GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 1-2 - Important Conventions 1-9 Indicates the presence of hazardous substance(s) above the maximum concentration value. Maximum concentration values for electronic information products, as set by the People’s Republic of China Electronic Industry Standard SJ/ T11364-2006, include the hazardous substances of lead, mercury, hexavalent chromium, cadmium, polybrominated biphenyl (PBB), and polybrominated diphenyl ether (PBDE). “10” indicates the number of years during which the hazardous substance(s) will not leak or mutate so that the use of this product will not result in any severe environmental pollution, bodily injury, or damage to any assets.Probe BT07 or later Indicates the presence of hazardous substance(s) above the maximum concentration value. Maximum concentration values for electronic information products, as set by the People’s Republic of China Electronic Industry Standard SJ/ T11364-2006, include the hazardous substances of lead, mercury, hexavalent chromium, cadmium, polybrominated biphenyl (PBB), and polybrominated diphenyl ether (PBDE). “20” indicates the number of years during which the hazardous substance(s) will not leak or mutate so that the use of this product will not result in any severe environmental pollution, bodily injury, or damage to any assets. Rear Panel, China Rating Plate BT07 or later Do not use the following devices near this equipment: cellular phone, radio receiver, mobile radio transmitter, radio controlled toy, broadband power lines, etc. Use of these devices near this equipment could cause this equipment to perform outside the published specifications. Keep power to these devices turned off when near this equipment. Rear Panel BT07 or later Table 1-5 Warnings Label/SymbolPurpose/MeaningLocationNote
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 1-10Section 1-2 - Important Conventions Section 1-3 Safety Considerations 1-3-1 Introduction The following safety precautions must be observed during all phases of operation, service and repair of this equipment. Failure to comply with these precautions or with specific warnings elsewhere in this manual, violates safety standards of design, manufacture and intended use of the equipment. 1-3-2 Human Safety Operating personnel must not remove the system covers. Servicing should be performed by authorized personnel only. Only personnel who have participated in a LOGIQ™ 7Training Seminar are authorized to service the equipment. 1-3-3 Mechanical Safety NOTE: Special Care should be taken when transporting the unit in a vehicle: • Secure the unit in an upright position. • Lock the wheels (brake). • DO NOT use the Control Panel as an anchor point. • Place the probes in the carrying case. • Eject any Magnet Optical disk from the MO Drive (if installed). WARNINGWARNINGWHEN THE UNIT IS RAISED FOR A REPAIR OR MOVED ALONG ANY INCLINE, USE EXTREME CAUTION SINCE IT MAY BECAUSE UNSTABLE AND TIP OVER. WARNINGWARNINGULTRASOUND PROBES ARE HIGHLY SENSITIVE MEDICAL INSTRUMENTS THAT CAN EASILY BE DAMAGED BY IMPROPER HANDLING. USE CARE WHEN HANDLING AND PROTECT FROM DAMAGE WHEN NOT IN USE. DO NOT USE A DAMAGED OR DEFECTIVE PROBE. FAILURE TO FOLLOW THESE PRECAUTIONS CAN RESULT IN SERIOUS INJURY AND EQUIPMENT DAMAGE. WARNINGWARNINGNEVER USE A PROBE THAT HAS FALLEN TO THE FLOOR. EVEN IF IT LOOKS OK, IT MAY BE DAMAGED. CAUTIONAlways lock the Control Console in its parking (locked) position before moving the scanner around. CAUTIONDisconnect all probes before moving the scanner around. CAUTIONThe LOGIQ™ 7 weights 225 kg or more, depending on installed peripherals, (496 lbs, or more) when ready for use. Care must be used when moving it or replacing its parts. Failure to follow the precautions listed below could result in injury, uncontrolled motion and costly damage. ALWAYS: • Be sure the path way is clear. • Use slow, careful motions. • Use two people when moving on inclines or lifting more than 23 kg (50 lb).
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 1-2 - Important Conventions 1-11 1-3-4 Electrical Safety To minimize shock hazard, the equipment chassis must be connected to an electrical ground. The system is equipped with a three-conductor AC power cable. This must be plugged into an approved electrical outlet with safety ground. If an extension cord is used with the system, make sure that the total current rating of the system does not exceed the extension cord rating. The power outlet used for this equipment should not be shared with other types of equipment. Both the system power cable and the power connector meet international electrical standards.
