    							Remote Controller3
    Document no. M1187342-004
    2 Functional description
    2.1 Remote Controller, K-REMCO/ K-CREMCO
    The Remote Controller consists of 12 direct function keys and the ComWheel.
    2.1.1 K-REMCO/ K-CREMCO PCBs
    The K-REMCO/ K-CREMCO has two PCBs located inside the Remote Controller. One board has 
    only the push button switches of the keys. The other board reads the status of the keys and the 
    ComWheel and forwards the in formation to the CPU board.
    2.1.2 External communication
    K-REMCO Rev. 00-01
    Two signals, TXD and RXD in RS232 format are in use. No handshaking is used. Serial 
    communication speed is 19.2 kbps. The 26-pin subminiature D-connector of the Remote 
    Controller is connected to the Display Controller Board, B-DISP. 
    K-REMCO Rev. 00 with optional Remote  Controller - Compact Monitor cable 
    or K-CREMCO Rev. 00
    Two signals, TXD and RXD in RS232 format are in use. No handshaking is used. Serial 
    communication speed is 19.2 kbps. 
    In Compact Monitors, the DIN 5 connector is  connected to the keyboard connector X9. 
    In S/5 FM and FM Light, the DIN 5 connector is connected to Multi IO adapter X7.
    2.1.3 ComWheel
    The ComWheel is used for menu selection. 
    						
    							Datex-Ohmeda E-Modules
    4
    Document no. M1187342-004
    3 Service procedures
    3.1 General service information
    The field service of the remote controllers is limited to replacing faulty circuit boards or 
    mechanical parts. The circuit boards should be returned to GE Healthcare for repair.
    GE Healthcare is always available for service advice. Please provide the unit serial number, full 
    type designation, and a detailed fault description. 
    WARNING Only trained personnel with appropriate equipment should perform the 
    tests and repairs outlined in this section. Unauthor ized service may void the 
    warranty of the unit.
    3.2 Service check
    These instructions include complete procedures for a service check. The service check is 
    mandatory after any service repair. However, th e service check procedures can also be used 
    for determining possible failures.
    The procedures should be performed in ascending order.
    The instructions in clude a check form  (“APPENDIX A:” ) which may be used when performing the 
    procedures. 
    The symbol 
    in the instructions indicate s that the check form contains space to record the 
    results of the particular procedure.
    3.2.1 Recommended tools
    3.2.2 Visual inspection
    Turn the monitor to STBY.
    Disconnect the remote controller cable from the monitor.
    Detach the remote controller upper cover and the keypad cover by removing the screws 
    (7 pcs) from the bottom.
    1. Internal parts Check that:
    •cables are connected properly
    •the remote controller cable is fasten ed to the bottom cover with screws
    •the keypad switches are intact
    •the software EPROM under the keypad is attached properly
    
    ToolOrder No.Notes
    Screwdriver 
    						
    							Remote Controller5
    Document no. M1187342-004
    2. External partsCheck that:
    •the upper and bottom covers are intact
    •the keypad cover is intact
    •the ComWheel cover is intact and attached properly
    
    Reassemble the remote controller.
    3. Cable Check the remote controller cable:
    •the cable is intact
    •the cable connector is intact
    •the connector pins are clean, straight and at about the same height
    •the locking screws inside the connector case are intact
    
    3.2.3 Functional inspection
    Reconnect the cable to the moni tor and turn the monitor ON.
    4. Command board software Enter the service menu:
    Menu (on the remote controller keypad) - Monitor Setup - Install/Service (password 
    16-4-34)  - Service (password 26-23-8)
    Take down the information regarding the remote controller software.
    
    Select the menu  KEYBOARD:
    Service - Keyboard
    5. Remote controller keys Press the keys one by one. Check that each key generates a sound from the loudspeaker.
    
    6. ComWheelTurn the ComWheel clockwise and counterclo ckwise and check that each step generates 
    a sound and the corresponding values at the bottom of the menu increase.
    Select DUMMY PRESS. Push the ComWheel and check that the push generates a sound 
    and the corresponding value in the menu increases.
     
    						
    							Datex-Ohmeda E-Modules
    6
    Document no. M1187342-004
    7. Electrical safety checkPerform an electrical safety check and a leakage current test.
    
    8. Functioning after electrical safety checkCheck that the remote controller functions no rmally after the electrical safety check.
    
    9. Final cleaning
    Clean the remote controller and the cable.
    
    •Fill in all necessary documents. 
    						
    							Remote Controller7
    Document no. M1187342-004
    3.3 Disassembly and reassembly 
    3.3.1 Before disassembly
    WARNINGWear a grounded, antistatic wristband when handling PC boards. 
    Electrostatic discharge may damage components on the board.
    Handle all PC boards by their edges.
    3.3.2 Tools needed
    3.3.3 To disassemble Remote Co ntroller K-REMCO / K-CREMCO
    To disassemble the Remote Controller (see  Figure 2):
    1. Disconnect the K-REMCO/ K-CREMCO cable from the monitor.
    2. Pull out the knob of the ComWheel.
    3. Open the nut on the shaft of the ComWheel.
    4. Open the three cross head screws on the bottom of the K-REMCO/ K-CREMCO.
    5. Remove the top cover.
    6. Open the four screws on the bottom of the K-REMCO/ K-CREMCO.
    7. Remove the keyboard cover.
    8. Disconnect the K-REMCO/ K-CREMCO cable and the wire set from the Comwheel.
    9. Remove the PCBs.
    For more information on the spare  parts, see the “Spare parts” slot.
    In reassembly, remember to put the reinforcing cord of the cable around the screw on the 
    metal bridge before tightening the screw. - screwdriver
    - flat blade screwdriver
    - antistatic wristband 
    						
    							Datex-Ohmeda E-Modules
    8
    Document no. M1187342-004
    Figure 2 K-REMCO, K-CREMCO disassembly and reassembly
    To reassemble the module, reverse the order of the disassembly steps.
    In reassembly, remember to put the reinforcing cord of the cable around the screw on the 
    metal bridge before tightening the screw.
    CAUTION When reassembling the remote controller make sure that the cables are 
    reconnected properly.
    Always perform the  “Service check” after reassembling the remote controller. 
    						
    							Remote Controller9
    Document no. M1187342-004
    4Troubleshooting
    4.1 K-REMCO, K-CREMCO
    See Keyboard Service Menu in the “Service Menu“ slot and perform tests available. If any of the 
    tests fail, see explanation below.Problem  Cause What to do
    ComWheel not working ComWheel leads broken or  connector loose. ComWheel 
    faulty. Check the items. Replace the 
    ComWheel if necessary.
    Membrane key not 
    working Switch cable loose or broken. 
    Keyboard cable loose or broken. 
    Cable connector pin failure. 
    RS232 communication failure 
    on CPU board. Check the items. Replace them if 
    necessary. 
    						
    							Datex-Ohmeda E-Modules
    10
    Document no. M1187342-004
    For your notes: 
    						
    							Appendix A, Service check form, Remote Controllers K-REMCO (Rev. 01) and K-CREMCO (Rev. 00)
    1(2)
                               Document no. M1187342-004
    APPENDIX A: Service check form, 
    Remote Controllers K-REMCO (Rev. 01) and 
    K-CREMCO (Rev. 00)
    OK = Test OK            N.A. = Test not applicable            Fail = Test failed
    Customer
     
    Service Keyboard type
    S/N
    Service engineer
     Date
      Visual inspection
           OK           N.A.          Fail      OK           N.A.          Fail
    1. Internal parts                             2. External parts
                               
    3. Cable                            
     Notes
     
      Functional inspection           
    4. Command board 
    software    KB
    5. Remote controller keys                             6. ComWheel
                               
    7. Electrical safety check                             8. Functioning after 
    electrical safety check
                               
    9. Final cleaning                                
    Notes 
     
     
    Used spare parts 