    							Frame for FM 49
    Document no. M1187335-004
    3.7 Adjustments and calibrations
    It is not necessary to perform calibrations or adjustments on S/5 Frame for FM. 
    						
    							Datex-Ohmeda S/5 FM
    50
    Document no. M1187335-004
    4Troubleshooting
    4.1 S/5 Frame for FM
    Table 6 S/5 Frame for FM troubleshooting chart
    ProblemCause What to do
    Monitor is not starting. 1. The batteries are empty. 2. Fuses may be blown.
    3. If power cord connected, AC/DC unit may be faulty.
    4. If AC/DC unit is working, the DC/ DC board may be faulty.
    5. On/Stby key may be faulty. 1. Connect the power cord.
    2. Replace fuses.
    3. Replace the AC/DC power unit.
    4. Replace the DC/DC unit.
    5. Replace the command board 
    panel keyboard foil.
    Monitor is not starting. 1. The connection between DC/DC  board and CPU board may be 
    faulty.
    2. Faulty CPU board. 1. Check the connector and the 
    interconnection board.
    2. Replace the CPU board.
    The monitor starts (alarm leds 
    are lit and a beep is heard), but 
    the display remains black. 1. The LCD display cables are loose.
    2. The backlights are not lit.  1. Check the LCD display connection 
    board connectors.
    2. Check backlight connector. Check  inverter cable. 
    Backlights may be faulty. Replace the 
    backlights.
    Backlight Inverter may be faulty. 
    Replace the Inverter board.
    Display and monitor operating 
    but alarm leds not illuminating 
    in start-up. 1. Loose connectors.
    2. Ribbon cable faulty.
    3. CPU not working. 1. Check all connectors from CPU to 
    led board.
    2. Replace ribbon cable.
    3. Replace CPU.
    Display and monitor operating 
    but no audible beep in start-up. 1. Loudspeaker connector or wires 
    loose or faulty. 1. Check loudspeaker connector and 
    wires.
    Display is too dim. 1. Incorrect brightness adjustment.
    2. Backlight faulty.
    3. Backlight inverter faulty. 1. Adjust display brightness higher.
    2. Replace backlight.
    3. Replace backlight inverter.
    Stripes or white areas on screen. 1. Loose faulty display connection  flexboard connectors in CPU and 
    display.1. Check display connection 
    flexboard connectors in CPU and 
    display.
    Module data disappears from 
    the screen. ‘Module power 
    supply overload’ message. Parameter module current (in 
    module bus) too high.
    Detach and change parameter 
    module.
    Module data disappears. 1. Module bus voltage or signals  path broken.1. DC/DC board module power 
    section may be faulty. Replace the 
    DC/DC board.
    2. Replace the Module bus cable  unit. 
    						
    							Frame for FM 51
    Document no. M1187335-004
    4.2 Memory option
    Table 3 Memory option troubleshooting chart
    ‘Battery failure’ message on the 
    screen. 1. Problem in communication 
    between battery and DC/DC 
    board.
    2. Battery too old or defected.
    3. DC/DC board may be faulty. 1. Replace battery.
    2. Replace battery.
    3. Replace the DC/DC board. 
    Frame temperature high 
    message. The temperature inside the frame is 
    too high. Check monitor ventilation holes.
    ‘Battery temperature high’ 
    message on the screen. Battery SMBus temperature is too 
    high. Check monitor ventilation holes. 
    Replace battery.
    After shut off, the trend and 
    patient data does not remain 15 
    min in the memory. 1. The On/Stby switch has been 
    pressed over 10 s. (= service 
    reset).
    2. Empty battery.
    3. Only one battery inserted. 1. When switching off, the On/Stby 
    switch has to be pressed less than 
    10 s. 
    2. Replace the battery.
    3. Use always two batteries.
    Keyboard not working, but 
    module communication is OK. Keyboard cables and connectors, or 
    interconnection board connectors 
    may be faulty. Check the interconnection board 
    connectors. 
    Check the keyboard cable connection 
    to the user interface board.
    Keyboard not working, and 
    module communication not 
    working. UPI section of the CPU board not 
    functioning normally.
    Restart the monitor.
    Replace the Central Processing 
    Board.
    Keyboard partly not working. 1. Keyboard faulty. 2. Keyboard cables and connectors, or interconnection board 
    connectors may be faulty. 1. Replace the keyboard foil.
    2. Check the interconnection board 
    connectors. 
    Check the keyboard cable connection 
    to the user interface board.
    Error message Cause What to do
    ‘No menus in Menu Card’ 
    message There are no menus in the Menu 
    card. Insert a Menu card with valid menu 
    configuration file
    s in the module.
    ‘Faulty Data Card - change card’ An error has occurred during Data  card read/write operation. Change Data card.
    ‘Faulty Menu Card - change card’ An error has occurred during Menu  card read/write operation. Change Menu card.
    Problem Cause What to do 
    						
    							Datex-Ohmeda S/5 FM
    52
    Document no. M1187335-004
    4.3 NET section
    Table 4 NET section troubleshooting chart
    ProblemCause What to do
    Monitor does not connect to 
    the network.
    Monitor connects to the 
    network, but disconnects 
    unexpectedly (‘Network 
    connection down’ message 
    on the monitor screen). Patch panel 
    Patch cable not connected to HUB or to panel.
    Patch cable  Patch cable or connector defective.
    HUB not connected to power supply.
    HUB port closed due to physical layer problems.
    HUB port temporarily closed and reopened due 
    to physical layer problems.
    Hubs not properly connected to each other.
    Monitor-Network cable Cable not pro perly connected to the wallplate 
    or to the monitor.
    Cable or connector defective.
    Network cable (inside the 
    monitor) defective Replace the Network cable.
    Network Connector Board 
    defective Replace the Network Connector Board Unit.
    Net section of the CPU board The NET section is defective. Replace the CPU  board.
    NET section memory on the 
    CPU board The SRAM of the NET section is defective or 
    uninitialized. The NET cannot be used. See 
    network service page for details.
    Wrong identification number Use the Virtual ID number with monitors that  have WLAN option installed.
    Identification plug There is no identification plug attached to the 
    monitor.
    The identification plug is defective or 
    uninitialized. The plug cannot be used.
    Multi I/O adapter faulty Replace multi I/O adapter. 
    						
    							Frame for FM 53
    Document no. M1187335-004
    ‘Check network connectors’ 
    message shows on the 
    monitor screen Monitor-Network cable Cable not pro
    perly connected to the wallplate 
    or to the monitor.
    Cable or connector defective.
    Network cable (inside the 
    monitor) defective Replace the Network cable.
    Network Connector Board 
    defective Replace the Network Connector Board Unit.
    Multi I/O adapter faulty Replace multi I/O adapter.
    Identification plug There is no identification plug properly attached  to the monitor.
    The identification plug is defective or 
    uninitialized. The plug cannot be used. See 
    network service page for details.
    ‘Network board error’ 
    message shows on the 
    monitor screen NET section on the CPU board
    NET section memory on the 
    CPU board The NET section is defective. The board cannot 
    be used. See network service page for details.
    The SRAM memory of the NET section is 
    defective or uninitialized. The board cannot be 
    used. See network service page for details.
    Other Site View shows no 
    waveforms  No waveforms are set up for 
    Monitor-to-Monitor 
    communication Run Network Setup to veri
    fy current Monitor-to-
    Monitor communication setup.
    Network printing fails Print server is busy
    Print queue is full
    Printer is off-line Network managers print server is busy at the 
    moment and cannot take more print jobs. Try 
    again after 15 seconds.
    There are too many unprinted documents 
    waiting in the print queue. Check the printer, as 
    it is not operating properly.
    Printer cable is loose, printer is out of paper, 
    there is a paper jam or the printer is simply 
    switched to off-line state.
    Record keeper menus are 
    blank There are no menus for the 
    record keeper Run Network Setup to verify the current setup.
    Problem Cause What to do 
    						
    							Datex-Ohmeda S/5 FM
    54
    Document no. M1187335-004
    For your notes: 
    						
    							Appendix A Service check form, Datex-Ohmeda S/5 Frame for FM, F-FM
    1(4)
                               Document no. M1187335-004
    APPENDIX A: Service check form, 
    Datex-Ohmeda S/5 Frame for FM, F-FM 
    OK = Test OK            N.A. = Test not applicable            Fail = Test failed
    Customer
     
    Service Module type
    S/N
    Service engineer
     Date
     Monitor Installation
     
    L- E- F-N-
    N-N-N-N-
    N-
     General
         OK           N.A.          Fail      OK           N.A.          Fail
    1. Internal parts                            2. External parts
                               
    3. Module installation                            4. Command Board LEDs
                               
    5. Start-up                            6. Module recognition
                               
    7. LCD display picture                            8. Time and date
                               
    9. Loudspeaker                            10. Display brightness
                               
    11. Monitor software                            12. Content of the service 
    log
                               
    13. Voltages                            14. Watchdog circuitry
                               
    15. Alarm LEDs                            16. Membrane keys
                               
    17. ComWheel                            18. Module 
    communication
                               
    19. Batteries                            20. Battery operation
                               
    21. Battery charging                             
    						
    							Datex-Ohmeda S/5 FM
    2(4)
    Document no. M1187335-004
    Recorder unit
         OK           N.A.          Fail
    22. Recorder messages                            23. Recording
                               
    24. Paper speed                            25. Quality of recording
                               
    Notes
     
     
    MemCard Option S/N
           OK           N.A.          Fail      OK           N.A.          Fail
    26. Module recognition                            27. Memories and PCMCIA 
    controller
                               
    28. Communication                            29. Data Card recognition
                               
    30. Menu Card recognition                            31. Menu card function
                               
                                                           
    Notes
     
     
     
    Network Option S/N
           OK           N.A.          Fail      OK           N.A.          Fail
    32. Connection to network                            33. Ethernet address
                               
    34. Data error counters                            35. Hardware error 
    counters
                               
    36. Disconnection 
    recognition                            37. ID-plug recognition
                               
                                                           
    Notes
      
    						
    							Appendix A Service check form, Datex-Ohmeda S/5 Frame for FM, F-FM
    3(4)
                               Document no. M1187335-004
    Wireless LAN option  S/N
          OK           N.A.          Fail      OK           N.A.          Fail
    38. Signal strength                            39. Configure access point
                               
    40. Connection to 
    Network                                                        
    Notes
     
     
    General 
     
          OK           N.A.          Fail      OK           N.A.          Fail
    41. Trends retaining                            42. Battery capacity
                               
    43. Service Log reset                            44. Electrical safety check
                               
    45. Final cleaning                            46. Fill in all necessary 
    documents
                               
    Notes
     
     
    Used spare parts 
    Signature   
    						
    							Datex-Ohmeda S/5 FM
    4(4)
    Document no. M1187335-004
    For your notes: 
    						