Philips Trim 1 Service Manual
Have a look at the manual Philips Trim 1 Service Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 619 Philips manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

Maintenance TestsTrim I Diagnostic and Performance Verification Tests PageWriter Trim Cardiograph Service Manual4-17 When the self-tests are completed, a status report is printed automatically. Figure 4-11 on page 4-17 shows a sample report. Figure 4-11 Self-Test Report Trim I Diagnostic and Performance Verification Tests Diagnostic tests are included in the Biomed Service Program and are used to identify and to troubleshoot a technical problem with the PageWriter Trim I cardiograph. CAUTIONEnsure that the AC power cord is plugged into the rear of the cardiograph and into a grounded electrical outlet before using the Biomed Service Program. If the cardiograph is powered by battery power only while using the Biomed Service Program, the battery will fail and will not recover. Launching the Trim I Biomed Service Program To launch the Biomed Service Program: 1 Turn off the cardiograph. NOTEEntering the Service Utility requires a full restart of the cardiograph. Any unsaved patient data will be lost. 2 Press the On/Off button to power on the unit. After approximately 20 seconds, the PageWriter Trim I screen appears, followed by a screen showing five triangles.

Maintenance TestsTrim I Diagnostic and Performance Verification Tests 4-18PageWriter Trim Cardiograph Service Manual 3Quickly, while the screen showing five triangles is displayed, press the Patient ID and Auto buttons at the same time. NOTEThe screen showing five triangles disappears after five seconds. If you miss it, you must restart the cardiograph again. An Access Code window appears. If the Access Code window does not appear, restart the cardiograph. Follow steps 1 to 3 again. 4Enter the access code by pressing the following buttons in sequence: Patient ID, Auto, Rhythm, Copy, Page. NOTEPress the Stop button to delete your last input if it is wrong. The Biomed Service Program Main Menu appears as shown in Figure 4-12. Figure 4-12 Service Utility Main Menu (Trim I) Diagnostic Tests in the Trim I Biomed Service Program The following tests may be completed in one session, or may be individually selected and executed. The Self-Test option will conduct all of the tests listed in the following table, with an option to print out the test results. MAIN MENU Settings Upgrade Test» Table 4-38 Biomed Service Program Diagnostic Tests Test Page Battery 4-19 Keyboard 4-19 LCD Screen 4-19 Onboard Flash 4-20 Options 4-20 PIM 4-20 Printer 4-21 RAM 4-21 Self-Test 4-23 Version 4-22 Voltage 4-23

Maintenance TestsTrim I Diagnostic and Performance Verification Tests PageWriter Trim Cardiograph Service Manual4-19 Battery Test This test analyzes if the battery voltage is below or above the voltage limits. To perform the keyboard test: 1 From the Biomed Service Program, turn the Trim Knob to highlight Test and push to select it. 2Turn the Trim Knob to highlight Test Battery and push to select it. 3Turn the Trim Knob to highlight Start and push to select it. The test results display on the screen. 4Press Patient ID to print out a report of the test results, or Stop to return to the main menu. Keyboard Test In a successful keyboard test, each key is properly recognized when pressed. To perform the keyboard test: 1 From the Biomed Service Program, turn the Trim Knob to highlight Test and push to select it. 2Turn the Trim Knob to highlight Test Keyboard and push to select it. 3Turn the Trim Knob to highlight Start and push to select it. The Keyboard Test window appears. 4Press each key on the keyboard. Push the Trim Knob, then turn it clockwise twice, and counterclockwise once to complete the test. A numeral appears to represent each key. 5After the numbers 0-8 all are displayed, the test is successful. Press Patient ID to pass the test or Stop to fail the test. The test is not successful if any keys are not recognized when you press them. LCD Screen Test The LCD Screen Test is used to verify that the display is working correctly. To perform the Screen Test: 1 From the Biomed Service Program, turn the Trim Knob to highlight Test and push to select it. 2Turn the Trim Knob to highlight Test LCD and push to select it. 3Turn the Trim Knob to highlight Start and push to select it. 4Characters scroll past on the screen. Verify that they appear correctly on the screen. Press Patient ID to pass the test and Stop to fail.

Maintenance TestsTrim I Diagnostic and Performance Verification Tests 4-20PageWriter Trim Cardiograph Service Manual 5If you pressed Patient ID to pass the test, now press Patient ID to print a report or Stop to return to the previous screen. If the screen test fails, try it again. Onboard Flash Test Verifies that the Onboard Flash is working properly. To test the Onboard Flash: 1 From the Biomed Service Program, turn the Trim Knob to highlight Test and push to select it. 2Turn the Trim Knob to highlight Test Flash and push to select it. 3Turn the Trim Knob to highlight Start and push to select it. 4The test results display on the screen. Press Patient ID to print a report or Stop to return to the previous screen. If the onboard flash card fails the test, try the test again. Options Test This test verifies that the localization options including the Kernel and Application language, the paper size (A or A4) and the lead standard (AAMI or IEC) are consistent and are properly configured. To perform the Options test: 1 From the Biomed Service Program, turn the Trim Knob to highlight Test and push to select it. 2Turn the Trim Knob to highlight Test Options and push to select it. 3Turn the Trim Knob to highlight Start and push to select it. 4The test results display on the screen. Press Patient ID to print a report or Stop to return to the previous screen. PIM Test Use this test to confirm that the Patient Interface Module (PIM) is communicating with the cardiograph. This test can be performed when the cardiograph displays PIM error messages, or when the cardiograph is unable to acquire data from the PIM. Be sure that the PIM patient data cable is securely attached to the PIM connector on the rear of the cardiograph before performing the test. If this test fails, it may indicate a problem with the PIM or with the PIM patient data cable. To perform the PIM Test: 1 From the Biomed Service Program, turn the Trim Knob to highlight Test and push to select it. 2Turn the Trim Knob to highlight Test PIM and push to select it.

