Home > GE > Medical Monitor > GE S5 Owners Manual

GE S5 Owners Manual

    Download as PDF Print this page Share this page

    Have a look at the manual GE S5 Owners Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    							Interfacing33
    Document no. M1187317-009
    4.1.1 Device Interfacing Solution components
    The Device Interfacing Solution consists of:
    a device specific interfacing module
     a device specific cable
     a bus cable
     a connector for another bus cable
     label specifying the external device
    4.1.2 Connections
    Connect the device specific cable to the external  device and the bus cable to the monitors DIS 
    connector or to another interfacing module. 
    Figure 8 Connection cables and LED indicators
    WARNING The monitor, interfacing modules and in terfaced devices must be situated 
    in the same patient environmen t (as defined in IEC 60601-1-1).
    WARNING Connecting electrical equipment togeth er or using the same extension cord 
    for more than one device may cause  their leakage currents to exceed the 
    limits specified in relevant  safety standards. Always make sure that the 
    combination complies with  the international safety standard IEC 60601-1-1 
    for medical electrical  systems and with the re quirements of local 
    authorities.
    WARNING The manufacturer guarantees a reliable functionin g of the devices with 
    tested software versions  only. Always refer to the Installation guide 
    accompanying the DIS module and ve rify the compatibility before use.
    (1)  label specifying the external device
    (2) LED indicators
    (3)  black bus cable from another 
    interfacing module, if needed
    (4)  gray device specific cable to the  communication port of the external 
    device
    (5)  black bus cable to the monitor’s DIS  connector (or to another interfacing 
    module) 
    						
    							Datex-Ohmeda S/5 FM
    34
    Document no. M1187317-009
    4.1.3 Mounting
    The DIS module can be mounted on the side panel of the external device. Also IV pole 
    placement is possible.
    NOTE: As the Device Interfacing Solution works on ly with the device specified in the label of the 
    interfacing module, it is recommended that the  interfacing module always travels along with 
    the external device.
    For mounting accessories, please refer to the  “Supplies and Accessories” catalog. See the 
    following figure for an example of a device interfacing.        
    Figure 9 An example of interfacing external devices with Device Interfacing  Solution
    (1)  Datex-Ohmeda S/5 FM
    (2) Interfacing module 
    (3) Monitor
    (4)  Blood gas analyzer
    NOTE:  Only N-DISxxx version 01 or  later is compatible with the S/5 FM.
    NOTE:  You can connect up to four (4) interf acing modules to one system simultaneously.
    Check the maximum number of modules: one me ter cable = max. four ten modules, three 
    meter cable = max. three modules, six meter cable = one module. 
    WARNING Make sure that the interfacing module is  always used in vertical position to 
    prevent water from entering the module.
    WARNING Make sure that you are connecting  the interfacing module to the device 
    specified in the label. Always veri fy the compatibility of the software 
    versions before use.
    4.1.4 Selecting th e external device
    1. Turn off the monitor.
    2. Turn off the external device.
    3. Connect the interfacing module to the mo nitor’s connector for N-DIS or to another 
    interfacing module.
    4. Connect the device specific c able to the external device and turn the external device on.
    5. Turn the monitor on. The monitor identi fies the connected device automatically.
    1
    2
    4
    3
    2 
    						
    							Interfacing35
    Document no. M1187317-009
    4.1.5 Functional check
    There are two ways to check the function of the Device Interfacing Solution:
    1. Press the 
    Monitor Setup key.
    2. Select  Interfacing and open the  Status Page menu. The status page shows you the 
    current communication status of  the interfacing module (1 - 4). 
    NOTE: The status message ‘Connected ’ appears on the Status Page after you have connected 
    the external device to the interfacing module an d turned it on. Note also that the monitor and 
    the interfacing module must be operational. 
    •Check the LED indicators on the interfacing module (the green LED indicates physical 
    connections, the yellow LED software selections)
    :
    4.1.6 Selecting the pa rameter data source
    Select the external device via Monitor Setup - Interfacing menu:
    •Select the desired measurement parameter (e.g., SvO2).
    •Select the desired source by name (e.g.,  Oxim3). 
    NOTE: The name of the device is visible on the list only if the device is correctly connected.
    NOTE: Detailed information abou t interfacing module related mountings, connections and 
    settings is included in  the installation guides that are delivered with the interfacing modules.
     GREEN  YELLOW  INDICATION
     lit              dark      Physical connections between the monitor, interfacing 
    module and external device are in order and the device has 
    been selected in the menu.
     dark
          
      lit            There is something wrong with the physical connections 
    between the monitor, interfacing module and external 
    device. The external device has not been selected in the 
    menu.
      lit                lit             Physical connections between the monitor, interfacing 
    module and external device are in order but the external 
    device has not been selected in the menu.
     dark          dark         The interfacing module is no
    t connected to the monitor. 
    						
