Home > GE > Medical Equipment > GE Logiq 7 Service Manual

GE Logiq 7 Service Manual

    Download as PDF Print this page Share this page

    Have a look at the manual GE Logiq 7 Service Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    Page
    of 394
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    Section 6-5 - Monitor and LCD Adjustments 6-19
    6-5-5-2 For Technart (cont’d)
    4.) “AutoAdjust OK” message will appear. If it does not appear. press  AUTO button again.
    Figure 6-23   AutoAdjust OK message
    Figure 6-24   Auto Button
    Auto Button 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    6-20Section 6-5 - Monitor and LCD Adjustments
    6-5-5-2 For Technart (cont’d)
    5.) Display the LCD monitor menu and visually check if auto adjustment succeeds.
    - If no character blurring exits, auto adjustment succeeds. Go to  6-5-5-2-3 Completing 
    adjustment .
    - If character blurring exits, auto adjustment fails. Perform  6-5-5-2-2 Manual Deblurring (For 
    Technart ONLY) .
    Figure 6-25   Blurring/No Blurring
    Blurring at the right side of the character Example: Blurring
    Example: No blurring 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    Section 6-5 - Monitor and LCD Adjustments 6-21
    6-5-5-2-2 Manual Deblurring (For Technart ONLY)
    Perform manual deblurring only when auto adjustment fails.
    1.) Refer to 4-2-2-3  Entering Maintenance Mode  to enter the windows desktop.
    2.) Select Program > Accessories > Paint  to run the paint software.
    3.) Phase Check: a.) Drag and drop the paint screen on the touch panel screen.
    b.) Click on square symbol at the top right corner of the paint screen to maximize the screen. Figure 6-26   Paint Software
    Figure 6-27   Moving Paint Screen 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    6-22Section 6-5 - Monitor and LCD Adjustments
    6-5-5-2 For Technart (cont’d)
    c.) Click on the line icon and select the thinest line.
    d.) Draw a vertical line at the right side of the screen below and check if it has no deblurring.
    - If deblurring exists, go to next step.
    - If no deblurring exists, go to step 4- HTotal Check Figure 6-28   Line
    Figure 6-29   Drawing Line 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    Section 6-5 - Monitor and LCD Adjustments 6-23
    6-5-5-2 For Technart (cont’d)
    e.) Display the LCD monitor menu and press the MENU button once. The following menu appears.
    f.) Repeat to press the MENU button until the “Phase” blinks.
    g.) Repeat to press the PLUS button to find the value with no character blurring. For example, this value is X.
    h.) Set the value ( X + 14) to Phase.
    i.) Press the AUTO button once to save the setting.
    j.) Repeat to press the MENU button until the “Exit” blinks
    k.) Press the AUTO button once to exit from the menu. Figure 6-30   Menu Screen
    NOTICEYou can select the number from 0-62. Values with deblurring cause are continuous seven or eight 
    values. So, tips for searching the best value is to avoid such a group. This depends on video card or 
    VGA cable of your scanner. 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    6-24Section 6-5 - Monitor and LCD Adjustments
    6-5-5-2 For Technart (cont’d)
    4.) Htotal Check:
    a.) Display the LCD monitor menu and press the MENU button once. The following menu appears.
    b.) Check HTotal value:
    - If it is 16, go to step5 - Hposition/VPosition check
    - If it is NOT 16, go to next step.
    c.) Repeat to press the MENU button until the “HTotal” blinks.
    d.) Repeat to press the PLUS or MINUS button to set  16.
    e.) Press the AUTO button once to save the setting.
    f.) Repeat to press the MENU button until the “Exit” blinks
    g.) Press the AUTO button once to exit from the menu.
    5.) Hposition/VPosition Check:
    a.) Close the paint software and display windows desktop. Figure 6-31   H Total
    Figure 6-32   Start-up Screen 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    Section 6-5 - Monitor and LCD Adjustments 6-25
    6-5-5-2 For Technart (cont’d)
    b.) Check the upper, lower, left, and right edge of the screen. They should be centered evenly in the monitor.
    - If they are vertically out of alignment, adjust VPosition.
    - If they are horizontally out of alignment, adjust HPosition.
    c.) Display the LCD monitor menu and press the MENU button once. The following menu appears.
    d.) Repeat to press the MENU button until the “HPosition” or “VPosition” blinks.
    e.) Repeat to press the PLUS or MINUS button until the screen should be centered evenly in the  monitor.
    f.) Press the AUTO button once to save the setting.
    g.) Repeat to press the MENU button until the “Exit” blinks
    h.) Press the AUTO button once to exit from the menu.
    6.) Go to  6-5-5-2-3 Completing adjustment .
    Figure 6-33   H Position or V Position 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    6-26Section 6-5 - Monitor and LCD Adjustments
    6-5-5-2-3 Completing adjustment
    1.) Display the LCD monitor menu and press the left button (MENU). The following menu appears.
    2.) Verify that:- Contrast: 210
    - ContrastRed: 145
    - ContrastGrn: 130
    - ContrastBlu: 160
    3.) Verify that: - The display does not move up, down, or sideways.
    - All menu are shown in the LCD display.
    - Turning the brightness knob clockwise fades in and turning it counterclockwise fades out. Figure 6-34   Contrast
    210
    37
    145
    130
    160 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    Section 6-6 - BW Printer Setting / Adjustment 6-27
    Section 6-6
    BW Printer Setting / Adjustment
    6-6-1 Parameters for UP-D897
    1.) When system started, Go to Utility > System >Peripherals, then hit Printers button in the Setup 
    group.
    2.) Click  SONY UP-D897  in the Printer List, then right click and select Properties .
    3.) Select  General tab, then hit  Printing Preferences ... 
    4.) Select  Layout tab, then ...
    -Set  1920x1280  into Paper
    -Check  off into Enlarge to Paper
    -Set  Bilinear into Interpolation Method
    -Set  Landscape  into Orientation
    Figure 6-35   Printing Preferences
    Bilinear 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    6-28Section 6-6 - BW Printer Setting / Adjustment
    6-6-1 Parameters for UP-D897 (cont’d)
    5.) Select  Density Adjust  tab, then ...
    -Set  TONE1 into GAMMA
    -Set  7 into Sharpness
    Figure 6-36   Printing Preferences 
    						
    All GE manuals Comments (0)