GE Logiq 7 Service Manual
Have a look at the manual GE Logiq 7 Service Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-15 8-2-7 Install Base System (cont’d) 4.) The following screen appears. Press B then Enter to initialize the C drive 5.) Check if the message appears as reloading C: partition only, then press Enter key. If the message appears as reloading the entire disk, press CTRL-ALT-DEL to reboot the sysytem and retry from procedures described in previous page. Figure 8-65 Select B CAUTIONIf you select [A], all the archive would be lost. Figure 8-66 Enter
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 8-16Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-2-7 Install Base System (cont’d) 6.) If the following screen appears, verify that OK buttton is selected, then press Enter key. (If it is not selected, select OK button using the arrow keys then press Enter key.) 7.) Approx. fifteen minutes later, the following screen appears. Eject the Base System Software Load Image DVD from the drive. 8.) To restart the system, press Ctrl + Alt + DEL key at the same time. Figure 8-67 OK Figure 8-68 Ejecting DVD NOTICEThe Base System Software Load Image DVD must be stored onto the PC box cover.
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-17 8-2-8 Modify System/Windows Settings After Base System is successfully installed, various parameters must be properly set before proceeding to Application load. The below outlines necessary system/windows setting. 8-2-8-1 UPS Setting UPS automatically setup by Ghost. You do not need to setup it manually. You can see the message below on the desktop. Please choose OK. Table 8-6 System/Windows Settings ItemMethodProper Value UPS Setting Automatic Setting (n/a – just load) (Restart System) Avance AC97 Audio Load from c:\windows\system32\drivers (n/a – just load) (Restart System) Drive LetterAutomatic Setting but check desktop icon if not.DVD Drive = G HW DeviceMy Computer Properties Various LAN Desktop My Network Places \ Properties Local Area Connection Double Click to activate TCP/IP Setting Desktop My Network Places \ Properties \ Local Area Connection \ Internet Protocol (TCP/IP) \ Properties \ Advanced \ Options TCP/IP Filter Values in ‘Permit Only’ Date/Time Start Settings \ Control Panel Date and Time Proper Time Zone Accurate Date / Time Figure 8-69 UPS Setting
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 8-18Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-2-8-2 Drive Letter Setting If you CANNOT see this dialog at the first boot up, this dialog appears at 2nd boot up. Please choose OK. You should check the Drive Letter manually by Windows Explorer for double check. - DVD Drive should be assigned to G: If Drive letter is wrong, please double-click Remap Drive Letter icon on the desktop. 8-2-8-3 Hardware Device Check: 1.) On the windows desktop, left-click on My Computer, then right-click on Properties. Figure 8-70 OK CAUTIONDrive F: (Swap) Letter is removed on R7.0.0EZ. (This is correct.) Figure 8-71 Properties
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-19 2.) The System Properties window appears. Select Hardware > Divice Manager... 3.) Verify that X or ! is not displayed. Figure 8-72 Divice Manager NOTICEThe Eusc (Mlultifunction adapters > Eusc) might contain “ !” mark, but it is no problem. Skip any of device driver installation procedures described below. “!” mark on Eusc is no problem.
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 8-20Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-2-8-3 Hardware Device Check: (cont’d) 4.) If X or ! is displayed: - Left-click on the device with x mark, then rigtht-click on Enable. - Left-click on the device with ! mark, then rigtht-click on Properties to perform the device driver reconfiguration. When some files are required, designate the file in the directories of C:\WINDOWS\SYSTEM32 or C:\WINDOWS\SYSTEM32\DRIVERS . If they are not found, search them using a windows search function. 5.) Close the window.
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-21 8-2-8-4 LAN Parameter Setting 1.) On the desktop, left-click on My Network Places, then right-click on Properties... 2.) Verify that two network icons appears. (If “Local Area Connection” icon does not appear, the hardware failure might occur.) 3.) Using the right trackball button, double-click on Local Area Connection. 4.) When the following window appears, it will be automatically closed later. 5.) When the following window appears, right-click on Cancel. 6.) Close all windows. Figure 8-73 Properties Figure 8-74 Local Area Connection Figure 8-75 Cancel
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 8-22Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-2-8-5 TCP/IP Filter Setup 1.) On the desktop, left-click on My Network Places, then right-click on Properties... 2.) Double-click Local Area Connection or Local Area Connection #2 (if you do NOT find Local Area Connection). 3.) Select Internet Protocol (TCP/IP) , then click Properties button. Figure 8-76 Local Area Connection Figure 8-77 Internet Protocol (TCP/IP)
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-23 8-2-8-5 TCP/IP Filter Setup (cont’d) 4.) Click Advanced... button. 5.) Click Options tab.Figure 8-78 Advanced Figure 8-79 Options