TroubleshootingTroubleshooting Cardiograph Issues PageWriter Trim Cardiograph Service Manual5-13 One or more leads show leads off (red dashed line) periodicallyPIM communications have been lost due to USB patient data cable connection failure, or defective leads are connected to the PIM. 1Check that PIM patient data cable is firmly seated to the PIM connector ( ) on the rear of the cardiograph. 2If the PIM is firmly attached, perform the PIM test as described in “Maintenance Tests (Trim II/III/Rx only)” on page 4-1, or for the Trim I as described in “Trim I Diagnostic and Performance Verification Tests” on page 4-17. If the PIM Test fails, replace the patient data cable. If the PIM Test continues to fail with a new patient data cable attached, replace the entire PIM PCA Set. Contact Philips Medical Systems for further assistance (see page 1-32). 3If the PIM test passes, ensure that all lead wires are fully seated in the PIM lead connectors. 4Ensure that all lead wires are intact and that there are no visible cracks in the lead wires. 5Connect all of the lead wires to a metallic conductive surface or to a shortening bar. – Check the Leads off indicator on the Status Bar. Ensure that no leads are being reported as leads off. – If any of the precordial leads report as leads off, with all limb leads connected, replace the precordial lead(s). – If all leads report as leads off, replace the RL/ R lead with a lead that known to function normally. If the new lead connected to the RL/ R connector functions normally, replace the defective RL lead. Once the RL/R lead is operable, proceed to replace any other leads that display as leads off. – If the RL/R lead still displays leads off, the cardiograph will require internal repair, contact Philips Medical Systems for further assistance (see page 1-32). Table 5-3 Signal Acquisition Issues (continued) SymptomPossible Cause & Investigation StepSolution
TroubleshootingTroubleshooting Cardiograph Issues 5-14PageWriter Trim Cardiograph Service Manual Real Time Screen Issues One or more leads show leads off (red dashed line) periodically6If replacing the lead wires does not resolve signal quality issues, check the PIM Kernel Revision field in the Service Utility (see page 4-7). – If the PIM Kernel version is G.001, upgrade the cardiograph software to version A.01.03. If the signal quality issues persist with software version A.01.03 installed, replace the PIM PCA sets. Contact Philips Medical Systems for further assistance (see page 1-32). – If the PIM Kernel version is D.008, and the cardiograph passes the PIM Test (page 4-1) or (page 4-17), replace the PIM PCA sets. Contact Philips Medical Systems for further assistance (see page 1-32). – After PIM PCA set replacement, upgrade the cardiograph to software version A.0.1.03 that includes PIM Kernel version G.003. Table 5-3 Signal Acquisition Issues (continued) SymptomPossible Cause & Investigation StepSolution Table 5-4 Real Time Screen Issues SymptomPossible Cause & Investigation StepConfirmation Solution Waveform display is no longer updating, system is unresponsiveMain controller board failure Action: Wait for 2 minutes or longer. The cardiograph will restart automatically, or you can restart it by pushing the restart button (page 5- 33). Restart the cardiograph to see if the cardiograph can operate normally. The cardiograph must be serviced to inspect and possibly upgrade kernel and application or replace main controller board. Contact Philips Medical Systems for further assistance (see page 1-32). System seems to be running slowly after boot-up Duplicate static IP configured for network Action: Confirm that the static IP assigned is not already in use on LAN. IP was already in use. It has been observed that boot up and operation can initially be slower when the PageWriter Trim is configured for an IP address that is already in use on the network.
TroubleshootingTroubleshooting Cardiograph Issues PageWriter Trim Cardiograph Service Manual5-15 Archive Screen Issues Table 5-5 Archive Screen Issues SymptomPossible Cause & Investigation StepConfirmation Solution After attempting an Archive screen operation, an error message appears: Command timed-out The TraceMaster ECG Management system is busy. Action: Retry TraceMaster operation. Operation succeeds on subsequent try. Transfer operations use network and server resource, which may occasionally be unable to service the requests. After attempting an Archive screen operation, an error message appears: Requested Function Failed Network or modem phone line was interrupted during operation. Action: Retry operation. Inspect cabling to network or analog phone line. Operation succeeds on subsequent try or damaged cabling or intermittent connection found.Transfer or query operations require reliable access to the network transport. Selected removable media (PC card, USB memory stick) was not fully inserted into the slot on the cardiograph. Action: Verify that selected removable media (PC card, USB memory stick) is inserted correctly. Removable media (PC card, USB memory stick) is not fully inserted or is not operational. Attempting to save ECGs to removable media (PC card, USB memory stick) that is not accessible will result in an error message. Retry the operation with fully functioning removable media that is fully inserted into the applicable slot on the cardiograph. An attempt to transfer an unsupported ECG report type was made to TraceMaster Action: Inspect ECG that failed the requested operation for selected format (report type). Format is Pan-12 or 12x1. Certain ECG report types are not supported by and will be rejected by the TraceMaster ECG Management System. A wireless LAN connection is being used to transfer ECGs to TraceMaster and LEAP credentials are not enabled.LEAP credentials are enabled and the ECG transfer subsequently completes.See “Enabling LEAP Credentials” on page C-10.
