Frame for FM 39 Document no. M1187335-004 To change the LCD backlight - Disconnect the connector from the inverter board. - Push the lid beside the backlight to release the backlight tube. Pull the backlight carefully out. NOTE: Do not bend the backlight tube. NOTE: Do not touch the backlight tube with your hands. To separate the LCD display from the frame unit - Remove the four (T10) screws holding the display to the frame unit. - Disconnect the connector from the inverter board. - Lift the display carefully up. - Disconnect the LCD display flat cable carefully from the display. NOTE: When reassembling the LCD display, be careful that no dirt or finger prints are left between the LCD display element and the protection glass window. NOTE: Do not use excessive force when fastening the display to the frame. Fastening the screws too tightly might bend the display too much and break it. The screws must also be fastened gradually, first fasten all screws slightly, then more tightly.
![](/img/blank.gif)
Datex-Ohmeda S/5 FM 40 Document no. M1187335-004 3.3.2 Handling and storage of LCD display component Handling of LCD display comp onent and protective window If the LCD Display component surface becomes dusty, wipe it gently with absorbent cotton, chamois or other soft material. If necessary, breathe onto the display surface and wipe immediately. The display surface may also be cleaned using a small amount of normal hexane. Do not use acetone, toluene or alcohol because they cause chemical damage to the polarizer. 1. Wipe off saliva and water drops as soon as possible. Their prolonged contact with the polarizer cause deformations and color fading. 2. Do not open the component case because inte rnal circuits are sensitive to electrostatic discharges. Taking a spare part LCD Display component or a display shield into use: Peel off the protective film slowly (in more th an 10 seconds) from the display or protective window surface. Fast peeling may generate enough static electricity to destroy the LCD Display component. Storaging an LCD display component as a spare part for a long period 1. Store the display in a dark place. Do not expose it to sunlight or fluorescent light. Keep the temperature between 5 °C and 35 °C / 41 °F and 95 °F at normal humidity. 2. The polarizer surface should not come into contact with any other object. It is recommended that the display unit is stored in the container in which it was originally shipped.
![](/img/blank.gif)
Frame for FM 41 Document no. M1187335-004 To remove the inverter board NOTE: The inverter board is sensitive to bending. - Remove the two (T10) screws and the insulator plate. - Disconnect the backlight connector(s) and the inverter cable connector. NOTE: When reassembling make sure that the connectors are fastened properly. Inverter Board F-FM(W)-01 Inverter Board F-FM(W)-00 NOTE: When replacing the Inverter board, fasten the inverter screws gently, using only as much force as necessary to fasten the board firmly enough in place. The screws must not be fastened so tightly that the board will be bent from around the screws. To remove the handle - Remove the two screws (T15) holding the handle to the handle guide unit.
![](/img/blank.gif)
Datex-Ohmeda S/5 FM 42 Document no. M1187335-004 To remove the EMC shield - Remove the 14 screws (T10) holding the shield to the frame unit. One of the screws is under the Network connection cable. Lift the shield carefully up. NOTE: When reassembling the monitor, remember to connect the grounding wire of the Sync connector unit to the EMC shield. Now you can access the following field replaceable spare parts. Follow the instructions below: - “To remove the handle guide unit” - “To remove the CPU board” - “To remove the loudspeaker unit” - To remove the Network cable, disconnect the cable connectors from the CPU board and the Network connector board. To remove the ha ndle guide unit - Remove the WLAN antenna from the handle guide. - Remove the two screws (T10) holding the handle guide unit to the frame.
![](/img/blank.gif)
Frame for FM 43 Document no. M1187335-004 To remove the CPU board - Disconnect the following cables from the CPU board: • loudspeaker cable • Sync connector cable • interconnection board connector cable • network connection cable - Remove the eight screws (T10) holding the CPU board to the frame. - Lift the CPU board carefully up and remove it from the frame. To remove the loudspeaker unit - Remove the two screws (T10) holding the loudspeaker unit to the frame. - Disconnect the loudspeaker cable from the CPU board.
![](/img/blank.gif)
Datex-Ohmeda S/5 FM 44 Document no. M1187335-004 To remove the interconnection board - Remove the three screws (T10) holding the plastic flex board cover. - Turn the frame around. Remove the two screws (T10) holding the flex board to the frame. - Disconnect the interconnection board connector from the module bus cable. NOTE: When reassembling the monitor, check that a cellular rubber sealing is attached to the module bus cable connector. - Remove the interconnection board carefully from the frame.
![](/img/blank.gif)
Frame for FM 45 Document no. M1187335-004 To remove the power supply frame - Remove the two screws (T15) holding the power supply frame to the inner frame. - Now you have access to replace the inner frame and Modulebus cable unit sub- assembly.
![](/img/blank.gif)
Datex-Ohmeda S/5 FM 46 Document no. M1187335-004 3.4 Batteries 3.4.1 Battery indicators The S/5 FM messages, screen symbols and front panel LED indicators tell the user about the status of the batteries. For screen symbols, see “Part 1 Symbols”. For LED indicators, consult the table below and for messages, see section “Troubleshooting.” Table 5 Battery indicators Screen symbol Explanation Front panel battery LED indicators Monitor is battery powered. Batteries are fully charged and the size of the green bar indicates the charging level.green lit orange dark Monitor is battery powered. Battery A is empty, battery B is ok. green lit orange dark Monitor is battery powered. Battery A failure, battery B is ok. green lit orange flashing Monitor is mains powered. Battery A is being charged (white bar), battery B is already charged. green dark orange lit Monitor is battery powered. Battery A is missing, battery B is ok. green lit orange flashing no screen symbol Monitor is mains powered. No battery backup: batteries have failed or they are not inserted. green dark orange flashing B BA AB B B B
![](/img/blank.gif)
Frame for FM 47 Document no. M1187335-004 3.4.2 To change the batteries Figure 14 Inserting a battery (1) Open the lid of the battery compartment located behind the handle by sliding it to the left. (2) Lift up the battery you want to change. Check the indicators and messages on the screen to make sure that you change the battery with lower charge. (3) Push in the new battery. Make sure that the charging indicator is facing forward and push the battery down all the way. Check the monitor indicators. CAUTION After having replaced a battery, alwa ys make sure that you close the battery compartment by sliding the lid bac k to the right until it clicks. WARNING Do not incinerate a battery or store at high temperatures as it will explode. 3.4.3 To check the battery Figure 15 Capacity indicator on the battery Check the battery charging level by pressing the test button on the battery as indicated in the drawing on the left. The green bar lights up and the number of lit segments indicates the charging level: the more lit segments, the higher the capacity level.
![](/img/blank.gif)
Datex-Ohmeda S/5 FM 48 Document no. M1187335-004 3.4.4 Conditioning the batteries Condition the batteries regularly to maintain their useful life. Condition a battery every six months or when the message ‘Condition Battery x’‘ appears on the screen. Always observe the messages and symbols on the screen to see the battery status. You can also check the status through Monitor Setup - Battery Setup . Conditioning a battery is best done on an external charger. If you do not have an external charger, proceed according to the following instructions. NOTE: You cannot condition batteries during patient monitoring. Always disconnect the modules first. 1. Continue normal battery use until the green bar of a battery charge indicator is less than 3/4 of the full height. After th is, remove the battery. Continue monitoring with one battery until its charge is less than 3/4 of the full capacity. 2. Insert both batteries and connect the moni tor to the power supply. The monitor starts charging both batteries, and the capacity indicators scroll accordingly. Keep charging the batteries until both capacity indicators are full height. 3. Continue charging for another two hours. Af ter this, check that the orange battery LED in the front panel is no longer on. If it is, continue charging until it goes off. 4. Disconnect the monitor from the power supply and leave it on until the batteries run out and the monitor switches off. Wait for another 15 minutes. 5. Reconnect the monitor to the power supply and turn it on. Continue charging the batteries until both capacity indicators are full height and no longer scrolling. 6. Keep charging for another two hours. After this, check that the orange battery LED in the front panel is no longer on. If it is, continue charging until it goes off to indicate that the battery conditioning is complete. NOTE: If only one battery is inserted during co nditioning, the orange LED will keep flashing and will not go off. 3.5 To replace the fuses 3.5.1 Primary fuses Pull out the fuse holder under the mains connec tor at the back of the monitor. Replace the fuses with fuses of exactly the same type and rating. 3.5.2 Fuse for external DC External DC power for future use only. Remove the rear cover. Replace the fuse that is located inside the fuse holder under the receptacle with a fuse of exactly the same type and rating. 3.6 To download the software on CPU board In case of a faulty CPU board refer first to se ction “To remove the CPU board“ and then follow the instructions for down loading service software. NOTE: All user settings will be lost afte r downloading of new monitor software. NOTE: During the downloading of software the serial number of the monitor is written on the software card. The software can then be down loaded again to the same monitor, but not to any other monitor. (For F-FM(W)-00 frame version).