GE S5 Owners Manual
Have a look at the manual GE S5 Owners Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.
![](/img/blank.gif)
Patient Side Module, E-PSM, E-PSMP (Rev. 01) 25 Document no. M1215098-002 3 Service procedures 3.1 General service information The field service of the hemodynamic modules is limited to replacing faulty printed circuit boards or mechanical parts. The circuit boards sh ould be returned to GE Healthcare for repair. GE Healthcare is always available for service ad vice. Please provide the unit serial number, full type designation, and a detailed fault description. WARNING Only trained personnel with appropriate equipment should perform the tests and repairs outlined in this section. Unauthorized se rvice may void warranty of the unit. 3.2 Service check These instructions include complete procedures for a service check. The service check is mandatory after any service repair. However, th e service check procedures can also be used for determining possible failures. The procedures should be performed in ascending order. The instructions incl ude a check form (“APPENDIX A:”) which may be used when performing the procedures. The symbol in the instructions indicate s that the check form contains space to record the results of the particular procedure. 3.2.1 Recommended tools NOTE: Use only properly maintained, calibrated and traceable measurement equipment for the specified calibrations and adjustments to ensure accuracy. Table 13 Recommended tools Tool Order No. For product(s) Hemodynamic patient simulator M1010831 E-PSM(P) Adapter cables for simulators - Dual temperature adapter cable 2016998 -001 Hemodynamic patient simulator and Medsim - Dual Inv.BP adapter cable 2005772- 001 Hemodynamic patient simulator - Temperature adapter cable M1010832 Medsim - Inv.BP adapter cable M1010858 Medsim - Temperature adapter cable M1010846 Lionheart & MPS450 - Inv.BP adapter cable M1010862 Lionheart & MPS450 Pressure manometer E-PSMP Accessories: Temperature test set 884515-HEL E-PSM(P)
![](/img/blank.gif)
E-modules 26 Document no. M1215098-002 3.2.2 Recommended parts Table 14 Recommended parts 3.2.3 Visual inspection Detach the module cover by removing the four screws from the bottom of the module. Check: 1. Internal parts−screws are tightened properly − connectors are connected properly − NIBP tubing is attached properly − there are no loose objects inside the module 2. External parts −the front cover and the front panel sticker are intact Multi-Link ECG accessories, IEC: - Multi-link 3-leadwire set 412682-003 E-PSM(P) - Multi-link 5-leadwire set 412681-003 E-PSM(P) - Multi-link 5-leadwire set, C2-C6 416467-004 E-PSM(P) - Multi-link 12-lead ECG trunk cable 416035-002 E-PSM(P) Multi-Link ECG accessories, AHA: - Multi-link 3-leadwire set 412682-001 E-PSM(P) - Multi-link 5-leadwire set 416681-001 E-PSM(P) - Multi-link 5-leadwire set, V2-V6 416467-003 E-PSM(P) - Multi-link 12-lead ECG trunk cable 416035-001 E-PSM(P) SpO 2 finger probe OXY-F-UN E-PSM(P) SpO 2 Interconnect Cable OXY-ES3 E-PSM(P) InvBP transducer 70077-001 E-PSMP Adult NIBP cuff hose with cuff ID 2021285-001 E-PSM(P) Adult NIBP cuff 2753EE-PSM(P) Infant cuff hose without cuff ID 414874-001 E-PSM(P) Screwdriver Part Order No. Notes E-PSM(P), Air Filter, FRU M1221481 Replace every 3 years Tool Order No. For product(s)
![](/img/blank.gif)
Patient Side Module, E-PSM, E-PSMP (Rev. 01) 27 Document no. M1215098-002 −all connectors are intact and attached properly − the module box and latch are intact 3.2.4 Functional inspection 3. NIBP pump filter Replace the NIBP pump filter, if necessary. Reattach the module cover and check that the latch is moving properly. Switch the monitor on and wait until the monitoring screen appears. Configure the monitor screen so that all the needed para meters are shown, for example as follows: Monitor Setup - Screen 1 Setup - Waveform Fields - Field 1 - ECG1 Field 2 - ECG2 Field 3 - P1 Field 4 - P2 Field 5 - Pleth Field 6 - Resp Digit Fields - Lower Field 2 - NIBP Lower Field 3 - T1+T2 4. Module installation Plug in the module. Check that it goes in smoothly and locks up properly. 5. Module recognitionCheck that the module is recognized, i.e. all the needed parameter information, except invasive blood pressure, starts to show on the screen. Preset ECG, Respiration, InvBP and SpO2 measurement settings: ECG - ECG Setup - Hr Source - Auto Pacemaker - Show Others - Resp Setup - Size - 1.0Resp Rate Source - Auto Measurement - On Detection Limit - Auto Invasive Pressures - P1 ‘Art’ Setup - Label - ArtP2 ‘Cvp’ Setup - Label - Cvp PulseOximetry - Pleth Scale - Auto
![](/img/blank.gif)
E-modules 28 Document no. M1215098-002 ECG measurement 6. Module software (serial numbers) Enter the service menu: Monitor Setup - Install/Service (password 16-4-34) - Service (password 26-23-8) Take down the information regarding the module software by selecting Scroll Vers and turning the ComWheel. 7. Communication and memories Enter the Parameters - ECG service menu. Check that the Time-outs, Bad checksums and Bad c-s by mod values are not increasing faster than by 5 per second. Check also th at the ECG/RESP board memories have passed the internal memory test, i.e. the RAM, ROM and EEPROM state all OK. 8. Power frequency Check that the power frequency value is set according to the current mains power frequency. Change the setting by selecting Power Freq, if necessary. 9. Cable recognition Connect a 12-lead ECG trunk cable without a lead set to the module. Check that the message ‘Leads off’ is displayed on the screen. 10. Lead detectionConnect both 5-leadwire sets to the trunk cable. Connect all the leads together, for example to a suitable screwdriver. Check th at all the electrodes show ON and the message ‘Asystole’ appears. Check that the Cable type shows 10 lead. Connect the 10-leadwire set to the simulator. Disconnect one of the leads and check that the corresponding electrode in the service menu shows OFF within 10 seconds of the disconnection, and then reconnect the lead. Check the rest of the leads using the same method. Disconnect the trunk cable. Connect a 3-leadwire set to a trunk cable and connect it to the module. Connect all the leads together, for example to a suitable screwdriver. Check that the cable type shows 3 lead. NOTE: When any of the limb le ads is disconnected, the measurement will automatically change to 3 electrode ECG measurement. NOTE: The asystole and different leads off messages are shown using certain priority. Even though one of the leads is disconnected, the related leads off message may not appear on the screen. NOTE: When RA, LA, LL or RL electrode is di sconnected, all six V electrodes show OFF.
![](/img/blank.gif)
Patient Side Module, E-PSM, E-PSMP (Rev. 01) 29 Document no. M1215098-002 11. Test with the patient simulatorConnect the leads to a patient simulator. Perform the settings and checks with Dynatech Nevada MedSim 300 Patient Simulator: ECG - BASE - BPM - 160 PACE - WAVE - NSR Check that a normal ECG waveform is shown, the HR value is 160 (±5) and the ‘Pacer count’ value is not increas ing in the service menu. ECG - PACE - WAVE - ASNC Check that pacemaker spikes are shown on the ECG waveform, the HR value changes to 75 (±5) and the Pacer count valu e is increasing according to the shown pacemaker spikes. Set the pacemaker option off: ECG - PACE - WAVE - NSR Respiration measurement 12. RESP measurement recognition Check that Resp Available and RESP Measurement both show ON in the ESTP: ECG service menu. 13. Test with patient simulatorCheck the respiration measurement with a patient simulator. The settings and checks with Dynatech Nevada MedSim 300 Patient Simulator: Simulator Cover: BASELINE IMPEDANCE -switch - 500 LEAD SELECT-switch - II/RL-LL Simulator Menu: RESP - WAVE - NORM RATE - 20 OHMS - 1.0 RATIO - 1/1 APNEA - OFF SHIFT - OFF Check that the RESP waveform is shown and the RR value is 20 (±5). Change the position of the BASELINE IMPEDANCE switch and check that appropriate RESP waveform and RR value are shown again within 30 seconds. RESP - APNEA - 32 S Check that the monitor activates the APNEA alarm. NOTE: Make sure that only the ECG leads are connected to the simulator during the apnea test. If other cables are connected at the same time, the respiration signal from the simulator may be disturbed, and therefore, the APNEA alarm may not be activated. NOTE: When you have the ECG service menu open, spikes will appear on the respiration waveform. These spikes represent the threshold level for detecting inspiration and expiration.
![](/img/blank.gif)
E-modules 30 Document no. M1215098-002 Temperature measurement 14. Communication and memoriesEnter the ESTP: STP service menu: Parameters - ESTP : STP Check that the Time-outs, Bad checksums and Bad c-s by mod values do not increase faster than by 5 per second. Check also that the STP board memories have passed the internal memory test, i.e. the RAM, ROM and EEPROM show all OK. 15. Temperature probe detection Check that the ‘Cable’ and ‘Probe’ show OFF for both channels, T1 and T2, when no probes are connected. Connect the temperature adapter cable to the module temperature connector and a temperature test plug to the adapter cable. Check that the Cable and Probe for T1 show ON and the corresponding temperature value appears on the monitor screen. Perform the same check also for the channel T2. 16. Calibration check Check the temperature calibrations using temperature test plugs. If the deviation on a temperature reading on the screen is more than 0.1°C, calibrate the temperature channels according to the instructions in chapter “Temperature calibration” on page 45.” 17. Temp test Activate the temperature test by selecting Temp Test from the menu and pressing the ComWheel twice. When the message ‘Performing temp test’ disappears from the digit field, check that no error messages appear and Temp error shows OFF for both channels in the service menu. 18. Module configuration Check that the module configuration has been set correctly. The configuration in use is shown beside the text Configuration in the service menu and it can be either STP or ST. Change the configuration in the Calibrations - Set Config menu, if necessary. To activate the change, reset the module communicatio n by removing and inserting the module.
![](/img/blank.gif)
Patient Side Module, E-PSM, E-PSMP (Rev. 01) 31 Document no. M1215098-002 Invasive blood pressure measurement 19. Membrane keysCheck the front panel membrane keys that are related to the InvBP measurement. Press each of the keys for at least one second. Check that the pressed key is identified, i.e. one of the texts for Buttons changes fr om OFF to ON in the service menu. 20. Cable and transducer detection Check that the Cable and Probe for P1 show OFF. Connect the InvBP adapter cable to the module, connect a cable with an invasive blood pressure transducer to the adapter cable and check that the Cable and Probe show ON and the corresponding pressure waveform appears on the screen. Perform the same check also for the InvBP channel P2. 21. Calibration Calibrate the InvBP channels P1 and P2 according to the instructions in chapter “Invasive pressure calibration” on page 45.“ 22. Test with patient simulatorCheck the InvBP channels with a patient simulator. The settings and checks with Dynatech Nevada MedSim 300 Patient Simulator: SENSITIVITY - switch - 5 µV/V/mmHg ECG - BASE - BPM - 60 - BP - 1 - WAVE - ATM 2 - WAVE - ATM Restore the normal monitoring screen by pressing the key Normal Screen. Connect cables from the channels BP1 and BP 2 to the module connectors. Zero the InvBP channels by pressing the keys ZERO P1 and ZERO P2 on the module front panel. BP - 1 - WAVE - ART 2 - WAVE - CVP Check that appropriate InvBP waveforms are shown and the InvBP values are approximately 120/80 (±3 mmHg) for the ch annel P1 and 15/10 (±2 mmHg) for the channel P2. Check that the HR value is calculated from P1, when ECG is not measured (ECG cable disconnected). SpO2 measurement 23. SpO2 probe detection Check that the message ‘No probe’ is shown, when no SpO 2 sensor is connected to the module. Connect an SpO 2 finger probe to the module (wit h the interconnection cable, if needed). Check that the message ‘Probe off’ is shown when the probe is not connected to a finger.
![](/img/blank.gif)
E-modules 32 Document no. M1215098-002 24. Test measurementConnect the SpO 2 probe onto your finger. Check that the reading of 95-99 and SpO2 waveform appears. Check that the HR value is calculated from SpO 2 when ECG and InvBP (P1) are not measured. Non Invasive Blood Pressure measurement 25. Communication and memories Enter the NIBP module service menu: Parameters - NIBP Check that the Time-outs, Bad checksums and Bad c-s by mod values are not increasing faster than by 5 per second. Check also that the NIBP board memories have passed the internal memory test, i.e. the RAM, ROM and EEPROM show all OK. 26. Membrane keys Check the front panel membrane keys. Select Buttons/Leds . Press each of the two NIBP related membrane keys for at least one second. Check that the pressed key is identified, i.e. the corresponding text changes from OFF to ON in the menu, when the key is released back up again. 27. Pump and valves Check the pump and valves. Select Pneumatics from the NIBP menu. Connect a pressure manometer to the NIBP module cuff connector. Select Start Pump and press the ComWheel. Check that the pump turns on and the pressure inside the tubing system starts to increase. Stop the pump by pressing the ComWheel again when the pressure reaches 280 mmHg. Select Open Exh2 . Press the ComWheel and check that the pressure inside the tubing system starts to drop, then press the ComWheel again. If necessary, turn the pump on again for a moment to increase the pressure inside the tubing system. Select Set Valve . Press the ComWheel and set the value under the text Pulse Valve to number 150 by turning the ComWheel. Press the ComWheel again and check that the pressure inside the tubing system starts to drop. Finish the test by selecting Previous Menu. 28. Leak test Check the NIBP tubing system for leakages. Select Calibrations from the NIBP service menu. Connect the pressure manometer to the NIBP mo dule cuff connector. Start the active leak test from the menu by pressing the ComWheel. The module pumps a pressure of about 290 mmHg and then the pump stops. The max pressure in Adult mode is about 290 mmHg, but in Infant mode only 140mmHg.
![](/img/blank.gif)
Patient Side Module, E-PSM, E-PSMP (Rev. 01) 33 Document no. M1215098-002 Wait for 15 seconds for the pressure to stabi lize then check that the pressure does not drop more than 6 mmHg per one minute. Release the pressure by pressing the ComWheel once more. 29. Calibration check Recalibrate the NIBP measurement accordin g to the instructions in section “NIBP calibrations” on page 44. Remember to set the calibration protection back on after the calibration. Disconnect the pressure manometer. Select Calibrations - Calibration Check . Press the ComWheel and take down the zero offset values for both pressure transducers, B1 and B2. The values should be within ±20 mmHg. Connect the pressure manometer to the cuff connector and check the calibration with pressures 100 mmHg, 200 mmHg and 260 mmHg. The zero offset value must be added to the displayed pressure value in order to determine the real pressure. 30. Safety valve functions Select Safety Valve from the NIBP service menu. Disconnect the pressure manometer from th e NIBP module cuff connector. Connect the NIBP hose and cuff to the NIBP module cuff connector. Perform the check with a standard adult cuff that is connected around some round object, for example a calibration gas bottle. Select Start Test . Start the adult safety valve test by pressing the ComWheel. Wait until the pump stops and the pressure is deflated. Open cuff connector or disconnect an d connect cuff connector from module Check the pressure values ‘Max press’ and ‘2 s after stop’ for both transducers. All the values should be within 270 - 330 mmHg. Select ADULT . Press the ComWheel and check that the text changes now to INFANT. Select Start Test and wait until the pump stops and the pressure values on the screen have been updated. Open cuff connector or disconnect an d connect cuff connector from module Check that the values ‘Max press’ and ‘2 s af ter stop’ are all now within 135 to 165 mmHg. Return to the normal monitoring mode by pressing Normal Screen. 31. Cuff related messages Connect an adult NIBP cuff to the cuff connector and disconnect one of its hoses. Start NIBP measurement by pressing the key Start/Cancel on the module and check that the message ‘Cuff loose’ appears on the screen within 70 seconds. Reconnect the hose and then bend it with your fingers. Restart the measurement and check that the message ‘Cuff occlusion’ appears on the screen within 70 seconds.
![](/img/blank.gif)
E-modules 34 Document no. M1215098-002 32. Test measurementCheck that the automatic inflation limits are in use: NIBP - NIBP Setup - Inflation Limits - Auto - Previous Menu Connect the cuff onto your arm, select Start Ven.Stasis in the NIBP menu and press the ComWheel. Check that the module identifies the cuff, i.e. the text Adult appears in the NIBP digit field for a short moment. Keep the pressure inside the cuff for about half a minute in order to find out that the cuff is not leaking, then press the ComWheel again. Select Normal Screen. Disconnect the cuff hose. 33. NIBP hose detection Press the Start/ Cancel module or side panel key and check that the ‘Cuff loose’ message appears in the NIBP digit field. Attach a NIBP cuff hose withou t cuff identification and check that the module identifies the hose: − The message ‘Select inflation limits’ appears in the NIBP digit field. − When you try to start the measurement, the monitor automatically opens the selections NIBP Setup - Inflation Limits. All modules 34. Electrical safety check Perform an electrical safety check and a leakage current test. 35. Functioning after electrical safety checkCheck that the module functions normally af ter the performed electrical safety check. 36. Final cleaning Clean the module with suitable detergent. Fill in all necessary documents.