Home > GE > Medical Monitor > GE S5 Owners Manual

GE S5 Owners Manual

    Download as PDF Print this page Share this page

    Have a look at the manual GE S5 Owners Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    							Patient Side Module, E-PSM, E-PSMP (Rev. 01) 25
    Document no. M1215098-002
    3 Service procedures
    3.1 General service information
    The field service of the hemodynamic modules is  limited to replacing faulty printed circuit 
    boards or mechanical parts. The circuit boards sh ould be returned to GE Healthcare for repair.
    GE Healthcare is always available for service ad vice. Please provide the unit serial number, full 
    type designation, and a detailed fault description.
    WARNING Only trained personnel with appropriate equipment should perform the 
    tests and repairs outlined  in this section. Unauthorized se rvice may void 
    warranty of the unit.
    3.2 Service check
    These instructions include complete procedures for a service check. The service check is 
    mandatory after any service repair. However, th e service check procedures can also be used 
    for determining possible failures.
    The procedures should be performed in ascending order.
    The instructions incl ude a check form (“APPENDIX A:”) which may be used when performing the 
    procedures. 
    The symbol 
    in the instructions indicate s that the check form contains space to record the 
    results of the particular procedure.
    3.2.1 Recommended tools
    NOTE: Use only properly maintained, calibrated and traceable measurement equipment for the 
    specified calibrations and adjustments to ensure accuracy.
    Table 13 Recommended tools
    Tool Order No. For product(s)
    Hemodynamic patient simulator M1010831 E-PSM(P)
    Adapter cables for simulators
    - Dual temperature adapter cable 2016998 -001 Hemodynamic patient simulator 
    and Medsim
    - Dual Inv.BP adapter cable 2005772- 001 Hemodynamic patient simulator
    - Temperature adapter cable M1010832 Medsim
    - Inv.BP adapter cable M1010858 Medsim
    - Temperature adapter cable M1010846 Lionheart & MPS450
    - Inv.BP adapter cable M1010862 Lionheart & MPS450
    Pressure manometer E-PSMP
    Accessories: 
    Temperature test set 884515-HEL E-PSM(P) 
    						
    							E-modules
    26
    Document no. M1215098-002
    3.2.2 Recommended parts
    Table 14 Recommended parts
    3.2.3 Visual inspection
    Detach the module cover by removing the four screws from the bottom of the module. 
    Check:
    1. Internal parts−screws are tightened properly
    − connectors are connected properly
    − NIBP tubing is attached properly
    − there are no loose objects inside the module
    
    2. External parts
    −the front cover and the front panel sticker are intact
    Multi-Link ECG accessories, IEC:
    - Multi-link 3-leadwire set 412682-003 E-PSM(P)
    - Multi-link 5-leadwire set 412681-003 E-PSM(P)
    - Multi-link 5-leadwire set, C2-C6 416467-004 E-PSM(P)
    - Multi-link 12-lead ECG trunk 
    cable 416035-002 E-PSM(P)
    Multi-Link ECG accessories, AHA:
    - Multi-link 3-leadwire set 412682-001 E-PSM(P)
    - Multi-link 5-leadwire set 416681-001 E-PSM(P)
    - Multi-link 5-leadwire set, V2-V6 416467-003 E-PSM(P)
    - Multi-link 12-lead ECG trunk  cable 416035-001 E-PSM(P)
    SpO
    2 finger probe OXY-F-UN E-PSM(P)
    SpO
    2 Interconnect Cable OXY-ES3 E-PSM(P)
    InvBP transducer 70077-001 E-PSMP
    Adult NIBP cuff hose with cuff ID 2021285-001 E-PSM(P)
    Adult NIBP cuff  2753EE-PSM(P)
    Infant cuff hose without cuff ID 414874-001 E-PSM(P)
    Screwdriver
    Part Order No. Notes
    E-PSM(P), Air Filter, FRU M1221481 Replace every 3 years
    Tool
    Order No. For product(s) 
    						
    							Patient Side Module, E-PSM, E-PSMP (Rev. 01) 27
    Document no. M1215098-002
    −all connectors are intact and attached properly
    − the module box and latch are intact
    
    3.2.4 Functional inspection
    3. NIBP pump filter
    Replace the NIBP pump filter, if necessary.
    
    Reattach the module cover and check that the latch is moving properly.
    Switch the monitor on and wait until the  monitoring screen appears. Configure the 
    monitor screen so that all the needed para meters are shown, for example as follows:
    Monitor Setup - Screen 1 Setup  - Waveform Fields - Field 1 - ECG1
    Field 2 - ECG2
    Field 3 - P1
    Field 4 - P2
    Field 5 - Pleth
    Field 6 - Resp
    Digit Fields - Lower Field 2 - NIBP
    Lower Field 3 - T1+T2
    4. Module installation Plug in the module. Check that it goes in smoothly and locks up properly.
    
    5. Module recognitionCheck that the module is recognized, i.e. all the needed parameter information, except 
    invasive blood pressure, starts to show on the screen.
    
    Preset ECG, Respiration, InvBP and SpO2 measurement settings:
    ECG - ECG Setup - Hr Source - Auto
    Pacemaker - Show
    Others - Resp Setup - Size - 1.0Resp Rate Source - Auto
    Measurement - On
    Detection Limit - Auto
    Invasive Pressures - P1 ‘Art’ Setup - Label - ArtP2 ‘Cvp’ Setup - Label - Cvp
    PulseOximetry - Pleth Scale - Auto 
    						
    							E-modules
    28
    Document no. M1215098-002
    ECG measurement
    6. Module software (serial numbers)
    Enter the service menu:
    Monitor Setup -  Install/Service  (password 16-4-34) - 
    Service  (password 26-23-8)
    Take down the information regarding the module software by selecting  Scroll Vers and 
    turning the ComWheel.
    
    7. Communication and memories
    Enter the   Parameters -  ECG  service menu.
    Check that the Time-outs, Bad checksums and Bad c-s by mod values are not increasing 
    faster than by 5 per second. Check also th at the ECG/RESP board memories have passed 
    the internal memory test, i.e. the RAM, ROM and EEPROM state all OK.
    
    8. Power frequency
    Check that the power frequency value is set according to the current mains power 
    frequency. Change the setting by selecting  Power Freq, if necessary.
    
    9. Cable recognition
    Connect a 12-lead ECG trunk cable without a lead set to the module. Check that the 
    message ‘Leads off’ is displayed on the screen.
    
    10. Lead detectionConnect both 5-leadwire sets to the trunk  cable. Connect all the leads together, for 
    example to a suitable screwdriver. Check th at all the electrodes show ON and the 
    message ‘Asystole’ appears. Check that the Cable type shows 10 lead.
    Connect the 10-leadwire set to the simulator. Disconnect one of the leads and check that 
    the corresponding electrode in the service  menu shows OFF within 10 seconds of the 
    disconnection, and then reconnect the lead. Check the rest of the leads using the same 
    method. Disconnect the trunk cable.
    Connect a 3-leadwire set to a trunk cable and  connect it to the module. Connect all the 
    leads together, for example to a suitable screwdriver. Check that the cable type shows 3 
    lead.
    NOTE: When any of the limb le ads is disconnected, the measurement will automatically 
    change to 3 electrode ECG measurement.
    NOTE: The asystole and different leads off  messages are shown using certain priority. 
    Even though one of the leads is disconnected, the related leads off message may not 
    appear on the screen.
    NOTE: When RA, LA, LL or RL electrode is di sconnected, all six V electrodes show OFF.
     
    						
    							Patient Side Module, E-PSM, E-PSMP (Rev. 01) 29
    Document no. M1215098-002
    11. Test with the patient simulatorConnect the leads to a patient simulator.
    Perform the settings and checks with Dynatech Nevada MedSim 300 Patient Simulator:
    ECG - BASE - BPM - 160
               PACE - WAVE - NSR
    Check that a normal ECG waveform is shown, the HR value is 160 (±5) and the ‘Pacer 
    count’ value is not increas ing in the service menu.
    ECG - PACE - WAVE - ASNC
    Check that pacemaker spikes are shown on  the ECG waveform, the HR value changes to 
    75 (±5) and the Pacer count valu e is increasing according to the shown pacemaker spikes.
    Set the pacemaker option off:
    ECG - PACE - WAVE - NSR
    
    Respiration measurement
    12. RESP measurement recognition Check that Resp Available and RESP Measurement both show ON in the ESTP: ECG service 
    menu.
    
    13. Test with patient simulatorCheck the respiration measurement with a patient simulator.
    The settings and checks with Dynatech Nevada MedSim 300 Patient Simulator:
    Simulator Cover:
    BASELINE IMPEDANCE -switch
     - 500
    LEAD SELECT-switch
     - II/RL-LL
    Simulator Menu:
    RESP - WAVE - NORM
    RATE - 20
    OHMS - 1.0
    RATIO - 1/1
    APNEA - OFF
    SHIFT - OFF
    Check that the RESP waveform is shown and the RR value is 20 (±5). Change the position 
    of the BASELINE IMPEDANCE switch and check that appropriate RESP waveform and RR 
    value are shown again within 30 seconds.
    RESP - APNEA - 32 S
    Check that the monitor activates the APNEA alarm.
    NOTE: Make sure that only the ECG leads are connected to the simulator during the apnea 
    test. If other cables are connected at the same time, the respiration signal from the 
    simulator may be disturbed, and therefore, the APNEA alarm may not be activated.
    NOTE: When you have the ECG service menu open, spikes will appear on the respiration 
    waveform. These spikes represent the threshold level for detecting inspiration and 
    expiration.
     
    						
    							E-modules
    30
    Document no. M1215098-002
    Temperature measurement
    14. Communication and memoriesEnter the ESTP: STP service menu:
    Parameters - ESTP : STP
    Check that the Time-outs, Bad checksums and Bad c-s by mod values do not increase 
    faster than by 5 per second. Check also that the STP board memories have passed the 
    internal memory test, i.e. the  RAM, ROM and EEPROM show all OK.
    
    15. Temperature probe detection
    Check that the ‘Cable’ and ‘Probe’ show OFF for both channels, T1 and T2, when no 
    probes are connected.
    Connect the temperature adapter cable to  the module temperature connector and a 
    temperature test plug to the adapter cable. Check that the Cable and Probe for T1 show 
    ON and the corresponding temperature value appears on the monitor screen.
    Perform the same check also for the channel T2.
    
    16. Calibration check
    Check the temperature calibrations using temperature test plugs.
    If the deviation on a temperature reading on the screen is more than 0.1°C, calibrate the 
    temperature channels according to the instructions in chapter “Temperature calibration” 
    on page 45.”
    
    17. Temp test Activate the temperature test by selecting  Temp Test from the menu and pressing the 
    ComWheel twice. When the message ‘Performing temp test’ disappears from the digit 
    field, check that no error messages appear and Temp error shows OFF for both channels 
    in the service menu.
    
    18. Module configuration 
    Check that the module configuration has been  set correctly. The configuration in use is 
    shown beside the text Configuration in the service menu and it can be either STP or ST. 
    Change the configuration in the  Calibrations - Set Config menu, if necessary. To activate 
    the change, reset the module communicatio n by removing and inserting the module.
     
    						
    							Patient Side Module, E-PSM, E-PSMP (Rev. 01) 31
    Document no. M1215098-002
    Invasive blood pressure measurement
    19. Membrane keysCheck the front panel membrane keys that are related to the InvBP measurement.
    Press each of the keys for at least one second. Check that the pressed key is identified, i.e. 
    one of the texts for Buttons changes fr om OFF to ON in the service menu.
    
    20. Cable and transducer detection
    Check that the Cable and Probe for P1 show  OFF. Connect the InvBP adapter cable to the 
    module, connect a cable with an invasive blood pressure transducer to the adapter cable 
    and check that the Cable and Probe show ON and the corresponding pressure waveform 
    appears on the screen.
    Perform the same check also for the InvBP channel P2.
    
    21. Calibration
    Calibrate the InvBP channels P1 and P2 according to the instructions in chapter “Invasive 
    pressure calibration” on page 45.“
    
    22. Test with patient simulatorCheck the InvBP channels with a patient simulator.
    The settings and checks with Dynatech Nevada MedSim 300 Patient Simulator:
    SENSITIVITY - switch
     - 5 µV/V/mmHg
    ECG - BASE - BPM - 60 - BP - 1 - WAVE - ATM  2 - WAVE - ATM
    Restore the normal monitoring screen by pressing the key 
    Normal Screen.
    Connect cables from the channels BP1 and BP 2 to the module connectors. Zero the InvBP 
    channels by pressing the keys ZERO P1 and ZERO P2 on the module front panel.
    BP - 1 - WAVE - ART
             2 - WAVE - CVP
    Check that appropriate InvBP waveforms are shown and the InvBP values are 
    approximately 120/80 (±3 mmHg) for the ch annel P1 and 15/10 (±2 mmHg) for the 
    channel P2.
    Check that the HR value is calculated from P1, when ECG is not measured (ECG cable 
    disconnected).
    
    SpO2 measurement
    23. SpO2 probe detection
    Check that the message ‘No probe’ is shown, when no SpO
    2 sensor is connected to the 
    module. Connect an SpO
    2 finger probe to the module (wit h the interconnection cable, if 
    needed). Check that the message ‘Probe off’ is  shown when the probe is not connected to 
    a finger.
     
    						
    							E-modules
    32
    Document no. M1215098-002
    24. Test measurementConnect the SpO
    2 probe onto your finger. Check that the reading of 95-99 and SpO2 
    waveform appears. Check that the HR value is calculated from SpO
    2 when ECG and InvBP 
    (P1) are not measured.
    
    Non Invasive Blood Pressure measurement
    25. Communication and memories
    Enter the NIBP module service menu:
    Parameters - NIBP
    Check that the Time-outs, Bad checksums and Bad c-s by mod values are not increasing 
    faster than by 5 per second. Check also that the NIBP board memories have passed the 
    internal memory test, i.e. the  RAM, ROM and EEPROM show all OK.
    
    26. Membrane keys
    Check the front panel membrane keys.
    Select  Buttons/Leds .
    Press each of the two NIBP related membrane keys for at least one second. Check that 
    the pressed key is identified, i.e. the corresponding text changes from OFF to ON in the 
    menu, when the key is released back up again.
    
    27. Pump and valves Check the pump and valves.
    Select  Pneumatics  from the NIBP menu. Connect a pressure manometer to the NIBP 
    module cuff connector.
    Select  Start Pump  and press the ComWheel. Check that the pump turns on and the 
    pressure inside the tubing system starts to increase. Stop the pump by pressing the 
    ComWheel again when the pressure reaches 280 mmHg.
    Select  Open Exh2 . Press the ComWheel and check that the pressure inside the tubing 
    system starts to drop, then press the ComWheel again. If necessary, turn the pump on 
    again for a moment to increase the pressure inside the tubing system.
    Select  Set Valve . Press the ComWheel and set the value under the text Pulse Valve to 
    number 150 by turning the ComWheel. Press the ComWheel again and check that the 
    pressure inside the tubing system starts to drop. Finish the test by selecting  Previous 
    Menu.
    
    28. Leak test
    Check the NIBP tubing system for leakages.
    Select  Calibrations  from the NIBP service menu.
    Connect the pressure manometer to the NIBP mo dule cuff connector. Start the active leak 
    test from the menu by pressing the ComWheel. The module pumps a pressure of about 
    290 mmHg and then the pump stops. The max pressure in Adult mode is about 290 
    mmHg, but in Infant mode only 140mmHg. 
    						
    							Patient Side Module, E-PSM, E-PSMP (Rev. 01) 33
    Document no. M1215098-002
    Wait for 15 seconds for the pressure to stabi lize then check that the pressure does not 
    drop more than 6 mmHg per one minute. Release the pressure by pressing the 
    ComWheel once more.
    
    29. Calibration check
    Recalibrate the NIBP measurement accordin g to the instructions in section “NIBP 
    calibrations” on page 44. Remember to set the calibration protection back on after the 
    calibration.
    Disconnect the pressure manometer. Select  Calibrations - Calibration Check . Press the 
    ComWheel and take down the zero offset values for both pressure transducers, B1 and 
    B2. The values should be within ±20 mmHg.
    Connect the pressure manometer to the cuff connector and check the calibration with 
    pressures 100 mmHg, 200 mmHg and 260 mmHg. The zero offset value must be added to 
    the displayed pressure value in order to determine the real pressure.
    
    30. Safety valve functions
    Select  Safety Valve  from the NIBP service menu.
    Disconnect the pressure manometer from th e NIBP module cuff connector. Connect the 
    NIBP hose and cuff to the NIBP module cuff connector.  Perform the check with a 
    standard adult cuff that is connected  around some round object, for example a 
    calibration gas bottle.  
    Select  Start Test . Start the adult safety valve test by pressing the ComWheel. Wait until 
    the pump stops and the pressure is deflated. 
    Open cuff connector or disconnect an d connect cuff connector from module
    Check the pressure values ‘Max press’ and ‘2 s after stop’ for both transducers. All the 
    values should be within 270 - 330 mmHg.
    Select  ADULT . Press the ComWheel and check that the text changes now to  INFANT. 
    Select  Start Test  and wait until the pump stops and the pressure values on the screen 
    have been updated. 
    Open cuff connector or disconnect an d connect cuff connector from module
    Check that the values ‘Max press’ and ‘2 s af ter stop’ are all now within 135 to 165 mmHg.
    Return to the normal monitoring mode by pressing 
    Normal Screen.
    
    31. Cuff related messages Connect an adult NIBP cuff to the cuff connector and disconnect one of its hoses.
    Start NIBP measurement by pressing the key
     Start/Cancel on the module and check 
    that the message ‘Cuff loose’ appears on the screen within 70 seconds.
    Reconnect the hose and then bend it with  your fingers. Restart the measurement and 
    check that the message ‘Cuff occlusion’ appears on the screen within 70 seconds.
     
    						
    							E-modules
    34
    Document no. M1215098-002
    32. Test measurementCheck that the automatic inflation limits are in use:
    NIBP - NIBP Setup - Inflation Limits - Auto - Previous Menu
    Connect the cuff onto your arm, select  Start Ven.Stasis in the NIBP menu and press the 
    ComWheel. Check that the module identifies  the cuff, i.e. the text Adult appears in the 
    NIBP digit field for a short moment.
    Keep the pressure inside the cuff for about half  a minute in order to find out that the cuff 
    is not leaking, then press the ComWheel again. Select 
    Normal Screen.
    Disconnect the cuff hose.
    
    33. NIBP hose detection
    Press the 
    Start/ Cancel module or side panel key and check that the ‘Cuff loose’ 
    message appears in the NIBP digit field.
    Attach a NIBP cuff hose withou t cuff identification and check that the module identifies 
    the hose: 
    − The message ‘Select inflation limits’ appears in the NIBP digit field.
    − When you try to start the measurement, the monitor automatically opens the 
    selections  NIBP Setup - Inflation Limits.
    
    All modules
    34. Electrical safety check
    Perform an electrical safety check and a leakage current test.
    
    35. Functioning after electrical safety checkCheck that the module functions normally af ter the performed electrical safety check.
    
    36. Final cleaning
    Clean the module with suitable detergent.
    
    Fill in all necessary documents. 
    						
    All GE manuals Comments (0)