GE S5 Owners Manual
Have a look at the manual GE S5 Owners Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.
Patient Side Module, E-PSM, E-PSMP (Rev. 01) 35 Document no. M1215098-002 3.3 Disassembly and reassembly 3.3.1 Before disassembly WARNINGWear a grounded, antistatic wristband when handling PC boards. Electrostatic discharge may damage components on the board. Handle all PC boards by their edges. 3.3.2 Tools needed CAUTION When reassembling the module, make sure to reconnect all cables properly. 3.3.3 To disassemble the module - torx screwdrivers; T6, T8 - flat blade screwdriver -pincers - antistatic wristband 1. Remove the four screws (T8) holding the module cover to the frame from the bottom of the module. 2. Hold the cover from the back corners, lift it about 45° to unlock the snaps from the front unit and pull the cover out backwards. NOTE: Be careful not to damage the seal. When reassembling the seal may stick to the cover.
E-modules 36 Document no. M1215098-002 3. To remove the NIBP board:NOTE: You may remove the NIBP filter cover and the filter before disconnecting the flex cable. - Disconnect the module bus co nnector, pump connector and NIBP flex connector. - Disconnect the hoses (2 pcs) coming from the manifold. NOTE: Note the positions of the hoses; mark them if necessary to ensure they are replaced correctly. - Remove the NIBP board. 4. Disconnect the air intake hose from the NIBP manifold. NOTE: The lips of the insulator plates secure the module bus connectors. While reassembling the insulator plates, ensure that the connector secure lips support the connectors correctly.
Patient Side Module, E-PSM, E-PSMP (Rev. 01) 37 Document no. M1215098-002 5. Lift the NIBP-STP insulator plate carefully up. 6. To remove the STP board - Carefully open the connector lock and then disconnect the STP input flex cable from the STP board. NOTE: When reassembling, ensure that the flex cable is aligned properly and the connector is locked. - Lift the STP board a little to disconnect the module bus connector. Remove the STP board. 7. Remove the STP-ECG insulator plate. Be careful not to damage the NIBP hoses. 8. Hold down the ECG board. Carefully open the connector lock and then disconnect the ECG input flex cable from the ECG board. NOTE: When reassembling, ensure that the flex cable is aligned properly and the connector is locked.
E-modules 38 Document no. M1215098-002 Reassemble the module in reverse order. Always perform the “Service check” after reassembling the module. 9. To remove the ECG board - While holding the ECG input unit out of the way, lift the ECG board a little and disconnect the module bus connector. - Remove the ECG board. 10. Remove the NIBP filter cover and the filter. (If not removed already.) 11. Remove the four screws (T6) with washers holding the NIBP pump to the frame. 12. Flip the module over and remo ve the two (T6) screws holding the lock unit to the frame. While pulling the tab push the lockers with a screwdriver to remove the lock unit. 13. Carefully lift up the front uni t together with the NIBP pump.
Patient Side Module, E-PSM, E-PSMP (Rev. 01) 39 Document no. M1215098-002 To remove the pump unit 14. Disconnect the hoses from the manifold. The hoses follow the pump. 3.3.4 To remove the manifold unit 15. Disconnect the two (T6) screws holding the manifold to the front cover unit. 16. Open the connector lock from the NIBP flex board and disconnect the membrane keyboard flex.
E-modules 40 Document no. M1215098-002 17. Lift the manifold carefully aside. Be careful not to damage the NIBP flex board. Disconnect the NIBP flex board connector from the STP input board. NOTE: When reassembling, make sure that the NIBP flex board connecto r is connected properly (all pins connected) to the STP input board. Reassemble the module in reverse order. Always perform the “Service check” after reassembling the module.
Patient Side Module, E-PSM, E-PSMP (Rev. 01) 41 Document no. M1215098-002 Reassemble the module in reverse order. Always perform the “Service check” after reassembling the module. 3.3.5 To remove the module bus connector 18. Use a flat blade screwdriver to unlock the module bus connecto r insulator cover. Put the screwdriver in the hole and move the blade backwards (away from the flex cable) until the insulato r cover unlocks. 19. Pull the module bus connector carefully through the hole in the frame.
E-modules 42 Document no. M1215098-002 3.3.6 To replace the NIBP filter Reassemble the module in reverse order. Always perform the “Service check” after reassembling the module.1. Follow the disassemble instruction steps 1 and 2. 2. Remove the NIBP filter cover and replace the filter.
