GE S5 Owners Manual
Have a look at the manual GE S5 Owners Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.
S/5 Extension Module for FM, N-FC, N-FCREC, N-FREC 21 Document no. M1187338-003 11. Remove the CO2 unit module bus flex cable from the REC board. NOTE: Be careful with the in sulation plate covering the cable connector on the REC board. 12. Lift the CO2 unit carefully up and remove the unit from the module cover. 13. Remove the screw (T8) holding the module bus flex cable to recorder unit metal plate and module cover. 14. Lift the REC board carefully up and disconnect the two module bus flex cables from the REC board. Remove the REC board with the recorder unit metal plate from the module cover. 15. Remove the four pozidrive screws holding the REC board to the recorder unit metal plate.
Datex-Ohmeda S/5 FM 22 Document no. M1187338-003 16. Use a flat blade screwdriver to unlock the module bus connector insulator cover. Put the screwdriver in the hole and move the blade backwards (away from the module bus flex cable) until the insulator cover unlocks. 17. By assisting with a screwdriver pull the module bus connector through the hole in the module cover. 18. Flip the module cover over and remove the two (T6) screws holding the lock un it to the frame. While pulling the tab push the lo ckers with a screwdriver to remove the lock unit.
S/5 Extension Module for FM, N-FC, N-FCREC, N-FREC 23 Document no. M1187338-003 3.4 Adjustments and calibrations 3.4.1 Calibrating The airway module should be calibrated once every six months or whenever there are indications of errors in the gas readings. Calibrate the gas measurement with the Datex-Ohmeda calibration gas. Do not use any other calibration gases. −Use the regulator 755534-HEL. − Use gas 755580 only and set the FiO2 le vel in the CO2 setup menu to 21-40% Airway Gas - CO2 setup. .O 2 concentration to 20%. Use only recommended calibration gases to guarantee a successful calibration. During gas calibration, % units are always used for CO 2 regardless of selected measuring units. NOTE: Ensure that the calibration gas and regulator are functioning properly before calibration. Perform the annual maintenance on the regulator as required. Figure 8 Attaching regulato r to the calibration can 1. Attach the regulator to the gas container. 2. Attach a new sampling line to the water trap. Connect the loose end of the sampling line to the regulator on the gas container.
Datex-Ohmeda S/5 FM 24 Document no. M1187338-003 Figure 9 Connecting sampling line to the gas valve and feeding gas 3. For maximum accuracy, let the monitor warm up for 30 minutes. The menu item Gas calibration remains gray as long as the message ‘Calibrating gas sensor’ is displayed. 4. Press the Airway Gas key and select Gas calibration . 5. Wait until the ‘Zero ok’ and then the ‘Feed gas’ messages appear after each gas on the screen. 6. Open the regulator and feed calibration gas until the message ‘Adjust’ appears, then close the valve. 7. Check that the displayed gas value matches the value on the calibration gas container. NOTE: Adjust the O 2 percentage according to the calibration gas (for 755580 the right O2 value is 20%). NOTE: If an error occurs during calibration or if no gas is fed, the text ‘Calibr. error’ appears. Push the ComWheel to perform a new calibration. If adjustments are required: •Turn the ComWheel to highlight the first gas to be adjusted and then push the ComWheel. Airway Gas
S/5 Extension Module for FM, N-FC, N-FCREC, N-FREC 25 Document no. M1187338-003 •Turn the ComWheel until the displayed value matches the desired value in the gas bottle and push it again. If the message ‘Zero error’ is displayed, press the Normal Screen key and repeat the calibration procedure. The time of the last calibration is shown at the bottom of the menu page. 3.4.2 Gas sampling system adjustment For flow rate measurements, a flow meter with a low flow resistance and the capability to measure low flow rates is required. A sampling line of no rmal length has to be connected to the monitor as it has a considerable effect on the flow. 3.4.3 Flow rate measurement If any flow rates are not correct, first replac e the Mini D-Fend water trap, then recheck the flows. The sampling flow rate is measured by a flow meter at the sampling line. The flow rate should be between 135 and 165 ml/min. The flow rate is adjusted in the Gases service menu with Sample gain adj . 3.4.4 Flow rate adjustment NOTE: Before adjusting the sampling flow, make sure there is no leakage in the sampling system. Refer to chapter 3.2 Service check, step 13. Gas calibration . Wait until the Sample Flow value is back to near 150 ml/min. Connect a flow meter to the 3 meter sampling line and check that the flow (the flow meter reading) is within the following range: Sampling flow (ml/min) 135...165 If necessary, readjust the sampling flow: Select Sample gain adj from the menu. To increase the sampling flow, turn the ComWheel counterclockwise. To decrease the flow, turn the ComWheel clockwise.
Datex-Ohmeda S/5 FM 26 Document no. M1187338-003 A change of 0.050 in the Gain value changes the flow approximately 7.5 ml/min. After you have changed the gain, wait until the Sample Flow value on the screen returns near to the original, then check the flow meter reading again. 3.4.5 Gas calibration Gas calibration is performed in the Airway Gas menu. Calibration gas regulator flow check Interval: every 12 months Regulator flow specification: REF 755583-HEL & 755534-HEL: 260 – 410 ml /min at 1-10 bar cylinder pressure REF 755530-HEL: 260 – 410ml/min at 5-7psi cylinder pressure Tools needed: calibration gas ca n, regulator, piece of silicon hose and flow meter. GE Healthcare recommends use of TSI 4140 Flow Meter. Insert the calibration gas regu lator on the gas cylinder. Connect a silicon hose between the regulator and the flow meter. Block the regulato r overflow port and open the regulator. Check the flow rate from the flow meter and verify that the flow is within the specification.
S/5 Extension Module for FM, N-FC, N-FCREC, N-FREC 27 Document no. M1187338-003 4Troubleshooting 4.1 Troubleshooting chart for recorder unit 4.2 Troubleshooting chart for CO 2 measurement Problem CauseWhat to do Module not responding to front panel key, but operates through Recorder menu. Membrane switch cable loose or broken. Check the cable. Replace the front panel if necessary. Recorder will not start. No error messages shown. Recorder: Module not properly inserted.Re-insert the module properly. Recorder: Flex-strip cable broken. Check the cable. Replace if necessary. Recorder board faulty. Replace the recorder board. Recorder unit faulty. Replace the recorder unit. Recorder works but nothing appears on the paper. Active side of the paper downwards. Turn the paper roll over. To test which side is active: Place the paper on a hard surface and draw a line with a fingernail - a dark line will appear on the active (thermal) side. Recorder unit faulty. Replace the recorder unit. Problem Cause/What to do No response to breathing Sampling line or water trap blocked or loose, or improperly attached. Water trap container full. See the gas sampling system troubleshooting. ‘SENSOR INOP.’ message The temperature is too low or high, check the temperature in the service menu. Supply voltage is too low or high, IR source current or voltage is too low or high, check current in the service menu. Pump is not working properly, check sample flow and pump voltage in the service menu. Ambient pressure too low or high, check the ambient pressure in the service menu. Zero valve not working properly, check the functionality by switching zero valve on and off in the service menu. ‘ZEROING ERROR’ message Gas zeroing failed. Condensation or residual gases are affecting the zero measurement. Allow the module to run drawing room air for half an hour and calibrate again. ‘CHECK D-FEND’ message Amb – Work pressure difference too small. Probably water trap or the sampling line is not attached properly. Gas zero valve failure. Pump failure or gas outlet blockage.
