    							TroubleshootingTroubleshooting Cardiograph Issues
    PageWriter Trim Cardiograph Service Manual5-13
    One or more leads 
    show leads off (red 
    dashed line) 
    periodicallyPIM communications 
    have been lost due to 
    USB patient data cable 
    connection failure, or 
    defective leads are 
    connected to the PIM. 1Check that PIM patient data cable is firmly seated 
    to the PIM connector ( ) on the rear of the 
    cardiograph. 
    2If the PIM is firmly attached, perform the PIM test 
    as described in “Maintenance Tests (Trim II/III/Rx 
    only)” on page 4-1, or for the Trim I as described 
    in “Trim I Diagnostic and Performance 
    Verification Tests” on page 4-17. If the PIM Test 
    fails, replace the patient data cable. If the PIM Test 
    continues to fail with a new patient data cable 
    attached, replace the entire PIM PCA Set. Contact 
    Philips Medical Systems for further assistance (see 
    page 1-32).
    3If the PIM test passes, ensure that all lead wires are 
    fully seated in the PIM lead connectors. 
    4Ensure that all lead wires are intact and that there 
    are no visible cracks in the lead wires. 
    5Connect all of the lead wires to a metallic 
    conductive surface or to a shortening bar. 
    – Check the Leads off indicator on the Status 
    Bar. Ensure that no leads are being reported as 
    leads off. 
    – If any of the precordial leads report as leads 
    off, with all limb leads connected, replace the 
    precordial lead(s). 
    – If all leads report as leads off, replace the RL/
    R lead with a lead that known to function 
    normally. If the new lead connected to the RL/
    R connector functions normally, replace the 
    defective RL lead. Once the RL/R lead is 
    operable, proceed to replace any other leads 
    that display as leads off. 
    – If the RL/R lead still displays leads off, the 
    cardiograph will require internal repair, 
    contact Philips Medical Systems for further 
    assistance (see page 1-32).
    Table 5-3 Signal Acquisition Issues (continued)
    SymptomPossible Cause & 
    Investigation StepSolution 
    						
    							TroubleshootingTroubleshooting Cardiograph Issues
    5-14PageWriter Trim Cardiograph Service Manual
    Real Time Screen Issues
    One or more leads 
    show leads off (red 
    dashed line) 
    periodically6If replacing the lead wires does not resolve signal 
    quality issues, check the 
    PIM Kernel Revision 
    field in the Service Utility (see page 4-7).
    – If the PIM Kernel version is 
    G.001, upgrade 
    the cardiograph software to version A.01.03. If 
    the signal quality issues persist with software 
    version A.01.03 installed, replace the PIM 
    PCA sets. Contact Philips Medical Systems for 
    further assistance (see page 1-32).
    – If the PIM Kernel version is 
    D.008, and the 
    cardiograph passes the PIM Test (page 4-1) or 
    (page 4-17), replace the PIM PCA sets. 
    Contact Philips Medical Systems for further 
    assistance (see page 1-32).
    – After PIM PCA set replacement, upgrade the 
    cardiograph to software version A.0.1.03 that 
    includes PIM Kernel version G.003.
    Table 5-3 Signal Acquisition Issues (continued)
    SymptomPossible Cause & 
    Investigation StepSolution
    Table 5-4 Real Time Screen Issues 
    SymptomPossible Cause & 
    Investigation StepConfirmation Solution
    Waveform display 
    is no longer 
    updating, system is 
    unresponsiveMain controller board 
    failure 
    Action: Wait for 2 
    minutes or longer. The 
    cardiograph will restart 
    automatically, or you can 
    restart it by pushing the 
    restart button (page 5-
    33). Restart the cardiograph to 
    see if the cardiograph can 
    operate normally. The cardiograph must be 
    serviced to inspect and 
    possibly upgrade kernel and 
    application or replace main 
    controller board. Contact 
    Philips Medical Systems 
    for further assistance (see 
    page 1-32).
    System seems to 
    be running slowly 
    after boot-up Duplicate static IP 
    configured for network
    Action: Confirm that the 
    static IP assigned is not 
    already in use on LAN. IP was already in use.  It has been observed that 
    boot up and operation can 
    initially be slower when the 
    PageWriter Trim is 
    configured for an IP 
    address that is already in 
    use on the network.  
    						
