    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-25
    8-2-8-5 TCP/IP Filter Setup (cont’d)
    7.) If your system has no TCP port numbers as follows, perform these procedure to add the TCP port numbers. If the TCP port already have been set, skip these procedures.  (This depends on the 
    network card installed in the PC box.)
    * At first check mark  Enable TCP/IP Filtering (All adapters) .
    *Check mark  Permit Only in TCP ports box
    *Click  Add... button.
    * Enter  21 (port number) as the TCP Port ,then click OK  button to apply it.
    * Repeat these procedures to register 21, 23, 80, 104, 512, 3003, 5800, 5801, 5900, and  5901 for TCP ports. Figure 8-81   TCP port number
    NOTICEUDP Ports and IP Protocols must be kept default settings.
    The TCP port numbers, 
    21, 23, 80, 104, 512, 3003,
    5800, 5801, 5900, and 5901 , have been set.No TCP port number is set. They must
    be set. 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    8-26Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later)
    8-2-8-5 TCP/IP Filter Setup (cont’d)
    8.) Verify that the proper TCP port numbers are entered in TCP Ports box, then click  OK buttons for all 
    opened window to complete the TCP/IP Filter Setup .
    9.) The following dialog box appears. Click  No to proceed the other settings. (After completion of all 
    system checks, the system is shut down.) Figure 8-82   TCP/IP Filter Setup
    Figure 8-83   No
    NOTICEIn case the last dialog box to restart the computer does NOT appear, check the TCP/IP Filter setup 
    again.
    Connectivity check will be performed in 8-2-10 System Settings. 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-27
    8-2-8-6 Date and Time Setting
    1.) Select Start > Settings > Control Panel > Date and Time. I
    2.) Click on  Time Zone tab.
    3.) Select the appropriate time zone.  Then click on  Apply.
    4.) Then click on Date & Time tab. Figure 8-84   Date and Time
    Figure 8-85   Time Zone 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    8-28Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later)
    8-2-8-6 Date and Time Setting (cont’d)
    5.) Set the current date and time.
    6.) Click on  Apply.
    7.) Click on  OK to close the Date and Time window.
    8.) Select Start > Shutdown > OK  to power OFF the scanner.
    Figure 8-86   Date & Time
    Month and Year
    Day Time 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-29
    8-2-9 Installing R8.x.x Application Software
    8-2-9-1 Installing R8.x.x Application Software
    1.) Power ON the scanner. The windows desktop appears. 
    2.) Insert the Application Software CD-ROM into the DVD drive. 
    NOTICEYou must install BT09 Base image before Application Software installation. If not, Application shows the 
    error message, “Service platform is out of date. You must reghost the C drive and reinstall the software.”
    NOTICEIf you are attempting to install only application software, you MUST first check current Regional Preset 
    Value.  Press  Utility > System > About  (Details in 4-2-6 - Archiving and Loading Presets for BT07 
    (Including R7.5.x)).
    Figure 8-87   Desktop
    NOTICEAt this time, the DVD writing software screen appears, but you can ignore the screen.
    NOTICEWait approx. 20 seconds until the disk can be read. 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    8-30Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later)
    8-2-9-1 Installing R8.x.x Application Software  (cont’d)
    3.) Double-click on  LoadSoftware.bat.
    4.) The Installation program will start, and the command prompt screen will be shown on the monitor  as shown.
    5.) Select TYPE A - BT09 with BEP4 does not have PRO version available. Figure 8-88   LoadSoftware.bat
    NOTICE1.If wrong application software installed, the system will not work properly. 
    2.You will ask confirmation twice as below. If you select wrong system (A or B) , please select N at the 
    confirmation. 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-31
    8-2-9-1 Installing R8.x.x Application Software (cont’d)
    6.) Type  Y key.
    NOTE: Press the N key to cancel this procedures and to power OFF the system.
    NOTE: When the Information Window appears asking you to select OK, do nothing. This window will  disappear automatically.
    7.) The installation procedures will start automatically. It will take 10 minutes.
    8.) After the installation procedures are completed, the system will power OFF automatically. Figure 8-89   Y
    NOTICEDo not operate with using mouse or keyboard during installation 
    procedures!!! The system shall be restarted automatically during 
    installation. 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    8-32Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later)
    8-2-10 System Settings
    After application is successfully installed, various LOGIQ™ 7 system parameters must be properly set.
    Table 8-7   System Settings
    ItemMethodProper Value
    Software Version Utility
        
    System
            
    About s/w version : R8.x.x
    Image Part No. : 5305206-3
    Regional Preset
    Utility
        
    System
            
    About
                 
    Additional About Information For the system other than “NONE”
    selected as Regional Preset, resetting
    MUST be required.
    Computer Name Utility
        
    Connectivity
            
    TCP/IP
                 
    Computer Name
                 
    Enable DHCP
                 
    IP Address
                 
    Subnet Mask
                 
    Default Gateway Same as Serial Number
    OFF
    restore customer setting
    restore customer setting
    restore customer setting
    Patient 
    Database CheckUtility
        
    Connectivity
            
    DataFlow
              \  
    Default Data Flow 
    Press Patient (Keyboard or Touch Panel)Local Archive – Int. HD
    (n/a – Automatic)
    LCD Monitor SettingPress ‘Toggle Button’ for few seconds(various)
    Software OptionUtility
        
    Admin
            
    System Admin
                 
    HW NumberSame as Option Dongle ID
    VCR Utility
        
    Peripherals
            
    Video Setting
            
    VCR  PAL/NTSC-depends on region
    (set accordingly)
    Language pre-work (For
    Japanese Only) Utility
        
    Regional Option
            
    Language (tab)
              \  
    Install Files to East Asian Language
    (Reboot) ON / Apply 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-33
    Language (For Japanese
    Only)Utility
     
    System 
    General
     
    Language
      
    Regional Option
      
    Regional Option (tab)
      
    Standard and formats
      
    Location
      
    Language (tab)
      
    Details   
    Settings (tab)
      
    Installed Service
      
    Default Input Language
      
    Advanced (tab)
      
    Language JPN
    Japanese
    Japan
    Japanese Keyboard
    Japanese
    Japanese
    Language (For Russian/
    Greek Only)Utility
     
    System
     
    General  
    Language
      
    Regional Option
      
    Regional Option (tab)
      
    Standard and formats
      
    Location
      
    Language (tab)
      
    Details   
    Settings (tab)
      
    Installed Service
      
    Default Input Language
      
    Advanced (tab)
      
    Language
    RUS/GRK
    Russian/Greek
    Russia/Greece
    Rus/Grk Keyboard
    Russian/Greek
    Russian/Greek
    Table 8-7   System Settings
    ItemMethodProper Value 
    						
    							GE HEALTHCARE
    DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL
    8-34Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later)
    8-2-10-1 Peripheral Device Connection/Setup
    1.) Connect all of the peripheral devices to the scanner.2.) Power ON all of the peripheral devices.
    3.) Verify that the system boots up with no error. (It will take a few minutes.) 4.) Set up customer data saved at 8-2-6 Pre-installation Procedures.
    8-2-10-2 Confirmation of the software version 1.) Touch the  Utility button on the Touch Panel.
    NOTE: The Operator Login window appears. Select 
    adm (Administrator) for the Operator field.
    Adm (Administrator) will be shown in the Operator field as default. 
    Enter the password which is configured already (or the password might not be necessary).
    Then click on 
    Log on. Figure 8-90   Utility
    Figure 8-91   Log on
    Press
    Utility
    button. 
    						