GE Logiq 7 Service Manual
Have a look at the manual GE Logiq 7 Service Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-35 8-2-10-2 Confirmation of the software version (cont’d) 2.) Touch the System button on the Touch Panel. 3.) Click on About tab. The software version will be shown on the monitor as shown. Confirm the following: - Software Version ( R 8.x.x) - Preset Resion (Americas, Europe, Asia, Japan, or None) - Image Part Number (described on the Base System Software Image DVD) Figure 8-92 System Figure 8-93 About Tab Press System button. Click on the About tab Software Version R 8.x..x Base System Software CD-ROM Part Number xxxxxxx Preset Region (See 8-2-10-10 Regional Preset.)
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 8-36Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-2-10-3 Computer Name and Connection Settings 1.) Touch Utility > Connectivity. 2.) Click on TCP/IP. 3.) Type Computer Name. 4.) Verify that Enable DHCP has no check mark. If checked, remove the mark. 5.) Check if Computer name, IP Address, Subnet Mask, and Default Gateway are proper ones which you wrote down in section 8-2-6-4, Saving Connectivity. 6.) Click on OK for confirmation dialog box. NOTICEThe computer name is unique. Check the scanner serial Number printed on the label located at the rear lower side of the scanner. When 123456YM1 (for example) is printed, L7-123456YM1 must be entered as a computer name. Figure 8-94 TCP/IP NOTICEIf the “Save settings” is NOT performed, you can NOT enter the new patient screen!! Figure 8-95 OK
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-37 8-2-10-3 Computer Name and Connection Settings (cont’d) 7.) Touch Scan on the Touch Panel to return to the scan screen. 8.) Shut down the scanner. 9.) Power ON the scanner, then press Patient key and verify that the following screen appears. Figure 8-96 Patient
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 8-38Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-2-10-4 Check LCD Monitor Hardware Setting For 19 inch Monitor To adjust the brightness: 1.) Press the Toggle button (2) or (3) Once for brightness adjustment. a.) Verify that Brightness is 80 (Dark Room) or 100 (Bright Room) . If it is not, increase brightness to press the adjustment button (3). If it is not, decrease brightness to press the adjustment button (2). b.) Press the button (1) Once. The brightness display disappears. NOTE: 19 inch LCD is ONLY a brightness adjustment. 8-2-10-5 Software Option Check 1.) Touch Utility > Admin. 2.) Click on System Admin tab and verify that HW Number is the same as the Option Dongle ID. Figure 8-97 19 inch LCD Monitor Adjustment buttons Figure 8-98 HW Number 123
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-39 8-2-10-6 VCR Parameters Setting (for the VCR-equipped system ONLY) 1.) Touch Utility > System. 2.) Click on Peripherals tab. 3.) Set the proper parameters: -PAL or NTSC - VCR type 4.) Shut down the scanner, then restart the scanner. 5.) Verify that the VCR can be properly operated with no error (remote, play...) . 8-2-10-7 Regional Setup (For the system with language other than English ONLY) For Japanese Languages You can set up the keyboard for Japanese. If you select Japanese, the following items display or enter by Japanese. • Patient name • Exam comment in Patient screen and Worksheet • Report/Direct Report free text, Insert Text • Annotation, Annotation Library • Error/Warning message • DICOM/Worklist, Query/Retrieve, Image Storage, Printer, MPPS Figure 8-99 Peripherals
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 8-40Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-2-10-7 Regional Setup (For the system with language other than English ONLY) (cont’d) 1.) Select Utility> System> General>Regional Option Menu , 2.) Click on Languages tab. 3.) Check Install files for East Asian languages. 4.) Click on Apply. 5.) The system requires the re-boot automatically, shut down the system. Figure 8-100 Regional Option NOTICEYou do not need to change “Language used in menus and dialogs” option. It should be English. Check Do not Change English.
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-41 8-2-10-7 Regional Setup (For the system with language other than English ONLY) (cont’d) 6.) In Utility> System> General , set the Language as JPN. Save this setting, but do not reboot the system yet. 7.) Press Regional Options , under Standards and Formats select Japanese, under Location select Japan . Press Apply .Figure 8-101 General Figure 8-102 Regional Options
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 8-42Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-2-10-7 Regional Setup (For the system with language other than English ONLY) (cont’d) 8.) Select the Language tab , press Details, under Installed Services select the Japanese keyboard , under Default input language select Japanese, press Apply , Press OK. 9.) Select the Advanced tab , then select Japanese in the Language for non-Unicode programs pull- down menu. Press Apply . Answer Yes to use files already loaded on the hard disk, then answer No to not reboot the system yet, press OK . Press Save and Exit the Utility screen. 10.) Reboot the system. 11.) In Utility--> System--> General , press Regional Options. Select Language Tab and press Details . Remove English under Installed Services and press Apply and OK. 12.) Select Utility--> System--> About-->Additional Info menu . Figure 8-103 Language tab NOTICETo have the settings take effect, you MUST turn off the system and turn it back on.
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-43 8-2-10-7 Regional Setup (For the system with language other than English ONLY) (cont’d) 13.) Select Automatic Setup for Japanese Language button. Then IME configuration shall be automatically done. This configuration shall setup the keyboard assignment (remove R
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 8-44Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-2-10-7 Regional Setup (For the system with language other than English ONLY) (cont’d) For Russian or Greek Languages 1.) In Utility--> System--> General , set the Language as Russian or Greek. Save this setting, but do not reboot the system yet. 2.) Press Regional Options , under Standards and Formats select Russian or Greek, under Location select Russia or Greece . Press Apply. 3.) Select the Language tab, press Details , under Installed Services select the Russian or Greek keyboard , under Default input language select Russian - Russian or Greek - Greek , press Apply, Press OK. Figure 8-105 General Figure 8-106 Regional Options