Home > Hitachi > Spectrophotometer > Hitachi F 2500 Manual

Hitachi F 2500 Manual

    Download as PDF Print this page Share this page

    Have a look at the manual Hitachi F 2500 Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 913 Hitachi manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    							5 - 3 
    (3) General  
     
    The parameters shown in Fig. 5-3 are provided under the 
    General tab.   
     
     
     
    Fig. 5-3  General Tab 
     
     
    (a) Measurement (mode)  
    Select a measurement mode from among the four 
    listed below.   
     Wavelength scan  
     Time scan  
     Photometry  
     3-D Scan  
    Here, select “3-D Scan.”    For Wavelength scan, time 
    scan and Photometry, refer to Sections 2, 3 and 4.   
     
    (b) Operator  
    Enter an operator name.     
     
    (c) Instrument  
    The model of the connected instrument is indicated.   
     
    (d) Sampling 
     Standard  
    When sample is exchanged manually.   
     
    (e) Comments  
    Enter a description or notes on measuring conditions.   
     
    5.2  
    						
    							5 - 4 
    (f) Load  
    Click this button to read out saved analytical conditions.   
    The file opening window appears upon clicking, so 
    select an analytical conditions file.   
     
    (g) Save  
    By clicking this button after setting the parameters for 
    all five tabs, the analytical conditions are saved.   
    Clicking this button overwrites parameters on the 
    existing ones.   
     
    (h) Save As  
    Click this button for renaming the set conditions for 
    saving.    The “Save As” window will appear, so enter a 
    file name.   
     
    (4) Instrument  
     
    This window appears when 3-D Scan is selected for the 
    Measurement under the General tab of the Analysis 
    Method window.    When you click the Instrument tab, the 
    window in Fig. 5-4 appears.   
     
     
     
    Fig. 5-4  Instrument Tab 
    5.2    Creating an Analysis Method  
    						
    							5 - 5 
    (a) Data mode  
    Select either of the following data modes.   
    The Phosphorescence mode is for acquiring only 
    phosphorescence after extinction of excitation beam 
    while turning a chopper on the excitation side.     
     Fluorescence  
      Phosphorescence (with Model F-4500)   
     
    (b)  EX Start WL   
    This is the start wavelength for wavelength scan on the 
    excitation side.    Input the shorter wavelength side of 
    measuring wavelength range.   
     
    Model Input Range  
    F-4500  0, 200 to 850 nm 
    F-2500  0, 200 to 750 nm 
     
     
    (c)  EX End WL   
    This is the end wavelength for wavelength scan on the 
    excitation side.    Input the longer wavelength side of 
    measuring wavelength range.   
     
    Model Input Range  
    F-4500  0, 200 to 900 nm 
    F-2500  0, 200 to 800 nm 
     
     
    (d)  EX Sampling interval   
    In the 3-dimensional measurement mode, emission 
    spectrum measurement is repeated while shifting the 
    excitation wavelength.    Therefore, a shorter sampling 
    interval on the excitation side will result in a longer 
    measurement time.   
    Input range:    1 to 50 nm   
     
    (e)  EM Start WL   
    This is the start wavelength for wavelength scan on the 
    emission side.    Input the shorter wavelength side of 
    measuring wavelength range.     
     
    Model Input Range  
    F-4500  0, 200 to 850 nm 
    F-2500  0, 200 to 750 nm 
     
    5.2  
    						
    							5 - 6 
    (f)  EM End WL   
    This is the end wavelength for wavelength scan on the 
    emission side.    Input the longer wavelength side of 
    measuring wavelength range.     
     
    Model Input Range  
    F-4500  0, 200 to 900 nm 
    F-2500  0, 200 to 800 nm  
     
     
    (g)  EM Sampling interval   
    In the 3-D Scan mode, emission spectrum 
    measurement is repeated while shifting the excitation 
    wavelength.    Therefore, a shorter sampling interval on 
    the emission side will result in a longer measurement 
    time.  
    Input range:    1 to 10 nm   
     
    (h) Scan speed  
    Set a scan speed.  (Unit:  nm/min)  
     
    F-4500 F-2500 
    30000 3000 
    12000 1500 
    2400 300 
    1200 60 
    300 15 
    60  
    15  
     
     
    (i) EX Slit  
    Select a slit width for the excitation side.    Unit:    nm   
     
    F-4500 F-2500 
    1.0 2.5 
    2.5 5.0 
    5.0 10.0 
    10.0 20.0 
    20.0  
     
     
    5.2    Creating an Analysis Method  
    						
    							5 - 7 
    (j) EM Slit  
    Select a slit width for the emission side.    Unit:    nm   
     
    F-4500 F-2500 
    1.0 2.5 
    2.5 5.0 
    5.0 10.0 
    10.0 20.0 
    20.0  
     
     
    (k) PMT Voltage  
    A function for controlling the voltage of the 
    photomultiplier detector (photomultiplier tube).     
    Select one of the following values.    Changing from 
    400 V to 700 V increases the data value about two 
    digits.   
      950 V (only with F-4500)   
     700 V  
     400 V 
     
    (l) Response  
    Select a response time from the following.   
     
    F-4500 F-2500 
    0.04 0.04 
    0.01 0.08 
    0.05 0.4 
    0.1 2 
    0.5 8 
    2.0 Auto 
    8.0  
    Auto  
     
     
    By selecting Auto, a response time is set automatically 
    according to scan speed.    Select Auto usually.   
     
    5.2  
    						
    							5 - 8 
    (m) Corrected spectra  
    A function for determining the spectrum inherent to a 
    sample by correcting the photometer wavelength 
    characteristic using the saved instrument parameters, 
    following measurement with the instrument parameters 
    for photometer control.    If there are no instrument 
    parameters for the wavelength range to be measured, 
    only the wavelength range that allows correction is 
    scanned.    And if there is no wavelength area 
    measurable, an error will occur.     
    For using this function, a Spectra Correction Parts Set 
    (P/N 251-0126) is required.    The Instrument 
    Response command is provided in the Utility menu for 
    operating this function.    When this setting is at ON, 
    “Corrected spectra” appears on the monitor window.   
    (See Fig. 1-26 in 1.6.)   
     
    (n)  Shutter control (only with F-4500)   
    The shutter can be automatically closed in other than 
    measurement for suppressing sample deterioration 
    due to the energy of excitation beam and opened when 
    measurement starts.    When you put a check mark at 
    the head, the shutter will close and open at start of 
    measurement.    The shutter will close again when 
    measuring wavelength begins returning to the start 
    wavelength after measurement in the wavelength scan 
    mode.   
     
    5.2    Creating an Analysis Method  
    						
    							5 - 9 
    (5) Monitor  
     
    When selecting the Monitor tab, the window in Fig. 5-5 
    appears.   
     
     
     
    Fig. 5-5  Monitor Tab 
     
    Assign the following parameters.   
     
    (a)  Y-Axis Max (maximum)   
    Input a maximum value for Y-axis on the monitor 
    window.  
     
    (b)  Y-Axis Min (minimum)   
    Input a minimum value for Y-axis on the monitor 
    window.  
     
    (c) Contour interval  
    Function for setting intervals of drawing contour lines.   
    Input range:    0.01 to 1000   
     
    (d)  Open data processing window after data acquisition   
    Select whether or not to conduct data processing after 
    sample measurement.    When selected (check mark is 
    applied), an icon is displayed for the data processing 
    window at the end of measurement.    And by opening 
    this icon, data processing such as peak detection is 
    available.  
     
    5.2  
    						
    							5 - 10 
    NOTE:  When the remaining memory capacity of PC 
    in operation is very small (system resource is 
    less than 20%), the data processing window 
    will not open even if this item is turned ON.   
    Either close the presently opened data 
    processing window or restart the FL 
    Solutions program.   
     
    (e)  Print report after data acquisition   
    Select whether or not to print a report after sample 
    measurement.    When selected (check mark is 
    applied), printing will be made automatically after 
    measurement.    The items set under the Report tab 
    will be printed.    When this item is not selected, data 
    processing can be made by reading out a file via File-
    Open.  
     
    (6) Processing  
     
    On selection of the Processing tab, the window in Fig. 5-6 
    will be displayed.   
     
     
     
    Fig. 5-6  Processing Tab 
     
     
    (a) Peak Finding  
    Select a peak detection method indicated in the data 
    processing window.   
     
    5.2    Creating an Analysis Method  
    						
    							5 - 11 
    (b) Threshold  
    Set a detection limit for the photometric value axis of 
    peak and valley.    Peaks or valleys below the setting 
    will not be detected.    Input range is 0.001 to 1000.     
     
    (c) Sensitivity  
    Select the number of data points in the horizontal axis 
    direction (abscissa).    Select a sensitivity of “1” for 
    detecting sharp peaks.    Or select “8” for detecting 
    broad peaks.   
     1  
     2 
     4 
     8 
     
    (7) Report  
     
    Select the Report tab, and the window in Fig. 5-7 will 
    appear.  
     
     
     
    Fig. 5-7  Report Tab 
     
     
    (8) Output 
     
    (a) Print Report  
    A report will be printed.    For printing method, refer to 
    5.7.  
     
    (b)  Use Microsoft® Excel   
    Data will be transferred to Microsoft® Excel.     
    For transfer method, refer to 6.13.   
    5.2  
    						
    							5 - 12 
    (c)  Use print generator sheet   
    Select this method for using the report generator 
    (option).    For usage, refer to the instruction manual of 
    the report generator.   
     
    (9) Orientation  
    Selectable when Report is chosen for Print Method.   
    Select a printing orientation.    Either Portrait or Landscape 
    orientation is selectable.    When the Landscape orientation 
    is selected, date and method can be printed.   
     
    (a)  Printable (transferable) items   
    1) Include date  
    2) Include instrument 
    3) Include graph  
     Contour  
     Bird’s-eye view  
      Contour and bird’s-eye view 
      Contour and Spectrum   
    4)  Include peak table   
     
    (b) Printer Font  
    Click this button for changing the font to be used in the 
    report.  
    5.2    Creating an Analysis Method  
    						
    All Hitachi manuals Comments (0)