Datex-Ohmeda S/5 FM 28 Document no. M1187338-003 4.2.1 CO2 measurement ‘REPLACE D-FEND’ message Amb – Work pressure difference too big.Indicates residue build-up on the water trap membrane. This decreases air flow. Replace the D-fend. ‘SAMPLE LINE BLOCKED’ message Amb – Work pressure difference too big. Sampling line or water trap is occluded. Water trap container is full. If occlusion persists, check internal tubing for blockages. No response to any gas Check Sample Gas Out. Amb - Work pressure low, flow too small and pump voltage too high. Sampling line, water trap, or internal tubing is blocked or loose, or improperly attached. Gas out connector or tubing is blocked. Zero valve malfunction. Pump failure or pump is worn. Supply voltage missing. Serial communication error. Sudden increase in gas display Water trap malfunction. Check all internal tubing and the interior of the water trap for occlusions or leaks. Replace water trap. Check flow rates. Abnormally high (or abnormally low response to CO 2 or sudden occlusion warning Pressure transducer failure. Check the Ambient and Amb – Work pressures in Gases service menu. Strong drift in all gases Leak in sampling line or internal tubing (especially in conjunction with too low readings). Problem Cause What to do Action too low ETCO 2 value •sudden decrease in circulation •pulmonary embolism •hyperventilation •very large dead-space •large shunting •leak in sampling system •calibration error •high by-pass flow from ventilator •check all connections •check calibration too high ETCO 2•hypoventilation •increased metabolism •D-fend contaminated •calibration error •change D-fend •check calibration waveform clipped •incorrect scaling•change scale Problem Cause/What to do
S/5 Extension Module for FM, N-FC, N-FCREC, N-FREC 29 Document no. M1187338-003 4.3 Gas sampling system troubleshooting The faults which can occur in the sampling system are: leaks or blockages in the tubing, failure of the sampling pump or the magnetic valves, or diminishing of the flow rates because of dirt accumulating in the internal tubing. The following checks should help in localizing the fault. Whenever suspecting the sampling system and always after having done any work on the sampling system, check and if necessary adjust the flow rate. CAUTION The special internal sa mple tube is mechanically fragile. Sharp bends will cause leaks. NOTE: D-fend water trap should be replaced, when the ‘REPLACE D-FEND’ message appears during the monitor startup. NOTE: If any liquid has entered the miniC measuring unit due to water trap filter failure, leave the module running without a sa mpling line for several hours and check the functions after it has dried out. 4.3.1 Sampling system leak test 1. Connect a flow cassette with a high flow resistance value (50/1.1) to the end of the sampling line and start following the ‘Amb-Wor k’ value in the service menu. When the value exceeds 130 mmHg, connect the other port of the flow cassette to the sample gas out connector and switch off the pump. 2. Wait until the pressu re inside the sampling system is stabilized, then observe the shown Amb-Work value. The value, i.e. the pressure inside the sampling system, should not drop more than 6 mmHg in 20 seconds. 3. If the pressure drops more, first ensure the connections you have made and repeat the test. no response to breathing •apnea •(disconnection) •sampling line or water trap loose or blocked (air leak) •check all connections •sample gas outlet blocked•check that outlet is open ETCO 2 over scale >20% Shown until 32%, specified range 0...20%•abnormally high ETCO2 (permissive hypercapnia) •CO2 sensor contaminated •D-fend malfunction •let the module run without a sampling line until the CO 2 sensor has dried out •change D-fend ETCO 2>PaCO2•unit is mmHg or kPa and ETCO2 is close to arterial PCO 2 •“dry gas” as default •change to “wet gas” by using install/service menu Problem Cause What to do
Datex-Ohmeda S/5 FM 30 Document no. M1187338-003 4.4 MiniC unit troubleshooting WARNINGThe miniC measuring unit can only be repaired and calibrated at the factory. Attempts to repair/calibrate the unit elsewhere will adversely affect operation of the unit. In case of failure, the complete miniC unit should be sent to GE Heal thcare for factory exchange. 4.5 Error messages Message Explanation Occlusion or Sample Line Blocked The sample tube inside or outside the monitor is blocked or water trap is occluded. If occlusion persists, measured gas values disappear. Continuous occlusion. Check sample line and D-fend. Occlusion over 40 seconds. Check D-fend - The water trap is not connected - There is a leak in the sampling line inside the module. If air leak persists, measured gas values disappear. Check sample gas out. (Air leak detected.) Check water trap and sample gas out-flow. Press normal screen to continue. Air leak over 40 seconds. Replace D-fend (replace water trap) Indicates residue build-up on the water trap membrane. This decreases air flow. Gas calibration is not available during first 5 minutes/during occlusion/during air leak Entering calibration is not allowe d during 5 minutes after power up and during occlusion or air leak. Gas out blocked - Gas out connector on the front panel, or the exhaust line connected to it, is blocked. - If the sample gas is returned to th e patient circuit, the filter in the return kit may be occluded. - Make sure the sample gas outlet is connected to an open scavenging system only where gas is removed in room pressure. Recalibration Time out, fluctuating gases, gain adjusted “over”. Zero error Unsuccessful zeroing. Unstable, Calibr error Unsuccessful calibration. Menu messages during calibration : Zero error Unsuccessful zeroing. Adjust Calibration gas accepted and monitor is ready for adjusting the gas values to match the calibration gas concentration. Unstable Unsuccessful calibration.