Home > Hitachi > Spectrophotometer > Hitachi F 2500 Manual

Hitachi F 2500 Manual

    Download as PDF Print this page Share this page

    Have a look at the manual Hitachi F 2500 Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 913 Hitachi manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    							 A -
    13
    And, eventually the molecule returns to the ground state while 
    emitting fluorescent radiation.    Also, if radiationless transition to 
    the triplet state takes place, then phosphorescence is emitted 
    during triplet-to-singlet transition (from the excited triplet state to 
    the ground singlet state).    Generally phosphorescence persists 
    for 10
    -4 sec or longer due to the selection rule imposed on the 
    triplet-to-singlet transition.    In contrast, fluorescence persists for 
    a period of 10
    -8 to 10-9 sec in most cases. 
     
    As mentioned above, part of the radiation absorbed by the 
    substance is lost as vibration energy, etc.; therefore, the 
    fluorescence wavelength emitted from it is longer than the 
    excitation wavelength (Stokes’ law). 
    The ratio of the number of photons emitted during fluorescence 
    to the number of photons absorbed is called the quantum 
    efficiency of fluorescence.    The larger the quantum efficiency a 
    substance has, the more fluorescence it emits.    Also, the 
    intensity of florescence emitted from a substance is proportional 
    to the quantity of light absorbed by it.    When a dilute solution 
    sample is measured, the intensity of fluorescence is expressed 
    by  
    F = KI
    0clεφ 
    where, 
    F  :  Intensity of fluorescence 
    K :  Instrumental constant 
    I
    0  :  Intensity of exciting radiation 
    c  :  Concentration of substance 
    l  :  Cell path length (distance in the cell through which the   
      radiation travels) 
    ε  :  Absorptivity of substance 
    φ  :  Quantum efficiency of substance 
      
    						
    							 A -
    14
    E.2  Advantages of Fluorometry 
     
    As contrasted with fluorometry, absorptiometry for a 
    low-concentration sample is explained in the following : 
    A sample having 99% transmittance to blank is taken as an 
    example.    In the absorbance measurement of such a substance, 
    inaccuracies must always be taken into consideration.     
    Here, the inaccuracy is assumed to be 0.1%. 
    Since it has an effect on both the blank and sample, 
    Percent transmittance of blank  100.0 ± 0.1% 
    Percent transmittance of sample  99.0 ± 0.1% 
    Difference (proportional to  1.0 ± 0.2% 
    concentration of sample) 
     
    In this example, the uncertainty in concentration measurement is 
    ±20%.    While in the fluorometry, a difference from zero level 
    corresponds to the concentration of sample. 
    Accordingly, measurement accuracy is as follows : 
    Output signal level at measurement of sample  100 ± 0.1 
    Value corresponding to blank  0 ± 0.1 
    Difference (proportional to  100 ± 0.2 
    concentration of sample) 
     
    As can be seen from the above, the fluorometry is very 
    advantageous for analyzing a low-concentration sample since its 
    uncertainty is in most cases theoretically independent of the 
    concentration of sample.    Although actual practice may involve 
    some error factors which increase as the concentration of 
    sample becomes extremely low, the fluorometry is capable of 
    measuring low concentration with an accuracy 100 times higher 
    than in absorptiometry. 
    Figure E-2 is an explanatory illustration of the foregoing 
    description.    In absorptiometry, a difference between the 
    quantity of incident radiation Io and the quantity of transmitted 
    radiation is represented by signal Is.    A level at which the signal 
    Is becomes almost equal to a noise level is used as a detection 
    limit.   
    In fluorometry, however, since the quantity of fluorescence I
    F 
    itself is represented as a signal, just a small amount of 
    fluorescence can be amplified electrically for enabling detection.   
    Still more, since a fluorescence wavelength of a substance is 
    different from its excitation wavelength (incident light 
    wavelength), the fluorescence wavelength is not readily affected 
    by the exciting radiation, thereby contributing to ensuring high 
    sensitivity.   
    						
    							 A -
    15
     Amplified 
     
    Fig. E-2    Comparison between Absorptiometry and Fluorometry 
     
    In addition to high sensitivity, fluorometry is advantageous in that 
    more information is attainable.    A fluorescence spectrum is also 
    available besides an excitation spectrum which corresponds to 
    an absorption spectrum in absorptiometry. 
    The two kinds of wavelengths can be selected as desired, and a 
    fluorescence spectrum can be recorded using a properly 
    selected excitation wavelength (or vice versa).    Thus, 
    quantitative and qualitative analyses can be made for a sample 
    containing plural components. 
    Figure E-3 shows a simplified spectral graph of measurement of 
    a sample containing multiple components.    In absorptiometry, 
    since only the absorption spectrum can be measured, two or 
    more component wavelengths are presented.    If the absorption 
    wavelengths are similar to each other, each component cannot 
    be separated in measurement.    In fluorometry, even if the 
    absorption wavelengths are similar, a difference in fluorescence 
    makes it possible to select each fluorescence wavelength 
    properly.    Thus, each component can be separated in 
    measurement. 
     
