Hitachi F 2500 Manual
Have a look at the manual Hitachi F 2500 Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 913 Hitachi manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.
2 - 8 (l) EM Slit Select a slit width for the emission side. Unit: nm F-4500 F-2500 1.0 2.5 2.5 5.0 5.0 10.0 10.0 20.0 20.0 (m) PMT Voltage A function for controlling the voltage of the photomultiplier detector (photomultiplier tube). Select one of the following values. Changing from 400 V to 700 V increases the data value about two digits and changing from 700 V to 950 V increases it about 1 digit. 950 V (only with F-4500) 700 V 400 V (n) Response Select a response time from the following. F-4500 F-2500 0.0004 0.04 0.01 0.08 0.05 0.4 0.1 2 0.5 8 2.0 Auto 8.0 Auto By selecting Auto, a response time is set automatically according to scan speed. Select Auto usually. 2.2 Creating an Analysis Method
2 - 9 (o) Corrected spectra A function for determining the spectrum inherent to a sample by correcting the photometer wavelength characteristic using the saved instrument parameters, following measurement with the instrument parameters for photometer control. If there are no instrument parameters for the wavelength range to be measured, only the wavelength range that allows correction is scanned. And if there is no wavelength area measurable, an error will occur. For using this function, a Spectra Correction Parts Set (P/N 251-0126) is required. The Instrument Response command is provided in the Utility menu for operating this function. When this setting is at ON, “Corrected spectra” appears on the monitor window. (See Fig. 1-26 in 1.6.) (p) Shutter control (only with F-4500) The shutter can be automatically closed in other than measurement for suppressing sample deterioration due to the energy of excitation beam and opened when measurement starts. When you put a check mark at the head, the shutter will close and open at start of measurement. The shutter will close again when measuring wavelength begins returning to the start wavelength after measurement in the wavelength scan mode. (q) Replicates Set the number of repeat measurements. Input range is 1 to 99. An average spectrum can also be determined in repeat measurement. Setting is available on the Processing tab. (r) Cycle time Set a repetition interval. Input range is 0.0 to 180.0 min. 2.2
2 - 10 (3) Monitor When selecting the Monitor tab, the window in Fig. 2-5 appears. Fig. 2-5 Monitor Tab Assign the following parameters. (a) Y-Axis Max (maximum) Input a maximum value for Y-axis on the monitor window. (b) Y-Axis Min (minimum) Input a minimum value for Y-axis on the monitor window. (c) Open data processing window after data acquisition Select whether or not to conduct data processing after sample measurement. When selected (check mark is applied), an icon is displayed for the data processing window at the end of measurement. And by opening this icon, data processing such as peak detection is available. 2.2 Creating an Analysis Method
2 - 11 NOTE: When the remaining memory capacity of PC in operation is very small (system resource is less than 20%), the data processing window will not open even if this item is turned ON. Either close the presently opened data processing window or restart the FL Solutions program. (d) Print report after data acquisition Select whether or not to print a report after sample measurement. When selected (check mark is applied), printing will be made automatically after measurement. The items set under the Report tab will be printed. When this item is not selected, data processing can be made by reading out a file via File - Open. (e) Overlay Spectra can be overlaid in the monitor window. This setting cannot be selected for repeat measurement. (4) Processing On selection of the Processing tab, the window in Fig. 2-6 will be displayed. Fig. 2-6 Processing Tab 2.2
2 - 12 (a) CAT (Computing for Averaging Transient) Used for obtaining an average spectrum in repeat measurement. This is especially effective for measuring samples with which the emission intensity varies with time. (b) Processing choices A list of data processing (Savitsky-Golay, Mean smooth, Median smooth, Derivative) is shown. Select a data processing item, and click the rightward- pointing arrow key between the Processing choices and Processing steps display fields. Then, the selected method appears in the Processing steps field. (c) Processing steps The processing sequence is displayed. To delete a processing method, first select the method, and then click the leftward-pointing arrow key between the Processing choices and Processing steps display fields. Then, the selected method disappears from the Processing steps field. Also, in case you want to carry out any smoothing in Processing steps, double-click the relevant item and the smoothing parameters (order, data points, etc.) can be changed. (d) Peak Finding Select a peak detection method indicated in the data processing window. (e) Threshold Set a detection limit for the photometric value axis of peak and valley. Peaks or valleys below the setting will not be detected. Input range is 0.001 to 1000. (f) Sensitivity Select the number of data points in the horizontal axis direction (abscissa). Select a sensitivity of “1” for detecting sharp peaks. Or select “8” for detecting broad peaks. 1 2 4 8 2.2 Creating an Analysis Method
2 - 13 (5) Report Select the Report tab, and the window in Fig. 2-7 will appear. Fig. 2-7 Report Tab This window is used to select an output method and items of measurement result. (a) Output 1) Print Report A report will be printed. For printing method, refer to 2.6. 2) Use Microsoft® Excel Data will be transferred to Microsoft® Excel. For transfer method, refer to 5.13. 3) Use print generator sheet Select this method for using the report generator (option). For usage, refer to the instruction manual of the report generator. (b) Orientation Selectable when Report is chosen for Print Method. Select a printing orientation. Either Portrait or Landscape orientation is selectable. When the Landscape orientation is selected, date and method can be printed. 2.2
2 - 14 (c) Printable (transferable) items 1) Include date 2) Include method 3) Include graph 4) Include data listing Constant Data will be printed at fixed intervals between 2 wavelength points. Input Data interval, Start wavelength and End wavelength. Select data Data at the specified wavelength will be printed. Up to 12 wavelengths can be specified. 5) Include peak table Select output items in the peak table. Peak WL/Peak data Start WL/End WL Valley WL/Valley data Peak Area (d) Printer Font Click this button for changing the font to be used in the report. 2.2 Creating an Analysis Method
2 - 15 2.3 Defining Your Samples When “Use Sample Table” has been set for the Analysis Method, the samples must be defined before measurement. (1) Open the monitor window and click the button. (2) When the Sample Table window in Fig. 2-8 appears, fill in the blank section. For instance, if there are five samples, click the Insert button five times. Fig. 2-8 Sample Table Window (3) Input the Sample, Comments and File Name for each sample. The sample names input here appear on the monitor window also. And the sample names input here are not changeable. Maximum number of characters to be input Sample : 20 characters Comments : 255 characters 2.3
2 - 16 (4) When input is finished, check the entries and then click the OK button. If a sample table is not used, then click the button. The window in Fig. 2-9 will appear. Make entries for Sample and Comment. For automatic saving, put a check mark in the “Auto File” box and specify a file name for saving. Fig. 2-9 2.3 Defining Your Samples
2 - 17 2.4 Prescan Select Pre-scan command from the Spectrophotometer menu or click the (pre-scan) button. This function executes high-speed scan under the set measuring conditions, detects the peak wavelength and moves the system to that wavelength. Also, the upper limit photometric value is automatically set so that the peak-wavelength value becomes approximately 70% of that upper limit value. Scan ranges are as follows. Excitation spectra : Start wavelength to emission wavelength - 20 nm or shorter wavelength side of end wavelength Emission spectra : Excitation wavelength or longer wavelength side of start wavelength + 20 nm to end wavelength Synchronous spectra : Start wavelength to end wavelength After prescan, the upper limit photometric value is automatically set. 2.4