards after their measurements, click its data as shown in Fig. 8-6. Then click the Remeasure button. Fig. 8-6 A message box will open as in Fig. 8-7, so press the sipping lever. Fig. 8-7 When a standard is remeasured, the calibration curve is recalculated and sample data is also recalculated based on the curve and then displayed. To saved the recalculated data, click the End button or the (save) button on the toolbar. A message box is displayed in both standard and sample measurement sessions. So operate according to the messages. When the message box appears, press the sipping lever of the sample sipper. NOTE: When the sample sipper is used, blank measurement cannot be made. 8.1.2 When Using a Sample Table 8.1 Use of Sample Sipper
8 - 5 Select the Measure command from the Spectrophotometer menu or click the (measurement) button on the toolbar, and the window in Fig. 8-8 will open. Fig. 8-8 Press the sipping lever and measurement starts. For measurement of the 2nd and subsequent standards, a message box as in Fig. 8-9 will open. For remeasurement or for skipping measurement of the next standard, select the relevant item with the mouse (put a bullet in its radio button) and then click Yes. When clicking No, the current measurement sequence will be terminated. For restarting measurement, click the button. Measurement then starts from the 1st standard. Fig. 8-9 8.1.2
8 - 6 On completion of standard measurement, the window in Fig. 8-10 will open. Fig. 8-10 By clicking OK with the mouse, a calibration curve is drawn and then samples can be measured. By clicking Cancel, the measurement is terminated, and the standards’ data is not saved. For remeasurement of the final standard, select “Repeat the last measurement” and click OK. Set the final standard in place and press the sipping lever. When repeat measurement has been set, measurement is started from the 1st one (Repl.No.1) for that standard. In the case of sample measurement also, a message box as in Fig. 8-11 will appear. Press the sipping lever, and measurement of the 1st sample starts. Fig. 8-11 8.1 Use of Sample Sipper
8 - 7 For measurement of the 2nd sample, a message box as in Fig. 8-12 opens. Fig. 8-12 Press the sipping lever for measuring the 2nd sample. For skipping this sample (and measuring the 3rd sample), select “Skip this sample?” and click the Yes button with the mouse. For remeasurement, select “Repeat the last measurement?” and click Yes button with the mouse. When repeat measurement has been set, measurement is started from the 1st one (Repl.No.1) for the 1st sample. In case of interrupt measurement, select “Insert a new sample into the sequence” and click the Yes button with the mouse. A message box as in Fig. 8-13 opens, so click the OK button. Fig. 8-13 8.1.2
8 - 8 For remeasurement after a series of measurements, click the data to be remeasured as in Fig. 8-14. Then click the Remeasure button. Fig. 8-14 A message box will open as in Fig. 8-15. Press the sipping lever, and measurement starts. When remeasuring a standard, the calibration curve is recalculated and sample data is also recalculated based on the curve and then displayed. To save the recalculated data, click the (Save) button. Fig. 8-15 (1) When not Using Sample Table Set your analysis method, and when the monitor window appears, press the sipping lever. Measurement starts. (2) When Using Sample Table Select the Measure command from the Spectrophotometer menu or click the (measurement) button on the toolbar. A message box opens. Follow the messages given. 8.1.3 In Wavelength Scan or Time Scan Mode 8.1 Use of Sample Sipper
