Home > Hitachi > Spectrophotometer > Hitachi F 2500 Manual

Hitachi F 2500 Manual

    Download as PDF Print this page Share this page

    Have a look at the manual Hitachi F 2500 Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 913 Hitachi manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    							2 - 8 
    (l) EM Slit  
    Select a slit width for the emission side.    Unit:    nm   
     
    F-4500 F-2500  
    1.0 2.5  
    2.5 5.0 
    5.0 10.0 
    10.0 20.0 
    20.0  
     
     
    (m) PMT Voltage  
    A function for controlling the voltage of the 
    photomultiplier detector (photomultiplier tube).     
    Select one of the following values.    Changing from 
    400 V to 700 V increases the data value about two 
    digits and changing from 700 V to 950 V increases it 
    about 1 digit.     
      950 V (only with F-4500)   
     700 V  
     400 V 
     
    (n) Response  
    Select a response time from the following.   
     
    F-4500 F-2500  
    0.0004 0.04 
    0.01 0.08 
    0.05 0.4 
    0.1 2 
    0.5 8 
    2.0 Auto 
    8.0  
    Auto  
     
     
    By selecting Auto, a response time is set automatically 
    according to scan speed.    Select Auto usually.   
     
    2.2    Creating an Analysis Method  
    						
    							2 - 9 
    (o) Corrected spectra  
    A function for determining the spectrum inherent to a 
    sample by correcting the photometer wavelength 
    characteristic using the saved instrument parameters, 
    following measurement with the instrument parameters 
    for photometer control.    If there are no instrument 
    parameters for the wavelength range to be measured, 
    only the wavelength range that allows correction is 
    scanned.    And if there is no wavelength area 
    measurable, an error will occur.     
    For using this function, a Spectra Correction Parts Set 
    (P/N 251-0126) is required.    The Instrument 
    Response command is provided in the Utility menu for 
    operating this function.     
    When this setting is at ON, “Corrected spectra” 
    appears on the monitor window.    (See Fig. 1-26 in 
    1.6.)  
     
    (p)  Shutter control (only with F-4500)   
    The shutter can be automatically closed in other than 
    measurement for suppressing sample deterioration 
    due to the energy of excitation beam and opened when 
    measurement starts.    When you put a check mark at 
    the head, the shutter will close and open at start of 
    measurement.    The shutter will close again when 
    measuring wavelength begins returning to the start 
    wavelength after measurement in the wavelength scan 
    mode.   
     
    (q) Replicates  
    Set the number of repeat measurements.    Input range 
    is 1 to 99.    An average spectrum can also be 
    determined in repeat measurement.    Setting is 
    available on the Processing tab.   
     
    (r) Cycle time  
    Set a repetition interval.    Input range is 0.0 to   
    180.0 min.   
    2.2  
    						
    							2 - 10 
    (3) Monitor  
     
    When selecting the Monitor tab, the window in Fig. 2-5 
    appears.   
     
     
     
    Fig. 2-5  Monitor Tab 
     
     
    Assign the following parameters.   
     
    (a)  Y-Axis Max (maximum)   
    Input a maximum value for Y-axis on the monitor 
    window.  
     
    (b)  Y-Axis Min (minimum)   
    Input a minimum value for Y-axis on the monitor 
    window.  
     
    (c)  Open data processing window after data acquisition   
    Select whether or not to conduct data processing after 
    sample measurement.    When selected (check mark is 
    applied), an icon is displayed for the data processing 
    window at the end of measurement.    And by opening 
    this icon, data processing such as peak detection is 
    available.  
     
    2.2    Creating an Analysis Method  
    						
    							2 - 11 
    NOTE:  When the remaining memory capacity of PC 
    in operation is very small (system resource 
    is less than 20%), the data processing 
    window will not open even if this item is 
    turned ON.    Either close the presently 
    opened data processing window or restart 
    the FL Solutions program.     
     
    (d)  Print report after data acquisition   
    Select whether or not to print a report after sample 
    measurement.    When selected (check mark is 
    applied), printing will be made automatically after 
    measurement.    The items set under the Report tab 
    will be printed.    When this item is not selected, data 
    processing can be made by reading out a file via File - 
    Open.  
     
    (e) Overlay  
    Spectra can be overlaid in the monitor window.     
    This setting cannot be selected for repeat 
    measurement.  
     
    (4) Processing  
     
    On selection of the Processing tab, the window in Fig. 2-6 
    will be displayed.   
     
     
     
    Fig. 2-6  Processing Tab 
     
    2.2  
    						
    							2 - 12 
    (a)  CAT (Computing for Averaging Transient)   
    Used for obtaining an average spectrum in repeat 
    measurement.    This is especially effective for 
    measuring samples with which the emission intensity 
    varies with time.   
     
    (b) Processing choices  
    A list of data processing (Savitsky-Golay, Mean 
    smooth, Median smooth, Derivative) is shown.     
    Select a data processing item, and click the rightward-
    pointing arrow key between the Processing choices 
    and Processing steps display fields.    Then, the 
    selected method appears in the Processing steps field.   
     
    (c) Processing steps  
    The processing sequence is displayed.    To delete a 
    processing method, first select the method, and then 
    click the leftward-pointing arrow key between the 
    Processing choices and Processing steps display 
    fields.    Then, the selected method disappears from 
    the Processing steps field.   
    Also, in case you want to carry out any smoothing in 
    Processing steps, double-click the relevant item and 
    the smoothing parameters (order, data points, etc.) can 
    be changed.   
     
    (d) Peak Finding  
    Select a peak detection method indicated in the data 
    processing window.   
     
    (e) Threshold  
    Set a detection limit for the photometric value axis of 
    peak and valley.    Peaks or valleys below the setting 
    will not be detected.    Input range is 0.001 to 1000.     
     
    (f) Sensitivity  
    Select the number of data points in the horizontal axis 
    direction (abscissa).    Select a sensitivity of “1” for 
    detecting sharp peaks.    Or select “8” for detecting 
    broad peaks.   
     1  
     2 
     4 
     8 
     
    2.2    Creating an Analysis Method  
    						
    							2 - 13 
    (5) Report  
     
    Select the Report tab, and the window in Fig. 2-7 will 
    appear.  
     
     
     
    Fig. 2-7  Report Tab 
     
     
    This window is used to select an output method and items 
    of measurement result.     
     
    (a) Output  
    1) Print Report  
    A report will be printed.    For printing method, 
    refer to 2.6.   
    2)  Use Microsoft® Excel   
    Data will be transferred to Microsoft® Excel.     
    For transfer method, refer to 5.13.   
    3)  Use print generator sheet 
    Select this method for using the report generator 
    (option).    For usage, refer to the instruction 
    manual of the report generator.   
     
    (b) Orientation  
    Selectable when Report is chosen for Print Method.   
    Select a printing orientation.    Either Portrait or 
    Landscape orientation is selectable.    When the 
    Landscape orientation is selected, date and method 
    can be printed.   
    2.2  
    						
    							2 - 14 
    (c)  Printable (transferable) items   
    1) Include date  
    2) Include method  
    3) Include graph  
    4)  Include data listing   
     Constant  
    Data will be printed at fixed intervals between 2 
    wavelength points.    Input Data interval, Start 
    wavelength and End wavelength.     
     Select data 
    Data at the specified wavelength will be printed.   
    Up to 12 wavelengths can be specified.   
    5)  Include peak table 
    Select output items in the peak table.   
      Peak WL/Peak data   
      Start WL/End WL   
      Valley WL/Valley data   
     Peak Area  
     
    (d) Printer Font  
    Click this button for changing the font to be used in the 
    report.  
    2.2    Creating an Analysis Method  
    						
    							2 - 15 
    2.3  Defining Your Samples  
     
    When “Use Sample Table” has been set for the Analysis Method, 
    the samples must be defined before measurement.   
     
    (1)  Open the monitor window and click the          button.   
     
    (2)  When the Sample Table window in Fig. 2-8 appears, fill in 
    the blank section.    For instance, if there are five samples, 
    click the Insert button five times.   
     
     
     
    Fig. 2-8    Sample Table Window 
     
     
    (3)  Input the Sample, Comments and File Name for each 
    sample.    The sample names input here appear on the 
    monitor window also.    And the sample names input here 
    are not changeable.     
    Maximum number of characters to be input   
    Sample : 20 characters  
    Comments :  255 characters   
     
     
    2.3  
    						
    							2 - 16 
    (4)  When input is finished, check the entries and then click the 
    OK button.   
    If a sample table is not used, then click the          button.   
    The window in Fig. 2-9 will appear.    Make entries for 
    Sample and Comment.    For automatic saving, put a check 
    mark in the “Auto File” box and specify a file name for 
    saving.  
     
     
     
    Fig. 2-9 
     
    2.3  Defining Your Samples  
    						
    							2 - 17 
    2.4  Prescan  
     
    Select Pre-scan command from the Spectrophotometer menu or 
    click the     (pre-scan) button.  
    This function executes high-speed scan under the set measuring 
    conditions, detects the peak wavelength and moves the system 
    to that wavelength.    Also, the upper limit photometric value is 
    automatically set so that the peak-wavelength value becomes 
    approximately 70% of that upper limit value.     
    Scan ranges are as follows.   
     
    Excitation spectra  :  Start wavelength to emission 
    wavelength - 20 nm or shorter 
    wavelength side of end wavelength   
     
    Emission spectra  :  Excitation wavelength or longer 
    wavelength side of start wavelength + 
    20 nm to end wavelength   
     
    Synchronous spectra :  Start wavelength to end wavelength   
     
    After prescan, the upper limit photometric value is automatically 
    set.   
    2.4  
    						
    All Hitachi manuals Comments (0)