Home > Hitachi > Spectrophotometer > Hitachi F 2500 Manual

Hitachi F 2500 Manual

    Download as PDF Print this page Share this page

    Have a look at the manual Hitachi F 2500 Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 913 Hitachi manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    							3 - 7 
    (f) Delay  
    After pressing the Measure button, measurement is 
    started following the delay time set here.    It is used for 
    temperature stabilization, etc.    In repeat measurement, 
    it is the time until the start of the first measurement.   
    The delay time is invalid for the second and 
    subsequent measurements.   
    Input range:    0 to 9999 s   
     
    (g) EX Slit  
    Select a slit width for the excitation side.    Unit:    nm   
     
    F-4500 F-2500 
    1.0 2.5  
    2.5 5.0 
    5.0 10.0 
    10.0 20.0 
    20.0  
     
     
    (h) EM Slit  
    Select a slit width for the emission side.    Unit:    nm   
     
    F-4500 F-2500 
    1.0 2.5  
    2.5 5.0 
    5.0 10.0 
    10.0 20.0 
    20.0  
     
     
    (i) PMT Voltage  
    A function for controlling the voltage of the 
    photomultiplier detector (photomultiplier tube).     
    Select one of the following values.    Changing from 
    400 V to 700 V increases the data value about two 
    digits and changing from 700 V to 950 V increases it 
    about 1 digit.     
    •  950 V (only with F-4500)   
    • 700 V  
    • 400 V 
     
    3.2  
    						
    							3 - 8 
    (j) Response  
    Select a response time from the following.   
     
    F-4500 F-2500 
    0.04 0.04 
    0.01 0.08 
    0.05 0.4 
    0.1 2 
    0.5 8 
    2.0 Auto 
    8.0  
    Auto  
     
     
    By selecting Auto, a response time is set automatically 
    according to scan speed.    Select Auto usually.   
     
    (k)  Shutter control (only with F-4500)   
    The shutter can be automatically closed in other than 
    measurement for minimizing sample deterioration due 
    to the energy of excitation beam and opened when 
    measurement starts.    When you put a check mark at 
    the head, the shutter will close and open at start of 
    measurement.    The shutter will close again when 
    measuring wavelength begins returning to the start 
    wavelength after measurement in the time scan mode.     
     
    (l) Replicates  
    Set the number of repeat measurements.    Input range 
    is 1 to 99.     
     
    (m) Cycle time 
    Set a repetition interval.    Input range is 0.0 to   
    180.0 min.   
     
    3.2    Creating an Analysis Method  
    						
    							3 - 9 
    (3) Monitor  
     
    When selecting the Monitor tab, the window in Fig. 3-5 
    appears.   
     
     
     
    Fig. 3-5  Monitor Tab 
     
     
    Assign the following parameters.   
     
    (a)  Y-Axis Max (maximum)   
    Input a maximum value for Y-axis on the monitor 
    window.  
     
    (b)  Y-Axis Min (minimum)   
    Input a minimum value for Y-axis on the monitor 
    window.  
     
    (c)  X-Axis Max (maximum)   
    Input a maximum value for X-axis on the monitor 
    window.  
     
    (d)  X-Axis Min (minimum)   
    Input a minimum value for X-axis on the monitor 
    window.  
     
    NOTE:  X-Axis Max and Min correspond to a display 
    interval on the monitor window.   
    For example, when Max and Min are 60 and 
    30, display on the window is made within 30 
    to 60 s, 60 to 90 s, 90 to 120 s and so on.       
    3.2  
    						
    							3 - 10 
    (f)  Open data processing window after data acquisition   
    Select whether or not to conduct data processing after 
    sample measurement.    When selected (check mark is 
    applied), data is automatically saved after 
    measurement.  
     
    NOTE:  When the remaining memory capacity of PC 
    in operation is very small (system resource is 
    less than 20%), the data processing window 
    will not open even if this item is turned ON.   
    Either close the presently opened data 
    processing window or restart the FL 
    Solutions program.   
     
    (g)  Print report after data acquisition   
    Select whether or not to print a report after sample 
    measurement.    When selected (check mark is 
    applied), printing will be made automatically after 
    measurement.    The items set under the Report tab 
    will be printed.     
     
    (h) Overlay  
    Spectra can be overlaid in the monitor window.     
    This setting cannot be selected for repeat 
    measurement.   
     
    (4) Processing  
     
    On selection of the Processing tab, the window in Fig. 3-6 
    will be displayed.   
     