8 - 9 8.2 Use of AS-3000 Intelligent Autosampler Set the rack parameters, injection volume and other items with the AS-3000 intelligent autosampler. For details, refer to the instruction manual attached to the AS-3000. Select the Measure command from the Spectrophotometer menu or click the (measurement) button on the toolbar. Then, press the START key on the AS-3000 intelligent autosampler. Measurement will start. NOTE: In measurement without using a sample table, remeasurement, interrupt measurement and blank measurement cannot be done. NOTE: In measurement when using a sample table, remeasurement, interrupt measurement and blank measurement cannot be done. But remeasurement of standards is possible. The standard to be remeasured should be set in the initial test tube position. For measurement after completion of settings: (1) Select the Measure command from the Spectrophotometer menu in the FL Solutions program or click the (measurement) button on the toolbar. The window in Fig. 8-16 opens. Fig. 8-16 For Start from Calibration Curve Creation (2) Press the START key on the AS-3000. (3) Measurement is started. 8.2.1 When Not Using a Sample Table 8.2.2 When Using a Sample Table 8.2.1
8 - 10 8.3 For Analog Output For connection, use the analog output terminal. Select the Analog Output Gain command from the Utility menu. Fig. 8-17 This is a setting of data value corresponding to analog output of 1 V. Analog output value (V) becomes as follows. Model F-4500 Low : (Data value - photometric value lower limit)/1024 High : (Data value - photometric value lower limit)/8192 Model F-2500 Low : (Data value)/1024 High : (Data value)/8192 8.3 For Analog Output
A - 1 APPENDIX A DETAILS ON QUANTITATION A.1 Foreword In the Photometry mode of the F-2500 spectrophotometer, there are four calibration types. Linear working curve Quadratic working curve Cubic working curve Multiple segment working curve These are explained in more detail below. A.2 Linear Working Curve (1st order) Regression line is determined via least squares method from a maximum of 20 data. The calculation formula is as follows : x = A1· y + A0 A1 = () yx nyx y nyii i i ii− ∑⋅ ∑ ∑ − ∑ ∑1 1 22, A0 = x nAy nii∑ −⋅∑ 1 where, x : Concentration of each sample (input value) y : Data on each sample (measured value) n : Number of samples A.3 Quadratic Working Curve (2nd order) A quadratic curve is determined via least squares method from a maximum of 20 data. The calculation formula is as follows : x = A2 · y 2 + A1 · y + A0 A2 = ()() ()() ()()()} Sy xSyy SyxSyy SyySy y Syy22 22 22− − S(yy) = ()yy nii 22∑−∑ A1 = ()()()() ()()()} SyxSy y Sy xSyy SyySy y Syy22 2 2 22 22− − S(yx) = yxyx niiii∑−⋅ ∑ ∑ A0 = x nAx nAy nii i∑ −∑ −∑ 122 S(yy2) = yyy niii 32∑−⋅ ∑ ∑ S(y 2x) = yxyx niiii 22∑−⋅ ∑ ∑ S(y 2y2) = ()yy nii 422∑−∑
A - 2 where, x : Concentration of each sample (input value) y : Data on each sample (measured value) n : Number of samples A.4 Cubic Working Curve (3rd order) A cubic curve is determined via least squares method from a maximum of 20 data. The calculation formula is as follows : X = A3 · Y 3 + A2 · y2 + A1 · y + A0 where, x : Concentration of each sample (input value) y : Data on each sample (measured value) n : Number of samples A.5 Multiple Segment Working Curve (Segmented) The calibration curve is apt to bend when measuring turbid samples or the like. Use of a quantification program allows correcting the calibration curve by using up to 20 standards. Figure A-1 shows an example of curve correction. For the part which exceeds the measuring range of the standards, simply extend the line as it is. Fig. A-1 Correction of Curve Bending STD5 STD4 STD1 STD6 STD3 STD2 CONC Data
A - 3 (1) A correct calibration curve can be created only when the measured value increases or decreases monotonically versus the concentration value. Especially when the inclination is negative-going, be sure to measure a blank having a data value larger than the other standards. A curve not showing a monotonic increase will appear as in Fig. A-2. And when the data value of the blank is small, regardless of a negative-going inclination, the curve will appear as in Fig. A-3. Fig. A-2 Example of Curve without Monotonic Increase (unsuitable curve) STD5 CONC Data STD4 STD1 STD2 STD3 STD6 Data STD1 CONC STD2 STD3 STD4 STD5 Fig. A-3 Example of Negative-going Curve where Measured Value of Blank is Small (unsuitable curve)