    							Datex-Ohmeda S/5 FM
    36
    Document no. M1187317-009
    4.2 Interfacing printer
    It is possible to interface a laser printer (serial or parallel) with the S/5 FM via the monitors’ serial 
    port. The printer should be PCL5 compatible and should contain at least 2 MB of memory. 
    Parallel printers require the us e of Serial-to-Parallel Converter, order code 78030, model PI 
    130-R2.
    WARNING Always make sure that the combinat ion complies with the international 
    safety standard IEC 60601-1-1 for medical electrical systems and with the 
    requirements of local authorities.
    WARNING Connecting the power supply cord of  the printer to the wall power outlet 
    may cause the printer leakage current  to exceed the limit specified for 
    medical equipment. A printer must  be supplied from an additional 
    transformer providing at  least basic isolation (isolating or separating 
    transformer).
    4.2.1 Setting S/5 FM interface for printers
    1. Press the Print/Record key.
    2. Select  Printer Connection .
    3. Select  Serial.
    4. Press the 
    Normal Screen key.
    4.2.2 Connection to serial printers
    A serial printer is connected to the serial port connector X9 in the Multi I/O adapter at the back 
    of the frame unit.
    Contact your authorized GE Healthcare distributor for advice on suitable serial printers.
    4.2.3 Connection to  parallel printers
    A parallel printer is connected to  the serial interface port connector X9 on the frame unit via the 
    Serial-to-Parallel Converter, order code 78030, model PI130-R2.
    Contact your authorized GE Healthcare distributor for advice on suitable parallel printers.
    4.2.4 Installing the Serial -to-Parallel Converter
    Order code 78030, model PI130-R2
    A common serial modem cable, gender changer and parallel printer cable are required when 
    connecting the serial to parallel converter between the monitor and a parallel printer. The 
    converter gets power from the connected devices.
    1. Make sure that power is switched off on both devices.
    2. Connect a standard PC-to-par allel-printer cable (order code 713701) to the parallel 
    printer. 
    						
    							Interfacing 37
    Document no. M1187317-009
    3. Check the DIP-switch settin gs on the converter (order code 78030, model PI130-R2):
    4. Connect a DB25 female-to-female gender changer (order code 78032) between the  printer cable and converter.      
    Figure 10 Connecting S/5 FM monitor  to printer, converter model PI130-R2
    5. Connect a common RS-232 modem cable’s  (order code 78031) DB9 connector to the 
    serial port connector X9 on the monitor,  and the DB25 connector to the converter.
    The new Serial to Parallel Converter, PI1115A, is not compatible with the FM monitor. 
    4.2.5 Connection to printer Setting the printing parameters 
    Refer to the documentation provided with the printer.
    4.3 Interfacing computer
    A computer is connected to the serial port connector on the Multi I/O connector.
    Contact your authorized GE Healthcare distri butor for further advice on computer interface.
    WARNINGAlways make sure that the combinat ion complies with the international 
    safety standard IEC 60601-1-1 for medical electrical systems and with the 
    requirements of local authorities.
    WARNING Connecting the power supp ly cord of the computer to the wall power outlet 
    may cause the computer leakage current to exceed the limit specified for 
    medical equipment. A computer must  be supplied from an additional 
    transformer providing at least basic isolation (isolating or separating 
    transformer).
    Position
    Flow Control:
    hardware
    LED: Enabled
    Parity/D, Bits/S, Bit:
    None/8/1
    D.Rate (kbps):
    115,2
    1
    OFF
    2
    ON
    3
    OFF
    4
    OFF
    5
    OFF
    6
    OFF
    7
    ON
    8
    OFF
    DB9F end 
    plugs into FM DB25M end 
    plugs into 
    Centronics end plugs
    plugs into 
    Standard Parallel Printerplugs into FM
    Serial to Parallel
     
