GE S5 Owners Manual
Have a look at the manual GE S5 Owners Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.
Functional check43 Document no. M1187317-009 Preset the measurement settings for those parameters that are connected, for example: Print/Recorder - Record Waveforms - Waveform 1 - ECG1 - Waveform 2 - P1 - Waveform 3 - P2 Invasive Pressures - P1 ‘ART’ Setup -- Label – ART - P2 ‘CVP’ Setup -- Label – CVP Others -SPO2 Setup - Pleth Scale –AUTO Airway Gas - CO2 Setup - Resp Rate Source – AUTO - Measurement – ON - Detection Limit – AUTO 5.3.2 Display 1. Check that the picture on the screen is displayed properly. 5.3.3 Keyboard(s) Tests with the Command Bar: Press the Monitor Setup key. Turn the ComWheel in bo th directions and check that the cursor in the menu moves correspondingly. Select Normal Screen and check that the menu disappears from the screen. Check the rest of the menu keys by pressing them one by one. Tests with the Remote Controller: Enter the Keyboard service menu. Check the function of the ComWheel. Press all keys. Check that each key produces a sound from the loudspeaker, or the Message count value in the service menu increases. 5.3.4 Frame unit 1. Check that the clock on the screen shows correct time. Readjust the time and date, if necessary. 5.3.5 Extension Module with CO2 measurement Wait until the message ‘Calibrating gas sensor’ disappears from the screen. 1. Block the tip of the sampling line with your finger and check that the message ‘Sample line blocked’ appears on the moni tor screen within 30 seconds. 2. Detach the Mini D-fend and check that the message ‘Check D-fend’ appears on the monitor screen within 30 seconds.
Datex-Ohmeda S/5 FM 44 Document no. M1187317-009 Breathe to the sampling line briefly. Check that the CO2 information is updated on the screen. 5.3.6 Multiparameter Hemodynamic Modules ECG and RESP measurements 1. Connect an ECG cable to the module. Connect the cable leads to a patient simulator. Check that all ECG and impedance respiratio n information is shown on the monitor screen as configured on the simulator. Turn the simulator off. Check that the ‘Asystole’ and ‘Apnea’ messages appear on the screen. Temperature measurement 2. Check the temperature channels with a patient simulator. Check that temperature measurement informat ion is shown on the monitor screen as configured on the simulator. Invasive blood pre ssure measurement 3. Check the function of the module and side panel membrane keys. 4. Check the InvBP channels with a patient simulator. 5. Zero the InvBP channels and check that the values and waveforms correspond to the simulator settings. SpO2 measurement 6. Connect an SpO2 finger probe to the module. Check that the message ‘Probe off’ is shown when the probe is not connected to a finger. 7. Attach the SpO2 probe to your finger. Check that a reading of 95-99 and a pleth waveform appear on the screen Non invasive blood pr essure measurement 8. Check the function of the module and side panel membrane keys. 9. Attach an adult NIBP cuff onto your arm and check that the module identifies the cuff, i.e. the text ‘Adult’ appears in the NIBP digit field for a short time. Perform a NIBP measurement and check that the module gives a reasonable measured result.
Functional check45 Document no. M1187317-009 5.3.7 Data Card and Menu Card function 1. Insert a Data card or a Menu card to the slot.Check that the corresponding symbol appears on the monitor screen. 5.3.8 Recorder 1. Press the Start/Stop sidepanel key and check that the module starts recording the selected waveforms. Press the Start/Stop sidepanel key again to stop recording. 2. Check that the quality of the recordings is acceptable. 5.3.9 Network connection 1. Check that the Mon-Net cable connector is clean and intact, then connect it to the Network connector on the backside of the monitor. Check that the monitor connects to the netw ork, i.e. the network symbol appears on the upper right-hand corner of the screen. Also a message regarding the connected Central should appear in the message field on the screen. 5.3.10 Wireless Network Option 1. Check that the WLAN signal strength symbol scrolls between zero and full or stays fixed on the monitor screen. 2. Check that the wireless LAN network symbol appears on the upper right-hand corner of the screen when the monitor conne cts to the Datex-Ohmeda Network. NOTE: If the monitor does not connect to the Datex-Ohmeda Network, check the WLAN configuration on the monitor and on the network. 5.3.11 Device Interfacing Solution, N-DISxxx 1. Make sure that the monitor receives all necessary parameter data from the connected devices. Check the screen conf iguration and the related interfacing settings, if necessary. Check also via the Interfacing menu that th e connected DIS module status is correct: Monitor Setup - Interfacing - Status Page
Datex-Ohmeda S/5 FM 46 Document no. M1187317-009 5.3.12 General• Switch the monitor to standby •Perform final cleaning •Fill in all necessary documents
General troubleshooting47 Document no. M1187317-009 6 General troubleshooting Figure 11 S/5 FM general troubleshooting flowchart No picture on screen Main CPU board may be faulty. Yes No NIBP pump starts pumping and SpO2 probes red LED lit?No No FM_general_trbl_flowchart.vsd Disconnect the mains power cord and remove the batteries to make a cold start. Connect the mains power cord Check the batteries charging levelsby pressing the test buttons on the batteries. Insert only one battery with 2 or more capacity leds illuminating. Press the ON/Standby key. DC/DC board failure Front panel greenbattery LED lit? Front panel green mains power LED lit?Replace fuses Mains power LED lit? DC/DC SW functioning. Disconnect the mains power cord and batteries. Reconnect and press the ON/Standby key Normal start-up sound and the alarm LEDs turn onand off? Possible keyboard or CPU failure. See Troubleshooting in S/5 FM frame slot Wait for about 1 minute andpress the NIBP Auto ON/ OFF key. NOTE: NIBP cuff must be connected Display failure. See Troubleshooting in S/5 FM frame slot No Yes No Possible AC/DC unit failure.Replace the AC/DC unit. Mains power LED lit? Yes Yes No Yes Yes W ith only one battery inserted orange battery led flashing? Yes No
Datex-Ohmeda S/5 FM 48 Document no. M1187317-009 For your notes:
Appendix A Functional check form, Datex-Ohmeda S/5 FMA-1(2) Document no. M1187317-009 APPENDIX A: Functional check form, Datex-Ohmeda S/5 FM OK = Test OK N.A. = Test not applicable Fail = Test failed Customer Service Service engineer Date Measuring equipment / test gases used: Equipment / tool / gas: Manufacturer: Model/Type/Part Number: Serial Number / ID:Calibration Date: Monitor Installation L- E- N- F- Visual Inspection OK N.A. Fail 1. Check all units visually. Functional Inspection OK N.A. Fail 5.3.1. General 5.3.2. Display S/N 5.3.3. Keyboard(s) S/N 5.3.4. Frame unit S/N 5.3.5. Extension Module with CO2 measurement S/N
Datex-Ohmeda S/5 FM A-2(2) Document no. M1187317-009 5.3.6. Multiparameter Hemodynamic ModulesS/N . ECG and RESP measurements . Temperature measurement . Invasive blood pressure measurement . SpO2 measurement . Non invasive blood pressure measurement 5.3.7. Data Card and Menu Card function 5.3.8. Recorder 5.3.9. Network connection 5.3.10. Wireless Network Option 5.3.11. Device Interfacing Solution, N-DISxxx S/N 5.3.12. General Notes Signature Functional Inspection OK N.A. Fail
Appendix B ElectroMagnetic CompatibilityB-1(4) Document no. M1187317-009 APPENDIX B: ElectroMagnetic Compatibility Table 1 Guidance and manufacturer’s declaration – electromagnetic emissions (1 S/5 FM equipped with S/5 Frame for FM, F-FM (2 S/5 FM equipped with S/5 Frame for FM with WLAN, F-FMW Guidance and manufacturer’s declar ation – electromagnetic emissions The S/5 FM is intended for use in the electromagne tic environment specified below. The customer or the user of the S/5 FM should assure that it is used in such an environment. Emissions test Compliance Electromagnetic environment - guidance RF emissions CISPR 11 Group 1 The S/5 FM (1 uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment. Group 2 The S/5 FM (2 must emit electromagnetic energy in order to perform its intended function. Nearby electronic equipment may be affected. RF emissions CISPR 11 Class A The S/5 FM is suitable for us e in all establishments other than domestic and those directly conne cted to the public low-voltage power supply network that supplies buildings used for domestic purposes. Harmonic emissions IEC 61000-3-2 Class A Voltage fluctuations/ flicker emissions IEC 61000-3-3 Complies
Datex-Ohmeda S/5 FM B-2(4) Document no. M1187317-009 Table 2 Guidance and manufacturer’s declaration – electromagnetic immunity Guidance and manufacturer’s decl aration – electromagnetic immunity The S/5 FM is intended for use in the electromagnetic en vironment specified below. The customer or the user of the S/5 FM should assure that it is used in such an environment. Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance Electrostatic discharge (ESD) IEC 61000-4-2 ±6 kV contact ±8 kV air ±6 kV contact ±8 kV air Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%. Electrical fast transients/bursts IEC 61000-4-4 ±2 kV for power supply lines ±1 kV for input/output lines ±2 kV for power supply lines ±1 kV for input/output lines Mains power quality should be that of a typical commercial or hospital environment. Surge IEC 61000-4-5 ±1 kV differential mode ±2 kV common mode ±1 kV differential mode ±2 kV common mode Mains power quality should be that of a typical commercial or hospital environment. Voltage dips, short interruptions and voltage variations on power supply lines IEC 61000-4-11 95% dip in UT) for 0.5 cycle 40% U T (60% dip in UT) for 5 cycles 70% U T (30% dip in UT) for 25 cycles 95% dip in UT) for 5 sec 95% dip in UT) for 0.5 cycle 40% U T (60% dip in UT) for 5 cycles 70% U T (30% dip in UT) for 25 cycles 95% dip in UT) for 5 sec Mains power quality should be that of a typical commercial or hospital environment. If user of the S/5 FM requires continued operation during power mains interruptions, it is recommended that the S/5 FM be powered from an uninterruptible power supply or a battery. Power frequency (50/60 Hz) magnetic field IEC 61000-4-8 3 A/m 3 A/m Power frequency magnetic field should be at levels characteristic of a typical location in a typical commercial or hospital environment. NOTE U T is the a.c. mains voltage prior to application of the test level.