Hitachi F 2500 Manual
Have a look at the manual Hitachi F 2500 Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 913 Hitachi manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.
A - 13 And, eventually the molecule returns to the ground state while emitting fluorescent radiation. Also, if radiationless transition to the triplet state takes place, then phosphorescence is emitted during triplet-to-singlet transition (from the excited triplet state to the ground singlet state). Generally phosphorescence persists for 10 -4 sec or longer due to the selection rule imposed on the triplet-to-singlet transition. In contrast, fluorescence persists for a period of 10 -8 to 10-9 sec in most cases. As mentioned above, part of the radiation absorbed by the substance is lost as vibration energy, etc.; therefore, the fluorescence wavelength emitted from it is longer than the excitation wavelength (Stokes’ law). The ratio of the number of photons emitted during fluorescence to the number of photons absorbed is called the quantum efficiency of fluorescence. The larger the quantum efficiency a substance has, the more fluorescence it emits. Also, the intensity of florescence emitted from a substance is proportional to the quantity of light absorbed by it. When a dilute solution sample is measured, the intensity of fluorescence is expressed by F = KI 0clεφ where, F : Intensity of fluorescence K : Instrumental constant I 0 : Intensity of exciting radiation c : Concentration of substance l : Cell path length (distance in the cell through which the radiation travels) ε : Absorptivity of substance φ : Quantum efficiency of substance
A - 14 E.2 Advantages of Fluorometry As contrasted with fluorometry, absorptiometry for a low-concentration sample is explained in the following : A sample having 99% transmittance to blank is taken as an example. In the absorbance measurement of such a substance, inaccuracies must always be taken into consideration. Here, the inaccuracy is assumed to be 0.1%. Since it has an effect on both the blank and sample, Percent transmittance of blank 100.0 ± 0.1% Percent transmittance of sample 99.0 ± 0.1% Difference (proportional to 1.0 ± 0.2% concentration of sample) In this example, the uncertainty in concentration measurement is ±20%. While in the fluorometry, a difference from zero level corresponds to the concentration of sample. Accordingly, measurement accuracy is as follows : Output signal level at measurement of sample 100 ± 0.1 Value corresponding to blank 0 ± 0.1 Difference (proportional to 100 ± 0.2 concentration of sample) As can be seen from the above, the fluorometry is very advantageous for analyzing a low-concentration sample since its uncertainty is in most cases theoretically independent of the concentration of sample. Although actual practice may involve some error factors which increase as the concentration of sample becomes extremely low, the fluorometry is capable of measuring low concentration with an accuracy 100 times higher than in absorptiometry. Figure E-2 is an explanatory illustration of the foregoing description. In absorptiometry, a difference between the quantity of incident radiation Io and the quantity of transmitted radiation is represented by signal Is. A level at which the signal Is becomes almost equal to a noise level is used as a detection limit. In fluorometry, however, since the quantity of fluorescence I F itself is represented as a signal, just a small amount of fluorescence can be amplified electrically for enabling detection. Still more, since a fluorescence wavelength of a substance is different from its excitation wavelength (incident light wavelength), the fluorescence wavelength is not readily affected by the exciting radiation, thereby contributing to ensuring high sensitivity.