     
    Fig. E-3    Measurement of Mult-Component Sample 
    Low 
    concentrationHigh 
    concentrationLow 
    concentration
    (b)  Fluorometry (a)  Absorptiometry
    High 
    concentration
     
    Amplified
    Absorption spectra 
    Component B 
    (b)  Fluorometry 
    Fluorescence spectra
    (a)  Absorptiometry
    Component A
    Excitation spectraComponent A
    Component B 
    						
    							 A -
    16
    Table E-1 compares information attainable in absorptiometry and 
    that in fluorometry. 
     
    Table E-1  Comparison of Information Attainable in 
    Absorptiometry and Fluorometry 
    Absorptiometry Fluorometry 
      Absorption spectrum only 
    (Corresponding to excitation 
    spectrum in fluorometry)  Excitation spectrum 
     Fluorescence spectrum 
     
    E.3  Notes of Fluorescence Analysis Measurement 
     
    For most kinds of samples, an increase of 1°C in the 
    temperature of sample causes the intensity of fluorescence to 
    decrease by 1 to 2%.    It is also reported that for some kinds of 
    biochemical samples, the intensity of fluorescence decreases as 
    much as 10% as the temperature increases by 1°C. 
    When analyzing a sample having a temperature-dependent 
    property, it is advisable to use the constant-temperature cell 
    holder (P/N 650-0150).    Constant-temperature measurement 
    can be carried out by circulating constant-temperature water 
    through this cell holder. 
     
    Some kinds of samples may be susceptible to a chemical 
    change due to exciting radiation.    In analysis of such a sample, 
    keep the shutter closed to cut off an excitation beam until 
    measurement is started, and then open the shutter immediately 
    before measurement.    If any chemical change due to exciting 
    radiation is observed still, determine a signal level at the start 
    time through extrapolation according to variation in signal level. 
      E.3.1 Temperature 
    Dependency of 
    Fluorescence 
    Intensity 
    E.3.2 Chemical 
    Change in 
    Sample due to 
    Radiation
      
    						
    							 A -
    17
    In fluorescence measurement, spectra having different natures 
    from that of fluorescence may be observed.    These are called 
    Rayleigh scattering spectrum and Raman scattering spectra; the 
    former appearing at the same wavelength position as the 
    excitation spectrum, and the latter appearing at the 
    longer-wavelength side near Rayleigh scattering. 
    In a fluorescence spectrum, when the excitation wavelength is 
    shifted, only the peak height is changed while the peak 
    wavelength position remains intact.    In a Raman scattering 
    spectrum, when the excitation wavelength is shifted, the peak 
    wavelength position is also changed accordingly.    Both the 
    Rayleigh scattering and Raman scattering are caused by a 
    solvent which may be contained in the sample.     
    When examining the spectral plot, be careful not to mistake 
    these scattering effects of the fluorescence peak of interest. 
    Table E-3 presents the Raman spectral peak position at each 
    excitation wavelength for the purpose of reference. 
     
     
    Fig. E-4    Raman Spectrum of Water 
    E.3.3 Raman 
    Scattering
     
    Relative intensity
    Excitation wavelength
    Raman scattering 
    						
    							 A -
    18
    Table E-3  Raman Peak Positions at Respective Excitation 
    Wavelengths
     
     
    (Excitation 
    wavelength)
     WaterEthanolCyclohexane Carbon 
    Tetrachloride
     Chloroform 
    248  271  267  267                    
    313 350 344  344  320  346 
    365 416 405  408  375  410 
    405 469 459  458  418  461 
    Excitation 
    wavelength 
    and 
    Raman 
    peak 
    position 
    (nm)
     436 511 500  499  450  502 
     
     
    In measurement of a high-concentration sample, a variety of 
    error factors may be involved.    The most significant error factor 
    consists in that an excitation beam is absorbed at the entrance of 
    a cell to prevent a sufficient level of excitation at the center of 
    cell. 
    Figure E-5 illustrates an extreme case of this condition.     
    Although fluorescence is emitted in the vicinity of the entrance 
    for the excitation beam, it is not taken into the emission 
    monochromator. 
     
     
     
     
     
     
     
     
     
    Fig. E-5    Sample Having an Extremely High Concentration 
     
      E.3.4 Handling of 
    High- 
    Concentration 
    Samples
     
    Fluorescence is reflected here. Excitation beam
    Fluorescence
      
    						
    							 A -
    19
    If only the incident point of excitation beam is bright, it is 
    necessary to dilute the sample properly for measurement. 
    The second significant error factor consists in extinction due to 
    concentration.    This condition is caused by preventing activation 
    through interaction of molecules. 
    The third significant error factor consists in re-absorption of 
    fluorescence.    As shown in Figure E-6, this condition occurs 
    due to overlapping between the shot-wavelength side of 
    fluorescence spectrum and the long-wavelength side of 
    excitation spectrum.    Therefore, it seems that the fluorescence 
    spectrum has been shifted toward the long-wavelength side to 
    some extent.   
    In measurement of an ordinary kind of sample, however, this 
    condition will not impede quantitative determination significantly. 
     
