Remote Controller3 Document no. M1187342-004 2 Functional description 2.1 Remote Controller, K-REMCO/ K-CREMCO The Remote Controller consists of 12 direct function keys and the ComWheel. 2.1.1 K-REMCO/ K-CREMCO PCBs The K-REMCO/ K-CREMCO has two PCBs located inside the Remote Controller. One board has only the push button switches of the keys. The other board reads the status of the keys and the ComWheel and forwards the in formation to the CPU board. 2.1.2 External communication K-REMCO Rev. 00-01 Two signals, TXD and RXD in RS232 format are in use. No handshaking is used. Serial communication speed is 19.2 kbps. The 26-pin subminiature D-connector of the Remote Controller is connected to the Display Controller Board, B-DISP. K-REMCO Rev. 00 with optional Remote Controller - Compact Monitor cable or K-CREMCO Rev. 00 Two signals, TXD and RXD in RS232 format are in use. No handshaking is used. Serial communication speed is 19.2 kbps. In Compact Monitors, the DIN 5 connector is connected to the keyboard connector X9. In S/5 FM and FM Light, the DIN 5 connector is connected to Multi IO adapter X7. 2.1.3 ComWheel The ComWheel is used for menu selection.
Datex-Ohmeda E-Modules 4 Document no. M1187342-004 3 Service procedures 3.1 General service information The field service of the remote controllers is limited to replacing faulty circuit boards or mechanical parts. The circuit boards should be returned to GE Healthcare for repair. GE Healthcare is always available for service advice. Please provide the unit serial number, full type designation, and a detailed fault description. WARNING Only trained personnel with appropriate equipment should perform the tests and repairs outlined in this section. Unauthor ized service may void the warranty of the unit. 3.2 Service check These instructions include complete procedures for a service check. The service check is mandatory after any service repair. However, th e service check procedures can also be used for determining possible failures. The procedures should be performed in ascending order. The instructions in clude a check form (“APPENDIX A:” ) which may be used when performing the procedures. The symbol in the instructions indicate s that the check form contains space to record the results of the particular procedure. 3.2.1 Recommended tools 3.2.2 Visual inspection Turn the monitor to STBY. Disconnect the remote controller cable from the monitor. Detach the remote controller upper cover and the keypad cover by removing the screws (7 pcs) from the bottom. 1. Internal parts Check that: •cables are connected properly •the remote controller cable is fasten ed to the bottom cover with screws •the keypad switches are intact •the software EPROM under the keypad is attached properly ToolOrder No.Notes Screwdriver
Remote Controller5 Document no. M1187342-004 2. External partsCheck that: •the upper and bottom covers are intact •the keypad cover is intact •the ComWheel cover is intact and attached properly Reassemble the remote controller. 3. Cable Check the remote controller cable: •the cable is intact •the cable connector is intact •the connector pins are clean, straight and at about the same height •the locking screws inside the connector case are intact 3.2.3 Functional inspection Reconnect the cable to the moni tor and turn the monitor ON. 4. Command board software Enter the service menu: Menu (on the remote controller keypad) - Monitor Setup - Install/Service (password 16-4-34) - Service (password 26-23-8) Take down the information regarding the remote controller software. Select the menu KEYBOARD: Service - Keyboard 5. Remote controller keys Press the keys one by one. Check that each key generates a sound from the loudspeaker. 6. ComWheelTurn the ComWheel clockwise and counterclo ckwise and check that each step generates a sound and the corresponding values at the bottom of the menu increase. Select DUMMY PRESS. Push the ComWheel and check that the push generates a sound and the corresponding value in the menu increases.
Datex-Ohmeda E-Modules 6 Document no. M1187342-004 7. Electrical safety checkPerform an electrical safety check and a leakage current test. 8. Functioning after electrical safety checkCheck that the remote controller functions no rmally after the electrical safety check. 9. Final cleaning Clean the remote controller and the cable. •Fill in all necessary documents.
Remote Controller7 Document no. M1187342-004 3.3 Disassembly and reassembly 3.3.1 Before disassembly WARNINGWear a grounded, antistatic wristband when handling PC boards. Electrostatic discharge may damage components on the board. Handle all PC boards by their edges. 3.3.2 Tools needed 3.3.3 To disassemble Remote Co ntroller K-REMCO / K-CREMCO To disassemble the Remote Controller (see Figure 2): 1. Disconnect the K-REMCO/ K-CREMCO cable from the monitor. 2. Pull out the knob of the ComWheel. 3. Open the nut on the shaft of the ComWheel. 4. Open the three cross head screws on the bottom of the K-REMCO/ K-CREMCO. 5. Remove the top cover. 6. Open the four screws on the bottom of the K-REMCO/ K-CREMCO. 7. Remove the keyboard cover. 8. Disconnect the K-REMCO/ K-CREMCO cable and the wire set from the Comwheel. 9. Remove the PCBs. For more information on the spare parts, see the “Spare parts” slot. In reassembly, remember to put the reinforcing cord of the cable around the screw on the metal bridge before tightening the screw. - screwdriver - flat blade screwdriver - antistatic wristband
Datex-Ohmeda E-Modules 8 Document no. M1187342-004 Figure 2 K-REMCO, K-CREMCO disassembly and reassembly To reassemble the module, reverse the order of the disassembly steps. In reassembly, remember to put the reinforcing cord of the cable around the screw on the metal bridge before tightening the screw. CAUTION When reassembling the remote controller make sure that the cables are reconnected properly. Always perform the “Service check” after reassembling the remote controller.
Remote Controller9 Document no. M1187342-004 4Troubleshooting 4.1 K-REMCO, K-CREMCO See Keyboard Service Menu in the “Service Menu“ slot and perform tests available. If any of the tests fail, see explanation below.Problem Cause What to do ComWheel not working ComWheel leads broken or connector loose. ComWheel faulty. Check the items. Replace the ComWheel if necessary. Membrane key not working Switch cable loose or broken. Keyboard cable loose or broken. Cable connector pin failure. RS232 communication failure on CPU board. Check the items. Replace them if necessary.
Datex-Ohmeda E-Modules 10 Document no. M1187342-004 For your notes:
Appendix A, Service check form, Remote Controllers K-REMCO (Rev. 01) and K-CREMCO (Rev. 00) 1(2) Document no. M1187342-004 APPENDIX A: Service check form, Remote Controllers K-REMCO (Rev. 01) and K-CREMCO (Rev. 00) OK = Test OK N.A. = Test not applicable Fail = Test failed Customer Service Keyboard type S/N Service engineer Date Visual inspection OK N.A. Fail OK N.A. Fail 1. Internal parts 2. External parts 3. Cable Notes Functional inspection 4. Command board software KB 5. Remote controller keys 6. ComWheel 7. Electrical safety check 8. Functioning after electrical safety check 9. Final cleaning Notes Used spare parts Signature
Datex-Ohmeda E-Modules 2(2) Document no. M1187342-004 For your notes: