    							Cardiograph Care and MaintenanceCalibrating the Barcode Reader
    PageWriter Trim Cardiograph Service Manual3-11
    Calibrating the Barcode Reader
    The optional barcode reader is shipped with configured settings that provide optimal use with 
    the cardiograph. If the barcode reader operates with errors or missed data, follow the 
    procedure below.
    If the barcode reader incorrectly scans data into fields on the Patient ID screen, follow the 
    procedure “Removing the Carriage Return” on page 3-13. 
    The barcodes must be scanned in order from top to bottom in one session. 
    To calibrate the barcode reader:
    1
    Hold the barcode reader at a 45o angle and push the button (top of scanner). Scan the 
    barcode labeled 1 in Figure 3-2. Three beeps are heard.
    2Scan the barcodes labeled 2-8 in Figure 3-2. A single beep is heard after each barcode is 
    scanned.
    3After barcode 8 is scanned three beeps are heard. Calibration is complete.
    Perform the Barcode Test (see page 4-2) to verify performance. 
    						
    							Cardiograph Care and MaintenanceCalibrating the Barcode Reader
    3-12PageWriter Trim Cardiograph Service Manual
    Figure 3-2 Barcode Calibration Sequence
    1
    2
    3
    4 
    						
    							Cardiograph Care and MaintenanceCalibrating the Barcode Reader
    PageWriter Trim Cardiograph Service Manual3-13
    Removing the Carriage Return 
    Scan the following barcode when the barcode reader incorrectly scans data in any field on the 
    Patient ID screen. The following procedure removes a carriage return from the end of a data 
    scan that can cause errors when entering Patient ID information.
    6
    7
    8 5 
    						
    							Cardiograph Care and MaintenanceMaintenance Tests for Trim II, III and Rx
    3-14PageWriter Trim Cardiograph Service Manual
    To calibrate the barcode reader:
    1
    Hold the barcode reader at a 45o angle and push the button (top of scanner). Scan the 
    barcode in Figure 3-3. Three beeps are heard.
    Figure 3-3 Barcode scan to remove the automatic carriage return
    2
    Perform the Barcode Test (see page 4-2) to verify performance.
    Maintenance Tests for Trim II, III and Rx
    Use the diagnostic utilities and maintenance tests to verify or to optimize cardiograph 
    performance. The following tests can be executed during active normal cardiograph use, and 
    may be helpful in identifying a technical problem. 
    For information on maintenance tests that are available through the Biomed Service Utility, 
    see “About the Trim II/III/Rx Biomed Service Utility” on page 4-6. Using the Biomed Service 
    Utility requires a full restart of the cardiograph. 
    For further assistance with any of the maintenance tests described in this section, contact the 
    nearest Philips Response Center. See Contacting a Philips Response Center on page 1-32.
    To open the Diagnostics/Maintenance Test screen:
    1
    Press the Ta b key ( ) or turn the Trim Knob to highlight the Config button on the 
    Command Toolbar.
    2Press the space bar or the Trim Knob to select the button. The Configuration screen 
    appears.
    3Select the System tab by holding down the Alt key (on keyboard) and then pressing the S 
    key. The 
    Diagnostics section of the screen lists the five maintenance tests. 
    Appendix 4, “Maintenance Tests” provides instructions for each test. Refer to the page 
    numbers in the table below to locate a specific test. 
    Table 3-33 Diagnostic and Maintenace Tests available in Configuration 
    Test Page
    PIM Test 4-2
    Barcode Test 4-2
    Magcard Test 4-3
    Printer Test 4-3
    Network Test 4-6
    CONFIG 
    						