    							TroubleshootingTroubleshooting Cardiograph Issues
    PageWriter Trim Cardiograph Service Manual5-15
    Archive Screen Issues
    Table 5-5 Archive Screen Issues 
    SymptomPossible Cause & 
    Investigation StepConfirmation Solution
    After attempting 
    an Archive screen 
    operation, an error 
    message appears: 
    Command 
    timed-out
     The TraceMaster ECG 
    Management system is 
    busy. 
    Action: Retry 
    TraceMaster operation. Operation succeeds on 
    subsequent try. Transfer operations use 
    network and server 
    resource, which may 
    occasionally be unable to 
    service the requests. 
    After attempting 
    an Archive screen 
    operation, an error 
    message appears: 
    Requested 
    Function Failed
    Network or modem 
    phone line was 
    interrupted during 
    operation. 
    Action: Retry operation. 
    Inspect cabling to 
    network or analog phone 
    line. Operation succeeds on 
    subsequent try or 
    damaged cabling or 
    intermittent connection 
    found.Transfer or query 
    operations require reliable 
    access to the network 
    transport. 
     Selected removable 
    media (PC card, USB 
    memory stick) was not 
    fully inserted into the 
    slot on the cardiograph. 
    Action: Verify that 
    selected removable 
    media (PC card, USB 
    memory stick) is 
    inserted correctly. Removable media (PC 
    card, USB memory stick) 
    is not fully inserted or is 
    not operational. Attempting to save ECGs to 
    removable media (PC card, 
    USB memory stick) that is 
    not accessible will result in 
    an error message. Retry the 
    operation with fully 
    functioning removable 
    media that is fully inserted 
    into the applicable slot on 
    the cardiograph. 
    An attempt to transfer an 
    unsupported ECG report 
    type was made to 
    TraceMaster 
    Action: Inspect ECG that 
    failed the requested 
    operation for selected 
    format (report type). Format is Pan-12 or 
    12x1. Certain ECG report types 
    are not supported by and 
    will be rejected by the 
    TraceMaster ECG 
    Management System. 
    A wireless LAN 
    connection is being used 
    to transfer ECGs to 
    TraceMaster and LEAP 
    credentials are not 
    enabled.LEAP credentials are 
    enabled and the ECG 
    transfer subsequently 
    completes.See “Enabling LEAP 
    Credentials” on page C-10.  
    						
    							TroubleshootingTroubleshooting Cardiograph Issues
    5-16PageWriter Trim Cardiograph Service Manual
    When an error 
    occurs during 
    transfer of multiple 
    ECGs to a 
    TraceMaster 
    system, the transfer 
    process does not 
    continue and 
    transferred files are 
    not deleted. Application limitation 
    Action: None. Current system 
    limitation. The current PageWriter 
    Trim operations fail in a 
    known safe manner when 
    multiple ECG transfer 
    operations are interrupted. 
    Retry the transfer for the 
    remaining ECGs. 
    After pressing the 
    Archive button, it 
    takes longer than 
    40 seconds before 
    the Archive list is 
    displayed when the 
    main archive has 
    more than 130 
    ECGs stored. System slow-down from 
    full ECG database Action: Save or delete 
    ECGs from the Main 
    Archive. Entry into the Archive 
    Screen is faster.There is a known system 
    performance impact when 
    the near maximum number 
    of ECGs have been stored 
    in the main archive area. 
    The message 
    Requested 
    Function Failure
     
