Hitachi F 2500 Manual
Have a look at the manual Hitachi F 2500 Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 913 Hitachi manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.
1 - 2 (3) A calibration curve can be printed. A calibration curve can be drawn by inputting its factor value. (4) Measurement data and a calibration curve are displayed on the same screen window. It is therefore possible to carry out measurement while observing the calibration curve. (5) According to the coefficient of determination calculated automatically, judgment can be made on a degree of matching between standard values and a calibration curve. Two/three-wavelength calculation is permitted in preparation of a calibration curve. (6) Statistical calculation is possible. Mean value of measured data and standard deviation (SD) can be calculated. (7) Upper/lower limit judgment will be performed. (8) Spectrum display under photometry In the data measured by a quatification method; peak height, peak area, peak ratio or derivative, a spectrum due to wavelength scan is displayable. 3-D Scan (1) Spectral calculation can be performed using the measured data. (2) A peak table can be displayed. (3) A bird’s-eye view/synchronous spectrum can be displayed and printed. (4) Printing conditions are selectable. In use of the DDE function, the wavelength interval and start/end wavelengths of data list can be selected. 1.1 Functions
1 - 3 Secondary Processing of Data Measured data can be saved, and a variety of processings can be carried out on the saved data. (1) Automatic reading of data (spectrum tracing) By placing the cursor on the currently displayed spectral graph, a photometric value at the cursor position can be read out. (2) Peak detection Peaks and valleys of the currently displayed spectrum are detected automatically. (3) Change of scale A measured spectrum can be enlarged or reduced by changing the ordinate and abscissa scales. (4) Smoothing A measured spectrum can be smoothed for elimination of noise components. (5) Derivative Spectral derivative is obtainable. It is permitted to specify an order of derivative, that of smoothing and a count value of data. (6) Spectrum calculation Fundamental arithmetic operations can be made between spectra or a spectrum can be multiplied with a factor. (7) Area The area value of the specified peak will be calculated. The average area value is also calculable among peaks. (8) File conversion Measurement data can be converted into an ASCII text file (*.txt) or JCAMP-DX file (*.dx). It is also possible to save a spectrum and calibration curve graph in a Windows meta file (*.wmf). (9) Display/print font setting The font, style and size of characters to be displayed and printed can be changed as desired. 1.1
1 - 4 (10) Data acquisition Photometric value data on the data processing window and monitor window can be acquired. Acquisition is allowed within 100 data. (11) Printing items collectively specifiable Printing items on overlaid spectra can be specified collectively. (12) Specification of printing items Items in a peak table can be selected and printed. (13) Portrait or Landscape printing settable Printing can be made Portrait or Landscape. (14) Collective spectrum printing Printing can be carried out just by specifying a file. (15) Selection of spectrum to be displayed The spectrum to be displayed for overlaying is selectable. (16) Automatic saving mode ON/OFF settable ON/OFF can be set in the sample name/comment input window. Automatic saving is effective in all the modes that use a sample table. (17) Data in each mode can be transferred to Microsoft Excel. (This is valid when Microsoft Excel 97 has been installed in PC.) (18) After copying spectral data into commercial software such as Microsoft Word, its scale, line type, color, etc. are changeable on the software. (19) The report generator (separately available, P/N 250-0455) can be used. 1.1 Functions
1 - 5 1.2 Installing the FL Solutions Program For using the FL Solutions program, it must be installed using the attached system disks. Installation of the FL Solutions program is required when using it at the first time (not required later for each operation). In the following installation session, it is assumed that: • The floppy disk drive is assigned as drive A. • The destination of installation is assigned as drive C. (1) Using the mouse, click the Start button of the Windows 98 and select Settings - Control Panel. The Control Panel window will then appear as shown in Fig. 1-1. Fig. 1-1 (2) In the Control Panel window, double-click the Add/Remove Programs icon. The Add/Remove Programs dialog box will then appear as shown in Fig. 1-2. In this dialog box, click Install. Fig. 1-2 1.2
1 - 6 (3) Then, when the dialog box shown in Fig. 1-3 appears, push in the first one of the attached system disks and click Next. Fig. 1-3 (4) Then, when the dialog shown in Fig. 1-4 appears, click Finish. Fig. 1-4 Installation process will then begin. Take steps according to the instruction messages. 1.2 Installing the FL Solutions Program
1 - 7 (5) On completion of the installation process, restart the personal computer. (6) For startup of the FL Solutions program, refer to 1.3. NOTES: 1. Monitor resolution must be set at 800 × 600 at minimum. Otherwise, each window of the FL Solutions program may not be displayed fully. 2. When the large font size of display is selected, words may not be indicated normally on each window. In this case, font size should be changed to the small one. For change, select Control Panel - Display Properties - Settings - Font Size. 1.2
1 - 8 Reference: Re-installation of FL Solutions Program For re-installing this program, remove it first and then install it in the following procedure. Select Start - Settings - Control Panel. In the Control Panel window, double-click the Add/Remove Programs icon, and the dialog box shown in Fig. 1-5 will then appear. Fig. 1-5 In this dialog box, select “FL Solutions” and click the Add/ Remove button. When it is removed, click OK. Thereafter, install the FL Solutions program again. 1.2 Installing the FL Solutions Program
1 - 9 1.3 Startup of FL Solutions Program To start up the FL Solutions program, take the following procedure. (1) Connect the communication cable (standard accessory) between the fluorescence spectrophotometer and the personal computer. Then, turn on power to the spectrophotometer and the computer in this order. (2) The Windows 98 starts. (3) Click the Start button. Point to Programs and Hitachi Applications and select FL Solutions. (See Fig. 1-6.) Or double-click the FL Solutions program icon. If the FL Solutions program is set as a startup application in the Windows 98, the program automatically starts when turning on power supply of the personal computer. This setting can be selected at installation of the program. Fig. 1-6 Icon of FL Solutions program 1.3
1 - 10 (4) The fluorescence spectrophotometer connected to the communication COM port of computer is now recognized and the window shown in Fig. 1-7 is displayed. Then, the spectrophotometer is initialized. Fig. 1-7 (5) A gauge of initialization is displayed as shown in Fig. 1-8, so its progress can be visually checked. Fig. 1-8 1.3 Startup of FL Solutions Program
1 - 11 NOTE: If the spectrophotometer is not connected with the COM port or its power is not turned on, the monitor window (“F-2500 Fluorescence Spectrophotometer on COM1” indicated on its title bar) is not presented. In this status, functions for other than measurement are usable. For measurement, first confirm connection of the cable, and then turn on power supply of the spectrophotometer. Next, select the Instruments (I) command from the Tools (T) menu. When selecting and connecting the spectrophotometer, the window shown in Fig. 1-8 is presented. In execution of initialization, self-diagnosis is performed for the following items: • ROM • RAM • Wavelength drive system • Xenon lamp turn-on • EEPROM • Slit mechanism • Chopper • Backup battery voltage (only with Model F-4500) • Contents of wavelength and slit calibration values If an abnormality is found in the self-diagnosis, the relevant error message will be displayed. In this case, a corrective measure should be taken, referring to 4.7 of the maintenance edition. If necessary, notify our local repair service office. Fig. 1-9 1.3