    							4
    4-1
    Appendix DMaintenance Tests
    This appendix provides instructions for performing maintenance tests on the PageWriter Trim 
    II/III/Rx cardiograph from the normal cardiograph operating mode, and includes instructions 
    for using the Biomed Service Utility. Using the Biomed Service Utility requires a full restart 
    of the cardiograph. This appendix also includes instructions on using the PageWriter Trim I 
    cardiograph Biomed Service Utility. 
    The tests contained in this appendix help you to identify specific technical issues with the 
    cardiograph. 
    Maintenance Tests (Trim II/III/Rx only)
    You can use the maintenance tests and diagnostic utilities in the normal cardiograph operating 
    mode to verify or to optimize cardiograph performance. These tests can be used as the first 
    step in identifying a technical problem with the cardiograph. 
    To open the Maintenance Test screen:
    1
    Press the Ta b key ( ) or turn the Trim Knob to highlight the Config button on the 
    Command Toolbar.
    NOTEAccess to Configuration may be password controlled.
    2
    Press the space bar or the Trim Knob to select the button. The Configuration screen 
    appears.
    3Select System(S) by holding down the Alt key on the keyboard, then pressing the S key. 
    The System screen appears.
    The 
    Diagnostics section of the screen lists the five configuration/maintenance tests.
    CONFIG
    Table 4-34 Maintenance Tests
    Test Page
    PIM Test 4-2
    Barcode Test 4-2
    Magcard Test 4-3
    Printer Test 4-3
    Network Test 4-6 
    						
    							Maintenance TestsMaintenance Tests (Trim II/III/Rx only)
    4-2PageWriter Trim Cardiograph Service Manual
    Patient Interface Module (PIM) Test
    This test is used to confirm that the Patient Interface Module (PIM) is communicating with the 
    cardiograph. This test can be performed when the cardiograph displays PIM error messages, or 
    when the cardiograph is unable to acquire data from the PIM. Be sure that the PIM patient data 
    cable is securely attached to the PIM connector on the rear of the cardiograph before 
    performing the test. 
    If this test fails, it may indicate a problem with the PIM or with the PIM patient data cable.
    To perform the PIM Test:
    1
    Press the Ta b key ( ) or turn the Trim Knob to highlight the PIM Test button under 
    Diagnostics. 
    2Press the space bar or the Trim Knob to select the button. The message Accessing PIM... 
    appears. The PIM Test Result window appears and indicates that the PIM Test has passed, 
    or that the cardiograph is unable to communicate with the PIM. 
    3If the message PIM Test Passed appears, the PIM is communicating properly with the 
    cardiograph. The 
    OK button is highlighted. Push the Trim Knob to close the window. 
    4If the message Please Check the Patient Interface Module appears, check that the PIM 
    patient data cable is securely attached to the PIM connector on the rear of the cardiograph. 
    If the error message persists, contact Philips Medical Systems for assistance, see 
    “Contacting a Philips Response Center” on page 1-32.
    Barcode Reader Test
    NOTEYou must attach the barcode reader to the port on the rear of the cardiograph before powering on 
    the cardiograph.
    The Barcode Reader Test is used with the optional barcode reader. This test can be used to 
    confirm that the barcode reader is accurately scanning barcode data. Ensure that the barcode 
    reader is securely attached to the barcode reader port on the rear of the cardiograph before 
    performing the test.
    To perform the barcode reader test:
    1
    Press the Ta b key ( ) or turn the Trim Knob to highlight the Barcode Test button 
    under 
    Diagnostics. 
    2Press the space bar or the Trim Knob to select the button. The Barcode Reader test 
    window appears. 
    3Hold the barcode reader at a 45o angle and scan a test barcode.
    4The barcode data appears next to the Patient ID field. Review the barcode data to ensure 
    that it is accurate. The 
    OK button is highlighted. If the barcode data is accurate, press the 
    space bar or the Trim Knob to select the 
    OK button. Turn the Trim Knob to highlight the 
    Cancel button if the barcode data is incorrect. 
    5If the test fails, perform a barcode reader calibration and try the test again. See 
    Calibrating the Barcode Reader on page 3-11. 
    						