A - 15 Amplified Fig. E-2 Comparison between Absorptiometry and Fluorometry In addition to high sensitivity, fluorometry is advantageous in that more information is attainable. A fluorescence spectrum is also available besides an excitation spectrum which corresponds to an absorption spectrum in absorptiometry. The two kinds of wavelengths can be selected as desired, and a fluorescence spectrum can be recorded using a properly selected excitation wavelength (or vice versa). Thus, quantitative and qualitative analyses can be made for a sample containing plural components. Figure E-3 shows a simplified spectral graph of measurement of a sample containing multiple components. In absorptiometry, since only the absorption spectrum can be measured, two or more component wavelengths are presented. If the absorption wavelengths are similar to each other, each component cannot be separated in measurement. In fluorometry, even if the absorption wavelengths are similar, a difference in fluorescence makes it possible to select each fluorescence wavelength properly. Thus, each component can be separated in measurement. Fig. E-3 Measurement of Mult-Component Sample Low concentrationHigh concentrationLow concentration (b) Fluorometry (a) Absorptiometry High concentration Amplified Absorption spectra Component B (b) Fluorometry Fluorescence spectra (a) Absorptiometry Component A Excitation spectraComponent A Component B
A - 16 Table E-1 compares information attainable in absorptiometry and that in fluorometry. Table E-1 Comparison of Information Attainable in Absorptiometry and Fluorometry Absorptiometry Fluorometry Absorption spectrum only (Corresponding to excitation spectrum in fluorometry) Excitation spectrum Fluorescence spectrum E.3 Notes of Fluorescence Analysis Measurement For most kinds of samples, an increase of 1°C in the temperature of sample causes the intensity of fluorescence to decrease by 1 to 2%. It is also reported that for some kinds of biochemical samples, the intensity of fluorescence decreases as much as 10% as the temperature increases by 1°C. When analyzing a sample having a temperature-dependent property, it is advisable to use the constant-temperature cell holder (P/N 650-0150). Constant-temperature measurement can be carried out by circulating constant-temperature water through this cell holder. Some kinds of samples may be susceptible to a chemical change due to exciting radiation. In analysis of such a sample, keep the shutter closed to cut off an excitation beam until measurement is started, and then open the shutter immediately before measurement. If any chemical change due to exciting radiation is observed still, determine a signal level at the start time through extrapolation according to variation in signal level. E.3.1 Temperature Dependency of Fluorescence Intensity E.3.2 Chemical Change in Sample due to Radiation
A - 17 In fluorescence measurement, spectra having different natures from that of fluorescence may be observed. These are called Rayleigh scattering spectrum and Raman scattering spectra; the former appearing at the same wavelength position as the excitation spectrum, and the latter appearing at the longer-wavelength side near Rayleigh scattering. In a fluorescence spectrum, when the excitation wavelength is shifted, only the peak height is changed while the peak wavelength position remains intact. In a Raman scattering spectrum, when the excitation wavelength is shifted, the peak wavelength position is also changed accordingly. Both the Rayleigh scattering and Raman scattering are caused by a solvent which may be contained in the sample. When examining the spectral plot, be careful not to mistake these scattering effects of the fluorescence peak of interest. Table E-3 presents the Raman spectral peak position at each excitation wavelength for the purpose of reference. Fig. E-4 Raman Spectrum of Water E.3.3 Raman Scattering Relative intensity Excitation wavelength Raman scattering
A - 18 Table E-3 Raman Peak Positions at Respective Excitation Wavelengths (Excitation wavelength) WaterEthanolCyclohexane Carbon Tetrachloride Chloroform 248 271 267 267 313 350 344 344 320 346 365 416 405 408 375 410 405 469 459 458 418 461 Excitation wavelength and Raman peak position (nm) 436 511 500 499 450 502 In measurement of a high-concentration sample, a variety of error factors may be involved. The most significant error factor consists in that an excitation beam is absorbed at the entrance of a cell to prevent a sufficient level of excitation at the center of cell. Figure E-5 illustrates an extreme case of this condition. Although fluorescence is emitted in the vicinity of the entrance for the excitation beam, it is not taken into the emission monochromator. Fig. E-5 Sample Having an Extremely High Concentration E.3.4 Handling of High- Concentration Samples Fluorescence is reflected here. Excitation beam Fluorescence
