Home > GE > Medical Monitor > GE S5 Owners Manual

GE S5 Owners Manual

    Download as PDF Print this page Share this page

    Have a look at the manual GE S5 Owners Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    							 Functional check43
    Document no. M1187317-009
    Preset the measurement settings for those parameters that are connected, for example:
    Print/Recorder - Record Waveforms - Waveform 1 - ECG1
                     - Waveform 2 - P1
                   - Waveform 3 - P2
    Invasive Pressures  -  P1 ‘ART’ Setup -- Label – ART
    - P2 ‘CVP’ Setup -- Label – CVP
    Others -SPO2 Setup - Pleth Scale –AUTO
    Airway Gas - CO2 Setup - Resp Rate Source – AUTO
    - Measurement – ON
    - Detection Limit – AUTO
    5.3.2 Display
    1. Check that the picture on the screen is displayed properly. 
    
    5.3.3 Keyboard(s)
    Tests with the Command Bar:
     Press the 
    Monitor Setup key. Turn the ComWheel in bo th directions and check that 
    the cursor in the menu moves correspondingly. Select Normal Screen  and check 
    that the menu disappears from the screen.
    Check the rest of the menu keys by pressing them one by one.
    Tests with the Remote Controller:
     Enter the  Keyboard  service menu.
     Check the function of the ComWheel.
     Press all keys. Check that each key produces a sound from the loudspeaker, or the 
    Message count value in the  service menu increases.
    
    5.3.4 Frame unit
    1. Check that the clock on the screen shows correct time. Readjust the time and date, if 
    necessary.
    
    5.3.5 Extension Module with CO2 measurement
    Wait until the message ‘Calibrating gas sensor’ disappears from the screen.
    1. Block the tip of the sampling line with your finger and check that the message ‘Sample  line blocked’ appears on the moni tor screen within 30 seconds.
    2. Detach the Mini D-fend and check that the message ‘Check D-fend’ appears on the  monitor screen within 30 seconds. 
    						
    							Datex-Ohmeda S/5 FM
    44
    Document no. M1187317-009
    Breathe to the sampling line briefly. Check that the CO2 information is updated on the 
    screen.
    
    5.3.6 Multiparameter Hemodynamic Modules
    ECG and RESP measurements
    1. Connect an ECG cable to the module. Connect the cable leads to a patient simulator. 
    Check that all ECG and impedance respiratio n information is shown on the monitor 
    screen as configured on the simulator.
    Turn the simulator off. Check that the ‘Asystole’ and ‘Apnea’ messages appear on the screen.
    
    Temperature measurement
    2. Check the temperature channels with a patient simulator.
    Check that temperature measurement informat ion is shown on the monitor screen as 
    configured on the simulator.
    
    Invasive blood pre ssure measurement
    3. Check the function of the module and side panel membrane keys.
    4. Check the InvBP channels with a patient simulator.
    5. Zero the InvBP channels and check that  the values and waveforms correspond to the 
    simulator settings.
    
    SpO2 measurement
    6. Connect an SpO2 finger probe to the module. Check that the message ‘Probe off’ is shown  when the probe is not connected to a finger.
    7. Attach the SpO2 probe to your finger. Check that a reading of 95-99 and a pleth  waveform appear on the screen
    
    Non invasive blood pr essure measurement
    8. Check the function of the  module and side panel membrane keys.
    9. Attach an adult NIBP cuff onto your arm and  check that the module identifies the cuff, i.e. 
    the text ‘Adult’ appears in the NIBP digit field for a short time.
    Perform a NIBP measurement and check that the module gives a reasonable measured 
    result.  
     
    						
    							 Functional check45
    Document no. M1187317-009
    5.3.7 Data Card and Menu Card function
    1. Insert a Data card or a Menu card to the slot.Check that the corresponding symbol appears on the monitor screen.
    
    5.3.8 Recorder 
    1. Press the Start/Stop  sidepanel key and check that the module starts recording the 
    selected waveforms. Press the 
    Start/Stop  sidepanel  key again to stop recording.
    2. Check that the quality of the recordings is acceptable.
    
    5.3.9 Network connection
    1. Check that the Mon-Net cable connector is  clean and intact, then connect it to the 
    Network connector on the backside of the monitor.
    Check that the monitor connects to the netw ork, i.e. the network symbol appears on the 
    upper right-hand corner of the screen. Also  a message regarding the connected Central 
    should appear in the message field on the screen.
    
