Device Interfacing Solution7 Document No. M1187344-003 2.4 Connectors and signals 2.4.1 Male bus cable connector 2.5 Interfaced devices See the following table of DIS modules and devi ces that you can interface with the Device Interfacing Solution. For specific information on parameters transf erred from the interfaced device to the Datex- Ohmeda Monitor and the applicable software versio ns of the device, refer to the Installation guide accompanying. Pin No. Signal Color 1Data from UPI + brown 2 Data from UPI - red 3 VDD 9 V to 18 V (max 1 A) orange 4 GND yellow 5 VCC 7 V to 8 V (max 1 A) blue 6GND grey 7Data to UPI + white 8Data to UPI - black Device Monitors N-DISOXIM3 Oximetrix 3a a Trademark of Hospira In c. (previously trademark of Abbott Laboratories) N-DISQVUE QVue /Q2 a N-DISVIGIL Baxter-Vigilance b b Trademark of Edwards Li fesciences Corporation Device Blood gas analyzers N-DISOPT Opti CCAa a Trademark of Osmetech plc 12 345 76 8

Datex-Ohmeda S/5 FM 8 Document no. Document No. M1187344-003 3 Service procedures 3.1 General service information The field service of the Device Interfacing Solution is limited to replacing faulty cables or mechanical parts. The circuit boards should be returned to GE Healthcare for repair. GE Healthcare is always available for service advice. Please provide the unit serial number, full type designation, and a detailed fault description. WARNING Only trained personnel with appropriate equipment should perform the tests and repairs outlined in this section. Unauthor ized service may void the warranty of the unit. 3.2 Service check These instructions include complete procedures for a service check. The service check is mandatory after any service repair. However, th e service check procedures can also be used for determining possible failures. The procedures should be performed in ascending order. The instructions incl ude a check form (“APPENDIX A:”) which may be used when performing the procedures. The symbol in the instructions indicate s that the check form contains space to record the results of the particular procedure. 3.2.1 Recommended tools 3.2.2 Recommended parts No recommended parts. 3.2.3 Visual inspection Disconnect the DIS module from the DIS bus and from the interfaced external device. 1. Internal check •Disassemble the DIS module. •Make sure that there are no loose parts inside the DIS module. •Check that the screws holding the PC board are tightened properly. •Check that the cables are attached properly and the connectors are intact. Tool Order No. Notes Screwdrivers

Device Interfacing Solution9 Document No. M1187344-003 2. External check •Check that the DIS module case and label are clean and intact. •Reassemble the DIS module. •Check that the screws for the DIS module case are secured properly. •Check that the bus cable connector is intact. •Check that the DIS bus and device specific cables are intact. 3.2.4 Functional inspection 3. DIS module interface statusConnect the DIS module to the DIS bus and to the external device that is specified in the DIS module label. Turn on the interfaced external device. Check that no error messages are displayed on the monitor screen. Check via the Interfacing menu that the connected DIS module status is correct: Monitor Setup - Interfacing - Status Page Check that the waveforms and numeric data are transferred to the monitor according to the configuration. 4. Recognition of interface Disconnect the DIS bus cable and check that the ‘[device name] module removed’ message appears on the monitor screen. Reconnect the cable. Turn off the external device (if possible) and check that the ‘[device name] disconnected from module’ message appears on the screen. Turn the external device back on again. 5. DIS module service menuEnter the service menu: Monitor Setup – Install/Service (password 16-4-34) – Service (password 26-23-8) – Parameters – DIS Interfacing Check that the menu displays submenus for all connected DIS modules. Enter the corresponding DIS module service menu and check that the displayed information corresponds to the info rmation on the DIS module labels. Check that the DIS bus voltage is between 13.80...16.00 V. Check that the DIS module time-out and checksum error values do not increase more than by 5 per second. Check that the status of each DIS module memory indicates OK.

Datex-Ohmeda S/5 FM 10 Document no. Document No. M1187344-003 6. Electrical safety checkPerform the electrical safety test and leakage current test. 7. Functioning after electrical safety checkCheck that the DIS module func tions normally after the tests. 8. Final cleaning Clean the DIS module, bus cable and device specific cable with a suitable detergent. •Fill up all the necessary documents.

Device Interfacing Solution11 Document No. M1187344-003 3.3 Disassembly and reassembly 3.3.1 Before disassembly WARNINGWear a grounded, antistatic wristband when handling PC boards. Electrostatic discharge may damage components on the board. Handle all PC boards by their edges. 3.3.2 Tools needed 3.4 To disassemble the module To disassemble the N-DISxxx module (see the ex ploded view of the module in the “S/5 FM Spare parts” slot): 1. Remove the two screws fr om the back of the module. 2. Loosen the two strain-relief threads and af ter that disconnect the cables from the connectors. To remove the Module Front Cover from the mo dule, release the snaps that hold the front cover to the front chassis. To reassemble the module, reverse the order of the disassembly steps. CAUTION When reassembling the module, make sure to reconnect all cables properly. Always perform the “Service check” after reassembling the module. - screwdriver - flat blade screwdriver - antistatic wristband

Datex-Ohmeda S/5 FM 12 Document no. Document No. M1187344-003 4Troubleshooting 4.1 LED indicators Figure 5 LED indicators 4.1.1 Green LED The green LED indicates that the communication between the monitor and DIS module and the communication between the DIS module and external device is working properly. When all cables are connected and the connected devices are on, the green LED should be lit continuously. 4.1.2 Yellow LED The yellow LED alerts the user. The yellow LED is lit when any of the following conditions becomes true: 1. The DIS module is connected to the DIS bus, but the external device is not connected. 2. The external device is in power off state. 3. The external device is not selected from the interfacing menu as an active source of data. NOTE: The meaning of the yellow LED varies with some external devices. See the “Installation Guide” delivered with the DIS module. Yellow LEDGreen LED

Device Interfacing Solution13 Document No. M1187344-003 4.2 Quick functional check You have two ways for checking the functi on of the Device Interfacing Solution: •Press the Monitor Setup key, select Interfacing and open the Status Page menu. The status page shows you the current communica tion status of the interfacing modules connected to the bus (1...4 pcs). NOTE: The status message ‘Connected’ appears on the monitor screen after you have connected the external device to the DIS modu le and turned it on, if the monitor and DIS module have alread y been initialized. •Check the LED indicators on the DIS module (the green LED indicates the physical connections, the yellow LED software selections): GREEN YELLOW INDICATION lit dark Physical connections between the monitor, DIS module and external device are in order and the device has been selected in the menu. dark lit There is something wrong with the physical connections between the monitor, interf acing module and external device. The external device has not been selected in the menu. lit lit Physical connections between the monitor, DIS module and external device are in order, but the external device has not been selected in the menu (see the “User’s Reference Manual” of the monitor). dark dark The DIS module is not connected to the monitor.

Datex-Ohmeda S/5 FM 14 Document no. Document No. M1187344-003 5 Earlier revisions This manual supports the previous version of N-DISxx.

Appendix A, Service check form, DeviceInterfacing Solution, N-DISxxx (Rev. 01) A-1(2) Document no. M1187344-003 APPENDIX A: Service check form, Device Interfacing Solution, N-DISxxx (Rev. 01) OK = Test OK N.A. = Test not applicable Fail = Test failed Customer Service Module type S/N Service engineer Date Visual inspection OK N.A. Fail OK N.A. Fail 1. Internal check 2. External check Notes Functional inspection 3. DIS module interface status 4. Recognition of interface 5. DIS module service menu 6. Electrical safety check 7. Functioning after electrical safety check 8. Final cleaning Notes Used spare parts Signature

Datex-Ohmeda S/5 FM A-2(2) Document no. M1187344-003 For your notes: