Hitachi F 2500 Manual
Have a look at the manual Hitachi F 2500 Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 913 Hitachi manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.
3 - 17 (4) When automatic saving is set, data will be saved after completion of measurement. Upon opening its file, a data processing window will appear as in Fig. 3-12. Although a peak table can be displayed, simultaneous display of rate calculation (kinetics) result and peak table cannot be selected. Fig. 3-12 shows a spectrum and the result of rate calculation. This is selectable from the View menu. Unless automatic saving is set, select the Save As command from the File menu and save data in file. Fig. 3-12 Data Processing Window (time scan) (a) Spectrum The spectrum of a measured sample is traced. (b) Result of rate calculation The results of rate calculation (gradient, activity, R, R2) are displayed. Start time and end time are settable by numeric input or clicking Set with the line cursor aligned. After setting, click Update. For rate calculation, refer to APPENDIX B. (c) Peak table Peak wavelength, valley wavelength, peak start/end wavelengths and peak area are displayed. (a) Spectrum (b) Result of rate calculation 3.4
3 - 18 3.5 Data Printout Printout can be provided even when the function for automatic printout after measurement is not used. The following two procedures are available. (1) Select the Print command from the File menu. Or click the (print) button on the toolbar. The field active on the data processing window (profiled in blue when clicked, field (a) in Fig. 3-12) will be printed. The contents, that are displayed by selecting the Print Preview command from the File menu as shown in Fig. 3-13, will be printed. Fig. 3-13 3.5 Data Printout
3 - 19 (2) Select the Report command in the Data menu. Or click the (report) button on the toolbar. The Print Preview window will appear as shown in Fig. 3-14. Then, click Print. Select the items to be printed at the Report tab under Properties on the Edit menu. Fig. 3-14 NOTE: If characters overflow from the print frame in repeat measurement or in overlaying, then change the printing orientation from Portrait to Landscape. 3.5
3 - 20 3.6 Calculating the Phosphorescence Life This is the function for calculating phosphorescence life from the measurement result (phosphorescence attenuation curve). Replacing data value at T1 with I 0, I0/e time (T2) is found, and T1, T2 and τ (phosphorescence life) = T2 - T1 are displayed. Select the phosphorescence life command from the data (D) menu or click the button on the Processing toolbar. A window as in Fig. 3-15 will appear. Fig. 3-15 Move the line cursor to the start point (T1) of calculation by the mouse or arrow keys [←] and [→] on the keyboard. Right-click the mouse at the start point. Or input a numeral in the T1 field and click Calculate. The result of calculation will be displayed. Fig. 3-16 3.6 Calculating the Phosphorescence Life
3 - 21 For printout of the calculation result, click Print. For printout of the spectrum and calculation result, apply a check mark to “With spectrum record” and click Print. For scale expansion, specify the area to be enlarged with the mouse. Draw a rectangle with the mouse button held down as shown in Fig. 3-17. Fig. 3-17 When releasing the mouse button, the specified area will be enlarged as shown in Fig. 3-18. Fig. 3-18 Double-clicking on the spectrum selects auto scale. 3.6
4 - 1 4. PHOTOMETRY 4.1 Flow of Operation This section explains the method of determining sample concentration with a calibration curve prepared. Following is the operational flow. Fig. 4-1 Operational Flow (photometry) End Start Method setting Click button. Photometric method Measuring wavelength Standard table Sample name/comment input Click button. Measurement Measurement of standard and sample Print (See page 4-2.) (See page 4-18.) (See page 4-19.) Click (measurement) button. Click button. (See page 4-24.) 4.1
4 - 2 4.2 Creating an Analysis Method Set the analytical conditions (method) prior to measurement. From the Edit menu, select the Method command or click the button. The dialog box shown in Fig. 4-2 will appear. Fig. 4-2 Analysis Method The following tabs are provided on the Analysis Method window for the photometry mode. These tabs are explained on the subsequent pages. • General • Quantitation • Instrument • Standards • Monitor • Report 4.2 Creating an Analysis Method
4 - 3 (1) General The parameters shown in Fig. 4-3 are provided under the General tab. Fig. 4-3 General Tab (a) Measurement (mode) Select a measurement mode from among the four listed below. • Wavelength scan • Time scan • Photometry • 3-D Scan (displayed when the ‘3-D Scan program’ (option) is installed in combination with Model F-2500) Here, select “Photometry.” For Wavelength scan, Time scan and 3-D Scan, refer to Sections 2, 3 and 5. (b) Operator Enter an operator name. (c) Instrument The model of the connected instrument is indicated. (d) Use sample table (Set measurement sample) Apply a check mark here when a sample table is used. 4.2
4 - 4 (e) Sampling • Standard When sample is exchanged manually. (f) Comments Enter a description or notes on measuring conditions. (g) Load Click this button to read out saved analytical conditions. The file opening window appears upon clicking, so select an analytical conditions file. (h) Save By clicking this button after setting the parameters for all six tabs, the analytical conditions are saved. Clicking this button displays the “Save As” window in creation of a new analytical conditions file. So input a file name. (i) Save As Click this button for renaming the set conditions for saving. (2) Quantitation Upon selecting the Quantitation tab, the window shown in Fig. 4-4 appears. Fig. 4-4 Quantitation Tab 4.2 Creating an Analysis Method
4 - 5 (a) Quantitation type Select the method of creating a calibration curve. Wavelength, Peak area, Peak height or Derivative method is selectable. Spectra measured by the Peak area, Peak height and Derivative method can be checked. 1) Wavelength Upon selecting “None” for calibration curve type in this quantitation method, up to 6 wavelengths can be set. When any other than “None” is selected for quantitation type, up to 3 wavelengths can be set. (i) One wavelength calculation This is the most generally used method. As shown in Fig. 4-5 (a), one wavelength is specified, and the concentration is determined from the data I 1 at that time. (ii) Two wavelength calculation As shown in Fig. 4-5 (b), assuming I 1 and I2 are the absorbance values at two different wavelengths, concentration is calculated through determination of I by the equation below. I = I 2 - I1 (iii) Three wavelength calculation As shown in Fig. 4-5 (c), assuming I 1, I2 and I 3 are the data values at three different wavelengths, concentration is calculated through determination of I by the equation below. Photometry is made in the order of 1, 2, 3. I = I 2 - But the following relation must hold: 1 < 2 < 3 or 1 > 2 > 3 1 - 3 1 - 2 × I 3 + 2 - 3 × I1 4.2