    Standard Async.Modem Cable gender changer
    Gender changer
     into printer
    DB25M end 
    plugs into 
    gender changer
    converter
    Converter Gender changer 
    						
    							Datex-Ohmeda S/5 FM
    38
    Document no. M1187317-009
    4.4 Output signals
    Analog/digital output signals on the connectors X5 and X8 can be used for interface with other 
    devices. The pin assignments are illu strated in tables/pictures below.
    Table 5 NET ID connector, X8 on Multi I/O adapter 
    Table 6 Defib & IABP sync connector, X5 
    4.4.1 Digital outputs
    The digital output signals are as follows:
    Defibrillation Sync (X5 pin 1)
    Def ibrillation Sync indication is generated by ECG. When active, the signal is state 1. After 10 
    ms the signal is reset to state 0. New Defibrillati on Sync is not generated before the indication is 
    deactivated. The delay from the R wave peak to the start of the signal is maximally 35 ms.
    Nurse Call (X8 pin 8)
    Nurse Call  indication is generated by red and yellow alarms. When activated, it is set to state 1 
    and remains at that state until the alarm situation is over or 
    SILENCE ALARM key is pressed.
    The range of state 0 is from 0 to 0.5 V, and range of the state 1 is from 3.5 to 5 V.
    4.4.2 Analog outputs
    S/5 FM produces two analog real-time signals. Th e Direct ECG signal is available in Defib&IABP 
    sync connector X5 pin 8. The other signal is Invasive pressure output. It is available in 
    Synchronization connector X5 pin 7.
    NOTE: When source of the selected analog output  is invalid (e.g. invasive pressure channel is 
    not zeroed), or becomes invalid (e.g. ECG lead is disconnected), the signal on the output is 
    shown as flat line (0 VDC). 9 pin female D-connector Pin Signal
    1
    2
    3
    4
    5
    6
    7
    8
    9NET_ID_CS (chip delect)
    NET_ID_CLK (clock)
    NET_ID_DO (data out)
    NET_ID_DI (data in)
    GND
    +5V_out
    GND
    Nurse call
    GND
    Mini DIN7 connector Pin Signal 1
    2
    3
    5
    6
    7
    8Def ib_sync_out
    Reserved
    Analog GND
    Digital GND
    Reserved
    Pressure_out
    Direct_ECG_out
    5
    9 1
    6 
    						
    							Interfacing 39
    Document no. M1187317-009
    The pacemaker pulses have been replaced with 2 ms /5 V fixed digital pulses at the ECG analog 
    output for IABP or defibrillator  equipment. A device that fulfils  the requirements of the IEC 
    60601-1 standard can be connected to the defibrillator & IABP synchronization connector. 
    There are no other limitations, because the signals of the connector are galvanically isolated 
    from patient applied part of the ECG and invasive blood pressure measurements.
    Direct ECG (X5 pin 8)
    From first user lead (ECG1)
    Gain:  1 V/1mV ±20%
    Delay: < 15 ms
    DC offset: ±100 mV max.
    Output range: ±4 V
    Noise: 50 mVpp max.
    Frequency response:  0.05 Hz to 40 Hz
    The signal requires input impedance of 100 k
    Pressure out (X5 pin 7)
    Invasive pressure signal:  From pressure channel P1
    Gain: 10 mV/1 mmHg ±20%
    Delay: < 35 ms
    DC offset: ±100 mV
    Output range: -0.4 V to +3.2 V 
    Noise: 50  mVpp max. 
    The signal requires input impedance of 100 k
      
    						
    							Datex-Ohmeda S/5 FM
    40
    Document no. M1187317-009
    5 Functional check
    These instructions include procedures for a functional check for Datex-Ohmeda S/5 FM. The 
    functional check is mandatory after monitor installation.
    These instructions include a  “Functional check form, Datex-Ohmeda S/5 FM” which may be 
    used when performing the procedures. The symbol 
     in the instructions  indicates that the 
    check form contains space to record the results of the particular procedure. The procedures 
    should be performed in ascending order, bypassing those that are not applicable for a 
    particular monitor.
    All menu selections related to Datex-Ohmeda  products are written in following typeface: 
    e.g.  Parameters - Gas Unit.
    As you enter the service menus, you need the following passwords:
    Monitor Setup - Install/Service  (password 16-4-34) - Service  (password 26-23-8)
    In case you evaluate the measurement accuracy  with a patient simulator, add simulator’s 
    accuracy specification to the one of the monitor.
    An electrical safety check and a leakage current test are recommended to be performed prior 
    to the monitor installation. 
    						
