GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 5-2 - Hardware Compatibility Matrix 5-5 5-2-1-4 Front End Figure 5-29 Front End- FRU Compatibility R8.0.x R8.0.x R8.0.x R8.0.x Latest Software SW Upgraded BT09 (w BT02/BT03 console)SW Upgraded BT09 from BT04/BT06/BT07 19 inch LCD Upgraded BT09 from 17 inch LCD R8.0.X (BEP4) comment 2357804-2 P9536CA DDBF SVC NNNN32ch 2399312 P9535CL DDBF64 ASSY YYYN64ch 5182379 DDBF128 ASSY CSL L7 NNNY128ch/use with EBUS4 2315492 P9538JD NEST BD JIG YYYY 2264606-2 P9535AU M DBRG ASSY NNNN 2264606-3 P9535AU M DBRG ASSY NNNNNoise reduction on BT04 2264606-4 P9535AU M DBRG ASSY YY+ N N +Y if upgraded f rom BT04~BT06 but not BT06-2 2264606-5 P9535AU M DBRG ASSY YY+ N N +Y see c omment above 6T Iss ue solved f rom -4 or lower vers ion 5145999 M DBRG2S ASSY CSL L7 NY*YYMDBRG 2S only on BT06-2 cons ole (MDBRG2S = MDBRG+SINANO2, Insert to SINANO2 slot ) *Y if upgraded from BT06-2 5145999-2 M DBRG2S ASSY CSL L7 NY*YYImprovement for Memory Test false fail. Previous parts can still be used as FRU 2271702-2 P9535BA EBUS ASSY NNNN 2399317 P9535FA EBUS2 ASSY YNNN Only for Upgraded BT04 from BT01/02/03 5111348 EBUS3T ASSY CSL LOGIQ7 NYY NUpper 5112465 EBUS3B ASSY CSL LOGIQ7 NYY NBottom 5182215 EBUS4T ASSY CSL L7 NNNY 5182757 EBUS4B ASSY CSL L7 NNNY 2273295 P9535BC M DCON ASSY NNNNNo need in BT04system 2264608-3 P9535AY M OTHER ASSY NNNN 5111345-2 M OTHER2 ASSY CSL LOGIQ7 NYYY 2264596-2 P9535AE PREA ASSY NNNN32ch 5119039 PREA2F ASSY CSL LOGIQ7 NYYY 64ch 2393853 PREA2H ASSY CSL LOGIQ7 YNNN Only for Upgraded BT04 from BT01/02/03 2264604-2 P9535AS PROM P ASSY NNNNNo need in BT04 /BT06system 2304617-5 P9538HW QCON ASSY YYYYNoise reduction on BT04 2264602-3 P9535AQ SINANO ASSY NNNN 2374744 P9535BQ SINANO2 ASSY YYNN 2277244-4 P9535AG STCW ASSY YYYY 2268026 P9535BY TERM INATOR ASSY YYYY 2323353-4 P9535BJ TRAP2 ASSY NNNN32ch 5118342-3 BTRAPHV ASSY NYNN 64ch. Countermeasured against HVL current control with full heigh HVPS. 5154321 BTRAPP5AHV Assy YNYY P 5A 64ch. 2323450-4 P9535GA TRAPCW ASSY NNNNTRAP + TXCW 5111346 LOGIQ7 BTXCW ASSY YYYYBTXCW only (piggy board) 5244917 MOTHER2A ASSY CSL LOGIQ7 NNNN 2305111-3 ? TX CABLE ASSY YYYY FRU Part# Description FRU compatibility BT09~
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 5-6Section 5-2 - Hardware Compatibility Matrix 5-2-1-5 Backend Figure 5-30 Backend - FRU Compatibility R8.0.x R8.0.x R8.0.x R8.0.x Latest Software SW Upgraded BT09 (w BT02/BT03 console)SW Upgraded BT09 from BT04/BT06/BT07 19 inch LCD Upgraded BT09 from 17 inch LCD R8.0.X (BEP4) comment 5119906-2 BCOMPSW3 SVC CSL L7 (w/o Video Card) YY NNNo Video card mounted. Need to reuse Video Card when BECOM P3 replace to this new one. Pentium 4 (BEP3) PCMother +CPU+PC2IPB+ATX +Ethernet+H DD+Chas is +Memory+ Mis c Par ts 5119906-3 BCOMPSW3 SVC CSL L7 (w/o Video Card / PC2IP) YYY NNo Video card/PC2IP/FALCON mounted. Need to reuse Video Card / PC2IP/FALCON when BECOMP3 replace to this new one. Pentium 4 (BEP3) PCMother +CPU+ATX +Ethernet+HDD+Chasis +Memory+ Mis c P arts +B as eImageCD 5119906-4 BCOMPSW3 SVC CSL L7 (w/o Video Card / PC2IP) YYY NNo Video card/PC2IP/FALCON mounted. Need to reuse Video Card / PC2IP/FALCON when BECOMP3 replace to this new one. Pentium 4 (BEP3) PCMother +CPU+ATX +Ethernet+HDD+Chasis +Memory+ Mis c P arts +B as eImageCD 5308149 SVP-BECOMP4-L7 N N NYNo Video card/PC2IP/FALCON mounted. Need to reuse Video Card / PC2IP/FALCON when BECOMP4 replace to this new one. Core2 DUO (BEP4) PCMother +CPU+ATX +Ethernet+HDD+Chasis +Memory+ Mis c P arts 5316350 SVP BECOMP4 PARTS N N NYMisc. parts for BEP4. Parts include SATA signal/power cable for HDD, DVI-VGA connector, BEP4 internal USB cables and BEP4 Label 2304809-2 P9536RH BATTERY PACK YYYY 5118511 Extended M EM ORY3 LOGIQ7 YYYN Never be us ed on BE P2/ BEP 4 5306528 DDR2 M EM ORY 4 1GB NNNY 5118510 HD LOGIQ7 YYYN80GB PATA 5306526 SATA HDD 160GB NNNY160GB SATA type. To be used with BEP4 5179643 HDD For Consip L7 Console NNNN160GB PATA type FC200656 PC2IP II B YY* N N Only supported by BECOMPSW 3 *Y if BT04/BT06 / N if BT06-2/V65x/BT07~ FC200755 PC2IP 3 NNYYP C2IP3 need R620 application 2362887 P9536TK AGP BOARD ASSY YY+ N N G550 Video Card AGP Type for CRT console +Y if CRT console 5111298 DUAL DVI AGP NY$YNSONO64 Video Card AGP Type : Only for LCD System Y$ if LCD console 5121039 PCI VGA NY$YN G450 Video Card PCI Type : Only for LCD System 5306525 PCI Express Graphic board NNNYMatrox P690 Video Card for PCI-E slot 5142743 BECOM P3 Rear Fan Assy YYYYCommon for BEP3/BEP4 TBD SCSI Card YNNN Mounted only on Upgraded B T 04 /BT06 system in BECOMP3 for MOD Connection 5147460 F ALCON Capture Board NY%YYVideoCapture Board directly from REAN3 on BT06-2 console Y% if upgraded from BT06-2/V65x~ (with REAN3) 5169391 ? YC Cable (S-Video) NY%YYCable that connects VideoCapture Board between REAN3 on BT06-2 console Y% if upgraded from BT06-2/V65x~ (with REAN3) 2384944 P9536CE DGPCIO/VIC_SVC NNNNOnly suppoted by BEP2. Even if BEP2 mounted, if PCVIC is mounted, order PC2IO/VIC for BT01 console. 5133526 DGPCIO VIC 2-A ASSY YYYN Updated f or LC D 2389075 ? PCI CABLE YYYY 5301220-2 PCI-DGVIC2 Assembly NNNY 5257226 BulkHead for L7 and LS6 NNNY 5257231 Cable13 f or B ulkHead NNNY 5257233 Cable25 f or B ulkHead NNNY FRU Part# Description FRU compatibility BT09~
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 5-2 - Hardware Compatibility Matrix 5-7 5-2-1-6 Power Units Figure 5-31 Power Units- FRU Compatibility R8.0.x R8.0.x R8.0.x R8.0.x Latest Software SW Upgraded BT09 (w BT02/BT03 console)SW Upgraded BT09 from BT04/BT06/BT07 19 inch LCD Upgraded BT09 from 17 inch LCD R8.0.X (BEP4) comment 2304538-4 P9534WP AC POWER BOX 200V Y N N N Include AC BOX, AC Code, and extention cable 2384371 P9534XL AC POWER BOX ASSY JPN YNNN Include AC BOX, AC Code, and extention cable (Only for Japan) 2384370 P9534XP AC POWER BOX ASSY 100V YNNN Include AC BOX, AC Code, and extention cable (not for Japan) 2304624 P9538JC CONSOLE FAN SSY NNNN 5117493 CONSOLE FAN ASSY 2 CSL L7 N YYY 5122614 CONSOLE FAN ASSY-A