Frame for FM 49 Document no. M1187335-004 3.7 Adjustments and calibrations It is not necessary to perform calibrations or adjustments on S/5 Frame for FM.
Datex-Ohmeda S/5 FM 50 Document no. M1187335-004 4Troubleshooting 4.1 S/5 Frame for FM Table 6 S/5 Frame for FM troubleshooting chart ProblemCause What to do Monitor is not starting. 1. The batteries are empty. 2. Fuses may be blown. 3. If power cord connected, AC/DC unit may be faulty. 4. If AC/DC unit is working, the DC/ DC board may be faulty. 5. On/Stby key may be faulty. 1. Connect the power cord. 2. Replace fuses. 3. Replace the AC/DC power unit. 4. Replace the DC/DC unit. 5. Replace the command board panel keyboard foil. Monitor is not starting. 1. The connection between DC/DC board and CPU board may be faulty. 2. Faulty CPU board. 1. Check the connector and the interconnection board. 2. Replace the CPU board. The monitor starts (alarm leds are lit and a beep is heard), but the display remains black. 1. The LCD display cables are loose. 2. The backlights are not lit. 1. Check the LCD display connection board connectors. 2. Check backlight connector. Check inverter cable. Backlights may be faulty. Replace the backlights. Backlight Inverter may be faulty. Replace the Inverter board. Display and monitor operating but alarm leds not illuminating in start-up. 1. Loose connectors. 2. Ribbon cable faulty. 3. CPU not working. 1. Check all connectors from CPU to led board. 2. Replace ribbon cable. 3. Replace CPU. Display and monitor operating but no audible beep in start-up. 1. Loudspeaker connector or wires loose or faulty. 1. Check loudspeaker connector and wires. Display is too dim. 1. Incorrect brightness adjustment. 2. Backlight faulty. 3. Backlight inverter faulty. 1. Adjust display brightness higher. 2. Replace backlight. 3. Replace backlight inverter. Stripes or white areas on screen. 1. Loose faulty display connection flexboard connectors in CPU and display.1. Check display connection flexboard connectors in CPU and display. Module data disappears from the screen. ‘Module power supply overload’ message. Parameter module current (in module bus) too high. Detach and change parameter module. Module data disappears. 1. Module bus voltage or signals path broken.1. DC/DC board module power section may be faulty. Replace the DC/DC board. 2. Replace the Module bus cable unit.
Frame for FM 51 Document no. M1187335-004 4.2 Memory option Table 3 Memory option troubleshooting chart ‘Battery failure’ message on the screen. 1. Problem in communication between battery and DC/DC board. 2. Battery too old or defected. 3. DC/DC board may be faulty. 1. Replace battery. 2. Replace battery. 3. Replace the DC/DC board. Frame temperature high message. The temperature inside the frame is too high. Check monitor ventilation holes. ‘Battery temperature high’ message on the screen. Battery SMBus temperature is too high. Check monitor ventilation holes. Replace battery. After shut off, the trend and patient data does not remain 15 min in the memory. 1. The On/Stby switch has been pressed over 10 s. (= service reset). 2. Empty battery. 3. Only one battery inserted. 1. When switching off, the On/Stby switch has to be pressed less than 10 s. 2. Replace the battery. 3. Use always two batteries. Keyboard not working, but module communication is OK. Keyboard cables and connectors, or interconnection board connectors may be faulty. Check the interconnection board connectors. Check the keyboard cable connection to the user interface board. Keyboard not working, and module communication not working. UPI section of the CPU board not functioning normally. Restart the monitor. Replace the Central Processing Board. Keyboard partly not working. 1. Keyboard faulty. 2. Keyboard cables and connectors, or interconnection board connectors may be faulty. 1. Replace the keyboard foil. 2. Check the interconnection board connectors. Check the keyboard cable connection to the user interface board. Error message Cause What to do ‘No menus in Menu Card’ message There are no menus in the Menu card. Insert a Menu card with valid menu configuration file s in the module. ‘Faulty Data Card - change card’ An error has occurred during Data card read/write operation. Change Data card. ‘Faulty Menu Card - change card’ An error has occurred during Menu card read/write operation. Change Menu card. Problem Cause What to do
Datex-Ohmeda S/5 FM 52 Document no. M1187335-004 4.3 NET section Table 4 NET section troubleshooting chart ProblemCause What to do Monitor does not connect to the network. Monitor connects to the network, but disconnects unexpectedly (‘Network connection down’ message on the monitor screen). Patch panel Patch cable not connected to HUB or to panel. Patch cable Patch cable or connector defective. HUB not connected to power supply. HUB port closed due to physical layer problems. HUB port temporarily closed and reopened due to physical layer problems. Hubs not properly connected to each other. Monitor-Network cable Cable not pro perly connected to the wallplate or to the monitor. Cable or connector defective. Network cable (inside the monitor) defective Replace the Network cable. Network Connector Board defective Replace the Network Connector Board Unit. Net section of the CPU board The NET section is defective. Replace the CPU board. NET section memory on the CPU board The SRAM of the NET section is defective or uninitialized. The NET cannot be used. See network service page for details. Wrong identification number Use the Virtual ID number with monitors that have WLAN option installed. Identification plug There is no identification plug attached to the monitor. The identification plug is defective or uninitialized. The plug cannot be used. Multi I/O adapter faulty Replace multi I/O adapter.