Maintenance TestsTrim I Diagnostic and Performance Verification Tests PageWriter Trim Cardiograph Service Manual4-21 3Turn the Trim Knob to highlight Start and push to select it. 4The test results display on the screen. Press Patient ID to print a report or Stop to return to the previous screen. If the PIM test fails, try it again. If the failure persists, contact the nearest Philips Response Center for further assistance, see “Contacting a Philips Response Center” on page 1-32. Printer Test To perform the Printer Test: 1 From the Biomed Service Program, turn the Trim Knob to highlight Test and push to select it. 2Turn the Trim Knob to highlight Test Printer and push to select it. 3Turn the Trim Knob to highlight Start and push to select it. The Printer Test window appears with the message Testing Printer... 4 The cardiograph prints out the printer test page. Inspect the printer test page using the test points explained in “Printer Test” on page 4-3. If the printer test page matches the printer test points as described in Figure 4-35, “Printer Test Page Description,” the cardiograph has passed the printer test. If the printer test page does not match the printer test points, the cardiograph has failed the print test. 5Press Patient ID to print a report or Stop to return to the previous screen. If the printer fails the test, try the test again. If the failure persists, contact the nearest Philips Response Center for further assistance, see “Contacting a Philips Response Center” on page 1-32. RAM Test The RAM Test verifies that the internal storage is operational and is not corrupt. To perform the RAM Test: 1 From the Biomed Service Program, turn the Trim Knob to highlight Test and push to select it. 2Turn the Trim Knob to highlight Test RAM and push to select it. 3Turn the Trim Knob to highlight Start and push to select it. 4The test results display on the screen. Press Patient ID to print a report or Stop to return to the previous screen. If the RAM fails the test, try the test again. If the failure persists, contact the nearest Philips Response Center for further assistance, see “Contacting a Philips Response Center” on page 1-32.

Maintenance TestsTrim I Diagnostic and Performance Verification Tests 4-22PageWriter Trim Cardiograph Service Manual Self-Test The Self-Test option will run all of the tests listed in Table 4-37. For information on any specific test, see the page number referenced in the table. To run the self-test: 1 Select Test from the Main Menu. 2Select Self-Test and then select Start to start the self-test program. 3At the conclusion of the Self-Test, press the Patient ID button to print out a report of all test results. Figure 4-13 Trim I Service Diagnostics Status Report Version Test The Version Test verifies that the version levels of the Application, Kernel, Loader, PIM Kernel, and Lead Set (AAMI or IEC) are consistent. To perform the Version Test: 1 From the Biomed Service Program, turn the Trim Knob to highlight Test and push to select it. 2Turn the Trim Knob to highlight Test Version and push to select it. 3Turn the Trim Knob to highlight Start and push to select it.

Maintenance TestsTrim I Diagnostic and Performance Verification Tests PageWriter Trim Cardiograph Service Manual4-23 4The test results display on the screen. Press Patient ID to print a report or Stop to return to the previous screen. NOTEThe Version Test will fail if the PIM is not connected to the cardiograph. Voltage Test The Voltage Test verifies that the cardiograph is operating within acceptable voltage limits. To perform the Voltage Test: 1 From the Biomed Service Program, turn the Trim Knob to highlight Test and push to select it. 2Turn the Trim Knob to highlight Test Voltage and push to select it. 3Turn the Trim Knob to highlight Start and push to select it. 4The test results display on the screen. Press Patient ID to print a report or Stop to return to the previous screen.

Maintenance TestsTrim I Diagnostic and Performance Verification Tests 4-24PageWriter Trim Cardiograph Service Manual

5 5-1 Chapter 5 Troubleshooting This chapter provides information for localizing cardiograph problems to the subassembly level. This information is designed for use with the cardiograph’s diagnostic self-tests to help you efficiently repair the cardiograph with a minimum of equipment. Contacting a Philips Response Center The Philips Response Center can assist with product troubleshooting and provide technical expertise to help with any issue with the PageWriter Trim cardiograph or any of its accessories. For a full listing of contact phone numbers, see “Contacting a Philips Response Center” on page 1-32. Power On and Power Off Sequence The following sequence of events occur during a normal warm or cold cardiograph boot.

TroubleshootingPower On and Power Off Sequence 5-2PageWriter Trim Cardiograph Service Manual Figure 5-14 Trim I Power On Sequence