UPG L7 Y N N N Only for Upgraded BT04 from BT01/02/03 2372387 P9534XH L7HV-SVC YYYY 2334197 P9534WU LV-2 SERVICE UNIT Y N N N Compatible with 2283531 (LV) 2292209 P9534WF SSR PANEL Y NNN 2315585 P9534WR FUSE FOR SSR PANEL Y NNN 2283530 P9534WB M AIN TRANS Y NNN 2373331-2 M AIN TRANS2 ASSY CSL L7 YYYY 5117721 LV-04 UNIT M AIN PS L7 N YYY 5118824-2 LV-04 UNIT with Cables N YYY Nois e Improvement 5117723 NF INLET ASSY M AIN PS L7 N YYY 5117724 AC IN OUT 100V M AIN PS L7 N YYY 5117725 AC IN OUT 200V M AIN PS L7 N YYY 5118052 AC IN200 OUT100 ASSY M AIN PS L7 N YYY O nly f or Korea Sys tem 5117726 SSR PCB ASSY M AIN PS L7 N YYY 5117727 FUSE HOLDER M AIN PS L7 N YYY 5117728-3 CONN ASSY M AIN PS L7 N YYY 5119119 SSR-HLV CABLE N YYY 5119120 SSR-AC PANEL CABLE N YYY 5119121 VOLTAGE SEL CABLE N YYY 2301858-3 ? PW SW ASSY N YYY 5117493 CONSOLE FAN ASSY2 N YYY 5117491 FAN CABLE BRAKET ASSY N YYY FRU Part# Description FRU compatibility BT09~
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 5-8Section 5-2 - Hardware Compatibility Matrix 5-2-2 Hardware Identification Tip 5-2-2-1 LCD Monitor Differences The flowchart below describes how to differentiate LCD monitor type. Figure 5-32 LCD monitor- How to Differentiate GPN Identification Mark (Bottom of LCD monitor)
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 5-2 - Hardware Compatibility Matrix 5-9 5-2-2-1 LCD Monitor Differences (cont’d) LCD Variation Tips: • Bezel Shape - Curvy for 17 inch - Straight for 19 inch • GE Logo - Silver for 17 inch - Silver on black background for 19 inch Figure 5-33 LCD Variation Tips
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 5-10Section 5-2 - Hardware Compatibility Matrix 5-2-2-2 BT09_SV_BEP4 Configuration Variationos Figure 5-34 BT09 Configuration Variations N/A R8.x.x R8.x.x SoftwareNote Console Config N/A BT04/BT06/ BT07 base BT04/BT06/B T07/BT09 base MD-3000 USB Upgraded to BT09 (BEP3) BT09 Original (BEP4) ・Label “RS232C Not Available” on back ・ Utility Menu shows Ghost Part Number ・ NO Rear Panel markings ・ Utility Menu shows Ghost Part Number ・ REAR Panel have GPN printed : 5304518 : REAR CONN4 ASSY L7N/A R8.x.x R8.x.x Software Note Console Config N/A BT04/BT06/ BT07 base BT04/BT06/B T07/BT09 base MD-3000 USB Upgraded to BT09 (BEP3) BT09 Original (BEP4) ・Label “RS232C Not Available” on back ・ Utility Menu shows Ghost Part Number ・ NO Rear Panel markings ・ Utility Menu shows Ghost Part Number ・ REAR Panel have GPN printed : 5304518 : REAR CONN4 ASSY L7 RS232C Not Available BEP4 MD-3000 USB
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 5-3 - Peripheral Compatibility 5-11 Section 5-3 Peripheral Compatibility NOTE: For SONY D21MD (or D23MD) color printer ONLY: The D21MD (or D23MD) printer contains the L-sized ribbon cassette and paper as a standard. However, the S-sized paper is recommended for LOGIQ 7 system. If the S-sized paper is used with the L-sized ribbon cassette, “Incorrect paper size setting” error occurs with no print. So, when using the S-sized paper, change the ribbon cassette to the S-sized. Figure 5-35 Peripheral Compatibility
