    							Service Menu33
    Document no. M1187329-003
    4. When the values are equal, push the ComWheel to confirm the calibration. After a few seconds the ‘Calibrated’ message  will appear in the NIBP digit field, which means that the 
    calibration data has now been saved.
    NOTE:  When calibrating NIBP, always change the displayed pressure value slightly with the 
    ComWheel, even in cases where the value woul d be correct. For example, change the value 
    one step higher and then back one step lower. The ‘Calibrated’ text should appear in the 
    display. This ensures that the calibration procedure is correctly registered and stored by the 
    module.
    5. Use the module buttons again to enable  Protection setting and set it ON, and finally 
    disable  Protection  setting.
    4.4.3 NIBP Safety Valve
    Test: Start test  is for starting and  Stop test is for 
    stopping the Safety Valve test.
    Safety Valve Data :  
    For information on general items  Pressure, Zero, 
    Protect handle, Calibr. prot., +15 V power, AD0  to 
    AD7  as well as  Timeouts etc., see service data 
    descriptions in section “ 4.4 NIBP Module”.
    Max. press  and 2 s after stop  show the measured 
    values at Safety Valve test.
    Safety Valve Test Adult/Infant
    Wrap an adult cuff around a pipe and connect the 
    cuff to the module. Highlight  Start test and give the 
    ComWheel a push. The test ends automatically or 
    when  Stop test  (appears in place of  Start test) is 
    pushed. 
    Max. press  indicates the pressure at which the 
    safety valve opens and is normally 300  ±15 mmHg 
    for adult and 150 mmHg ±15 mmHg for infant. 
    2 s after stop  indicates the pressure at 2 seconds 
    after the pump has stopped and is normally > 270 mmHg  for adult and > 130 mmHg for infant. If the value is 
    less, check leakage by the active leak test. 
    						
    							Datex-Ohmeda S/5 FM
    34
    Document no. M1187329-003
    4.4.4 NIBP Pulse ValveHow to check Pulse Valve
    Wrap an adult cuff around a pipe and connect the cuff to the module. Select  Start test and 
    push the ComWheel. The pressure rises beyond 240 mmHg and stops. The pulse valve opens. 
    The module counts the time it takes for the pressure to go down from 240 mmHg to 50 mmHg 
    and displays it on the screen. The test can be manually stopped by selecting  Stop test.
    The valve can be adjusted between 0 and 255 (0  for fully closed and 255 for fully open). First 
    select  Set Valve  and push the ComWheel. See the pulse valve value and adjust it by turning the 
    ComWheel. Then push the ComWheel to confirm the value.
    The  Interval 240 mmHg -> 50 mmHg  time should be less than 60 seconds when the valve is 
    150 and less than 10 when fully opened (255). When fully closed (0), the system should be 
    airtight and the pressure does not drop. Depending on an individual, the pulse valve may 
    remain closed up to approx. value 100.
    If the measured time deviates much from those above, then the pulse valve or its tubes are 
    faulty.
    Start test  is for starting and  Stop test is for 
    stopping the test.
    Set Valve  lets you adjust the opening of the pulse 
    valve.
    Pulse Valve Data
    For information on general items  Pressure, Zero, 
    Protect handle, Calibr. prot., +15 V power, AD0  to 
    AD7  as well as  Timeouts etc., see section “ 4.4 NIBP 
    Module ”.  
    						
    							Service Menu35
    Document no. M1187329-003
    4.4.5 NIBP Buttons/Leds
    The selections  Auto ON/OFF, Manual ON/OFF , STAT ON/OFF , and Measur. ON/OFF .
    Buttons/Leds Data
    For information on general items  Pressure, Zero, 
    Protect handle, Calibr. prot., +15 V power, AD0  to 
    AD7  as well as  Timeouts etc., see section “ 4.4 NIBP 
    Module ”. 
    Buttons Checking
    The front panel keys function is confirmed by 
    pressing and releasing the key and observing that 
    OFF turns to  ON at Auto On/Off , and Start Cancel . 
    Set Cycle Time and STAT On/Off are not in use with 
    E-PSM(P). 
    						
    							Datex-Ohmeda S/5 FM
    36
    Document no. M1187329-003
    4.4.6 NIBP PneumaticsHow to check Interval 20 mmHg -> 185 mmHg 
    Select Start pump  at different combinations of the valves open/closed and push the 
    ComWheel. The module counts the time it takes for the pressure to go up from 20 mmHg to 
    185 mmHg and displays it. When all the valves are closed, the pump should be able to pump 
    the pressure in about 1 to 4 seconds into an adult cuff wrapped around a pipe. The pump does 
    not stop without selecting  Stop Pump by pushing the ComWheel.
    Start Pump/Stop Pump:  A manual control for the 
    pump. The selection changes to  Stop Pump when the 
    pump turns on.
    Open Exh1/Close Exh1:   This function is not used 
    with E-PSM(P) modules.
    Open Exh2/Close Exh2:  A manual control for the 
    exhaust valve 2. The selection changes to  Close Exh2 
    when the valve is opened.
    Open Zero valve:  This function is not in use with E-
    PSM(P) modules.  
    Set Valve:  The opening of the pulse valve is adjusted 
    between 0 and 255 (0 for fully  closed and 255 for fully 
    open). First push the ComWheel, then turn it to adjust 
    the value on the screen and finally push to set the 
    value.
    Reset Clock  is not in use with E-PSM(P) modules.
    Pneumatics Data field
    For information on general items  Pressure, Zero, 
    Protect handle, Calibr. prot., +15 V power, AD0  to AD7  as well as  Timeouts etc., see section “ 4.4 NIBP 
    Module ”.  
    Pump, Exh1 Valve , and Exh2 Valve  show their states. NOTE:   Exh1 Valve has no effect on the module.
    Pulse Valve shows how much the valve is opened (0 to 255) during Valve Setting.
       
    						
    							Service Menu37
    Document no. M1187329-003
    4.4.7 NIBP Watchdog
    Test ADULT:  No effect on the module.
    Test INFANT:  No effect on the module.
    Stop Test: No effect on the module.
    Watchdog Data field
    For information on general items  Pressure, Zero, 
    Protect handle, Calibr. prot., +15 V power, AD0  to 
    AD7  as well as  Timeouts etc., see section “ 4.4 NIBP 
    Module ”.
    Watchdog Interval:  No effect on the module.
    Adult watchdog time testing
    No effect on the module.
    Infant watchdog time testing
    No effect on the module. 
    						
    							Datex-Ohmeda S/5 FM
    38
    Document no. M1187329-003
    4.5 DIS Interfacing
    4.5.1 InterfacingHow to interface
    To select the parameter data source:
    •Press the Monitor Setup and select  Interfacing.
    •Select the desired measurement parameter.
    •Select the desired source by name. 
    NOTE: The name of the device is visible on the list only if the device is correctly connected to the 
    module. 
    						
    							Service Menu39
    Document no. M1187329-003
    4.5.2 Status Page
    4.5.3 DIS Interfacing service menu
    Access via Monitor setup - Interfacing.
    The selection  Next page is available, if more than 8 
    DIS modules are connected to the DIS bus 
    simultaneously.
    Interfaces
    The menu displays a list of all connected DIS modules 
    and the statuses of the corresponding external 
    devices.
    If the bus voltage is too low, you cannot add more 
    devices. ‘DIS module bus voltage low. Do not add 
    more devices or reduce cable length’ message 
    appears.
    Access via Monitor setup - Install/Service-
    Service
    The menu displays submenus for all connected DIS 
    modules. 
    						
    							Datex-Ohmeda S/5 FM
    40
    Document no. M1187329-003
    4.5.4 DIS Module specific pageService Data
    Product name: DIS module name.
    Product type: DIS module type.
    Driver sw id:  DIS module software and its release 
    date.
    Short product name:  DIS module name.
    Module serial number: DIS module serial number.
    HW card type:  PCB type.
    HW id:  DIS module PCB identification number.
    HW test date:  DIS module PCB testing date.
    Unit serial number:  DIS module PCB serial 
    number.
    Comment field:  Indicates the status of the 
    external device.
    bus voltage : DIS bus voltage.
    tout : DIS module timeouts, seen by the monitor. The value should not increase more than by 5 per second.
    cse : DIS module checksum errors, seen by the monitor.  The value should not increase more than by 5 per 
    second.
    rx: The number of data packets from the external device received by the DIS module.
    rx err:  The number of data errors from the exte rnal device received by the DIS module.
    tx:  The number of data packets to the external device sent by the DIS module.
    Ram:  Status of DIS module RAM memory.
    Rom:  Status of DIS module ROM memory.
    EEPROM:  Status of DIS module EEPROM memory.  
    						
    							Service Menu41
    Document no. M1187329-003
    5 Set/Test
    The power supply unit contains a watchdog 
    circuitry, which needs refreshment at every 1.5 
    seconds. If the refreshm ent did not occur, the 
    watchdog circuitry will re set the main CPU. In 
    normal operation, the main CPU refreshes the 
    watchdog circuitry at every 0.2 seconds.
    The purpose of the watchdog circuitry is to restart 
    the monitor, if there was a serious malfunction. 
    This feature is useful in two cases: when the main 
    CPU is not able to contro l the monitor, or when 
    the CPU controls the monitor but detects a 
    serious malfunction. Watchdog tests check 
    proper functionality of the watchdog circuitry in 
    various conditions.
    Watchdog  test ensures directly that the 
    watchdog of the power unit functions properly. 
    Choosing this test prevents the watchdog 
    circuitry from refreshing and shows running 
    seconds with an accuracy of 0.1 seconds.
    The test should have the following result when 
    the watchdog circuitry is working properly: The monito r will restart after 1.5 seconds from the start of the 
    test. In malfunction: ‘>20 s’ is disp layed, and the test will be interrupted. In this case, the fault is in the 
    watchdog circuitry of the power unit.
    WD by Overload  test ensures the functionality of a feature,  where the software controls the monitor, but 
    detects an overload situation in the main CPU.
    The test should have the following re sult when the feature is working properly: The mo nitor will restart after 
    15 seconds from the start of the test.
    Factory Reset restores the factory default settings and clears the data memories. Factory reset should be 
    run if the monitor software is replaced or  if the SRAM/Timekeeper battery is replaced. 
    						
    							Datex-Ohmeda S/5 FM
    42
    Document no. M1187329-003
    5.1 Country Settings
    Language
    Select the software language.
    National Reqs
    Select the software features, which include national 
    requirements.
    Power Frequency
    Set the power frequency (50 Hz/60 Hz). This setting is 
    used to filter out possible power frequency 
    interference from the parameter measurements.
    Time format
    Set the time format of the real time clock (24 h/12 h). 
    						