Home > GE > Medical Monitor > GE S5 Owners Manual

GE S5 Owners Manual

    Download as PDF Print this page Share this page

    Have a look at the manual GE S5 Owners Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    							Patient Side Module, E-PSM, E-PSMP (Rev. 01) 15
    Document no. M1215098-002
    2.2.3 ECG board in 12-lead measurement
    The 12-lead ECG measurement consists of the  functions shown in Figure 7 on page 15. All 
    functions are located in the ECG board except the ECG input unit.
    Figure 7 12-lead ECG measurement block diagram
    ECG input unit
    The ECG input unit consists of the front panel connector and the ECG input connector board 
    with the high voltage protection resistors. Th e connector for the 12-lead ECG cable is a green 
    11-pin rectangle shaped connector.
    Input protection and filtering
    The input protection is implemented with high voltage protection resistors in the ECG input unit 
    and with protection diodes in the ECG board.  The input filtering for ECG measurement is done 
    with passive RC filtering.
    PATIENT AND ECG ELECTRODES
    ECG CABLE - ECG LEAD SET
    - ECG TRUNK CABLE
    ECG INPUT UNIT- ECG CONNECTOR
    - INPUT PROTECTION RESISTORS
    INPUT PROTECTION DIODES FOR ECG & RESPIRATION MEASUREMENT
    INPUT FILTERING FOR ECG & RESPIRATION MEASUREMENTS
    RESPIRATION
    MEASUREMENT AMPLIFIERS
    BASELINE
    RESTORATION
    ECG CPU
    RESPIRATION
    MEASUREMENT
    CURRENT SUPPLYECG
    PREAMPLIFIERS &
    RLD CIRCUITLEADS OFF &PACER &
    DEFIBRILLATION DETECTION
    RS 485
    COMMUNICATIONPOWER SUPPLYNV
    MEMORY
    ISOLATION                                 ISOLATION
    MODULE BUS CONNECTORECG BLOCK DIAGRAM12_lead_ECG_meas_blck_dgrm.vsd 
    						
    							E-modules
    16
    Document no. M1215098-002
    ECG preamplifiers
    The buffer amplifiers are used for each lead. The “Leads off” detection is implemented by 
    measuring the output level of the input buffer amplifiers with the A/D converter of the CPU. The 
    ECG signals are measured using differential amplifiers.
    ECG amplifiers and baseline restoration
    The function of the ECG amplifiers and baseline restoration is to amplify the signal and to 
    restore the baseline of the signal in the middle of the display after the change of the signal 
    level, e.g. after the change of the DC offset voltage.
    Pacer detection
    Pacer detection has been made by using four slew rate detector circuits. The pacer detection 
    amplifiers have been realized at the front of th e slew rate detectors independently of the ECG 
    measuring channels.
    Respiration impedance supply
    The 31.25 kHz sine wave generator is used as the respiration measurement signal supply. 
    Analog switches are used for connecting the sine wave to the ECG leads to be measured.
    Respiration impedance amplifiers
    Buffer amplifiers are used in respiration measurement. Analog switches are used for selecting 
    the measurement leads. There are also additional amplifiers for increasing the respiration 
    signal gain. When ECG measurement is 5/12-lead, the respiration measurement is always done 
    between R and F, independently on the ECG lead selection. When ECG measurement is 3-lead, 
    then the respiration measurement happens at the same lead as the ECG measurement (I, II or 
    III).
    ECG CPU
    The CPU is a 16 bit H8/3052 single-chip microcompu ter. It contains 128 kbytes of flash memory 
    and 4 kbytes of RAM. The clock frequency is 16 MHz.
    RS485 communication
    The communication to the CPU board of the mo nitor uses RS485 protocol. The RS485 driver 
    circuits are optically isolated from the processor of the module. 
    Power supply
    The ECG board has a driver-controlled half-bridge  switching power supply with 5 kV isolation. 
    The supply voltages have been regulated with linear regulators.
    2.2.4 ECG filtering
    Datex-Ohmeda S/5 monitors have three ECG filtering modes:
    MONITORING 0.5 to 30 Hz (with 50 Hz reject filter)
    0.5 to 40 Hz (with 60  Hz reject filter)
    DIAGNOSTIC 12-lead ECG 0.05 to 150 Hz
    ST FILTER 0.05 to 30 Hz (with 50 Hz reject filter)
    0.05 to 40 Hz (with 60 Hz reject filter)
    The purpose of filtering is to reduce high frequency noise and low frequency (e.g. respiratory) 
    movement artifacts.
    The monitor filter is used in normal monitoring.  The diagnostic filter is used if more accurate 
    diagnostic information is needed. The ST filter  gives more accurate information of ST segment, 
    but reduces high frequency noise. 
    						
    							Patient Side Module, E-PSM, E-PSMP (Rev. 01) 17
    Document no. M1215098-002
    The high-pass filters 0.5 Hz and 0.05 Hz are done with software. The monitor sends a command 
    to the hemodynamic module determining which of  the corner frequencies 0.5 Hz or 0.05 Hz is 
    to be used.
    The 50 Hz and 60 Hz reject filters are both lo w-pass filters with zero at 50 Hz or 60 Hz 
    correspondingly. They are software based filters used for the mains supply filtering. With these 
    filters the 3 dB value for low-pass filter is 30 Hz or 40 Hz.
    In diagnostic mode the upper frequency is 150 Hz and it is limited by software.
    2.2.5 STP board
    Figure 8 STP board block diagram
    Microprocessor unit
    The CPU is a 16 bit H8/3052 single-chip microcompu ter. It contains 128 kbytes of flash memory 
    and 4 kbytes of RAM. The clock frequency is 16 MHz.
    High speed I/O is used to obtain a pulse control sequence necessary for pulse oximetry 
    measurement. Timing for the clock is from the oscillator.
    STP_brd_blck_dgrm.vsd
    PATIENT AND SpO2 PROBE
    SpO2 TRUNK CABLE
    SpO2 CONNECTOR
    A/D CONVERSION
    STP CPU
    STP BLOCK DIAGRAM
    PATIENT AND TEMPERATURE SENSOR
    TEMPERATURE CONNECTOR
    PATIENT AND INVASIVE CANNULA OR CATHETER
    FLUSHING KIT & INVASIVE PRESSURE
    SENSOR WITH DOME
    INV.PRESSURE CONNECTORINPUT PROTECTION CIRCUITRY
    SpO2 PROBE RECOGNITION & LEDDRIVE SELECTION MATRIX
    TEMPERATURE
    MEASUREMENT
    AMPLIFIER
    SpO2
    AMPLIFIERSpO2 LED
    DRIVE
     SENSOR SIGNAL CURRENT
    SOURCESENSOR SIGNAL VOLTAGE
    SOURCEINV: PRESSURE
    MEASURE MENT
    AMPLIFIER
    RS 485
    COMMUNICATIONPOWER SUPPLYNV
    MEMORY
    ISOLATION                                 ISOLATION
    MODULE BUS CONNECTOR 
    						
    							E-modules
    18
    Document no. M1215098-002
    Temperature measurement unit
    The NTC-resistor value in the probe depends on the patient’s temperature. It is measured with 
    the following principle described below.
    The constant current source is supplied about 38 μA current through the temperature sensor 
    (YSI 400-series NTC resistor). The constant current is caused a voltage over the temperature 
    sensor (NTC resistor). The voltage over the temperature sensor is amplified in a differential 
    amplifier stage. The amplified voltage is transfe rred to a controller of the STP board through an 
    A/D converter.
    Figure 9 Temperature measurement principle
    Invasive blood pressure measurement unit
    An isolated +5 V voltage is supplied to the pressure  transducer. The differential voltage, which 
    depends on the pressure and the supplied voltage, is calculated from the bridge connection 
    (see the formula below).
    U
    out = Uin  ×  pressure  × 5  μV, where Uin is 5 V
    ⇒  U
    out = 25  μV  × pressure [mmHg]
    d/dt
    0
    0
    0 0
    R
    R
    R
    R
    Temperature
    sensors T1, T2 constant
    currentsource
    Differential amplifier
    Defibrillation/ESD protection
    resistors and diodes
    To A/D
    converter
    0C:7k36
    15C:3k54
    25C:2k53
    38C:1k30 45C: 984
    PSM_temp_meas_principle.vsd
    Ref2
    Ref1
    T2T1
    RR 
    						
    							Patient Side Module, E-PSM, E-PSMP (Rev. 01) 19
    Document no. M1215098-002
    Pressure amplification is realized in the instrumentation amplifier. The gain of the amplifier is 
    set to keep the level of the signal transferred to the A/D converter within the measurement 
    range even when there are circumstantial offsets or offsets caused by the transducer. There is 
    a filter before the amplifier to attenuate high frequency disturbances.
    Figure 10 Pressure measurement principle
    Pulse oximetry measurement section
    LED control signals
    The D/A converters of the microcontroller on the STP board set the LED intensity adjustment 
    values for the infrared and red LEDs of the SpO
    2 probe. The microcontroller on the STP board 
    switches ON (to the adjusted intensity) and OFF the SpO
    2 probe LEDs according to the 
    predetermined sequence.
    LED driving circuit
    Differential amplifiers measure the LED currents (LED current indication) of the SpO2 probe over 
    the shunt resistors placed in the LED current paths. The LED driving voltages (LED voltage 
    indication) are measured from the driver circuitr y. The LED driving circuits also have MOSFET 
    transistor matrix to enable the use  of different probe configurations.
    Measured signal preamplification
    The preamplifier is a bipolar/single-ended current-to-voltage converter with adjustable gain. A 
    higher gain is used for measuring thin tissue.  The preamplification stage has also ambient light 
    reduction and a second amplifier stage.
    Input filter
    Instrum entation
    amplifier
    to AD converter
    Vout
    G
    PSM_pressure_meas_principle.vsd
    Current
    measurementG
    to AD converter
    Vin
    Pressure
    transducer 
    						
    							E-modules
    20
    Document no. M1215098-002
    Figure 11 Pulse oximetry measurement block diagram
    Red and infrared channel separation
    It is possible to multiplex the detector signal to four different channels depending on the 
    content of the signal. The detector signal must at least multiplex into infrared and red signals. 
    Other channels are e.g. for diagnostic purposes.
    Serial communication
    An RS485 type bus driver makes the serial communication between the module and the frame. 
    The data transmission rate is 500kbps.
    LED Intensity adjustment 1
    LED Intensity adjustment 2
    LED ON/OFF control 1
    LED ON/OFF control 2
    LED Driving circuit 1
    LED Driving circuit 2
    LED voltage indication 1
    LED current indication 1
    LED voltage indication 2
    LED current  indication 2
    Preamplifier:
    Current-to-voltage type
    Bipolar/single-ended modes
    Adjustable gain
    Ambient reduction
    Amplifier:
    Gain = 2
    DE-MUX
    LP
    LP
    LP
    LP
    Gain=7.5
    Gain=7.5
    Gain=7.5
    Amplifiers
    DC-
    suppression
    DC-
    suppression
    DC-
    suppression
    Oximeter channel 1
    Oximeter channel 2
    Oximeter channel 3
    Oximeter channel 4
    SpO2
    Probe
    AnalogDigitalSpo2_measurement_blck_diagr.vsd 
    						
    							Patient Side Module, E-PSM, E-PSMP (Rev. 01) 21
    Document no. M1215098-002
    Figure 12 Serial communication of E-PSM(P) module
    Signals and isolation barrier
    The communication signals transfer over the isol ation barrier by using high isolation voltage 
    (6kV) opto isolators. 
    Power supply section
    The power for the electronics on the floating pa rt of the STP and the ECG boards is made on 
    each board with the switching power supplie s connected to a high voltage isolated 
    transformer. The switching power supplies on the STP and ECG boards are synchronized to the 
    frequency, about 340kHz of the switching power  supply on the NIBP board. The NIBP board 
    supplies non-isolated 5 V to the ECG and STP boards. The module uses only Vmod 13.8 - 16 V 
    voltage of the frame. The other voltages of th e measuring boards are made by the switching 
    power supplies and regulators or the linear regulators. Each measuring board is protected 
    against overloading with PTC type automatic fuses.
    RS 485
    DRIVER
    NIBP BOARD CONNECTOR
    DATA
    NDATARECEIVE DATA SEND DATA
    SEND/RECEIVE
    NIBP BOARD CONTROLLERS
    NIBP BOARD
    RS 485
    DRIVER
    STP BOARD CONNECTOR
    RECEIVE DATA
    SEND DATA
    SEND/RECEIVE
    STP BOARD CONTROLLER
    STP BOARD
    PATIENT ISOLATION
    RS 485
    DRIVER
    ECG BOARD CONNECTOR
    DATA
    NDATA
    ECG BOARD CONTROLLER
    ECG BOARD
    PATIENT ISOLATION
    DIRECT ECG (PW M)
    PSM_serial_communication.vsd
    Module bus connector
    DATA
    NDATA
    synchronization signal
    +5 V
    +5 V
    +5 V
    RECEIVE DATA SEND DATA
    SEND/RECEIVE 
    						
    							E-modules
    22
    Document no. M1215098-002
    2.3 Connectors and signals
    2.3.1 Module bus connector
    Table 8 Module bus connector description
    2.3.2 Front panel connectors
    Table 9 ECG connector
    5 pin connectorPin No.Signal
    1
    2
    3
    4
    5 GND
    Vmod 13.8 - 16 V
    Data +
    Data -
    Shield
    ECG Connector Pin No. Signal Name
    1
    2
    3
    4
    5
    6
    7
    8
    9
    10
    11 R/RA; Right arm electrode
    C2/V2; Chest electrode
    C3/V3; Chest electrode
    L/LA; Left arm electrode
    N/RL; Neutral/Right Leg Drive electrode
    C1/V1; Chest electrode
    C4/V4; Chest electrode
    F/LL; Left Leg electrode
    C6/V6; Chest electrode
    C5/V5; Chest electrode
    Cable Shield 
    						
    							Patient Side Module, E-PSM, E-PSMP (Rev. 01) 23
    Document no. M1215098-002
    Table 10 SpO2 connector
    Table 11 Invasive blood pressure connectors (P1, P2)
    SpO2 connector  Pin No. Signal Description
    1 DET_A Photodiode anode
    2 DET_C Photodiode cathode
    3DATA-
    4Wire 1/3LED connection
    5 IR_C IR LED cathode
    6OUTER SHIELD
    7 DET_SHIELD
    8 PRB_ID Bin/ID Resistor+
    9 Wire 3/5 LED Connection
    10 RED_C RED LED cathode
    11 DATA+
    Invasive blood pressure 
    connectors (Dual BP) Pin 
    No.Signal Description
    1BP_+V
    REFBP transducer excitation voltage, channel 1
    2 BP SIG+ BP transducer signal positive (+), channel 1
    3BP_+V
    REFBP transducer excitation voltage, channel 2
    4 AGND Analog ground
    5 BP SIG+ BP transducer signal positive (+), channel 2
    6 SHIELD BP cable shield
    7 AGND Analog ground
    8 BP SIG1 BP transducer signal negative (-), channel 1
    9 BP SIG2 BP transducer signal negative (-), channel 2
    10 BP1_ID BP1 probe identification
    11 NC Not connected 
    						
    							E-modules
    24
    Document no. M1215098-002
    Table 12 Temp connector (T1, T2)
    Temp connector Pin No Signal
    1 Sensor drive current
    2 Input from temperature sensor, channel 1
    3Not connected
    4Not connected
    5 Thermistor ID (LOW= Temperature error,  HIGH=YSI 400 series)
    6 Cable shield
    7 Analog ground
    8 Input from temperature sensor, channel 2
    9Not connected
    10 Not connected
    11 Digital ground 
    						
    All GE manuals Comments (0)