Patient Side Module, E-PSM, E-PSMP (Rev. 01) 43 Document no. M1215098-002 3.4 Adjustments and calibrations NOTE: Use only properly maintained, calibrated and traceable measurement equipment for the specified calibrations and adjustments to ensure accuracy. 3.4.1 NIBP calibrations The electronics of the NIBP pressure measurement is calibrated at the factory. The processor automatically maintains the zeroing pressure. If the zero point of the pressure transducer drifts more than specified, an error message is given and the NIBP board should be recalibrated or replaced. Recalibrate the NIBP measurement once a year. The checking and recalibration can be done in the NIBP service menu. The calibration of the primary pressure channel can also be checked from the NIBP setup menu ( NIBP - NIBP Setup - Calibration Check). In this case, the auto zeroing is performed at start - remove the hose before entering to ensure atmospheric pressure to the pressure transducers - the primary pressure is displayed. The zero-offset value should then be zero. Check the intake air filter as part of the calibration check. Change the filter if it is visibly dirty. Calibration check 1. Enter Calibration menu: Monitor Setup - Install/Service (password 16-4-34) - Service (26-23-8) - Parameters - NIBP - Calibrations 2. Select Calibration Check and push the ComWheel. 3. Connect an external precision manometer to the module. 4. Pump the following pressures to manometer and check the difference between the manometer and monitor pr essure display (The zeroing offset is automatically subtracted from the pressure readings). PRESTN_calib_menu.vsdPRESTN_calib_menu.vsd
E-modules 44 Document no. M1215098-002 Table 3 NIBP calibration check pressures If the error of pressure channel B1 is larger than specified above, the module should be recalibrated. The error of B2 is allowed to be even twice as large because it has no effect on blood pressure measurement accuracy. Howeve r, we recommend recalibrating the module when the error of B2 is larger than spec ified above to ensure best possible operation. Calibration 1. Enter Calibration menu. 2. Remove the hoses from the front pane l connector to enable proper zeroing. 3. Select Calibration. If it is not available, perform the steps a, b, and c. NOTE: Do not pull out the hemodynamic module from the monitor frame. The module must be in the frame during the whole procedure. a. Press the hemodynamic module buttons Auto ON/OFF and Start Cancel simultaneously for 3 seconds to enable the calibration. This enables menu selection Protection . The message ‘Calibration switch ON!’ is displayed. b. Select Protection OFF in the Calibration menu and push the ComWheel. c. Press the buttons again for 3 seconds. Menu selection Calibration is now enabled, and Protection is disabled. When the calibration is enabled, a message ‘Calibration not protected’ is displayed. •Start calibration by pushing the ComWheel . Messages ‘Zeroing’ and ‘Zeroed’ will be displayed in the NIBP message field. After this , a pressure bar and text ‘Calibrating’ will be displayed. •Connect an external mercury manometer with a pump to the module through the both tubes of the hose - both transducers B1 and B2 must be calibrated simultaneously. Pump up to a pressure of about 200 mmHg according to the manometer. Calibration is possible in the range of 150 to 250 mmHg. •Verify that both pressure values in the prom pt field match the manometer reading. If not, adjust by turning the ComWheel. When the values of the pressure bar and the manometer are equal, push the ComWheel to confirm the calibration. The message ‘Calibrated’ will be displayed on the NIBP digit field after a few seconds, which means that the calibration succeeded, and the new calibration data is saved in EEPROM. NOTE: When calibrating NIBP, always change th e displayed pressure value slightly with the ComWheel, even in cases where the value woul d be correct. For example, change the value one step higher and then back one step lower. ‘Calibrated’ text should appear in the display. This ensures that the calibration procedure is correctly registered and stored by the module. •To set the protection on: Press NIBP module buttons Auto ON/OFF and Start Cancel simultaneously for 3 seconds. Select Protection ON and push the ComWheel. Then press the buttons again for three seconds. Pressure Max. error Example 0 mmHg ±5 mmHg (=zero offset) -1 100 mmHg 100 ±2 mmHg 100 ±2 200 mmHg 200 ±3 mmHg 200 ±3 260 mmHg 260 +/-3 mmHg 200 ±3