    							 Functional check41
    Document no. M1187317-009
    5.1 Recommended tools
    NOTE: Use only properly maintained, calibrated and traceable measurement equipment for the 
    specified calibrations and adjustments to ensure accuracy. 
    For details on recommended accessories s ee the “Supplies and Accessories” catalog.
    Table 7 Patient simulators’ compatibility with each hemodynamic module
    For product(s)ToolOrder No.
    Airway modules
    N-FC(REC) Calibration gas755580
    Regulator 755534-HEL
    N-FC(REC) CO
    2 Sampling line 3m/10 ft 733163
    Hemodynamic modules
    E-PSM(P) Multi-Link ECG accessories, IEC
    - Multi-link 3-leadwire set 412682-003
    - Multi-link 5-leadwire set 412681-003
    - Multi-link 5-leadwire set, C2-C6 416467-004
    or Multi-Link ECG accessories, AHA
    E-PSM(P) - Multi-link 3-leadwire set 412682-001 - Multi-link 5-leadwire set 416681-001
    - Multi-link 5-leadwire set, V2-V6 416467-003
    E-PSM(P) SpO
    2 finger probe OXY-F-UN
    SpO
    2 Interconnect Cable OXY-ES3
    Temperature test set 884515-HEL
    Adult NIBP cuff hose with cuff ID 2021285-001
    NIBP cuff 2753E
    Infant cuff hose with out cuff ID 414874-001
    MemCard – Data or Menu
    Module Parameter Patient simulator
    M1010831 MedSim Lionheart & MPS450
    E-PSM(P) ECG Cable included Mult ilink ECG acc. Multilink ECG acc.
    T 2016998-001 2016998-001  and 
    M1010832 2016998-001
     and 
    M1010846 
    InvBP Cable included M1010858 and 
    2005772-001 M1010862 and 
    2005772-001 
    						
    							Datex-Ohmeda S/5 FM
    42
    Document no. M1187317-009
    Table 8 Adapter cables for hemodynamic patient simulators
    5.2 Visual inspection
    Make sure that the monitor is switched to standby.
    Disconnect the mains power cord from the monitor. 
    If the monitor is connected to the Datex-Ohmeda Network, disconnect the Mon-Net cable 
    from the monitor.
    1. Check all units visually. Check that all parts are intact and that the cables and screws are connected and 
    tightened properly.  Especially check the following parts:
    sampling line is connected to the extension module.
    Check that modules go in smoothly and lock up properly.
    
    5.3 Functional inspection
    WARNING Handle the water trap and its conten ts as you would any body fluid. 
    Infectious hazard may be present. 
    5.3.1 General
    1. Connect the mains power cord.
    Check that the Mains power LED is lit.
    2. Switch the monitor on. Check that the monitor starts up properly, i.e.  a normal start-up sound is heard from the 
    loudspeaker, the alarm LEDs turn on and off, and the monitoring screen appears.
    No error messages should appear on the screen.
    3. Configure the screen for the parameters that are connected.
    4. Enter the  Service Menu. 
    5. When applicable, check from the corresponding  Parameters submenu that the Timeouts, 
    Bad checksums and Bad c-s by mod values of inserted modules are not increasing faster 
    than by 5 per second. Check also that the module memories have passed the internal 
    memory test, i.e. RAM, ROM and EEPROM all state OK.
    If connected, the recorder should record two lines of start-up information.
    
    Patient simulator Adapter cables for simulators
    Hemodynamic patient simulator - Dual temperature adapter cable 2016998-001
    Hemodynamic patient simulator - Dual Inv.BP adapter cable 2005772-001
    Medsim - Temperature adapter cable M1010832
    Medsim - Inv.BP adapter cableM1010858
    Lionheart & MPS450 - Temperature adapter cable M1010846
    Lionheart & MPS450 - Inv.BP adapter cableM1010862 
    						
    All GE manuals Comments (0)