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 1-12Section 1-2 - Important Conventions 1-3-5 Label Locations (For BT07 or later, including R7.5.x) NOTE: For the symbols shown in the illustration below, refer to previous pages in this chapter. Figure 1-1 OUTSIDE MARKING S OF LOGIQ™ 7 For BT07 or later, including R7.5.x (Back Side) 7™ Japan/USA/Asia Console(100V) (9) (10) (11) (8) (2) (1) (3) (4) (6) (7) 7™ GE Yokogawa Medical Systems ETL LISTEDCONFORMS TO UL STD. 2601-1464 77RETL TESTING LABORATORIES INC. CORTLAND, NEW YORK 13043LISTEDRLISTEDCERTIFIED TO CAN/CSA C22.2 NO. 601.1C ~ (5) (12) (13) (14) (15) (16) (17) (21) Except for R7.5.x Label for China (2) (1) (3) (6) (7) (5) CAUTION CAUTION LCD model ONLY
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 1-2 - Important Conventions 1-13 1-3-5 Label Locations (For BT07 or later, including R7.5.x) (cont’d) 1.) Possible Shock Hazard 2.) Caution for devices near by the equipment 3.) Caution for Transportation 4.) Prescription Devices (For USA Only) 5.) CE Marking of conformity 6.) WEEE mark 7.) CISPR 8.) Voltage Range 9.) Signal Ground Point Label 10.) Power Indication Label 11.) Caution for Grounding Reliability (For USA, Canada and Japan) 12.) ETL Label 13.) Identification and Rating Plate (USA/Asia 120V console) 14.) Identification and Rating Plate (Europe/Asia/USA 220V console) 15.) Identification and Rating Plate (Jpan 100V console) 16.) Identification and Rating Plate (China) 17.) Label that indicates the presence of hazardous substance(s) above the maximum concentration value. 18.) Identification and Rating Plate (Korean) 19.) Caution label on the monitor 20.) Caution label for the range of motion 21.) This machine should be used in compliance with law. Some jurisdictions restrict certain uses, such as gender determination. - Korea Console
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 1-14Section 1-2 - Important Conventions 1-3-6 Label Locations (For BT04, BT06, and V65x) NOTE: For the symbols shown in the illustration below, refer to previous pages in this chapter. 1.) Possible Shock Hazard 2.) Caution for devices near by the equipment 3.) Caution for Transportation 4.) Prescription Devices (For USA Only) 5.) CE Marking of conformity and WEEE mark 6.) CISPR 7.) Voltage Range 8.) Signal Ground Point Label 9.) Power Indication Label 10.) Caution for Grounding Reliability (For USA, Canada and Japan)Figure 1-2 OUTSIDE MARKING S OF LOGIQ™ 7 For BT04 or later (Back Side) 1. 3. 2. 4. 5. 6. 7 9. 8. 7. 10.
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 1-2 - Important Conventions 1-15 1-3-6 Label Locations (For BT04, BT06, and V65x) (cont’d) 1.) ETL Label 2.) Identification and Rating Plate 3.) SFDA or SDA Label (For China ONLY) Figure 1-3 OUTSIDE MARKINGS OF LOGIQ™ 7 For BT04 or later ~ /05(çùù÷ç 65ï0./;ç0+,ð 2. 1. 3.
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 1-16Section 1-2 - Important Conventions 1-3-7 Label Locations (For BT03 or lower) NOTE: For the symbols shown in the illustration below, refer to previous pages in this chapter. 1.) Possible Shock Hazard 2.) Caution for devices near by the equipment 3.) Caution for Transportation 4.) Prescription Devices (For USA Only) 5.) CE Marking of conformity 6.) CISPR 7.) Voltage Range 8.) Signal Ground Point Label 9.) Power Indication Label 10.) Caution for Grounding Reliability (For USA, Canada and Japan)Figure 1-4 OUTSIDE MARKING S OF LOGIQ™ 7 For BT03 or lower (Back Side) 1. 3. 2. 4. 5. 6. 10. 9. 8. 7.
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 1-2 - Important Conventions 1-17 1-3-7 Label Locations (For BT03 or lower) (cont’d) 1.) ETL Label 2.) Identification and Rating Plate 3.) SDA Label (For China ONLY) Figure 1-5 OUTSIDE MARKINGS OF LOGIQ™ 7 For BT03 or lower (Left Side) ~ $IJOB7$POTPMF 3JHIU4JEF 2. 1. 3.
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 1-18Section 1-2 - Important Conventions 1-3-8 Dangerous Procedure Warnings Warnings, such as the example below, precede potentially dangerous procedures through our this manual. Instructions contained in the warnings must be followed. 1-3-9 Lockout/Tagout Requirements (For USA Only) Follow OSHA Lockout/Tagout requirements by ensuring you are in total control of the electrical Mains plug. 1-3-10 Returning/Shipping Probes and Repair Parts Equipment being returned must be clean and free of blood and other infectious substances. GEMS policy states that body fluids must be properly removed from any part or equipment prior to shipment. GEMS employees, as well as customers, are responsible for ensuring that parts/equipment have been properly decontaminated prior to shipment. Under no circumstance should a part or equipment with visible body fluids be taken or shipped from a clinic or site (for example, body coils or an ultrasound probe). The purpose of the regulation is to protect employees in the transportation industry, as well as the people who will receive or open this package. NOTE: The US Department of Transportation (DOT) has ruled that “items that were saturated and/or dripping with human blood that are now caked with dried blood; or which were used or intended for use in patient care” are “regulated medical waste” for transportation purposes and must be transported as a hazardous material. DANGERDANGEROUS VOLTAGES, CAPABLE OF CAUSING DEATH, ARE PRESENT IN THIS EQUIPMENT. USE EXTREME CAUTION WHEN HANDLING, TESTING AND ADJUSTING. WARNINGWARNINGEXPLOSION WARNING: DO NOT OPERATE THE EQUIPMENT IN AN EXPLOSIVE ATMOSPHERE. OPERATION OF ANY ELECTRICAL EQUIPMENT IN SUCH AN ENVIRONMENT CONSTITUTES A DEFINITE SAFETY HAZARD. WARNINGWARNINGDO NOT SUBSTITUTE PARTS OR MODIFY EQUIPMENT: BECAUSE OF THE DANGER OF INTERDICTING ADDITIONAL HAZARDS, DO NOT INSTALL SUBSTITUTE PARTS OR PERFORM ANY UNAUTHORIZED MODIFICATION OF THE EQUIPMENT. NOTICEEnergy Control and Power Lockout for LOGIQ™ 7 When servicing parts of the system where there is exposure to voltage greater than 30 Volts: Unplug the system Maintain control of the system power plug There are no test points to verify isolation, you must wait for at least 20 seconds for capacitors to discharge Beware that the AC Control Box, Front End Processor and Back End Processor may be energized even if the power is turned off when the cord is still plugged into the AC Outlet.