Service Menu33 Document no. M1187329-003 4. When the values are equal, push the ComWheel to confirm the calibration. After a few seconds the ‘Calibrated’ message will appear in the NIBP digit field, which means that the calibration data has now been saved. NOTE: When calibrating NIBP, always change the displayed pressure value slightly with the ComWheel, even in cases where the value woul d be correct. For example, change the value one step higher and then back one step lower. The ‘Calibrated’ text should appear in the display. This ensures that the calibration procedure is correctly registered and stored by the module. 5. Use the module buttons again to enable Protection setting and set it ON, and finally disable Protection setting. 4.4.3 NIBP Safety Valve Test: Start test is for starting and Stop test is for stopping the Safety Valve test. Safety Valve Data : For information on general items Pressure, Zero, Protect handle, Calibr. prot., +15 V power, AD0 to AD7 as well as Timeouts etc., see service data descriptions in section “ 4.4 NIBP Module”. Max. press and 2 s after stop show the measured values at Safety Valve test. Safety Valve Test Adult/Infant Wrap an adult cuff around a pipe and connect the cuff to the module. Highlight Start test and give the ComWheel a push. The test ends automatically or when Stop test (appears in place of Start test) is pushed. Max. press indicates the pressure at which the safety valve opens and is normally 300 ±15 mmHg for adult and 150 mmHg ±15 mmHg for infant. 2 s after stop indicates the pressure at 2 seconds after the pump has stopped and is normally > 270 mmHg for adult and > 130 mmHg for infant. If the value is less, check leakage by the active leak test.
![](/img/blank.gif)
Datex-Ohmeda S/5 FM 34 Document no. M1187329-003 4.4.4 NIBP Pulse ValveHow to check Pulse Valve Wrap an adult cuff around a pipe and connect the cuff to the module. Select Start test and push the ComWheel. The pressure rises beyond 240 mmHg and stops. The pulse valve opens. The module counts the time it takes for the pressure to go down from 240 mmHg to 50 mmHg and displays it on the screen. The test can be manually stopped by selecting Stop test. The valve can be adjusted between 0 and 255 (0 for fully closed and 255 for fully open). First select Set Valve and push the ComWheel. See the pulse valve value and adjust it by turning the ComWheel. Then push the ComWheel to confirm the value. The Interval 240 mmHg -> 50 mmHg time should be less than 60 seconds when the valve is 150 and less than 10 when fully opened (255). When fully closed (0), the system should be airtight and the pressure does not drop. Depending on an individual, the pulse valve may remain closed up to approx. value 100. If the measured time deviates much from those above, then the pulse valve or its tubes are faulty. Start test is for starting and Stop test is for stopping the test. Set Valve lets you adjust the opening of the pulse valve. Pulse Valve Data For information on general items Pressure, Zero, Protect handle, Calibr. prot., +15 V power, AD0 to AD7 as well as Timeouts etc., see section “ 4.4 NIBP Module ”.
![](/img/blank.gif)
Service Menu35 Document no. M1187329-003 4.4.5 NIBP Buttons/Leds The selections Auto ON/OFF, Manual ON/OFF , STAT ON/OFF , and Measur. ON/OFF . Buttons/Leds Data For information on general items Pressure, Zero, Protect handle, Calibr. prot., +15 V power, AD0 to AD7 as well as Timeouts etc., see section “ 4.4 NIBP Module ”. Buttons Checking The front panel keys function is confirmed by pressing and releasing the key and observing that OFF turns to ON at Auto On/Off , and Start Cancel . Set Cycle Time and STAT On/Off are not in use with E-PSM(P).
![](/img/blank.gif)
Datex-Ohmeda S/5 FM 36 Document no. M1187329-003 4.4.6 NIBP PneumaticsHow to check Interval 20 mmHg -> 185 mmHg Select Start pump at different combinations of the valves open/closed and push the ComWheel. The module counts the time it takes for the pressure to go up from 20 mmHg to 185 mmHg and displays it. When all the valves are closed, the pump should be able to pump the pressure in about 1 to 4 seconds into an adult cuff wrapped around a pipe. The pump does not stop without selecting Stop Pump by pushing the ComWheel. Start Pump/Stop Pump: A manual control for the pump. The selection changes to Stop Pump when the pump turns on. Open Exh1/Close Exh1: This function is not used with E-PSM(P) modules. Open Exh2/Close Exh2: A manual control for the exhaust valve 2. The selection changes to Close Exh2 when the valve is opened. Open Zero valve: This function is not in use with E- PSM(P) modules. Set Valve: The opening of the pulse valve is adjusted between 0 and 255 (0 for fully closed and 255 for fully open). First push the ComWheel, then turn it to adjust the value on the screen and finally push to set the value. Reset Clock is not in use with E-PSM(P) modules. Pneumatics Data field For information on general items Pressure, Zero, Protect handle, Calibr. prot., +15 V power, AD0 to AD7 as well as Timeouts etc., see section “ 4.4 NIBP Module ”. Pump, Exh1 Valve , and Exh2 Valve show their states. NOTE: Exh1 Valve has no effect on the module. Pulse Valve shows how much the valve is opened (0 to 255) during Valve Setting.
![](/img/blank.gif)
Service Menu37 Document no. M1187329-003 4.4.7 NIBP Watchdog Test ADULT: No effect on the module. Test INFANT: No effect on the module. Stop Test: No effect on the module. Watchdog Data field For information on general items Pressure, Zero, Protect handle, Calibr. prot., +15 V power, AD0 to AD7 as well as Timeouts etc., see section “ 4.4 NIBP Module ”. Watchdog Interval: No effect on the module. Adult watchdog time testing No effect on the module. Infant watchdog time testing No effect on the module.
![](/img/blank.gif)
Datex-Ohmeda S/5 FM 38 Document no. M1187329-003 4.5 DIS Interfacing 4.5.1 InterfacingHow to interface To select the parameter data source: •Press the Monitor Setup and select Interfacing. •Select the desired measurement parameter. •Select the desired source by name. NOTE: The name of the device is visible on the list only if the device is correctly connected to the module.
![](/img/blank.gif)
Service Menu39 Document no. M1187329-003 4.5.2 Status Page 4.5.3 DIS Interfacing service menu Access via Monitor setup - Interfacing. The selection Next page is available, if more than 8 DIS modules are connected to the DIS bus simultaneously. Interfaces The menu displays a list of all connected DIS modules and the statuses of the corresponding external devices. If the bus voltage is too low, you cannot add more devices. ‘DIS module bus voltage low. Do not add more devices or reduce cable length’ message appears. Access via Monitor setup - Install/Service- Service The menu displays submenus for all connected DIS modules.
![](/img/blank.gif)
Datex-Ohmeda S/5 FM 40 Document no. M1187329-003 4.5.4 DIS Module specific pageService Data Product name: DIS module name. Product type: DIS module type. Driver sw id: DIS module software and its release date. Short product name: DIS module name. Module serial number: DIS module serial number. HW card type: PCB type. HW id: DIS module PCB identification number. HW test date: DIS module PCB testing date. Unit serial number: DIS module PCB serial number. Comment field: Indicates the status of the external device. bus voltage : DIS bus voltage. tout : DIS module timeouts, seen by the monitor. The value should not increase more than by 5 per second. cse : DIS module checksum errors, seen by the monitor. The value should not increase more than by 5 per second. rx: The number of data packets from the external device received by the DIS module. rx err: The number of data errors from the exte rnal device received by the DIS module. tx: The number of data packets to the external device sent by the DIS module. Ram: Status of DIS module RAM memory. Rom: Status of DIS module ROM memory. EEPROM: Status of DIS module EEPROM memory.
![](/img/blank.gif)
Service Menu41 Document no. M1187329-003 5 Set/Test The power supply unit contains a watchdog circuitry, which needs refreshment at every 1.5 seconds. If the refreshm ent did not occur, the watchdog circuitry will re set the main CPU. In normal operation, the main CPU refreshes the watchdog circuitry at every 0.2 seconds. The purpose of the watchdog circuitry is to restart the monitor, if there was a serious malfunction. This feature is useful in two cases: when the main CPU is not able to contro l the monitor, or when the CPU controls the monitor but detects a serious malfunction. Watchdog tests check proper functionality of the watchdog circuitry in various conditions. Watchdog test ensures directly that the watchdog of the power unit functions properly. Choosing this test prevents the watchdog circuitry from refreshing and shows running seconds with an accuracy of 0.1 seconds. The test should have the following result when the watchdog circuitry is working properly: The monito r will restart after 1.5 seconds from the start of the test. In malfunction: ‘>20 s’ is disp layed, and the test will be interrupted. In this case, the fault is in the watchdog circuitry of the power unit. WD by Overload test ensures the functionality of a feature, where the software controls the monitor, but detects an overload situation in the main CPU. The test should have the following re sult when the feature is working properly: The mo nitor will restart after 15 seconds from the start of the test. Factory Reset restores the factory default settings and clears the data memories. Factory reset should be run if the monitor software is replaced or if the SRAM/Timekeeper battery is replaced.
![](/img/blank.gif)
Datex-Ohmeda S/5 FM 42 Document no. M1187329-003 5.1 Country Settings Language Select the software language. National Reqs Select the software features, which include national requirements. Power Frequency Set the power frequency (50 Hz/60 Hz). This setting is used to filter out possible power frequency interference from the parameter measurements. Time format Set the time format of the real time clock (24 h/12 h).