TroubleshootingTroubleshooting Cardiograph Issues 5-16PageWriter Trim Cardiograph Service Manual When an error occurs during transfer of multiple ECGs to a TraceMaster system, the transfer process does not continue and transferred files are not deleted. Application limitation Action: None. Current system limitation. The current PageWriter Trim operations fail in a known safe manner when multiple ECG transfer operations are interrupted. Retry the transfer for the remaining ECGs. After pressing the Archive button, it takes longer than 40 seconds before the Archive list is displayed when the main archive has more than 130 ECGs stored. System slow-down from full ECG database Action: Save or delete ECGs from the Main Archive. Entry into the Archive Screen is faster.There is a known system performance impact when the near maximum number of ECGs have been stored in the main archive area. The message Requested Function Failure appears after fax ECG transmission has completed. Fax application time- out.Action: Close the Fax dialog box after transmission completes by touching OK. Message no longer appears after FAX ECG transmission. The FAX application requires the user to close the dialog box after successful transmission. Otherwise, the operation times out and indicates a failure. Transfer of ECGs to an external PC card or USB memory stick fails and there are still less than 200 ECGs on the removable media device. P C c a rd o r U S B m e m o r y stick file system limitation Action: Delete extra non- ECG files from the PC card or USB memory stick.ECGs can now be transferred. Delete unwanted files from the PC card or USB memory stick. Table 5-5 Archive Screen Issues (continued) SymptomPossible Cause & Investigation StepConfirmation Solution
TroubleshootingTroubleshooting Cardiograph Issues PageWriter Trim Cardiograph Service Manual5-17 Configuration Screen Issues Some ECGs retrieved from TraceMaster and printed on the cardiograph look different from TraceMaster prints or XLi printouts. Older or non-PageWriter Trim source ECG files were retrieved. Action: Inspect the printed report date and information ECG was not created by a PageWriter Trim cardiograph.Retrieving and printing older ECG files that were not sourced by the PageWriter Trim cardiograph can have the following differences when printed: „Algorithm version appears as HPxxx on TraceMaster and PH on PageWriter Trim. „Pacer tick marks are not present on PageWriter Trim printout. „Frank lead system generates 3 rhythm traces with flat line and no lead label on PageWriter Trim printout. „Custom lead names do not appear on PageWriter Trim printout. „PageWriter Trim prints full interpretations regardless of original Xli ECG settings. Table 5-5 Archive Screen Issues (continued) SymptomPossible Cause & Investigation StepConfirmation Solution Table 5-6 Configuration Screen Issues SymptomPossible Cause & Investigation StepConfirmation Solution Can not enter the configuration windowForgot password. Contact Philips Medical Systems for further assistance (page 1-32).
TroubleshootingTroubleshooting Cardiograph Issues 5-18PageWriter Trim Cardiograph Service Manual Printer Issues Table 5-7 PageWriter Trim I Printer Issues Software Version A.00.03 and lower Error Code Possible Cause & Investigation Step Solution 0201 No battery or dead battery.1Replace battery. 2Replace motherboard, contact Philips Medical Systems for further assistance (page 1-32). 3Replace power supply cord. 0202UPR running: a new print job has been sent to the printer and the printer cannot accept a new print job.Follow the procedure to restart the cardiograph (see page 5-33). 0203„Paper drawer is not fully closed „Out of paper, or paper loaded incorrectly „Application error 1Open and firmly close the printer drawer. 2Open the printer drawer and check to make sure that the paper is properly inserted and aligned. 3Follow the procedure to restart the cardiograph (see page 5-33). 4It has been observed that on rare occasions a Printer Door Open message cannot be dismissed because of overlapping simultaneous error status reporting. 0204 „Paper jam „Paper is not loaded correctly „Defective TOF sensor assembly „Defective paper tray assembly 1Open and firmly close the printer drawer. 2Open the printer drawer and check to make sure that the paper is properly inserted and aligned. 3Replace TOF sensor assembly, contact Philips Medical Systems for further assistance (page 1-32). 4Replace paper tray assembly. 0205 No paper is in drawer Insert paper into drawer.