A - 19 If only the incident point of excitation beam is bright, it is necessary to dilute the sample properly for measurement. The second significant error factor consists in extinction due to concentration. This condition is caused by preventing activation through interaction of molecules. The third significant error factor consists in re-absorption of fluorescence. As shown in Figure E-6, this condition occurs due to overlapping between the shot-wavelength side of fluorescence spectrum and the long-wavelength side of excitation spectrum. Therefore, it seems that the fluorescence spectrum has been shifted toward the long-wavelength side to some extent. In measurement of an ordinary kind of sample, however, this condition will not impede quantitative determination significantly. Fig. E-6 Explanatory Illustration of Re-absorption In any case, if there is a possibility of a measurement error due to high concentration of a sample, dilute the sample properly or carry out surface fluorescence measurement using a solid sample holder. Wavelength Fluorescence spectrum Excitation spectrum Re-absorption occurs here. Relative intensity
A - 20 Where the excitation and emission wavelengths are plotted near each other, care should be exercised not to mistake the Raman and Rayleigh scattering for the fluorescence spectrum as mentioned in E.3.3. Where the excitation and emission wavelengths are plotted apart from each other, care should be exercised not to mistake the second-order and third-order scattered radiations for the fluorescence spectrum. The second-order scattered radiation appears at a wavelength two times longer than the excitation wavelength, and the third-order scattered radiation occurs at a wavelength three times longer. For instance, if the excitation wavelength is 240 nm, the second-order and third-order radiations take place at 480 and 720 nm, respectively. For eliminating these scattered radiations, insert a short-wavelength cutoff filter in the path of fluorescing radiation (before the emission monochromator). It is advisable to use the filter set (P/N 650-0157) which is available as an optional accessory. Since the fluorescence spectrophotometer provides high sensitivity, just a slight amount of contamination on a cell may have an adverse effect on results of measurement. To prevent this, treat the cell properly after its use. Do not leave the cell containing say sample. In evaporation of a solvent, a residue of sample may adhere to the wall of the cell to cause contamination. In measurement of a very dilute sample, contamination on the inner and outer walls of the cell may cause a problem. If a droplet of sample solution is put on the outer wall of the cell in sample injection into it, wipe off the cell with tissue paper and then set it on the cell holder. E.3.5 Second-Order Scattered Radiation E.3.6 Contamination of Cell
A - 21 Figure E-7 shows a measurement example of fluorescence spectrum. (1) Scattering of exciting radiation (2) Raman spectrum of solvent (3) Fluorescence of impurities, solvent, etc. (4) Fluorescence of sample (5) Second-order spectrum of exciting radiation Fig. E-7 Measurement Example of Fluorescence Spectrum As shown in Figure E-7, other peaks than a fluorescence peak of sample appear in fluorescence spectral measurement. With reference to this example, it is necessary to identify a fluorescence peal of sample. E.3.7 Measurement Example of Fluorescence Spectrum (1) (5) (4) (3) (2) Relative intensity EX EX × 2 Wavelength
A - 22 APPENDIX F MEASUREMENT OF INSTRUMENTAL RESPONSE (CORRECTED SPECTRA) Spectrum correction is performed to enable measuring a true spectrum by eliminating instrumental response such as wavelength characteristics of the monochromator or detectors. The measurement of instrumental response is needed to perform spectrum correction. “Instrumental Response” is the function to measure and save the instrumental response. F.1 Measurement of Instrumental Response on Excitation Side This is the function to obtain the instrumental response on the excitation side such as wavelength characteristics of the excitation monochromator using Rhodamine B as a standard (quantum counter). The instrumental response is automatically read with a single wavelength scan operation. A spectrum is correctable within a range of 200 to 600 nm. (1) Handling of Rhodamine B Pour Rhodamine B into a triangular cell in the procedure illustrated in Fig. F-1. The triangular cell filled with Rhodamine B should in principle be stored at a dark place. Fig. F-1 Handling of Rhodamine B Syringe F649090 Rhodamine B Cut the supplied ampoule of Rhoda mine B with a cutter. Suck the solution into a syringe. Open the cover of triangular cell and pour the solution into it. Fill the cell with the solution in a volume at least half the capacity and close the cover.