GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 3-5 - Installation Paperwork 3-23 3-5-1-1 Rear Panel Connector (cont’d) Pin Assignment for Camera B/W NOTE: Output level of control signals indicated in the above tables are TTL level. Pin Assignment of Insite 3-5-1-2 Front Connector Panel Located on the front panel are Microphone, LED, and Reset. Table 3-16 Pin Assignment of Mini-Jack for Controlling B/W Camera Pin No Output Signal 1P RIN T 2 Signal GND Table 3-17 Pin Assignment of Mini-Jack for Controlling Color Camera Pin NoOutput Signal 1 SHUTTER 2 Signal GND Figure 3-9 Front Connector Panel 7 8 Reset Microphone LED (HD access) Note:USB1.1. Some systems do NOT have this port. If your system has this port, do NOT use this port! Instead, use the USB port located at the rear connector panel. Serial For BT04 or later For BT03 or lower You can use the serial port. USB2.0
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 3-24Section 3-5 - Installation Paperwork 3-5-1-3 B/W Printer Connector Panel 3-5-1-4 Footswitch Connector Panel Round 5 pin connector. NOTE: Output level of control signals indicated in the above tables are TTL level. Table 3-18 Pin Assignment of Mini-Jack for Controlling B/W Printer Pin No Output Signal 1P RIN T 2 Signal GND Table 3-19 Pin Assignment of Mini-Jack for Footswitch Pin NoOutput Signal 1S W1-W H 2S W2-R D 3S W3-G N 4 SW1-BK, SW2-BK, SW3-BK 5F ra m e G ND 12 3 4 5
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 3-5 - Installation Paperwork 3-25 This page was intentionally left blank.
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 3-26Section 3-5 - Installation Paperwork
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 4-1 - Overview 4-1 Chapter 4 Functional Checks Section 4-1 Overview 4-1-1 Purpose for Chapter 4 This chapter provides procedures for quickly checking major functions of the LOGIQ™ 7 console, diagnostics by using the built-in service software, and power supply adjustments. 4-1-2 Special Equipment Required • An empty (blank) MO Disk. • At least one transducer. (normally you should check all the transducers used on the system.) Table 4-1 Contents in chapter 4 SectionDescriptionPage Number 4-1 Overview 4-1 4-2 General Procedure 4-2 4-3 Functional Checks 4-20 4-4 Application Turnover Check List 4-27 4-5 Diagnostics 4-28 4-6 Power Supply 4-29 4-7 Site Log 4-30 NOTICEMost of the information pertaining to this Functional Checks chapter is found in the LOGIQ™ 7 Quick Guide (Direction Number 2291859-100). Look for the letters (QG) after a section in the Table of Contents to determine if the information is in this chapter or in the Quick Guide.
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 4-2Section 4-2 - General Procedure Section 4-2 General Procedure 4-2-1 Lockout/Tagout Requirements Follow OSHA Lockout/Tagout requirements by ensuring you are in total control of the plug. 4-2-2 Power On/Boot Up NOTE: After turning off the system, wait at least ten seconds before turning it on again. The system may not be able to boot if power is recycled too quickly. 4-2-2-1 Power Up 1.) Connect the Main Power Cable at the rear of the System. 2.) Connect the Main Power cable to an appropriate mains power outlet. 3.) Switch ON the Main Circuit Breaker at the rear of the System. When power is applied to the Scanner, and the Rear Circuit breaker is turned ON, Power is distributed to the Fans, Control panel, Monitor, Internal and External I/O’s, Cage Boards, Peripherals and the Backend Processor. When the Power ON/OFF key is pressed once, the Backend Processor starts and its software code is distributed to initiate the scanner. CAUTIONSYSTEM REQUIRES ALL COVERS Operate this unit only when all board covers and frame panels are securely in place. The covers are required for safe operation, good system performance and cooling purposes. WARNINGWARNINGPROTECTIVE EARTH MUST BE TAKEN WHEN CONNECTING AC POWER CABLE (200V) WITHOUT ITS PLUG TO WALL OUTLET. Figure 4-1 Circuit Breaker Main Circuit Breaker Main Power Cable For BT04 or lator
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 4-2 - General Procedure 4-3 4-2-2-1 Power Up (cont’d) 4.) Press the ON/OFF key at the front of the System once. 4-2-2-2 Power Up Sequence NOTE: For consoles with BEP4 (BT09), power switch lamp is darker compared to existing BT04-BT07 consoles with BEP3. This is NOT abnormality and system will function normally regardless of the brightness of the power button. 1.) The Start Up Screen will be shown on the Monitor display when the system is turned ON. Figure 4-2 Power On/Off Standby Switch Location Figure 4-3 Start Up Screen
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 4-4Section 4-2 - General Procedure 2.) After initialization is complete, all lighted buttons on the Control Panel light and the default B-Mode screen or Patient screen (no probes are connected) is displayed on the monitor display. 4-2-2-3 Entering Maintenance Mode 4-2-2-3-1 R6.2.x or later 1.) Insert the service dongle into any USB port. 2.) Wait for scanner boot-up. 3.) Tap ON/OFF the power switch. 4.) Click on Exit. 5.) Enter the proper password to enter the maintenance mode. 6.) Click on OK. Figure 4-4 Clicking on Exit Figure 4-5 Clicking on OK
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 4-2 - General Procedure 4-5 4-2-2-3 Entering Maintenance Mode (cont’d) 7.) Click on maintenance . 8.) Click on Exit to window . Figure 4-6 Clicking on Maintetance Figure 4-7 Clicking on Exit to Window 7 7
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 4-6Section 4-2 - General Procedure 4-2-2-3 Entering Maintenance Mode (cont’d) 4-2-2-3-2 R 4.x.x to R6.0.x 1.) The Start Up Screen will be shown on the Monitor display when the system is turned ON. 2.) The “Probing for maintenance access” dialog appears. Enter Password, then click on OK. 3.) Then Start Loader display will be shown on the Monitor display.Click on Maintenance. 4.) Click on Exit to Windows in maintenace diaglog window.Then the scanner is booted up automatically. 4-2-2-3-3 R2.x.x to R4.x.x 1.) Insert the service dongle in the Service port located at the rear panel. 2.) The Start Up Screen will be shown on the Monitor display when the system is turned ON. 3.) Then Start Loader display will be shown on the Monitor display. Then the scanner is booted up automatically. NOTE: To enter the Maintenance Mode, select Maintenance button. NOTE: Start is selected automatically when it time out. Figure 4-8 Clicking on Mainttenance Figure 4-9 Start Application Window