Appendix A Planned maintenance check form, Datex-Ohmeda S/5 FM1(4) Document no. M1187318-003 APPENDIX A: Planned maintenance check form, Datex-Ohmeda S/5 FM OK = Test OK N.A. = Test not applicable Fail = Test failed Customer Service Service engineer Date Measuring equipment / test gases used: Equipment / tool / gas: Manufacturer: Model/Type/Part Number: Serial Number / ID:Calibration Date: Monitor Installation L- E- N- F- N-N- N- N- K- Visual Inspection OK N.A. Fail 2.1.1. General 2.1.2. Extension module, N-F(C)(REC) S/N . MiniC unit 2.1.3. Recorder unit 2.1.4. E-PSM(P) Module S/N Notes
Datex-Ohmeda S/5 FM 2(4) Document no. M1187318-003 Functional Inspection OK N.A. Fail 2.2.1. General 2.2.2. Display 2.2.3. Keyboard(s) 2.2.4. N-FC(REC) Notes 2.2.5. Multiparameter Hemodynamic Modules . ECG and RESP measurements . Temperature measurement . Invasive blood pressure measurement . SpO2 measurement . Non invasive blood pressure measurement Notes 2.2.6. Memory 2.2.7. Recorder 2.2.8. Network 2.2.9. Wireless Network Option 2.2.10. Device Interfacing Solution, N-DISxxx S/N S/N S/N S/N Notes
Appendix A Planned maintenance check form, Datex-Ohmeda S/5 FM3(4) Document no. M1187318-003 2.2.11. General 1. Storing trend data 2. Service reset 3. Watchdog 4. Service Log reset 5. Electrical safety check 6. Save information into th e Maintenance service menu 7. Final cleaning Notes Used Spare Parts Notes Signature Functional Inspection OK N.A. Fail
Datex-Ohmeda S/5 FM 4(4) Document no. M1187318-003 For your notes:
Datex-Ohmeda S/5 TM FM Service Menu Conformity according to the Council Directive 93/42/EEC concerning Medical Devices amended by 2007/47/EC CAUTION: U.S. Federal law rest ricts this device to sale by or on the order of a li censed medical practitioner. Outside the USA, check local laws for any restriction that may apply. All specifications subject to change without notice. Document number M1187329-003 February 2, 2010 GE Healthcare Finland Oy Kuortaneenkatu 2 FI-00510 Helsinki, Finland Tel: +358 10 39411 Fax: +358 9 1463310 www.gehealthcare.com Copyright © 2010 General Electric Com pany. All rights reserved.
Table of contentsi Document no. M1187329-003 Table of contents Table of contents i Introduction1 1Frame 3 1.1 Memory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 1.2 Network . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 1.2.1 Network Status. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 1.2.2 Network Conf ig . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 1.2.3 Ethernet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 1.2.4 WLAN. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 1.2.5 Socket . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 1.3 MemCards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 1.3.1 Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 1.3.2 Communication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 1.3.3 PCMCIA Board. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 1.4 Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 1.4.1 WPM Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 2Display 19 3 Keyboard 20 3.1 Keyboard Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21 4 Parameters 22 4.1 Gas Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 4.1.1 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23 4.1.2 Gases . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24 4.2 ECG Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 4.2.1 ECG Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26 4.3 PSM STP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27 4.3.1 Calibrations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 4.4 NIBP Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30 4.4.1 NIBP Demo. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31 4.4.2 NIBP Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32 4.4.3 NIBP Safety Valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33 4.4.4 NIBP Pulse Valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34 4.4.5 NIBP Buttons/Leds . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35 4.4.6 NIBP Pneumatics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36 4.4.7 NIBP Watchdog . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37 4.5 DIS Interfacing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38 4.5.1 Interfacing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38 4.5.2 Status Page . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39 4.5.3 DIS Interfacing service menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39 4.5.4 DIS Module specif ic page . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Datex-Ohmeda S/5 FM ii Document no. M1187329-003 5 Set/Test41 5.1 Country Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42 6 Service Log 43 6.1 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44 6.1.1 Planned Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44 6.1.2 Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45 6.1.3 Upgrade . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45 7 Record Data 46 8SW Download 47 8.1 Remote Access . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Service Menu1 Document no. M1187329-003 Introduction The monitor has a Service Menu, which is a useful tool to examine monitor functions and to troubleshoot in case a fault occurs. Service Menu structure NOTE: The Service Menu structure can vary depending on the software version. Service Menu Frame Power SupplyWPM Battery Set/Test Keyboard ParametersGas Unit ECG STP Calibrations Safety valve Service Log Record Data MaintenancePlanned Maint. Repair Upgrade Country Settings Network MemCards Display NIBP demo Pulse valve Config Ethernet Pneumatics servmenu_struct09_FM.vsd Buttons/Leds Remote Access Watchdog General ECG Setup Gases Memory Keyboard Log DIS Interfacing NIBP Calibrations Network Status WLAN PCMCIA Board SW Download
Datex-Ohmeda S/5 FM 2 Document no. M1187329-003 Service Menu NOTE: The Service Menu pictures are for reference only. Details on the menu page can vary depending on the software version and the module type in use. 1. Press the Monitor Setup key. 2. Select Install/Service (password 16-4-34). 3. Select Service (password 26-23-8).