Home > Hitachi > Spectrophotometer > Hitachi F 2500 Manual

Hitachi F 2500 Manual

    Download as PDF Print this page Share this page

    Have a look at the manual Hitachi F 2500 Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 913 Hitachi manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    							4 - 16 
    (f) PMT Voltage  
    A function for controlling the voltage of the 
    photomultiplier detector (photomultiplier tube).     
    Select one of the following values.    Changing from 
    400 V to 700 V increases the data value about two 
    digits and changing from 700 V to 950 V increases it 
    about 1 digit.     
    •  950 V (only with F-4500)   
    • 700 V  
    • 400 V 
     
    (g)  Auto statistic cal. number   
    Calculation is made for mean value, SD (standard 
    deviation) and CV (coefficient of variation).    Set the 
    number of samples for this purpose.     
    Input range:    2 to 4000   
     
    (h) Replicates  
    Set the number of measurements.     
    Input range:    1 to 20   
     
    (i) Integration time  
    A function for obtaining data averaged over the 
    specified time for the purpose of acquiring stabilized 
    data.   
    F-4500:    0.1 to 10.0 s   
    F-2500:    1, 2, 5, 10 s 
     
    (j) Delay  
    After pressing the Measure button, measurement is 
    started following the delay time set here.    It is used for 
    temperature stabilization, etc.    In repeat measurement, 
    it is the time until the start of the first measurement.   
    The delay time is invalid for the second and 
    subsequent measurements.   
    Input range:    0 to 9999 s   
     
    4.2    Creating an Analysis Method  
    						
    							4 - 17 
    (4) Standards  
     
    Clicking the Standards tab opens the window shown in Fig. 
    4-16.   
     
     
     
    Fig. 4-16  Standards Tab 
     
     
    (a) Sample table  
    Sample table gives a list of standards for sample 
    measurement or calibration curve preparation.     
    This table contains the items listed below.     
     
    NOTE:  The size of the sample table columns (cells) 
    is changeable.    Bring the cursor to a 
    boundary line between left and right table 
    headers, and it changes into the one shown 
    below.   
     
     
    Now click the mouse button on the line and 
    drag it to the left or right to obtain a suitable 
    column size. 
     
    1) No.  
    This is the number given to each sample.   
    2) Sample  
    Enter a sample name in up to 20 characters.   
    3) Comments  
    Enter comments on each sample.   
    4) Concentration  
    Enter a sample concentration.   
    4.2  
    						
    							4 - 18 
    (b) Update  
    By clicking this button, sample numbers are set by the 
    entered number of samples.    The displayed sample 
    names, comments, etc. are all cleared.   
     
    (c) Insert  
    When the initial screen is opened, the Insert button 
    becomes active.    Click this button to insert data at the 
    end of the sample list.   
     
    (d) Delete  
    Click the column of the sample No. to be deleted, and 
    it becomes active.    Now click the Delete button and 
    the item is deleted.     
     
    (5) Monitor  
     
    When selecting the Monitor tab, the window in Fig. 4-17 
    appears.   
     
     
     
    Fig. 4-17  Monitor Tab 
     
     
    Assign the following parameters.   
     
    (a)  Y-Axis Max (maximum)   
    Input a maximum value for Y-axis on the monitor 
    window.  
     
    (b)  Y-Axis Min (minimum)   
    Input a minimum value for Y-axis on the monitor 
    window.  
    4.2    Creating an Analysis Method  
    						
    							4 - 19 
    (c)  Open data processing window after data acquisition   
    Select whether or not to conduct data processing after 
    sample measurement.    When selected (check mark is 
    applied), the data is saved automatically after 
    measurement and its file is displayed as an icon on the 
    window.    And by opening this icon, data processing 
    such as peak detection is available.    When this 
    function is not selected, data processing can be 
    achieved by reading out the file via File-Open.  
     
    (d)  Print report after data acquisition   
    Select whether or not to print a report after sample 
    measurement.    When selected (check mark is 
    applied), printing will be made automatically after 
    measurement.    The items set under the Report tab 
    will be printed.   
     
    (6) Report  
     
    Select the Report tab, and the window in Fig. 4-18 will 
    appear.  
     
     
     
    Fig. 4-18  Report Tab 
     
     
    On this window, select the items to be contained in an analysis 
    report.   
     
    4.2  
    						
    							4 - 20 
    (7) Output  
     
    (a) Print Report  
    A report will be printed.    For printing method, refer to 
    4.5.  
     
    (b)  Use Microsoft® Excel   
    Data will be transferred to Microsoft® Excel.     
    For transfer method, refer to 6.13.   
     
    (c)  Use print generator 
    Select this method for using the report generator 
    (option).    For usage, refer to the instruction manual of 
    the report generator.   
     
       
    • Include date  
    • Include method  
    •  Include calibration curve   
    •  Include standards data   
    •  Include calibration data   
    •  Include sample data   
     
    (8) Printer Font  
     
    Click this button for changing the font to be used in the 
    report.  
    4.2    Creating an Analysis Method  
    						
    							4 - 21 
    4.3  Defining Your Samples  
     
    When “Use Sample Table” has been set for the Analysis Method, 
    the samples must be defined before measurement.   
     
    (1)  Open the monitor window and select the Sample table 
    command in the Edit menu or click the          button.   
     
    (2)  When the Sample Table window in Fig. 4-19 appears, fill in 
    the blank section.    For instance, if there are five samples, 
    click the Insert button five times.    Or enter “5” for Number 
    of Samples and click the Update button.   
     
     
     
    Fig. 4-19    Sample Table Window 
     
     
    (3)  Input the sample name of each sample and comments on it.   
    Since sample names cannot be changed after 
    measurement, input them here.   
    Maximum number of characters to be input   
    Sample : 20 characters  
    Comments :  255 characters   
     
    (4)  Input the name of the data file after measurement.     
     
    (5)  When input is finished, check the entries and then click the 
    OK button.   
    4.3  
    						
    							4 - 22 
    4.4  Conducting Measurement  
     
    First, create a working curve, and then proceed to sample 
    measurement.   
     
    (1)  Set the analysis method, and the window in Fig. 4-20 will 
    appear provided a sample table is not used.   
     
     
     
    Fig. 4-20 
     
     
    When a sample table is used, the window in Fig. 4-21 will 
    appear.  
     
     
     
    Fig. 4-21 
    4.4  Conducting Measurement  
    						
    							4 - 23 
    (2)  There are two methods of starting measurement.     
     
    (a)  Select the Measure command from the 
    Spectrophotometer menu.   
     
    (b)  Click the     (measurement) button on the 
    measurement toolbar.   
     
    (3)  Standards will be measured.    When finished, the window 
    in Fig. 4-22 will appear.    So click the OK button. 
     
     
     
    Fig. 4-22 
     
     
    (4)  Now measure the samples.    If a sample table is not used, 
    click the Sample button.    To interrupt the measurement for 
    other samples, click the Interrupt button.   
    The window in Fig. 4-23 will appear, so enter a sample No.   
     
     
     
    Fig. 4-23 
     
     
    Click the OK button and measurement starts.     
    When a sample table is used, a message box is displayed, 
    so conduct operation according to the messages.   
     
    (5)  When sample measurement is finished and a sample table 
    is not used, click the End button.    From the File menu, 
    select the Save As command and save the data.     
    When a sample table is used, the data will be saved 
    automatically.  
    4.4  
    						
    							4 - 24 
    (6)  For conducting remeasurement after completing a series of 
    measurements, click the data area to be remeasured as 
    shown in Fig. 4-24 and then click the Remeasure button.   
    For saving data after remeasurement, click the End button 
    or     button.  The procedure is the same when a sample 
    table is used.    Both standards and samples can be 
    remeasured.   
     
     
     
    Fig. 4-24 
     
     
    (7)  For conducting a blank measurement, click the Blank 
    button.  
     
    (8)  Data is saved on completion of measurement.    When its 
    file is opened, a data processing window shown in Fig. 4-25 
    appears.   
     
     
     
     
     
    Fig. 4-25    Data Processing Window (photometry) 
    (a) Standard  data   
    (b) Calibration  curve    (d) Calibration curve data   
    (c) Sample data   
    4.4  Conducting Measurement  
    						
    							4 - 25 
    (a) Standard data  
    The measurement results of standards used for 
    preparing a calibration curve are displayed.   
     
     
     
    Fig. 4-25-1  Standard Data 
     
     
    (b) Calibration curve  
    A calibration curve obtained from standard data is 
    displayed.    The curve can also be traced.   
    Concentration and photometric value are indicated at 
    the top left of this field.   
     
    (c) Sample data  
    Sample data such as photometric value (Abs), 
    concentration (Conc), mean value (Avg Conc), 
    standard deviation (SD), coefficient of variation (CV) 
    and upper/lower limit judgment results (HIGH/LOW) 
    are displayed.   
     
     
     
    Fig. 4-25-2  Sample Data 
     
     
    4.4  
    						
    All Hitachi manuals Comments (0)