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 5-12Section 5-4 - Block Diagrams and Theory Section 5-4 Block Diagrams and Theory 5-4-1 Block Diagram (R7.5.x or later, 19 inch LCD model) Figure 5-36 System Block Diagram (R7.5.x or later) 1 SPCF $PO1 SPCF $PO 1$0/ 48 48 5*OU FSGBDF&64$ *OUF SGBD F 14&- 2$0/ L NA TGC 13& 1 VT MFS 4*$64ch EU SC I/F & Pul se gener ator %SJWFS 1 VT MFS 4*$1 VT MFS 4*$)74&-#5311)7#531$8%ESJWFS #59$8 %SJWF S%SJWF S%SJWF S%SJWF S%SJWFS 128ch&64$ *OUF SGBD F&64$ *OUF SGBD F45$816bit A/ D PSN %FU LNA TGC L NA TGC LNA TGC LNA TGC L NA TGC &64$ *OUFS GBDF Am p 8chAD CAA FAm p AD CAA F OQX 2 Am p 8chAD CAA FAm p AD CAA F OQX 2 Test Si gnal 8bl ockClo ck I/F B¡ d SU M I/ F Am p 8chAD CAA FAm p AD CAA F OQX 2 &64$ *OUFS GBDF %%# TG C Control Test Siganl Clock Distribu ter M aster Clock (160MHz) TI-DSP PC I I/F CHACOM I / F TIDS P I/ F ADI DSP I/ F EUSC I/F DDR SDRAMCHACOMCHACO M 1( 41 ** -JOL1P SU -JOL1PSU &.* MID BUS(PCI)D opple r A udio PC I-P CI Bridge Intel 21152 PC I Cabl e to B EP SDRAMSH 4-PC I Br idge SH BUSRO MSH4 EUSC I/FSafe ty6T P rob eHV C TL4D BoxPS Observat ion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
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 5-4 - Block Diagrams and Theory 5-13 5-4-2 General Information • LOGIQ™ 7 is a phased and linear array ultrasound imaging scanner. It has provisions for analog input sources like ECG and phono. A dedicated Doppler probe may also be connected and used. • The system can be used for: - 2D Black and White imaging -2D Color Flow - M-Mode Black and White imaging - Color M-Mode - Doppler - a number of combinations of the above • LOGIQ™ 7 is a digital beam forming system that can handle up to 192 element linear probes through multiplexing. • Signal flow from the Probe Connector Panel to the Front End, to the Mid Processors and Back End Processor and finally to the monitor and peripherals. • System configuration is stored on a hard disk and all necessary software is loaded from the hard disk on power up.
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 5-14Section 5-4 - Block Diagrams and Theory 5-4-3 CPU/Back End Processor (For BEP4 for BT09 or later) 5-4-3-1 BEP4 Signal Connections Figure 5-37 BEP4 Signal Cable Connection CPU PW2 Keyboard, Mouse ( Not Used ) VGA ( Not Used ) USB Ports (2 ports) MIC, Line in/out OPMA Conn ( Not Used ) PCI-Express slot - Graphics Card PCI Slot - PCIDGVIC2 PCI-Express slot (Not Used ) PCI Slot - FALCON PCI Slot - PC2IP3 PCI Slot ( Not Used ) FDD / IDE ( Not Used ) USB 4 USB5 USB6 USB3 CPU Fan Fan2 ( Not Used ) Fan 6 Fan 4 ( Not Used ) SATA LAN (one of three is used) PW1 Memory DIMM1 DIMM2 DIMM3 DIMM4 Not Used Not Used SATA HDDBulk Head Extended USB Ports (2 ports) Pin Header ( Power SW,Reset SW, HDD LED, Power LED ) Connector Panel ECG unit [ Signal Cable connection in BECOMP4 for L7 ] Extended USB Ports (2 ports) Extended USB Ports (4 ports) USB Ports (2 ports) Fan 5 Fan 3 ( Not Used )