GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 3-4 - Completing the Installation 3-13 3-4-2 Optional Peripherals/Peripheral Connection (cont’d) Check if the fixing belts are loosen. If it is, tighten the belts again to secure the peripheral(s). If necessary, remove the peripheral from the device room to tighten the belts. 3-4-2-1 Approved on-board peripherals Refer to Section 5-3 Peripheral Compatibility . Figure 3-6 How to Hold Peripherals Belts Belts BeltsConnect the related cables between the peripheral and LOGIQ 7 if they are NOT con- nected each other.
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 3-14Section 3-4 - Completing the Installation 3-4-2-1 Approved on-board peripherals (cont’d) Connecting Cables 3-4-2-2 Reference off-board peripherals and options None. CAUTIONEquipment damage possibility. Be sure to use the following recommended connecting cables to connect recording devices and a network with LOGIQ™ 7 console. Table 3-4 List of Connecting Cables NamePart No.FigureNOTE Power Supply Cable P9509EE Connected to power USB Serial Bridge Cable 2304621 For converting the signal of RS232C cable to USB cable: connected to VCR1 on the Rear Panel AV Cable 2119874 Connected to Video-In/Out on the Rear Panel The followings are the cables fo r BT04 ore lower system ONLY. Mini-Plug Cable P9509BE Shutter control signals: connected to B/W Printer RS232C Cable Cross 2305550 For control signals: connected to Serial Bridge Cable RS232C Cable Straight 2305549 For control signals: connected to Serial Bridge Cable BNC Cable 2297053 For control signal: connected to Composite B/W USB Cable 2324360 Connected to USB port. SCSI cables (UP-D50 ONLY) 2375479 Connected toSCSI port and SCSI Cable Connector.
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 3-4 - Completing the Installation 3-15 3-4-3 Available Probes See in specification in the LOGIQ™ 7 Reference Manual for Probes and intended use. See Chapter 9 - Renewal Parts for Part Numbers to be used when ordering new or replacement probes. Table 3-5 List of Transducers supported (All models) Probe Name Material of Headshell Area of UsingTYPECatalog NumberPart Number 3C PESABDOMINAL CONVEX H79802P H40412LB 2286353 2286354 5C PESABDOMINAL CONVEX H79822P H40412LA 2294515 2294516 8C PBT NEONATAL PEDIATRICS MICRO-CONVEX H79792P H40412LJ 2348093 2348094 E8C NORYL PBT TRANSVAGINAL MICRO-CONVEX H79852P H40412LE 2294640 2294641 M7C PBTABDOMINAL CONVEXH79832P H40412LC 2294513 2294514 M12L PBTSMALL PARTS LINEARH79842P H40412LD 2294510 2294511 7L NORYL ABDOMINAL SUPERFICIAL LINEAR H79862P H40412LF 2294520 2294521 10L NORYL SUPERFICIAL LINEARH79872P H40412LG 2294522 2294523 10S ABSPEDIATRIC SECTORH79922P H4901PC 2309478 2298589 3.5C NORYL ABDOMINAL CONVEXH79812P H4901PE 2303215 2050357 3.5CS NORYL ABDOMINAL CONVEXH78042P H40412LK 2380854 2051858 3S NORYL CARDIOLOGY SECTORH79632P H4701SZ 2348878 2323337 i12L ABSINTRAOPERATIVE LINEARH79322P H4012L 2270556 2264883 M3S PBTCARDIOLOGY SECTORH79892P H45011SZ 2293726 2378099 6T-OR-TEE PU: PolyUrethane TRASOPHAGEAL FOR ADULT CARDIOLOGY SECTOR H79932P H45521DX 5131947 KN100068 7S PBTCARDIOLOGY SECTORH78082P H40422LB 2355698 2347471 BE9C PBTTRANSRECTAL MICRO-CONVEX H40412LW2389381 2389382 T739 NORYLINTRAOPERATIVE LINEARH76572SR H40212LM2259245 2259246 4D3C_L PBTABDOMINAL CONVEX Volume H44801G 5121652 KTZ195893