     
     
    Fig. 3-6  Processing Tab 
    3.2    Creating an Analysis Method  
    						
    							3 - 11 
    (a) Processing choices  
    A list of data processing (Savitsky-Golay smooth, 
    Mean smooth, Median smooth, Derivative) is shown.     
    Select a data processing item, and click the rightward-
    pointing arrow key between the Processing choices 
    and Processing steps display fields.    Then, the 
    selected method appears in the Processing steps field.   
     
    (b) Processing steps  
    The processing sequence is displayed.    To delete a 
    processing method, first select the method, and then 
    click the leftward-pointing arrow key between the 
    Processing choices and Processing steps display 
    fields.    Then, the selected method disappears from 
    the Processing steps field.   
    Also, in case you want to carry out any smoothing in 
    Processing steps, double-click the relevant item and 
    the smoothing parameters (order, data points, etc.) can 
    be changed.   
     
    (c) Peak Finding  
    Set threshold and sensitivity values for peak detection.   
     
    (d) Kinetics  
    Input a start time, end time and K-factor for kinetics 
    (rate calculation).   
     
    3.2  
    						
    							3 - 12 
    (5) Report  
     
    Select the Report tab, and the window in Fig. 3-7 will 
    appear.  
     
     
     
    Fig. 3-7  Report Tab 
     
     
    This window is used to select an output method and items 
    of measurement result.     
     
    (a) Output  
    1) Print Report  
    A report will be printed.    For printing method, 
    refer to 3.6.   
    2)  Use Microsoft® Excel   
    Data will be transferred to Microsoft® Excel.     
    For transfer method, refer to 6.13.   
    3)  Use print generator sheet 
    Select this method for using the report generator 
    (option).    For usage, refer to the instruction 
    manual of the report generator.   
     
    (b) Print format  
    Selectable when Report is chosen for Print Method.   
    Select a printing orientation.    Either Portrait or 
    Landscape orientation is selectable.    When the 
    Landscape orientation is selected, date and method 
    can be printed.   
     
    3.2    Creating an Analysis Method  
    						
    							3 - 13 
    (c)  Printable (transferable) items   
    1) Include data 
    The date and time of analysis can be printed.   
    Printing time will always be printed.       
    2) Include method  
    3) Include graph  
    4)  Include data listing   
    • Constant  
    Data will be printed at fixed intervals between 2 
    wavelength points.    Input Data interval, Start 
    wavelength and End wavelength.     
    • Select data  
    Data at the specified wavelength will be printed.   
    Up to 12 wavelengths can be specified.   
    5)  Include peak table   
    Select output items in the peak table.   
    •  Peak time/Peak data 
    •  Start time/End time 
    •  Valley time/Valley data 
    • Peak Area  
    6) Include kinetics  
     
    (d) Printer Font  
    Click this button for changing the font to be used in the 
    report.  
    3.2  
    						
    							3 - 14 
    3.3  Defining Your Samples  
     
    When “Use Sample Table” has been set for the Analysis Method, 
    the samples must be defined before measurement.   
     
    (1)  Open the monitor window and click the          button.   
     
    (2)  When the Sample Table window in Fig. 3-8 appears, fill in 
    the blank section.    For instance, if there are five samples, 
    click the Insert button five times.   
     
     
     
    Fig. 3-8    Sample Table Window 
     
     
    (3)  Input the Sample, Comments and File Name for each 
    sample.    The sample names input here appear on the 
    monitor window also.     
    Maximum number of characters to be input   
    Sample : 20 characters  
    Comments :  255 characters   
     
    3.3  Defining Your Samples  
    						
    							3 - 15 
    (4)  When input is finished, check the entries and then click the 
    OK button.   
    If a sample table is not used, then click the          button.   
    The window in Fig. 3-9 will appear.    Make entries for 
    Sample and Comment.    For automatic saving, put a check 
    mark in the “Auto file saving after measurement” box and 
    specify a file name for saving.   
     
     
     
    Fig. 3-9 
    3.3  
    						
    							3 - 16 
    3.4    Conducting a Time Scan   
     
    (1)  There are two methods of starting the scan.     
     
    (a)  Select the Measure command from the 
    Spectrophotometer menu.   
     
    (b)  Click the     (measurement) button on the 
    measurement toolbar.   
     
    For halting the measurement midway, click the          button 
    on the toolbar.   
     
    (2)  Upon clicking the          button, the window shown in   
    Fig. 3-10 appears when using a sample table.    Now click 
    Yes.  
     
     
     
    Fig. 3-10 
     
     
    (3)  The dialog box as in Fig. 3-11 will open.    Set each sample 
    and click OK.  
     
     
     
    Fig. 3-11 
    3.4    Conducting a Time Scan  
    						
    All Hitachi manuals Comments (0)