GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 6-6 - BW Printer Setting / Adjustment 6-29 6-6-2 Parameters for UP-D895 1.) When system started, Go to Utility > System >Peripherals, then hit Printers button in the Setup group. 2.) Click SONY UP-D895 in the Printer List, then right click and select Properties . 3.) Select General tab, then hit Printing Preferences ... 4.) Set Landscape into Orientation, then click on Advanced... 5.) Set 1920x1280 into PaperFigure 6-37 Landscape Figure 6-38 Paper
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 6-30Section 6-6 - BW Printer Setting / Adjustment 6-6-2 Parameters for UP-D895 (cont’d) 6.) Click on Scaling..., then check off into Enlarge to Paper and set Bilinear into Interpolation Method. 7.) Click on Density Adjust... , set TONE1 into GAMMA and set 7 into Sharpness. 6-6-3 Service Tips - Print Quality When printer is set to “Nearest Neighbor” interepolation method, comments and body pattern may become hard to read. Re-adjust interpolation method to “Bi-Linear” to avoid this situation. Figure 6-39 Scaling Figure 6-40 Density Adjust Bilinear
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 6-6 - BW Printer Setting / Adjustment 6-31 6-6-4 Service Tips - Print Speed 1.) Make sure that the UP-D897 is switched ON. (The back light of the LCD light is in green.) If not, press the ON/OFF switch. 2.) Confirm that “READY” is displayed on the LCD, then press the jog dial. The UP-D897 enters the menu mode. When you perform a menu operation for the first time after you purchase the unit , the top item of the “BEEP” menu will be displayed. If you already have performed a menu operation, the items that you set last is displayed. NOTICEProblem: Compared to the UP-D895, it takes a few seconds longer for the UP-D897 to receive data from L7 system. All that time, an user cannot operate the L7 system and must wait until the UP-D897 finished receiving data from the L7. Cause: This is due to driver software difference between UP-D895 and UP-D897. It seems the UP-D897 driver software has been modified from the UP-D895. Solution: The following actions must be required: • UP-D897 printer: Change from UP-D897 to UP-D895 on “COMD” menu using the printer jog dial. This solution allows the UP-D897 to be configured to use the UP-D895 interface and driver. When the “COMD” is selected to “CO:895”, the response time becomes the same as the UP- D895. Perform this section. • LOGIQ7: Change from UP-D897 to UD-D895 using Connectivity > Service > BW Printer to use UP-D895 driver software. Jog Dial (PRESS) READY
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 6-32Section 6-6 - BW Printer Setting / Adjustment 6-6-4 Service Tips - Print Speed (cont’d) 3.) Turn the jog dial up or down until “ COMD” is displayed on the LCD. 4.) Press the jog dial. “ CO:897” is displayed. The UP-D897 enters the mode in which you can select the driver setting. CO:897, which is the driver setting for D897, is displayed on the LCD. You need to change to CO:895 which is the driver setting for the D895. 5.) Turn the jog dial up or down until “CO:895” is displayed on the LCD. 6.) Press the jog dial. The LCD display changes to “COMD”. The UP-D897 completes to be set to the UP-D895 mode. Notice: With setting the UP-D897 COMD to UP-D895, the following items cannot be selected or operated from the jog dial. Item No.ItemsDescriptions 1GAMMATONE4(VERY SOFT) cannot be selected. 2GAMMA Cannot be selected from jog dial. It can change only from Properties of UP-D895 menu. Corresponding type of GAMMA between UP-D895 and UP-D897 is as follows: UP-D897 -> UP-D895 TONE1 -> SOFT (TONE3) TONE2 -> NORMAL (TONE2) TONE3 -> HARD (TONE1) TONE4 (VERY SOFT) -> Not Supported 3QTYCannot be selected from jog dial. It can change only from Properties of UP-D895 menu. 4 RSIZE Cannot be selected from jog dial. It can change only from Properties of UP-D895 menu. 5 SHARP Cannot be selected from jog dial. It can change only from Properties of UP-D895 menu. Jog Dial COMD Jog Dial CO:895