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 3-16Section 3-4 - Completing the Installation NOTE: PES: Polyethersulfone NORYL: Modified Polyphenylene Oxide PU: Polyurethane PBT: Polybutylene Terephthalate ABS: Acrylonitrile Butadiene Styrene NOTE: Some probes indicated on the table above have two different part numbers. The upper row shows the part numbers of probes for Japan. The lower row shows the part numbers of probes for regions other than Japan. Probes which have only one part number are not available in Japan. The following transducers, additional to ones listed in Table 3-6, are supported by V65x and later LOGIQ 7 models. The following transducers, additional to ones listed in Table 3-7 , are supported by R7.5.x and later LOGIQ 7 models. The following transducers, additional to ones listed in Table 3-8 , are supported by BT09 and later LOGIQ 7 models. 4D10L PBDSUPERFICIAL LINEAR Volume H44801GB 5121651 KTZ156836 4C PBTABDOMINAL CONVEXH4904PC5131944 5123455 12L NorylSUPERFICIAL LINEARH40412LH2295375 2295377 P2D ABSCARDIOLOGY CWDH4830JETE100024 P6D ---SUPERFICIAL CWDH4830JGTQ100002 Table 3-6 List of Transducers supported by V65x and later Probe Name Material of Headshell Area of UsingTYPECatalog NumberPart Number 9L PBTVASCULAR SMALL PARTS LINEAR H40412LT5131433 4DE7C PBTOB Gyn Urology CONVEX Volume H44801GA 5121650 Table 3-7 List of Transducers supported by R7.5.x and later Probe Name Material of Headshell Area of UsingTYPECatalog NumberPart Number 3CRF PU: PolyUrethane ABDOMINALMICRO-CONVEX H40442LP 5214820 5196216 Table 3-8 List of Transducers supported by BT09 and later Probe Name Material of Headshell Area of UsingTYPECatalog NumberPart Number 11L PBTSUPERFICIAL LINEARH78842P H40412LY 5251881 5171885 Table 3-5 List of Transducers supported (All models) Probe Name Material of Headshell Area of UsingTYPECatalog NumberPart Number
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 3-4 - Completing the Installation 3-17 3-4-4 Video Specification 3-4-5 Software Option Configuration 3-4-5-1 Onsite check and configuration Select Utility > Admin > System Admin and check the option software to be installed. Table 3-9 Video Specifications CRTLCD Timing Parameter 800x600 75Hz Timing Parameter 1280 x 1024 60Hz Horizontal Rate [kHz] 46.88Horizontal Rate [kHz] 64 Horizontal Period [ µs] 21.33 Horizontal Period [µs] 15.625 Pixel Clock [MHz] 49.50Pixel Clock [MHz] 108 H Blank Width [µs] 5.17 H Blank Width [µs] 3.778 (408dots) H Sync Width [µs] 1.62 H Sync Width [µs] 1.037 (112dots) H Front Porch [µs] 0.32 H Sync Front Porch [µs] 0.444 (48dots) Active Horizontal Period [µs] 16.16 Active Horizontal Period [µs] 11.852 (1280dots) Vertical Rate [Hz] 75.00Vertical Rate [Hz] 60.0 Vertical Period [ms] 13.33Vertical Period [ms] 16.67 (1066lines) V Sync Width [lines=ms] 25=0.53V Sync Front Porch [ms] 0.016 (1 line) V Front Porch [lines=µs] 3=64.00 Equalization Porch None Equalization Gate [lines=µs] 1=21.3 Lines: Field/Frame 1066 Lines: Field/Frame 625Active Lines/Frame 1024 Active Lines/Frame 600
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 3-18Section 3-5 - Installation Paperwork Section 3-5 Installation Paperwork NOTE: During and after installation, the documentation (i.e. Users Manual, Installation Manuals...) for the peripheral units must be kept as part of the original system documentation. This will ensure that all relevant safety and user informations are available during the operation and service of the complete system. 3-5-1 Peripherals/Accessories Connector Panel LOGIQ™ 7 peripherals and accessories can be properly connected using the rear connector panel located behind the rear door, front connector panel located next to the video printer, and Footswitch connector located bottom of the OP panel.
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 3-5 - Installation Paperwork 3-19 3-5-1-1 Rear Panel Connector Located on the rear panel are video input and output connectors, audio input and output, camera expose connectors, footswitch connector power connector and control connections for VCR, printer, and service tools. This section indicates the pin assignment for each connector. Figure 3-7 Rear Pannel Connector (BT04 or later) Video In Rec/Expose VGA S-Video Out S-Video In CmpstOut Insite Audio In LR LR AudioOut Ethernet VCR2 VCR1 Service CameraB/W Color B/W Color Video Out Power AC Line Input 1200VA 220-240V Circuit Breaker Circuit Breaker 50 /60Hz Power AC Line Input 1200VA 50 /60Hz 100-120V Circuit Breaker Circuit Breaker 100-120V 350VA Max Rear Pannel Connector for BT04 and later models 5 6 87 2 3 1
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 3-20Section 3-5 - Installation Paperwork 3-5-1-1 Rear Panel Connector (cont’d) NOTE: Each outer (case) ground line of peripheral/accessory connectors are protectively grounded. Signal ground lines are not isolated, except the Service port (3). All of signal lines (include signal GND) of the Service port are isolated. The specified peripherals/accessories only can be connected to the USB ports.Rear Pannel Connector for BT03 or lower Figure 3-8 Rear Panel Connector (BT03 or lower) USB1.1 (USB compatible) or (USB + USB Serial Bridge + RS232C) (Reserved for future use) 2 1 3 4 5 6 87 USB1.1 (USB compatible) or (USB + USB Serial Bridge + RS232C) USB1.1 (USB compatible) or (USB + USB Serial Bridge + RS232C)
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 3-5 - Installation Paperwork 3-21 3-5-1-1 Rear Panel Connector (cont’d) Pin Assignment of S-Video Connector Connector: S-Terminal, 4-pin Pin Assignment of VGA Connector Connector: Shrank D-Sub, 15-pin Pin Assignment of Service/VCR 1/VCR 2 Connector, USB1.1 Connector: 4 pin Table 3-10 Pin Assignment of S-Video Connector Pin No Output/Input Signal Description 1SVIDEO OUT/IN YG Y (Luma) GND 2 SVIDEO OUT/IN CG C (Chroma) GND 3 SVIDEO OUT/IN Y Y (Luma) SIGNAL 4 SVIDEO OUT/IN C C (Chroma) SIGNAL Table 3-11 Pin Assignment of VGA Connector Pin NoOutput Signal Description 1IO VGA OUT1 R Red 2 IO VGA OUT1 G Green 3 IO VGA OUT1 B Blue 6 IO VGA OUT1 RG Reg GND 7 IO VGA OUT1 GG Green GND 8 IO VGA OUT1 BG Blue GND 13 IO VGA OUT1 HS H Sync 14 IO VGA OUT1 VS V Sync Others GND GND Table 3-12 Pin Assignment of Service/VCR 1/VCR 2 Connector Pin NoOutput Signal Description 1VBUSn Power Supply 2 Dn Data (-) 3 Dn Data (+) 4 GNDn Power Ground 1 24 3 1 2 1 6 115 10 15 3 1 2 3 4
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 3-22Section 3-5 - Installation Paperwork 3-5-1-1 Rear Panel Connector (cont’d) Pin Assignment of Digital Int. Connector Connector: 6-pin Pin Assignment of Ethernet Connector: RJ-45 Modular, 8-pin Pin Assignment of Insite Connector: RJ-11 Modular, 6-pin Table 3-13 Pin Assignmen t of IEEE-1394 Connector) Pin NoOutput Signal Description 1VP 1394 Power Supply 2 VG 1394 Power Ground 3 TPB 1394 Data B (-) 4 TPB 1394 Data B (+) 5 TPA 1394 Data A (-) 6 TPA 1394 Data A (+) Table 3-14 Pin Assignment of Ethernet Connector Pin NoOutput Signal Description 1ETHER TD Ethernet TD+ 2 ETHER TD Ethernet TD- 3 ETHER RD Ethernet RD+ 6 ETHER RD Ethernet RD- Others NC Non-connection Table 3-15 Pin Assignment of Insite Connector Pin NoOutput Signal Description 2TEL L4 Telephone L4 3 TEL L2 Telephone L2 4 TEL L1 Telephone L1 5 TEL L3 Telephone L3 Others NC Non-connection 4 1 3 5 2 4 6 5 8 1 6 8 1