    							Maintenance TestsMaintenance Tests (Trim II/III/Rx only)
    PageWriter Trim Cardiograph Service Manual4-3
    Magnetic Card Reader Test
    NOTEYou must attach the magnetic card reader to the port on the rear of the cardiograph before powering 
    on the unit.
    The Magnetic Card Reader Test is used with the optional magnetic card reader. This test can 
    be used to confirm that the magnetic card reader is correctly reading data from a magnetic 
    card. Ensure that the magnetic card reader is securely attached to the magnetic card reader 
    connector on the rear of the cardiograph before performing the test.
    To perform the magnetic reader test:
    1
    Press the Ta b key ( ) or turn the Trim Knob to highlight the Magcard Test button. 
    2Press the space bar or the Trim Knob to select the button. The Magnetic Card Reader 
    window appears. 
    3Insert the magnetic card into the slot. 
    4Leave the magnetic card in the slot for five seconds and then pull it out.
    5The magnetic card data appears next to the Patient ID field. 
    6Review the data to ensure that it is correct. The OK button is highlighted. If the magnetic 
    card data is accurate, push the Trim Knob to select the 
    OK button. Turn the Trim Knob to 
    highlight the 
    Cancel button if the magnetic card data is incorrect and push to select it. 
    7If the test fails, try it again.
    Printer Test
    The Printer Test is used to verify that the cardiograph printer is able to correctly print the test 
    page. Use this test to verify proper printer performance or when reports appear to have print 
    quality errors.
    To perform the print test:
    1
    Press the Ta b key ( ) or turn the Trim Knob to highlight the Printer Test button under 
    Diagnostics. 
    2Press the space bar or the Trim Knob to select the button. The message Printing Test 
    Page... 
    appears and the test page prints out. 
    3Review the printer test page at points A, B, C, and D as shown on Figure 4-4, “Printer Test 
    Page” on page 4-5.
    Table 4-35 Printer Test Page Description 
    Test 
    PointDescription
    A The stepped bars are sharp edged and printed cleanly without 
    distortion or missing segments 
    B The spacing between the vertical lines is 25 mm with a discrepancy 
    of no more or less than 2% 
    						
    							Maintenance TestsMaintenance Tests (Trim II/III/Rx only)
    4-4PageWriter Trim Cardiograph Service Manual
    Test 
    PointDescription
    C The diagonal lines should be straight and printed cleanly without 
    distortion or breaks in the lines
    D The character set is printed cleanly without distortion or missing 
    characters, and all characters are clearly legible
    Table 4-35 Printer Test Page Description (continued) 
    						
    							Maintenance TestsMaintenance Tests (Trim II/III/Rx only)
    PageWriter Trim Cardiograph Service Manual4-5
    Figure 4-4 Printer Test Page
    4
    If the printer test page matches the printer test points as described in Table 4-35, “Printer 
    Test Page Description,” on page 4-3, the cardiograph has passed the printer test. If the 
    printer test page does not match the printer test points, the cardiograph has failed the print 
    test. 
    AA
    B
    C
    D 
    						
    							Maintenance TestsTrim II/III/Rx Diagnostic and Performance Verification Tests
    4-6PageWriter Trim Cardiograph Service Manual
    Network Test
    The Network Test displays the configured network information for the cardiograph. All 
    cardiograph network configuration is completed on the Network screen. From any 
    Configuration screen, press and hold down the Alt key on the keyboard, and then press the N 
    key to select the 
    Network tab and to open the Network screen. Before performing the 
    Network test, ensure that the LAN card is fully inserted into the PC Card slot on the rear of the 
    cardiograph, and that the LAN cable is fully inserted into the cable connector on the card. 
      CAUTIONOnly use the shielded LAN cable provided with the PageWriter Trim cardiograph (Philips part number 
    989803138021). Do not use LAN cables with the cardiograph that have not been approved by Philips 
    Medical Systems.
    To perform the network test:
    1
    Press the Ta b key ( ) or turn the Trim Knob to highlight the Network button.
    2Press the space bar or the Trim Knob to select the button. The Network Connection 
    window appears. The configured network information for the cardiograph appears on the 
    window. 
    3Review the network data to ensure that it is correct. The Retry button is highlighted. If the 
    network data does not display, press the Trim Knob again.
    4If the network data displays correctly, turn the Trim Knob to highlight the Close button. 
    Push the Trim Knob to select the button. 
    Trim II/III/Rx Diagnostic and Performance Verification 
    Tests
    More comprehensive diagnostic tests and operating statistics are provided in the Biomed 
    Service Utility. These tests and utilities are used to identify and to troubleshoot a technical 
    problem with the cardiograph. 
    CAUTIONEnsure that the AC power cord is plugged into the rear of the cardiograph and into a grounded 
    electrical outlet before using the Biomed Service Utility. If the cardiograph is powered by battery 
    power only while using the Biomed Service Utility, the battery will fail and will not recover.
    About the Trim II/III/Rx Biomed Service Utility
    The Biomed Service Utility screen displays important information and statistics about the 
    cardiograph and is a valuable diagnostic tool when troubleshooting the cardiograph.
    Using the Biomed Service Utility, you can: 
    „Check important unit status
    „Check the current software revisions 
    						