    							Device Interfacing Solution7
     Document No. M1187344-003
    2.4 Connectors and signals
    2.4.1 Male bus cable connector 
     
    2.5 Interfaced devices
    See the following table of DIS modules and devi ces that you can interface with the Device 
    Interfacing Solution.
    For specific information on parameters transf erred from the interfaced device to the Datex-
    Ohmeda Monitor and the applicable software versio ns of the device, refer to the Installation 
    guide accompanying.   Pin No. Signal Color
    1Data from UPI + brown
    2 Data from UPI - red
    3 VDD 9 V to 18 V (max 1 A) orange
    4 GND yellow
    5 VCC 7 V to 8 V (max 1 A) blue
    6GND grey
    7Data to UPI + white
    8Data to UPI - black
    Device Monitors
    N-DISOXIM3 Oximetrix 3a
    a Trademark of Hospira In c. (previously trademark of Abbott Laboratories)
    N-DISQVUE
    QVue /Q2
    a                           
    N-DISVIGIL Baxter-Vigilance
    b 
    b Trademark of Edwards Li fesciences Corporation
    Device Blood gas analyzers
    N-DISOPT
    Opti CCAa
    a Trademark of Osmetech plc 
    12
    345
    76
    8 
    						
    							Datex-Ohmeda S/5 FM
    8
    Document no. Document No. M1187344-003
    3 Service procedures
    3.1 General service information
    The field service of the Device Interfacing Solution is limited to replacing faulty cables or 
    mechanical parts. The circuit boards should be returned to GE Healthcare for repair.
    GE Healthcare is always available for service advice. Please provide the unit serial number, full 
    type designation, and a detailed fault description. 
    WARNING Only trained personnel with appropriate equipment should perform the 
    tests and repairs outlined in this section. Unauthor ized service may void the 
    warranty of the unit.
    3.2 Service check
    These instructions include complete procedures for a service check. The service check is 
    mandatory after any service repair. However, th e service check procedures can also be used 
    for determining possible failures.
    The procedures should be performed in ascending order.
    The instructions incl ude a check form (“APPENDIX A:”) which may be used when performing the 
    procedures. 
    The symbol 
    in the instructions indicate s that the check form contains space to record the 
    results of the particular procedure.
    3.2.1 Recommended tools
    3.2.2 Recommended parts
    No recommended parts.
    3.2.3 Visual inspection
    Disconnect the DIS module from the DIS bus and from the interfaced external device.
    1. Internal check
    •Disassemble the DIS module.
    •Make sure that there are no loose parts inside the DIS module.
    •Check that the screws holding the PC board are tightened properly. 
    •Check that the cables are attached properly and the connectors are intact.
    
    Tool Order No. Notes
    Screwdrivers 
    						
    							Device Interfacing Solution9
     Document No. M1187344-003
    2. External check
    •Check that the DIS module case and label are clean and intact.
    •Reassemble the DIS module.
    •Check that the screws for the DIS module case are secured properly.
    •Check that the bus cable connector is intact.
    •Check that the DIS bus and device specific cables are intact.
    
    3.2.4 Functional inspection
    3. DIS module interface statusConnect the DIS module to the DIS bus and to the external device that is specified in the 
    DIS module label. Turn on the interfaced external device.
    Check that no error messages are displayed on the monitor screen.
    Check via the  Interfacing menu that the connected DIS module status is correct:
    Monitor Setup - Interfacing - Status Page 
    Check that the waveforms and numeric data are transferred to the monitor according to 
    the configuration.
    
    4. Recognition of interface Disconnect the DIS bus cable and check that the ‘[device name] module removed’ 
    message appears on the monitor screen. Reconnect the cable.
    Turn off the external device (if possible) and check that the ‘[device name] disconnected 
    from module’ message appears on the screen. Turn the external device back on again.
    
    5. DIS module service menuEnter the service menu:
    Monitor Setup – Install/Service   (password 16-4-34)  – 
    Service    (password 26-23-8)  – 
    Parameters – DIS Interfacing
    Check that the menu displays submenus for all connected DIS modules.
    Enter the corresponding DIS module service menu and check that the displayed 
    information corresponds to the info rmation on the DIS module labels.
    Check that the DIS bus voltage is between 13.80...16.00 V.
    Check that the DIS module time-out and checksum error values do not increase more 
    than by 5 per second.
    Check that the status of each DIS module memory indicates OK.
     
    						
    							Datex-Ohmeda S/5 FM
    10
    Document no. Document No. M1187344-003
    6. Electrical safety checkPerform the electrical safety test and leakage current test. 
    
    7. Functioning after electrical safety checkCheck that the DIS module func tions normally after the tests.
    
    8. Final cleaning
    Clean the DIS module, bus cable and device  specific cable with a suitable detergent.
    
    •Fill up all the necessary documents. 
    						
    							Device Interfacing Solution11
     Document No. M1187344-003
    3.3 Disassembly and reassembly
    3.3.1 Before disassembly
    WARNINGWear a grounded, antistatic wristband when handling PC boards. 
    Electrostatic discharge may damage components on the board. 
    Handle all PC boards by their edges.
    3.3.2 Tools needed
    3.4 To disassemble the module
    To disassemble the N-DISxxx module (see the ex ploded view of the module in the “S/5 FM 
    Spare parts” slot):
    1. Remove the two screws fr om the back of the module. 
    2. Loosen the two strain-relief threads and af ter that disconnect the cables from the 
    connectors.
    To remove the Module Front Cover from the mo dule, release the snaps that hold the front 
    cover to the front chassis.  
    To reassemble the module, reverse the order of the disassembly steps.
    CAUTION When reassembling the module, make  sure to reconnect all cables properly.
    Always perform the “Service check” after reassembling the module.
    - screwdriver
    - flat blade screwdriver
    - antistatic wristband 
    						
    							Datex-Ohmeda S/5 FM
    12
    Document no. Document No. M1187344-003
    4Troubleshooting
    4.1 LED indicators
    Figure 5 LED indicators
    4.1.1 Green LED
    The  green LED indicates that the communication between the monitor and DIS module and 
    the communication between the DIS module and external device is working properly. When all 
    cables are connected and the connected devices are on, the green LED should be lit 
    continuously.
    4.1.2 Yellow LED
    The yellow LED alerts the user. The yellow LED is lit when any of the  following conditions 
    becomes true:
    1. The DIS module is connected to the DIS bus,  but the external device is not connected.
    2. The external device is in power off state.
    3. The external device is not selected from the interfacing menu as an active source of data.
    NOTE: The meaning of the yellow LED varies with some external devices. See the “Installation 
    Guide” delivered with the DIS module.
    Yellow LEDGreen LED 
    						
    							Device Interfacing Solution13
     Document No. M1187344-003
    4.2 Quick functional check
    You have two ways for checking the functi on of the Device Interfacing Solution:
    •Press the Monitor Setup key, select Interfacing  and open the  Status Page menu. The 
    status page shows you the current communica tion status of the interfacing modules 
    connected to the bus (1...4 pcs). 
    NOTE: The status message ‘Connected’ appears on the monitor screen after you have 
    connected the external device to the DIS modu le and turned it on, if the monitor and DIS 
    module have alread y been initialized.
    •Check the LED indicators on the DIS module (the green LED indicates the physical 
    connections, the yellow LED software selections):
    GREEN YELLOW INDICATION
    lit             dark          Physical connections between the monitor, DIS module 
    and external device are in order and the device has been 
    selected in the menu.
    dark       lit            There is something wrong with the physical connections 
    between the monitor, interf acing module and external 
    device. The external device has not been selected in the 
    menu.
    lit              lit             Physical connections between the monitor, DIS module 
    and external device are in order, but the external device 
    has not been selected in the menu (see the “User’s 
    Reference Manual” of the monitor).
    dark         dark         The DIS module is not connected to the monitor. 
    						
    							Datex-Ohmeda S/5 FM
    14
    Document no. Document No. M1187344-003
    5 Earlier revisions
    This manual supports the previous version of N-DISxx. 
    						
    							Appendix A, Service check form, DeviceInterfacing Solution, N-DISxxx (Rev. 01)
    A-1(2)
                               Document no. M1187344-003
    APPENDIX A: Service check form, 
    Device Interfacing Solution, N-DISxxx (Rev. 01)
    OK = Test OK            N.A. = Test not applicable            Fail = Test failed
    Customer
     
    Service Module type
    S/N
    Service engineer
     Date
      Visual inspection
           OK           N.A.          Fail      OK           N.A.          Fail
    1. Internal check                             2. External check
                               
    Notes
     
      Functional inspection         
    3. DIS module interface 
    status                             4. Recognition 
    of interface
                               
    5. DIS module service 
    menu                             6. Electrical safety check
                               
    7. Functioning after 
    electrical safety check                             8. Final cleaning
                               
    Notes
     
    Used spare parts 
    Signature  
    						
    							Datex-Ohmeda S/5 FM
    A-2(2)
    Document no. M1187344-003
    For your notes: 
    						
    All GE manuals Comments (0)