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 6-7 - Cleaning the Trackball 6-33 Section 6-7 Cleaning the Trackball 1.) Power OFF the scanner. 2.) Place your fingers onto the notches of the trackball retainer ring. 3.) Rotate the retainer ring counterclockwise until it can be removed from the keyboard. 4.) Lift off the inner retainer and trackball from the keyboard.Figure 6-41 Rotating the Retainer Ring Figure 6-42 Removing Inner Retainer and Trackball Retainer Ring Inner RetainerTrackball
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 6-34Section 6-7 - Cleaning the Trackball Section 6-7 Cleaning the Trackball (cont’d) 5.) Wipe off any oil or dust from the trackball using a cleaner or dry cloth. 6.) Wipe off any oil or dust from the trackball housing, rollers, and small ball, using a cleaner or cotton bud. NOTICEWhen cleaning the trackball housing, make sure not to spill or spray any liquid into the trackball housing (Keyboard or system). Avoid organic solvents that may damage the mechanical parts of the trackball assembly. Do not apply much force to the small ball. Figure 6-43 Cleaning TrackBall and Housing Trackbal Housing Cotton Bud Roller Roller Small Ball
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 6-7 - Cleaning the Trackball 6-35 Section 6-7 Cleaning the Trackball (cont’d) 7.) Wipe off any oil or dust from the two rollers using a cleaner or cotton bud. NOTICEWhen cleaning the roller, make sure not to spill or spray any liquid into the trackball housing (Keyboard or system). Use either ethanol, isopropyl alcohol or VCR head cleaner to clean the trackball assembly. Avoid other solvents that may damage the mechanical parts of the trackball assembly. Figure 6-44 Cleaning Rollers Rollers Cotton Bud
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 6-36Section 6-7 - Cleaning the Trackball Section 6-7 Cleaning the Trackball (cont’d) 8.) Insert the trackball into the housing. 9.) Place the trackball and inner retainer into the housing with its stopper facing down.Lift off the inner retainer and trackball from the keyboard. 10.) Install the trackball retainer ring onto the inner retainer, then rotate it clockwise until its notches are set in the horizontal position. Figure 6-45 Removing Inner Retainer and Trackball Figure 6-46 Rotating the Retainer Ring Inner RetainerTrackball Inner RetainerStopper Retainer Ring Notches (Set Horizontally)
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 6-8 - Jumper and Dip Switch Setting 6-37 Section 6-8 Jumper and Dip Switch Setting 6-8-1 Dip Switch Setting Normally the dip switches shall not be adjusted. In case of special needs, such as scanner upgrade, refer to each instructions/documentations for proper setting. The dip switch must be set properly according to the table below. The value in the table (Dip Switch setting) represents the bit location as shown. Figure 6-47 Dip Switch Setting Figure 6-48 Bit Setting (1 to 4) 4 (lower 4 bits) E (Upper 4 bits) Example: “E4 (h)” is set as shown. O N 1 2 3 4 5 6 7 8 O N 1 2 3 4 5 6 7 8O N 1 2 3 4 5 6 7 8O N 1 2 3 4 5 6 7 8O N 1 2 3 4 5 6 7 8O N 1 2 3 4 5 6 7 8O N 1 2 3 4 5 6 7 8O N 1 2 3 4 5 6 7 8O N 1 2 3 4 5 6 7 8 0 (h) 1 (h) 2 (h) 3 (h) 4 (h) 5 (h) 6 (h) 7 (h) O N 1 2 3 4 5 6 7 8O N 1 2 3 4 5 6 7 8O N 1 2 3 4 5 6 7 8O N 1 2 3 4 5 6 7 8O N 1 2 3 4 5 6 7 8O N 1 2 3 4 5 6 7 8O N 1 2 3 4 5 6 7 8O N 1 2 3 4 5 6 7 8 8 (h) 9 (h) A (h) B (h) C (h) D (h) E (h) F (h) Bit Setting for DIP Switch 1 to 4 (lower 4 bits)
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 6-38Section 6-8 - Jumper and Dip Switch Setting 6-8-1 Dip Switch Setting (cont’d) Figure 6-49 Bit Settings (5 to 8) 0 (h) 1 (h) 2 (h) 3 (h) 4 (h) 5 (h) 6 (h) 7 (h) O N 1 2 3 4 5 6 7 8O N 1 2 3 4 5 6 7 8O N 1 2 3 4 5 6 7 8O N 1 2 3 4 5 6 7 8O N 1 2 3 4 5 6 7 8O N 1 2 3 4 5 6 7 8O N 1 2 3 4 5 6 7 8O N 1 2 3 4 5 6 7 8 O N 1 2 3 4 5 6 7 8O N 1 2 3 4 5 6 7 8O N 1 2 3 4 5 6 7 8O N 1 2 3 4 5 6 7 8O N 1 2 3 4 5 6 7 8O N 1 2 3 4 5 6 7 8O N 1 2 3 4 5 6 7 8O N 1 2 3 4 5 6 7 8 8 (h) 9 (h) A (h) B (h) C (h) D (h) E (h) F (h) Bit Setting for DIP Switch 5 to 8 (Upper 4 bits)