     
     
    Fig. E-6    Explanatory Illustration of Re-absorption 
     
    In any case, if there is a possibility of a measurement error due 
    to high concentration of a sample, dilute the sample properly or 
    carry out surface fluorescence measurement using a solid 
    sample holder. 
     
    Wavelength 
    Fluorescence spectrum
         Excitation spectrum
    Re-absorption occurs here.
     
    Relative intensity
      
    						
    							 A -
    20
    Where the excitation and emission wavelengths are plotted near 
    each other, care should be exercised not to mistake the Raman 
    and Rayleigh scattering for the fluorescence spectrum as 
    mentioned in E.3.3.    Where the excitation and emission 
    wavelengths are plotted apart from each other, care should be 
    exercised not to mistake the second-order and third-order 
    scattered radiations for the fluorescence spectrum. 
    The second-order scattered radiation appears at a wavelength 
    two times longer than the excitation wavelength, and the 
    third-order scattered radiation occurs at a wavelength three 
    times longer. 
    For instance, if the excitation wavelength is 240 nm, the 
    second-order and third-order radiations take place at 480 and 
    720 nm, respectively.    For eliminating these scattered radiations, 
    insert a short-wavelength cutoff filter in the path of fluorescing 
    radiation (before the emission monochromator).    It is advisable 
    to use the filter set (P/N 650-0157) which is available as an 
    optional accessory. 
     
    Since the fluorescence spectrophotometer provides high 
    sensitivity, just a slight amount of contamination on a cell may 
    have an adverse effect on results of measurement.    To prevent 
    this, treat the cell properly after its use. 
    Do not leave the cell containing say sample.    In evaporation of a 
    solvent, a residue of sample may adhere to the wall of the cell to 
    cause contamination. 
    In measurement of a very dilute sample, contamination on the 
    inner and outer walls of the cell may cause a problem.    If a 
    droplet of sample solution is put on the outer wall of the cell in 
    sample injection into it, wipe off the cell with tissue paper and 
    then set it on the cell holder. 
     E.3.5 Second-Order 
    Scattered 
    Radiation
     
    E.3.6 Contamination 
    of Cell
      
    						
    							 A -
    21
    Figure E-7 shows a measurement example of fluorescence 
    spectrum. 
     
     
     
     
    (1)  Scattering of exciting radiation 
    (2)  Raman spectrum of solvent 
    (3)  Fluorescence of impurities, solvent, etc. 
    (4)  Fluorescence of sample 
    (5)  Second-order spectrum of exciting radiation 
    Fig. E-7    Measurement Example of Fluorescence Spectrum 
     
    As shown in Figure E-7, other peaks than a fluorescence peak of 
    sample appear in fluorescence spectral measurement.     
    With reference to this example, it is necessary to identify a 
    fluorescence peal of sample. 
     
     
    E.3.7 Measurement 
    Example of 
    Fluorescence 
    Spectrum
     
    (1) 
    (5)  
    (4)
     
    (3)
     
     
    (2)
     Relative intensity
    EX EX × 2 Wavelength 
    						
    							 A -
    22
    APPENDIX F  MEASUREMENT OF INSTRUMENTAL 
    RESPONSE (CORRECTED SPECTRA) 
     
     
    Spectrum correction is performed to enable measuring a true 
    spectrum by eliminating instrumental response such as 
    wavelength characteristics of the monochromator or detectors.     
    The measurement of instrumental response is needed to perform 
    spectrum correction.    “Instrumental Response” is the function to 
    measure and save the instrumental response.     
     
     
    F.1  Measurement of Instrumental Response on Excitation Side 
     
    This is the function to obtain the instrumental response on the 
    excitation side such as wavelength characteristics of the 
    excitation monochromator using Rhodamine B as a standard 
    (quantum counter).   
    The instrumental response is automatically read with a single 
    wavelength scan operation.    A spectrum is correctable within a 
    range of 200 to 600 nm.   
     
    (1)  Handling of Rhodamine B 
     
    Pour Rhodamine B into a triangular cell in the procedure 
    illustrated in Fig. F-1.    The triangular cell filled with   
    Rhodamine B should in principle be stored at a dark place.     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
     
    Fig. F-1    Handling of Rhodamine B 
    Syringe F649090 
     
    Rhodamine B 
     Cut the supplied 
    ampoule of 
    Rhoda mine B 
    with a cutter.     Suck the solution 
    into a syringe.   Open the cover 
    of triangular cell 
    and pour the 
    solution into it.   Fill the cell with the 
    solution in a volume 
    at least half the 
    capacity and close 
    the cover.    
    						
    All Hitachi manuals Comments (0)

    Related Manuals for Hitachi F 2500 Manual