    5.3.10 Wireless Network Option
    1. Check that the WLAN signal strength symbol  scrolls between zero and full or stays fixed 
    on the monitor screen.
    2. Check that the wireless LAN network symbol appears on the upper right-hand corner of  the screen when the monitor conne cts to the Datex-Ohmeda Network.
    NOTE: If the monitor does not connect to  the Datex-Ohmeda Network, check the WLAN 
    configuration on the monitor and on the network.
    
    5.3.11 Device Interfacing Solution, N-DISxxx
    1. Make sure that the monitor  receives all necessary parameter data from the connected 
    devices. Check the screen conf iguration and the related interfacing settings, if necessary. 
    Check also via the Interfacing menu that th e connected DIS module status is correct:
    Monitor Setup - Interfacing - Status Page
     
    						
    							Datex-Ohmeda S/5 FM
    46
    Document no. M1187317-009
    5.3.12 General•
    Switch the monitor to standby
    •Perform final cleaning
    •Fill in all necessary documents
     
    						
    							 General troubleshooting47
    Document no. M1187317-009
    6 General troubleshooting    
    Figure 11 S/5 FM general  troubleshooting flowchart
    No picture on screen
    Main CPU board may be
    faulty.
    Yes
    No
    NIBP pump starts
    pumping and SpO2
    probes red LED lit?No
    No
    FM_general_trbl_flowchart.vsd
    Disconnect the mains power cord and
    remove the batteries to make a cold start.
    Connect the mains power cord
    Check the batteries charging levelsby pressing the test buttons on the
    batteries.
    Insert only one battery with 2 or
    more capacity leds illuminating.
    Press the ON/Standby key.
    DC/DC board failure
    Front panel greenbattery LED lit?
    Front
    panel green mains power LED lit?Replace fuses
    Mains
    power LED lit?
    DC/DC SW functioning.
    Disconnect the mains
    power cord and batteries. Reconnect and press the ON/Standby key
    Normal start-up
    sound  and the alarm LEDs turn onand off?
    Possible keyboard or CPU
    failure. See Troubleshooting in S/5 FM frame slot
    Wait for about 1 minute andpress the NIBP Auto ON/
    OFF key. NOTE: NIBP cuff
    must be connected
    Display failure. See
    Troubleshooting in S/5 FM frame slot
    No
    Yes
    No
    Possible AC/DC unit failure.Replace the AC/DC unit.
    Mains
    power LED lit?
    Yes
    Yes
    No
    Yes
    Yes
    W ith only one
    battery inserted
    orange battery led flashing?
    Yes
    No 
    						
    							Datex-Ohmeda S/5 FM
    48
    Document no. M1187317-009
    For your notes: 
    						
    							Appendix A Functional check form, Datex-Ohmeda S/5 FMA-1(2)
                               Document no. M1187317-009
    APPENDIX A: Functional check form, 
    Datex-Ohmeda S/5
     FM 
    OK = Test OK            N.A. = Test not applicable            Fail = Test failed
    Customer
    Service
    Service engineer
    Date
    Measuring equipment / test gases used:
    Equipment / tool / gas:  Manufacturer: Model/Type/Part 
    Number: Serial Number / 
    ID:Calibration 
    Date:
    Monitor Installation
    L- E- N-
    F-
    Visual Inspection          OK           N.A.          Fail
    1. Check all units visually.                            
    Functional Inspection      OK           N.A.          Fail
    5.3.1. General                            
    5.3.2. Display S/N
                               
    5.3.3. Keyboard(s) S/N
                               
    5.3.4. Frame unit S/N
                               
    5.3.5. Extension Module with CO2 measurement S/N
                                
    						
    							Datex-Ohmeda S/5 FM
    A-2(2)
    Document no. M1187317-009
    5.3.6. Multiparameter Hemodynamic ModulesS/N
                               
    . ECG and RESP measurements                            
    . Temperature measurement                            
    . Invasive blood pressure measurement                            
    . SpO2 measurement                            
    . Non invasive blood pressure measurement                            
    5.3.7. Data Card and Menu Card function                            
    5.3.8. Recorder                            
    5.3.9. Network connection                            
    5.3.10. Wireless Network Option                            
    5.3.11. Device Interfacing Solution, N-DISxxx S/N
                               
    5.3.12. General                            
    Notes
     
     
     
     
    Signature Functional Inspection       OK          N.A.          Fail 
    						
    							Appendix B ElectroMagnetic CompatibilityB-1(4)
                               Document no. M1187317-009
    APPENDIX B: ElectroMagnetic Compatibility
    Table 1 Guidance and manufacturer’s  declaration – electromagnetic 
    emissions
    (1  S/5 FM equipped with S/5 Frame for FM, F-FM 
    (2  S/5 FM equipped with S/5 Frame for FM with WLAN, F-FMW
    Guidance and manufacturer’s declar ation – electromagnetic emissions
    The S/5 FM is intended for use in the electromagne tic environment specified below. The customer or the 
    user of the S/5 FM should assure that it is used in such an environment.
    Emissions test Compliance Electromagnetic environment - guidance
    RF emissions
    CISPR 11 Group 1
    The S/5 FM
    (1 uses RF energy only for its internal function. 
    Therefore, its RF emissions are very low and are not likely to 
    cause any interference in nearby electronic equipment.
    Group 2 The S/5 FM
    (2 must emit electromagnetic energy in order to 
    perform its intended function. Nearby electronic equipment may 
    be affected.
    RF emissions
    CISPR 11 Class A The S/5 FM is suitable for us
    e in all establishments other than 
    domestic and those directly conne cted to the public low-voltage 
    power supply network that supplies buildings used for domestic 
    purposes.
    Harmonic emissions
    IEC 61000-3-2 Class A
    Voltage fluctuations/ 
    flicker emissions
    IEC 61000-3-3 Complies 
    						
    							Datex-Ohmeda S/5 FM
    B-2(4)
    Document no. M1187317-009
    Table 2 Guidance and manufacturer’s declaration – electromagnetic 
    immunity
    Guidance and manufacturer’s decl aration – electromagnetic immunity
    The S/5 FM is intended for use in the electromagnetic en vironment specified below. The customer or the user of 
    the S/5 FM should assure that it is used in such an environment.
    Immunity test IEC 60601 test level Compliance level Electromagnetic environment -  guidance
    Electrostatic 
    discharge (ESD)
    IEC 61000-4-2 ±6 kV contact
    ±8 kV air ±6 kV contact
    ±8 kV air Floors should be wood, concrete or 
    ceramic tile. If floors are covered with 
    synthetic material, the relative humidity 
    should be at least 30%.
    Electrical fast 
    transients/bursts
    IEC 61000-4-4 ±2 kV for power 
    supply lines
    ±1 kV for input/output 
    lines ±2 kV for power 
    supply lines
    ±1 kV for input/output 
    lines Mains power quality should be that of a 
    typical commercial or hospital 
    environment.
    Surge
    IEC 61000-4-5 ±1 kV differential 
    mode
    ±2 kV common mode ±1 kV differential 
    mode
    ±2 kV common mode Mains power quality should be that of a 
    typical commercial or hospital 
    environment.
    Voltage dips, short 
    interruptions and 
    voltage variations on 
    power supply lines
    IEC 61000-4-11 95% dip in UT)
    for 0.5 cycle
    40% U
    T
    (60% dip in UT)
    for 5 cycles
    70% U
    T
    (30% dip in UT)
    for 25 cycles
    95% dip in UT)
    for 5 sec 95% dip in UT)
    for 0.5 cycle
    40% U
    T
    (60% dip in UT)
     for 5 cycles
    70% U
    T
    (30% dip in UT)
    for 25 cycles
    95% dip in UT)
    for 5 sec Mains power quality should be that of a 
    typical commercial or hospital 
    environment. If user of the S/5 FM 
    requires continued operation during 
    power mains interruptions, it is 
    recommended that the S/5 FM be 
    powered from an uninterruptible power 
    supply or a battery.
    Power frequency 
    (50/60 Hz) magnetic 
    field
    IEC 61000-4-8 3 A/m 3 A/m Power frequency magnetic field should 
    be at levels characteristic of a typical 
    location in a typical commercial or 
    hospital environment.
    NOTE U
    T is the a.c. mains voltage prior to application of the test level. 
    						
    All GE manuals Comments (0)