TroubleshootingTroubleshooting Cardiograph Issues PageWriter Trim Cardiograph Service Manual5-19 0206 Temperature shutdown: the temperature of the print head is too high1Shut down cardiograph and wait 20 minutes. 2If error persists, send cardiograph for repair, contact Philips Medical Systems for further assistance (page 1-32). 0207Unknown error 1Shut down cardiograph 2Remove battery 3Reinsert battery and press On/ Standby button 4If error persists, contact Philips Medical Systems for further assistance (see page 1-32). Table 5-8 PageWriter Trim I Printer Issues Software Version A.01.00 and higher Error Code Possible Cause & Investigation Step Solution 0101 Start Up CRC Failure Reinstall the cardiograph software (see page A-1). 0201USB printer - No USB connection Restart the cardiograph. 0202USB printer - running (busy) Restart the cardiograph. 0203USB printer - door open1Open and firmly close the printer drawer. 2Open the printer drawer and check to make sure that the paper is properly inserted and aligned. 3Follow the procedure to restart the cardiograph (see page 5-33). 4It has been observed that on rare occasions a Printer Door Open message cannot be dismissed because of overlapping simultaneous error status reporting. Table 5-7 PageWriter Trim I Printer Issues Software Version A.00.03 and lower (continued) Error Code Possible Cause & Investigation Step Solution
TroubleshootingTroubleshooting Cardiograph Issues 5-20PageWriter Trim Cardiograph Service Manual 0204 USB printer - paper jam or out of paper1Open and firmly close the printer drawer. 2Insert paper into drawer, if necessary. 3Open the printer drawer and check to make sure that the paper is properly inserted and aligned. 4Replace TOF sensor assembly, contact Philips Medical Systems for further assistance (page 1-32). 5Replace paper tray assembly. 0207No PIM or defective PIM is connected to cardiograph when powered up with software version A.01.03 installed, or some other unknown error (software version A.01.02 or lower)If software version A.01.03 is installed, follow the directions on page 5-5. If software version A.01.02 or lower is installed, perform the following steps: 1Shut down cardiograph. 2Remove battery. 3Reinsert battery and press On/ Standby button. 4If error persists, contact Philips Medical Systems for further assistance (see page 1-32). All other errors Unknown error Contact Philips Medical Systems for further assistance (see page 1-32). Table 5-9 PageWriter Trim II/III/Rx Printer Issues SymptomPossible Cause & Investigation StepConfirmation Solution After the paper drawer was opened during printing, the print image was corrupted or incorrect.Printer controller error Action: NonePrinting continued on second page when only one page was expected.Occasional printer controller status errors can occur when opening the paper drawer during printing. Table 5-8 PageWriter Trim I Printer Issues Software Version A.01.00 and higher (continued) Error Code Possible Cause & Investigation Step Solution
TroubleshootingTroubleshooting Cardiograph Issues PageWriter Trim Cardiograph Service Manual5-21 After the paper drawer was opened during printing, the print image was corrupted or incorrect and the paper does not stop feeding.Printer controller error Action: Press the rear restart button (see page 5-33). The cardiograph restarts and boots correctly. Paper stops feeding.Occasional printer controller status errors can occur when opening the paper drawer during printing. The message Printer Door is Open appears, and the cardiograph is not responding. Application error Action: Press the rear restart button (see page 5-33). The cardiograph restarts and boots correctly. It has been observed that on rare occasions a Printer Door Open message cannot be dismissed because of overlapping simultaneous error status reporting. The message Check Printer appears with no apparent printer problem present. Printer controller error Action: Open and close printer paper drawer. Message does not reoccur.Infrequently, the printer controller may indicate a printer mechanism status error when no error exists. This has been noted specifically when A4 paper is in use. Printed report looks too small or compressed. Printer controller error Action: Press the rear restart button (see page 5-33).Cardiograph restarts and boots correctly. Reports print correctly.On very rare occasions, reports with A4 paper show a compressed image. Restart the cardiograph to resolve the issue. After opening and closing the paper drawer rapidly many times, the cardiograph was unresponsive. Printer controller error Action: Press the rear restart button (see page 5-33). Cardiograph restarts and boots correctly. Reports print correctly.A rapid and sustained repeating open-and-close action on the paper drawer can cause an application error related to printer status handling. Table 5-9 PageWriter Trim II/III/Rx Printer Issues (continued) SymptomPossible Cause & Investigation StepConfirmation Solution
TroubleshootingTroubleshooting Cardiograph Issues 5-22PageWriter Trim Cardiograph Service Manual Printer Error: Out of Paper/Paper Jam/ Door Open„Paper drawer is not fully closed „Out of paper, or paper loaded incorrectly „Defective TOF sensor assembly „Defective paper tray assembly Error does not occur again.„Open and firmly close the paper drawer. „Open paper drawer and check to make sure paper is aligned correctly. „Replace TOF sensor assembly, contact Philips Medical Systems for further assistance (page 1- 32). „Replace paper tray assembly. Printer not detected message „Application error „Unseated or defective main harness assembly „Defective printer control board Message no longer appears.„Press the rear restart button (see page 5-33). „Check seating of main harness assembly. Replace if necessary, contact Philips Medical Systems for further assistance (page 1-32). „Replace the printer control board, contact Philips Medical Systems for further assistance (page 1-32). Paper does not advance „Application error „Defective TOF sensor assembly „Defective printer gearbox assembly Paper advances correctly.„Press the rear restart button. „Replace the TOF sensor assembly, contact Philips Medical Systems for further assistance (page 1- 32). „Replace the printer gearbox assembly, contact Philips Medical Systems for further assistance (page 1-32). Table 5-9 PageWriter Trim II/III/Rx Printer Issues (continued) SymptomPossible Cause & Investigation StepConfirmation Solution