    appears after fax 
    ECG transmission 
    has completed. Fax application time-
    out.Action: Close the Fax 
    dialog box after 
    transmission completes 
    by touching 
    OK. Message no longer 
    appears after FAX ECG 
    transmission. The FAX application 
    requires the user to close 
    the dialog box after 
    successful transmission. 
    Otherwise, the operation 
    times out and indicates a 
    failure. 
    Transfer of ECGs 
    to an external PC 
    card or USB 
    memory stick fails 
    and there are still 
    less than 200 
    ECGs on the 
    removable media 
    device. P C  c a rd  o r  U S B  m e m o r y  
    stick file system 
    limitation 
    Action: Delete extra non-
    ECG files from the PC 
    card or USB memory 
    stick.ECGs can now be 
    transferred. Delete unwanted files from 
    the PC card or USB 
    memory stick.
    Table 5-5 Archive Screen Issues (continued)
    SymptomPossible Cause & 
    Investigation StepConfirmation Solution 
    						
    							TroubleshootingTroubleshooting Cardiograph Issues
    PageWriter Trim Cardiograph Service Manual5-17
    Configuration Screen Issues
    Some ECGs 
    retrieved from 
    TraceMaster and 
    printed on the 
    cardiograph look 
    different from 
    TraceMaster prints 
    or XLi printouts. Older or non-PageWriter 
    Trim source ECG files 
    were retrieved.
    Action: Inspect the 
    printed report date and 
    information
    ECG was not created by a 
    PageWriter Trim 
    cardiograph.Retrieving and printing 
    older ECG files that were 
    not sourced by the 
    PageWriter Trim 
    cardiograph can have the 
    following differences when 
    printed:
    „Algorithm version 
    appears as 
    HPxxx on 
    TraceMaster and 
    PH on 
    PageWriter Trim.
    „Pacer tick marks are not 
    present on PageWriter 
    Trim printout.
    „Frank lead system 
    generates 3 rhythm 
    traces with flat line and 
    no lead label on 
    PageWriter Trim 
    printout.
    „Custom lead names do 
    not appear on 
    PageWriter Trim 
    printout.
    „PageWriter Trim prints 
    full interpretations 
    regardless of original 
    Xli ECG settings.
    Table 5-5 Archive Screen Issues (continued)
    SymptomPossible Cause & 
    Investigation StepConfirmation Solution
    Table 5-6 Configuration Screen Issues 
    SymptomPossible Cause & 
    Investigation StepConfirmation Solution
    Can not enter the 
    configuration 
    windowForgot password. Contact Philips Medical 
    Systems for further 
    assistance (page 1-32). 
    						
    							TroubleshootingTroubleshooting Cardiograph Issues
    5-18PageWriter Trim Cardiograph Service Manual
    Printer Issues
    Table 5-7 PageWriter Trim I Printer Issues Software Version A.00.03 and lower 
    Error Code Possible Cause & Investigation Step Solution
    0201
    No battery or dead battery.1Replace battery.
    2Replace motherboard, contact Philips 
    Medical Systems for further 
    assistance (page 1-32).
    3Replace power supply cord.
    0202UPR running: a new print job has 
    been sent to the printer and the printer 
    cannot accept a new print job.Follow the procedure to restart the 
    cardiograph (see page 5-33).
    0203„Paper drawer is not fully closed
    „Out of paper, or paper loaded 
    incorrectly
    „Application error
    1Open and firmly close the printer 
    drawer.
    2Open the printer drawer and check to 
    make sure that the paper is properly 
    inserted and aligned.
    3Follow the procedure to restart the 
    cardiograph (see page 5-33).
    4It has been observed that on rare 
    occasions a 
    Printer Door Open 
    message cannot be dismissed 
    because of overlapping simultaneous 
    error status reporting.
    0204 „Paper jam
    „Paper is not loaded correctly
    „Defective TOF sensor assembly
    „Defective paper tray assembly 
    1Open and firmly close the printer 
    drawer.
    2Open the printer drawer and check to 
    make sure that the paper is properly 
    inserted and aligned.
    3Replace TOF sensor assembly, 
    contact Philips Medical Systems for 
    further assistance (page 1-32).
    4Replace paper tray assembly.
    0205 No paper is in drawer Insert paper into drawer. 
    						
    							TroubleshootingTroubleshooting Cardiograph Issues
    PageWriter Trim Cardiograph Service Manual5-19
    0206 Temperature shutdown: the 
    temperature of the print head is too 
    high1Shut down cardiograph and wait 20 
    minutes.
    2If error persists, send cardiograph for 
    repair, contact Philips Medical 
    Systems for further assistance (page 
    1-32).
    0207Unknown error 1Shut down cardiograph
    2Remove battery
    3Reinsert battery and press On/
    Standby button
    4If error persists, contact Philips 
    Medical Systems for further 
    assistance (see page 1-32).
    Table 5-8 PageWriter Trim I Printer Issues Software Version A.01.00 and higher 
    Error Code Possible Cause & Investigation Step Solution
    0101
    Start Up CRC Failure Reinstall the cardiograph software (see 
    page A-1).
    0201USB printer - No USB connection Restart the cardiograph.
    0202USB printer - running (busy) Restart the cardiograph.
    0203USB printer - door open1Open and firmly close the printer 
    drawer.
    2Open the printer drawer and check to 
    make sure that the paper is properly 
    inserted and aligned.
    3Follow the procedure to restart the 
    cardiograph (see page 5-33).
    4It has been observed that on rare 
    occasions a 
    Printer Door Open 
    message cannot be dismissed 
    because of overlapping simultaneous 
    error status reporting.
    Table 5-7 PageWriter Trim I Printer Issues Software Version A.00.03 and lower (continued)
    Error Code Possible Cause & Investigation Step Solution 
    						
    							TroubleshootingTroubleshooting Cardiograph Issues
    5-20PageWriter Trim Cardiograph Service Manual
    0204 USB printer - paper jam or out of 
    paper1Open and firmly close the printer 
    drawer.
    2Insert paper into drawer, if necessary.
    3Open the printer drawer and check to 
    make sure that the paper is properly 
    inserted and aligned.
    4Replace TOF sensor assembly, 
    contact Philips Medical Systems for 
    further assistance (page 1-32).
    5Replace paper tray assembly.
    0207No PIM or defective PIM is 
    connected to cardiograph when 
    powered up with software version 
    A.01.03 installed, or some other 
    unknown error (software version 
    A.01.02 or lower)If software version A.01.03 is installed, 
    follow the directions on page 5-5. 
    If software version A.01.02 or lower is 
    installed, perform the following steps:
    1Shut down cardiograph.
    2Remove battery.
    3Reinsert battery and press On/
    Standby button.
    4If error persists, contact Philips 
    Medical Systems for further 
    assistance (see page 1-32).
    All other errors
    Unknown error Contact Philips Medical Systems for 
    further assistance (see page 1-32).
    Table 5-9 PageWriter Trim II/III/Rx Printer Issues 
    SymptomPossible Cause & 
    Investigation StepConfirmation Solution
    After the paper 
    drawer was opened 
    during printing, the 
    print image was 
    corrupted or 
    incorrect.Printer controller error
    Action: NonePrinting continued on 
    second page when only 
    one page was 
    expected.Occasional printer controller 
    status errors can occur when 
    opening the paper drawer 
    during printing.
    Table 5-8 PageWriter Trim I Printer Issues Software Version A.01.00 and higher (continued)
    Error Code Possible Cause & Investigation Step Solution 
    						
    							TroubleshootingTroubleshooting Cardiograph Issues
    PageWriter Trim Cardiograph Service Manual5-21
    After the paper 
    drawer was opened 
    during printing, the 
    print image was 
    corrupted or 
    incorrect and the 
    paper does not stop 
    feeding.Printer controller error
    Action: Press the rear 
    restart button (see page 
    5-33).
    The cardiograph 
    restarts and boots 
    correctly. Paper stops 
    feeding.Occasional printer controller 
    status errors can occur when 
    opening the paper drawer 
    during printing.
    The message 
    Printer 
    Door is Open 
    appears, and the 
    cardiograph is not 
    responding.
    Application error
    Action: Press the rear 
    restart button (see page 
    5-33).
    The cardiograph 
    restarts and boots 
    correctly. It has been observed that on 
    rare occasions a 
    Printer Door 
    Open 
    message cannot be 
    dismissed because of 
    overlapping simultaneous 
    error status reporting.
    The message 
    Check 
    Printer 
    appears with 
    no apparent printer 
    problem present. 
    Printer controller error
    Action: Open and close 
    printer paper drawer.
    Message does not 
    reoccur.Infrequently, the printer 
    controller may indicate a 
    printer mechanism status error 
    when no error exists. This has 
    been noted specifically when 
    A4 paper is in use.
    Printed report looks 
    too small or 
    compressed. Printer controller error
    Action: Press the rear 
    restart button (see page 
    5-33).Cardiograph restarts 
    and boots correctly. 
    Reports print correctly.On very rare occasions, 
    reports with A4 paper show a 
    compressed image. Restart the 
    cardiograph to resolve the 
    issue.
    After opening and 
    closing the paper 
    drawer rapidly many 
    times, the 
    cardiograph was 
    unresponsive. 
    Printer controller error
    Action: Press the rear 
    restart button (see page 
    5-33). 
    Cardiograph restarts 
    and boots correctly. 
    Reports print correctly.A rapid and sustained 
    repeating open-and-close 
    action on the paper drawer 
    can cause an application error 
    related to printer status 
    handling.
    Table 5-9 PageWriter Trim II/III/Rx Printer Issues (continued)
    SymptomPossible Cause & 
    Investigation StepConfirmation Solution 
    						
    							TroubleshootingTroubleshooting Cardiograph Issues
    5-22PageWriter Trim Cardiograph Service Manual
    Printer Error: Out of 
    Paper/Paper Jam/
    Door Open„Paper drawer is not 
    fully closed
    „Out of paper, or 
    paper loaded 
    incorrectly
    „Defective TOF 
    sensor assembly
    „Defective paper 
    tray assembly
    Error does not occur 
    again.„Open and firmly close the 
    paper drawer.
    „Open paper drawer and 
    check to make sure paper 
    is aligned correctly.
    „Replace TOF sensor 
    assembly, contact Philips 
    Medical Systems for 
    further assistance (page 1-
    32).
    „Replace paper tray 
    assembly.
    Printer not 
    detected
     message
    „Application error
    „Unseated or 
    defective main 
    harness assembly
    „Defective printer 
    control board
    Message no longer 
    appears.„Press the rear restart 
    button (see page 5-33).
    „Check seating of main 
    harness assembly. 
    Replace if necessary, 
    contact Philips Medical 
    Systems for further 
    assistance (page 1-32).
    „Replace the printer 
    control board, contact 
    Philips Medical Systems 
    for further assistance 
    (page 1-32).
    Paper does not 
    advance
    „Application error
    „Defective TOF 
    sensor assembly
    „Defective printer 
    gearbox assembly
    Paper advances 
    correctly.„Press the rear restart 
    button.
    „Replace the TOF sensor 
    assembly, contact Philips 
    Medical Systems for 
    further assistance (page 1-
    32).
    „Replace the printer 
    gearbox assembly, contact 
    Philips Medical Systems 
    for further assistance 
    (page 1-32).
    Table 5-9 PageWriter Trim II/III/Rx Printer Issues (continued)
    SymptomPossible Cause & 
    Investigation StepConfirmation Solution 
    						