GE Logiq 7 Service Manual
Have a look at the manual GE Logiq 7 Service Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 6-8 - Jumper and Dip Switch Setting 6-39 6-8-2 Jumper Setting 6-8-2-1 HDD Always set the jumper between MASTER pins. 6-8-3 Printer Dip Switch Setting 6-8-3-1 Mitsubishi MD-3000 VCR Dip Switch SettingThis section describes Mitsubishi MD-3000 VCR communication setting. Field Service Engineer (FSE) is not allowed to modify dip switch setting. Following setting is documented for purpose of troubleshooting by FSE. Figure 6-50 HDD Jumper Figure 6-51 MD-3000 Dip Switch Master 1 2 3 4 5 6 7 8 ON OFF S W50 01 1 2 3 4 5 6 7 8 ON OFF S W50 01 Serial Communicatiton USB Communicatiton
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 6-40Section 6-8 - Jumper and Dip Switch Setting 6-8-3-2 Accessing Dip Switch Module for inspection 1.) Remove the two screws of RS232C board plate. 2.) Remove the RS232C board from the VCR to access the dip switch. Figure 6-52 Opening Board Plate Figure 6-53 Removing Board Screws RS232C board plate Dip Switch RS232C Board
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 6-8 - Jumper and Dip Switch Setting 6-41 6-8-3-3 CP30D Dip Switch Setting This section describes CP30D Dip Switch setting. Field Service Engineer (FSE) is not allowed to modify dip switch setting. Following setting is documented for purpose of troubleshooting by FSE. NOTICE D Do NOT change the DIP SW setting!! Figure 6-54 CP30D Dip Switch (All OFF) 1 2 3 4 5 6 7 8 ON OFF
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 6-42Section 6-8 - Jumper and Dip Switch Setting
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 7-1 - Overview 7-1 Chapter 7 Diagnostics/Troubleshooting Section 7-1 Overview 7-1-1 Purpose of Chapter 7 This section describes how to setup and run the tools and software that help maintain image quality and system operation. Very basic host, system and board level diagnostics are run whenever power is applied. Some Service Tools may be run at the application level. However most software tests are required. Table 7-1 Contents in Chapter 7 SectionDescriptionPage Number 7-1 Overview 7-1
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 7-2Section 7-1 - Overview 7-1-2 Diagnostic Procedure Summary Although Diagnostics can be run in any order, the Bottom-up Confidence-Building Order outlined in this section: - Provides a framework from which overall diagnostic testability can be discussed. - Provides a top-level model that describes the confidence-building aspect of the diagnostics. - Provides a logical step-by-step approach to system check-out and fault isolation. There are two levels of diagnostic: board-level and system level. - Board-level diagnostics are intended to test functionality of a single circuit board. - System-level diagnostics are intended to test functionary on more than one circuit board. Unused system components (board or signals) for each diagnostic test are drawn in gray (ghosted). NOTE: In this document, the Host includes all hardware upstream of the PCI cable. The diagnostics in this chapter do NOT test anything upstream of the PCI cable. Therefore, any upstream hardware or software must be functional before running these diagnostics.
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 8-1 - Overview 8-1 Chapter 8 Replacement Procedures Section 8-1 Overview 8-1-1 Purpose of Chapter 8 This chapter describes replacement procedures for the following modules and subsystems. 8-1-2 Returning/Shipping Probes and Repair Parts Equipment being returned must be clean and free of blood and other infectious substances. GEMS policy states that body fluids must be properly removed from any part or equipment prior to shipment. GEMS employees, as well as customers, are responsible for ensuring that parts/equipment have been properly decontaminated prior to shipment. Under no circumstance should a part or equipment with visible body fluids be taken or shipped from a clinic or site (for example, body coils or an ultrasound probe). The purpose of the regulation is to protect employees in the transportation industry, as well as the people who will receive or open this package. NOTE: The US Department of Transportation (DOT) has ruled that “items that were saturated and/or dripping with human blood that are now caked with dried blood; or which were used or intended for use in patient care” are “regulated medical waste” for transportation purposes and must be transported as a hazardous material. Table 8-1 Contents in Chapter 8 SectionDescriptionPage Number 8-1 Overview 8-1 8-2 Software Loading Procedures for BT09 (R8.x.x or later) 8-3
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 8-2Section 8-1 - Overview This page was intentionally left blank.
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-3 Section 8-2 Software Loading Procedures for BT09 (R8.x.x or later) 8-2-1 General This describes software loading procedures for BT09 System with BEP4 only. For the consoles upgraded BT09 systems (contains BEP3 inside), follow procedure in Section 8-7 Software Loading Procedures (R7.5.x or later). Also, do not attempt to apply this procedure if you are upgrading older BTs to BT09/R8.x.x. NOTE: Be sure no images are in the clipboard or opened exams before loading base image software. NOTE: Before loading base image software, if you are troubleshooting loss of image or patient data, review troubleshooting to collect information for future analysis. 8-2-2 Parts Required - LOGIQ 7 R8.x.x Application CD -LOGIQ7 Ghost DVD for BECOMP4: 5305206-x 8-2-3 Time Required Approximately 1 hour CAUTIONThis software can be installed for the system with software version R8.x.x. CAUTIONILoading R8.x.x application software procedures requires basically C drives to be ghosted. In this case, patient image data, preset parameters, and customer data will be kept. However , in case full ghost will be required, you MUST back up before starting this procedure as follows. For customer data safe, we recommend you to back up even for the C ghost only.
GE HEALTHCARE DIRECTION 2286865, REVISION 14LOGIQ™ 7 SERVICE MANUAL 8-4Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later) 8-2-4 Software Loading Work-flow and Check-list This check sheet is prepared to assist a FE during software loading. Customer NameCustomer Contact Console #Unit ID Unit Type SectionItemCheck ParameterResult 5-2-2-2 BT09_SV_BEP4 Configuration Varia- tionosCheck current system version before ghost/application loadingShould be R8.x.x or laterVer. 8-2-6-1 - Disabling USB DevicesCheck if all USB printers are OFF. Check if SCSI UP-D50 is ON.Should be OFF. OK N/A 8-2-6-2 - Save Cus- tomer Data / SettingSave Customer SettingUser and Password OK N/A Save Customer SettingConnectivity Parameter OK N/A Save Customer SettingOption Key String OK N/A Regional Preset OK N/A 8-2-7 - Install Base SystemInstall Base SystemSelect fullghost or C ghost Full C only 8-2-8 - Modify Sys- tem/Windows Set- tingsUPS SettingRun UPSSETUP.reg OK NG Drive Letter settingCD/DVD to be Drive ‘G’ OK NG Hardware Divice OK NG LAN Parameter OK NG TCP/IP SettingVe ri f y OK NG Date/Time SettingVe ri f y OK NG 8-2-9 - Installing R8.x.x Application SoftwareApplication CD correctly recognizedApp loading OK NG