Frame for FM 53 Document no. M1187335-004 ‘Check network connectors’ message shows on the monitor screen Monitor-Network cable Cable not pro perly connected to the wallplate or to the monitor. Cable or connector defective. Network cable (inside the monitor) defective Replace the Network cable. Network Connector Board defective Replace the Network Connector Board Unit. Multi I/O adapter faulty Replace multi I/O adapter. Identification plug There is no identification plug properly attached to the monitor. The identification plug is defective or uninitialized. The plug cannot be used. See network service page for details. ‘Network board error’ message shows on the monitor screen NET section on the CPU board NET section memory on the CPU board The NET section is defective. The board cannot be used. See network service page for details. The SRAM memory of the NET section is defective or uninitialized. The board cannot be used. See network service page for details. Other Site View shows no waveforms No waveforms are set up for Monitor-to-Monitor communication Run Network Setup to veri fy current Monitor-to- Monitor communication setup. Network printing fails Print server is busy Print queue is full Printer is off-line Network managers print server is busy at the moment and cannot take more print jobs. Try again after 15 seconds. There are too many unprinted documents waiting in the print queue. Check the printer, as it is not operating properly. Printer cable is loose, printer is out of paper, there is a paper jam or the printer is simply switched to off-line state. Record keeper menus are blank There are no menus for the record keeper Run Network Setup to verify the current setup. Problem Cause What to do
Datex-Ohmeda S/5 FM 54 Document no. M1187335-004 For your notes:
Appendix A Service check form, Datex-Ohmeda S/5 Frame for FM, F-FM 1(4) Document no. M1187335-004 APPENDIX A: Service check form, Datex-Ohmeda S/5 Frame for FM, F-FM OK = Test OK N.A. = Test not applicable Fail = Test failed Customer Service Module type S/N Service engineer Date Monitor Installation L- E- F-N- N-N-N-N- N- General OK N.A. Fail OK N.A. Fail 1. Internal parts 2. External parts 3. Module installation 4. Command Board LEDs 5. Start-up 6. Module recognition 7. LCD display picture 8. Time and date 9. Loudspeaker 10. Display brightness 11. Monitor software 12. Content of the service log 13. Voltages 14. Watchdog circuitry 15. Alarm LEDs 16. Membrane keys 17. ComWheel 18. Module communication 19. Batteries 20. Battery operation 21. Battery charging
Datex-Ohmeda S/5 FM 2(4) Document no. M1187335-004 Recorder unit OK N.A. Fail 22. Recorder messages 23. Recording 24. Paper speed 25. Quality of recording Notes MemCard Option S/N OK N.A. Fail OK N.A. Fail 26. Module recognition 27. Memories and PCMCIA controller 28. Communication 29. Data Card recognition 30. Menu Card recognition 31. Menu card function Notes Network Option S/N OK N.A. Fail OK N.A. Fail 32. Connection to network 33. Ethernet address 34. Data error counters 35. Hardware error counters 36. Disconnection recognition 37. ID-plug recognition Notes
Appendix A Service check form, Datex-Ohmeda S/5 Frame for FM, F-FM 3(4) Document no. M1187335-004 Wireless LAN option S/N OK N.A. Fail OK N.A. Fail 38. Signal strength 39. Configure access point 40. Connection to Network Notes General OK N.A. Fail OK N.A. Fail 41. Trends retaining 42. Battery capacity 43. Service Log reset 44. Electrical safety check 45. Final cleaning 46. Fill in all necessary documents Notes Used spare parts Signature
Datex-Ohmeda S/5 FM 4(4) Document no. M1187335-004 For your notes: