GE S5 Owners Manual
Have a look at the manual GE S5 Owners Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.
Patient Side Module, E-PSM, E-PSMP (Rev. 01) 45 Document no. M1215098-002 •Remove the module from the frame and plug it back again. Then perform “Calibration check” (see the preceding page) to verify the new calibration. 3.4.2 Temperature calibration NOTE: For the temperature calibration, separate, accurate test plugs (25 °C and 45 °C) are needed. A test set of two plugs is available from GE Healthcare, order code 884515-HEL. A Dual temperature adapter cable, order code 2016998-001 is also required for the temperature calibration. Calibrate the temperature, when the measured test values differ for more than ±0.1 °C, and always after STP board replacement . 1. Enter the STP service menu. ( Monitor Setup - Install/Service (password 16-4-34) - Service (password 26-23-8) - Parameters ). 2. Enter Calibrations menu. 3. Choose Protection OFF in protect mode. 4. Select Calibrate T1/Calibrate T2 . 5. Insert calibration plug ( 25 °C) into T1/T2 connector. 6. Push the ComWheel. 7. Insert calibration plug ( 45 °C) into T1/T2 connector. 8. Push the ComWheel. 9. Choose Protection ON in protect mode. 3.4.3 Invasive pressure calibration NOTE: Before starting invasive pressure calibrati on, disconnect all patient cables and discharge the patient. NOTE: For the Invasive pressure calibration a Dual InvBP adapter cable, order code 2005722-001, is needed. Calibrate the invasive pressure when the pressure transducer (probe) is replaced with a different type of transducer, and when the STP board is replaced. 1. Enter the STP service menu. ( Monitor Setup - Install/Service (password 16-4-34) - Service (password 26-23-8) - Parameters ). 2. Enter Calibrations menu. 3. Connect a pressure transducer with a pr essure manometer to the P1/P2 connector. Choose Calibrate P1 or Calibrate P2 . Leave the transducer to room air pressure. 4. Push the ComWheel to start zeroing. 5. Supply a pressure of 100 mmHg to 300 mmHg to the transducer. The recommended pressure is 200 mmHg. 6. Set the pressure on the display to match the pressure reading on the manometer and push the ComWheel. A tolerance of ±1 mmHg is allowed. 7. The message ‘Calibrated’ will be displayed on the display.
E-modules 46 Document no. M1215098-002 4Troubleshooting 4.1 Troubleshooting charts See also the “User’s Reference Manual” for more troubleshooting procedures. 4.1.1 NIBP ProblemCause What to do No NIBP value displayed NIBP not selected on screen. Check monitor setup. NIBP menu fading No E-PSM(P) module, module not properly connected. Plug in the module. ‘Artifacts’ message Unsuccessful measurement due to patient movement, shivering, external artifact or weak signal. Check the patient status. ‘Weak pulsation’ message Weak or unstable oscillation pulses due to: •artifacts •weak pulse pressure due to arrhythmias •improper cuff position or attachment •too few pulses detected •weak or unusual blood circulation •obese patient Check patient condition and retry. Check any leaks and retry. Use proper size of cuff. Check attachment. Call service ‘Error X’ message NIBP hardware error. X = error number. See the description of the error message code, the causes and the solutions listed in the “NIBP error code explanation” chapter. ‘Cuff loose’ message 1. Hose and/or cuff not connected. 1. Connect the hose and the cuff. 2. Hose and cuff connected. Reasons: - cuff loosely wrapped - Tighten the cuff. - leakage inside the shield, in the Patient connector panel or tubings connecting to the module - Check the tubings inside the shield and Patient connector panel, fix if necessary. - leakage in cuff or hose - Replace cuff/hose. - leakage inside module - Check internal tubing and fix if necessary. - pump does not work - Check pump connector; if OK, replace the NIBP Pump Unit.
Patient Side Module, E-PSM, E-PSMP (Rev. 01) 47 Document no. M1215098-002 Cuff ID not working 1. Defective cuff ID holes in the NIBP cuff hose - Replace NIBP cuff hose. 2. NIBP flex board connector wrongly connected - Check that the NIBP flex board connector is properly connected to the STP input board: all pins have to be connected. 3. Cuff ID switches defective - To check the switches, attach a NIBP cuff hose without the cuff ID and check that the message ‘Select inflation limit’ appears. If not, replace the Front Panel Unit. 4. NIBP ID switch cable (between patient connector panel and the module) broken or poorly connected - Check the cable, fix if necessary. ‘Air leakage’ message 1. Hose or cuff leaking. Reasons: 1. Replace cuff - cuff damaged - Replace cuff. - cuff connector damaged - Replace cu ff connector (if the fault is in hose connector). - O-ring damaged or missing - Replace O-ring. - hose double connector damaged - Replace NIBP cuff hose. 2. Hose and cuff OK. Reasons: 2. Connect or replace tube - leakage in the tubes connecting the patient connector panel and the module - Check the tubes. - leakage inside the module - Replace the whole tubing. - tube disconnected or damaged - Fix connections. - manifold leaking - Replace the manifold. - tubes or valve(s) damaged - Replace tubes/valve(s). ‘Unable to measure Sys’ message Systolic blood pressure probably higher than the inflation pressure or artifacts. Automatic retrial with increased pressure. Problem Cause What to do
E-modules 48 Document no. M1215098-002 ‘Cuff occlusion’ message 1. Cuff and/or hose occluded. Reason: - cuff tube kinked - Straighten tube. - tubes inside the shield kinked - Straighten tubes. - tubes inside module kinked - Straighten tubes. - occlusion inside/outside module - Remove occlusion. 2. Cuff, hose, and tubes OK. Reason: - fault in pressure transducer - Replace the NIBP board. - fault in A/D converter - Replace the NIBP board. - faulty calibration - Check calibration. ‘Calibration switch on’ message EEPROM protection has been handled by pressing module buttons Auto ON/OFF and Start/Cancel simultaneously for 3 seconds. Enables setting the protection OFF in the Calibration menu. Press the buttons again if you are not going to calibrate. ‘Calibration not protected’ message. Calibration protection is set to OFF. Set the protection ON in the NIBP Calibration menu. Problem Cause What to do
Patient Side Module, E-PSM, E-PSMP (Rev. 01) 49 Document no. M1215098-002 4.1.2 NIBP error code explanation Code Problem What to do 0 RAM failure; memory failure Change the NIBP board. 1 ROM checksum error; memory failure Change the NIBP board. 2 Pump current failure Check short circuits. Change the NIBP board. 3 Safety CPU internal test failure or pressure sensor reference voltage failure Change the NIBP board. 4 EEPROM protection error Press module buttons Auto ON/OFF and Start/ Cancel simultaneously for 3 seconds. 5 Calibration not protected Protect calibra tion by selecting Protection ON in the NIBP calibration menu. 6 Pressure sensors give different readings Try to remeasure. If the problem persists, recalibrate. If the problem still persists, change the NIPB board. 7 Calibration failure Reset the module and recalibrate. If this does not help, change the NIBP board. 8 Exhaust Valve occlusion Check and clean the tubing and air chamber. If this does not help, change the NIBP board. 9 Measurement related error Automatic recovery. 10 EEPROM checksum error; memory failure Change the NIBP board. 11 Auto zero range exceeded Calibrate the NIBP. 12 Communication break; temporal break down of communication from monitor detected Automatic recovery. 13 Illegal neonate cuff with identifying magnet connected Remove the cuff. 14 Not in use Not in use 15 Safety CPU pressure calibration error Recalibrate. If this does not help, change the NIBP board. 16 Communication error between CPUs Change the NIBP board. 17 Safety CPU has cut down power from pneumatics due to repeating safety limit violations Reset the module. If the problem persists, change the NIBP board.
E-modules 50 Document no. M1215098-002 4.1.3 Pulse oximetry (SpO2) ProblemCause What to do Message ‘NO PROBE’ No sensor connected to the module SpO2 connector. Check sensor connections. Sensor faulty. Change the sensor. Flat cable connecting the SpO2 connector to the STP board loosen or broken. Check the Flat cable, replace if necessary. Message ‘PROBE OFF’ though sensor properly attached to the patient Unsuitable site. Try another site. Sensor faulty. Try another sensor. Sensor connection cable not connected to sensor. Connect the cable to sensor. Finger sensor falls off Sensor is slippery. Wipe with 70% isopropyl alcohol and allow drying. Finger is too thin or thick. Try other fingers, or other sensor types. Weak signal artifacts Poor perfusion. Try another place. Movement artifacts. Shivering. Message ‘NO PULSE’ Pulse search > 20 sec. and low SpO 2 or low pulse rate. Try other fingers. Message ‘ARTIFACT’ Pulse modulation exceeds the present scale. Try another place or another sensor. Message ‘CHECK PROBE’ DC value not in balance. Try another sensor. Message ‘POOR SIGNAL’ Poor perfusion. Modulation (Red or Ired) < 0.25% Check that the sensor is positioned correctly to the patient. Message ‘FAULTY PROBE’ Sensor is faulty. Change the sensor. No SpO 2No waveform selected on screen. Check the selected SpO2 waveforms by pressing Monitor Setup key and selecting Screen 1 Setup - Waveform Fields . Wrong configuration setting. Check the configuration settings from the STP/Calibrations menu ( Monitor Setup - Install/Service - Service - Parameters )
Patient Side Module, E-PSM, E-PSMP (Rev. 01) 51 Document no. M1215098-002 4.1.4 Invasive blood pressure ProblemCause What to do Abnormally low pressure Transducer wrongly positioned. Check mid-heart level and reposition transducer. No pressure Defective transducer. Check transducer. No pressure module plugged in. Check the module. No waveform selected on screen. Check the selected pressure waveforms by pressing Monitor Setup key and selecting Screen 1 Setup - Waveform Fields . Check that the pressure transducer is open to the patient. Wrong configuration setting Check the configuration setting from the STP/Calibrations menu ( Monitor Setup - Install/Service - Service - Parameters ). Flat cable connecting the patient connector panel to the module loosen or broken. Check the Flat cable, replace if necessary. ‘Not zeroed’ message Measurement on, channel not zeroed. Zero the channel. ‘Zeroing failed’ message Unsuccessful zeroing of P1 /P2 (number field). Possibly due to pulsating pressure waveform. Open the transducer to air and zero the channel. Offset is > 150 mmHg. Open the transducer to air and zero the channel. Defective transducer. Replace it and zero the channel. ‘Calibration failed’ message Unsuccessful calibration of P1/P2 (number field), possibly due to a pulsating waveform Turn the transducer to sphygmomanometer and try again (zeroing takes place first). Gain is beyond the limits (± 20% of the default gain). Replace the transducer. Out of range < 40 mmHg Measurement pressure is beyond the measurement range. Check the transducer level. Zero the channel. Out of range > 320 mmHg Measurement pressure is beyond the measurement range. Check the transducer level. Zero the channel. The patient may also have high pressure. Zero adj. > 100 mmHg Offset when zeroing is > 100 mmHg (but < 150 mmHg) from the absolute zero of the module (with default gain). Check the transducer. The waveform may hit the top and the numeric display not shown. Out of range Measured pressure is beyond the internal measurement range of the module. The waveform hits the top and the numeric display not shown. Check the transducer and its level. Zero the channel.
E-modules 52 Document no. M1215098-002 4.2 Troubleshooting flowcharts 4.2.1 Troubleshooting for NIBP parameter Figure 13 Troubleshooting flowchart for NIBP parameter Insert new module and turn power on. Does fault still appear? Module ID on screen?Does another module work in same slot? Error message on screen? See error code explanation inservice manual and fix it. Do they work? Check tubes and connectors. Find leak and fix it.Leak test in Service Menu OK? Pump check in Service Menu OK? Key check in Service Menu OK?Check keyboard connector. IfOK, change front panel. OK? NO YES YES NO NO NO YESNO NO YES YES YES NO Check pumpconnector. If OK, change pump. YES START NO OK? Replace the module. Check and replace NIBP board. NO PSM_module_troubles_flowch.vsd NO Fault not in original module Start NIBP without hoseCuff loose message appears on screen? Check and replace - STP Input Board - NIBP flex board - connectors YES NO Cuff ID OK but measurement does not work
Patient Side Module, E-PSM, E-PSMP (Rev. 01) 53 Document no. M1215098-002 4.2.2 Troubleshooting for ESP parameters Figure 14 Troubleshooting flowchart for ESP Parameters Is it correct?Select the correctconfiguration YES NO Does another module work in same slot? ECG module ID on screen? Remove ECG boardSTP module ID on screen? Replace STP board Replace ECG board Check front panel key functions in Service menu Do they wo rk? Che ck front panel unit, ECG board, and STP board to find culprit. Replace if necessary. NO NO YES Has SpO 2 measurement failed? Re place STP board Replace STP bo ard YES NO YES Has Temp measu reme nt failed? Replace ECG boardYES NO Has ECG measurement failed?Replace STP board YES Has STP board failed?Replac e STP board NO NO YESHa s i nv BP measurement failed? YES YES Enter the Service menu PSM_NESTPR_MESTPR_trouble s_flowch.vsd Possibly fault in module with ESTPR parameters. Insert module and turn power on Fault not inmodule Does fault still appear? NIBP module ID on screen? STP module ID on screen? YES YES YES YES NO NO NO Does another module work in same slot?NORemove NIBP boardYES NO Check module configuration in Service menu (ESTPR/S/TP/Calibrations)
E-modules 54 Document no. M1215098-002 5 Earlier revisions Patient Side Modules E-PSM, E-PSMP (Rev. 00)