Home > Philips > Cardiograph > Philips Trim 1 Service Manual

Philips Trim 1 Service Manual

    Download as PDF Print this page Share this page

    Have a look at the manual Philips Trim 1 Service Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 619 Philips manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    							TroubleshootingTroubleshooting Cardiograph Issues
    PageWriter Trim Cardiograph Service Manual5-23
    PC Card/USB Memory Stick Issues
    Paper does not tear 
    off cleanly, or paper 
    stops at the wrong 
    spot„Wrong paper size 
    setting. 
    „Wrong method is 
    being used to tear 
    off paper.Paper tears cleanly and 
    advances correctly.
    „Check that paper size is 
    correctly set in the 
    cardiograph 
    configuration.
    „Refer to the PageWriter 
    Trim Instructions for Use 
    for proper method for 
    tearing paper.
    Blank pages printed 
    after ECG report
    Wrong paper size 
    settingNo blank pages print 
    after changing paper 
    size.Check that paper size is 
    correctly set in the 
    cardiograph configuration.
    On Chinese 
    cardiograph, test 
    report is missing 
    Chinese characters
    Only the Chinese 
    application is updated 
    on an English language 
    option cardiograph.ECG report print 
    correctly after the 
    software application is 
    reinstalled.Reinstall the software 
    application and kernel using 
    the 
    Install All option (see 
    page B-1).
    All other errors
    Unknown error Contact Philips Medical 
    Systems for further assistance 
    (see page 1-32).
    Table 5-9 PageWriter Trim II/III/Rx Printer Issues (continued)
    SymptomPossible Cause & 
    Investigation StepConfirmation Solution
    Table 5-10 PC Card / USB Memory Stick Issues 
    SymptomPossible Cause & 
    Investigation StepConfirmation Solution
    Occasionally, 
    copying ECGs from 
    archive to a PC card/
    USB Memory Stick 
    fails.
     
    PC card or USB 
    memory stick 
    subsystem time-out 
    failure
    Action: Retry the 
    transfer operation.Operation succeeds 
    without error.Write operations occasionally 
    fail for PC card/USB memory 
    stick.
    If this condition reoccurs 
    persistently, replace the PC 
    card/USB Memory Stick.
    Modem is 
    unavailable after 
    removing and 
    reinserting the PC 
    card or reconnecting 
    USB Memory Stick 
    into the cardiograph. 
    Main controller error
    Action: Press the On/
    Standby button to put 
    the cardiograph into 
    Standby. Then, press 
    the restart button (see 
    page 5-33).
    Modem is now 
    available.Hot-swapping the modem 
    after system boot is not 
    supported. You must restart 
    the cardiograph when 
    reinserting the modem PC 
    card or reconnecting the USB 
    Memory Stick. 
    						
    							TroubleshootingWireless Troubleshooting
    5-24PageWriter Trim Cardiograph Service Manual
    Software Installation Issues 
    Wireless Troubleshooting
    All wireless LAN troubleshooting topics are included in this section. For further assistance 
    with configuring the cardiograph for wireless transmission to a TraceMasterVue ECG 
    Management System, see the Installing TraceMasterVue and Configuring Communication 
    Guide available for download from the Philips InCenter web site (
    incenter.medical.philips.com). 
    Table 5-11 Software Installation Issues 
    SymptomPossible Cause & 
    Investigation StepPossible Solutions
    An error message 
    appears during 
    software installation 
    that an incorrect 
    model, incorrect 
    token, or invalid 
    token key is being 
    installed.The PC card used for 
    software installation is 
    not fully inserted into 
    the PC card slot before 
    the cardiograph was 
    powered up.
    The software 
    installation files 
    contained on the PC 
    card are for the 
    incorrect product 
    model. 
    The token key is 
    invalid.
    The cardiograph serial 
    number is invalid.Ensure that the PC card contains the 
    necessary files and is not defective.
    Ensure that the PC card is fully inserted 
    into the slot on the cardiograph before the 
    On/Standby button is pressed. 
    Check the PC card slot to ensure that 
    there are no broken pins on the 
    mainboard connector.
    Click the 
    Refresh from Source button on 
    the Software Installation screen. If no 
    proposed revisions are found, the PC card 
    was not fully inserted into the slot before 
    On/Standby button was pressed.
    The Software Application files saved to 
    the PC card are not for the correct 
    product model. For example, you can 
    only install PageWriter Trim III software 
    on a PageWriter Trim III cardiograph. 
    Download from InCenter the correct 
    software files applicable for your 
    cardiograph model.
    On the 
    Please Input Serial Number and 
    Token 
    screen, ensure that the entered 
    information is accurate. If the displayed 
    serial number is not identical to the serial 
    number of the cardiograph (displayed on 
    the rear label next to the 
    SN text, contact 
    Philips Medical Systems for further 
    assistance (see page 1-32).  
    						
    							TroubleshootingWireless Troubleshooting
    PageWriter Trim Cardiograph Service Manual5-25
    Table 5-12 Aironet Client Utility (ACU) Error Messages 
    Error Message Possible Cause Recommended Solution
    „Cisco 
    Wireless LAN 
    Adapter Not 
    Found! 
    „Client 
    Adapter Not 
    Found
    „The Aironet Client Utility was 
    launched without the wireless 
    adapter being inserted into the 
    cardiograph
    „The wireless adapter was 
    ejected from the cardiograph 
    1Select OK to close the error 
    message.
    2Press the On/Standby button 
    on the cardiograph.
    3Insert the wireless adapter 
    into the PC card slot.
    4Press the On/Standby button.
    WEP Key X Must 
    Be 10 Hex Digits!
    „An invalid number of 
    characters or an incorrect 
    character was entered for the 
    WEP key indicated. 
    „40-bit keys must have 10 total 
    characters.
    1Select OK to close the error 
    message.
    2Reenter the characters for 
    the invalid key.
    WEP Key X Must 
    Be 26 Hex Digits!
    „An invalid number of 
    characters or an incorrect 
    character was entered for the 
    WEP key indicated. 
    „128-bit keys must have 26 
    characters.
    1Select OK to close the error 
    message.
    2Reenter the characters for 
    the invalid key.
    You must enter a 
    WEP Key!A WEP key was not entered on the 
    WEP keys window.
    1Select OK to close the error 
    message.
    2Enter a WEP key in the 
    WEP Keys window.
    350 Series Radio 
    is Not AssociatedThe wireless adapter is not 
    associated to an access point.
    „There are several reasons 
    why the wireless adapter 
    may not be associated to an 
    access point. 
    „Follow the steps in the 
    procedure “Checking the 
    Wireless Adapter 
    Association to an Access 
    Point” on page 5-31. 
    						
    							TroubleshootingWireless Troubleshooting
    5-26PageWriter Trim Cardiograph Service Manual
    Table 5-13 PageWriter Trim Cardiograph Wireless Error Messages
    Error Message Possible Cause  Recommended Solution
    Unidentified PC 
    card Adapter: 
    Enter the name of 
    the PC card 
    driver. Driver 
    name:________„The wireless adapter is 
    inserted incorrectly into the PC 
    card slot on the cardiograph
    „The wireless adapter was 
    inserted into the cardiograph 
    when the cardiograph was in 
    active use (not in Standby 
    mode)
    „An unsupported wireless 
    adapter was inserted into the 
    cardiograph
    1Press the Esc key to close 
    the dialog. 
    2Press the On/Standby button 
    on the cardiograph.
    3Insert the wireless adapter 
    card into the PC card slot.
    4Press the Reset button on the 
    rear panel of the 
    cardiograph. For more 
    information, see Chapter 7 
    of the PageWriter Trim 
    Cardiograph Instructions 
    for Use.
    Table 5-14 PageWriter Trim Cardiograph ECG Transmission Error Messages
    Error Message Possible Cause Recommended Solution
    No Reply 
    Received From 
    Remote Site 
    Server„The wireless connection to the 
    access point does not exist
    „The network connection to the 
    TraceMasterVue Remote Site 
    server does not exist or is not 
    configured properly
    „The cardiograph did not 
    receive a response from the 
    TraceMasterVue Remote Site 
    server within a specified 
    period of timeFollow the steps in the procedure 
    “Checking the Remote Site 
    Server Connection” on 
    page 5-27.
    Unexpected 
    Reply Received 
    from Remote SiteA connection to the 
    TraceMasterVue Remote Site 
    server exists, but the server is not 
    properly acknowledging receipt of 
    the packet data. Follow the steps in the procedure 
    “Resolving an Unexplained 
    Reply Received from the 
    Remote Site” on page 5-30. 
    						
    							TroubleshootingWireless Troubleshooting
    PageWriter Trim Cardiograph Service Manual5-27
    Checking the Remote Site Server Connection
    Use the following procedure to check that the wireless adapter is communicating with a 
    TraceMasterVue Remote Site.
    To check the remote site server connection:
    1
    Check the LEDs on the wireless adapter. See Table 5-15 for further steps.
    2Press the On/Standby button for three seconds, then release it to shut down the 
    cardiograph.
    3Connect a USB mouse to the USB port ( ) on the rear of the cardiograph.
    4Press the On/Standby button. 
    Table 5-15 LED Indicators 
    LED Indicators Recommended Solution
    Red LED: blinking 
    quickly
    Green LED: blinking 
    quickly„Move the cardiograph to an area with a strong wireless 
    signal. The green LED should then blink slowly or turn off 
    completely indicating that the wireless adapter is 
    associated with an access point.
    „If the green LED continues to blink quickly indicating that 
    the wireless adapter is still not associated to an access 
    point, follow the steps in the procedure “Checking the 
    Wireless Adapter Association to an Access Point” on 
    page 5-31.
    Red LED: off
    Green LED: off
    „Ensure that the wireless adapter is securely inserted into 
    the PC card slot on the rear of the cardiograph. 
    „Push the On/Standby button to put the cardiograph into 
    Standby.
    „Push the On/Standby button again.
    „Check the status LEDs on the wireless adapter. If the 
    LEDs are still not illuminated, contact the Philips 
    Response Center for further assistance. 
    Red LED: blinking 
    quickly
    Green LED: off or 
    blinking slowlyThe wireless adapter is associated to an access point. Go to 
    Step 2 in this procedure. 
    						
    							TroubleshootingWireless Troubleshooting
    5-28PageWriter Trim Cardiograph Service Manual
    5When the software identification screen appears, press the Ctrl () + Shift () 
    + 6 ( ) keys at the same time.
    Figure 5-18 PageWriter Trim Software Identification Screen
    6
    The PageWriter Trim Wireless LAN Installation screen appears. Click on the Configure 
    Wireless LAN Card
     button. 
    Figure 5-19 PageWriter Trim Wireless LAN Installation screen
    7
    The Aironet Client Utility (ACU) window appears. Click the Status tab. If green or 
    yellow bars appear on the 
    Signal Strength or Signal Quality fields, this indicates an active 
    connection. If the bars are red, try moving the cardiograph to an area with a stronger 
    wireless signal. For information on performing additional diagnostic tests from the Status 
    tab, see the Cisco Aironet Wireless LAN Client Adapters Installation and Configuration 
    Guide for Windows CE available for download from: 
    www.cisco.com.
    8
    Verify that the IP address information on the cardiograph network screen is correct (DHCP 
    or Static IP) and that the computer name is also correct.The DHCP IP address is displayed 
    on the ACU screen. If DHCP is being used, the DHCP lease may have expired. To 
    reacquire an IP address, press the On/Standby button on the cardiograph to enter Standby, 
    then press the button again to return the cardiograph to active use.  
    						
    							TroubleshootingWireless Troubleshooting
    PageWriter Trim Cardiograph Service Manual5-29
    9Ping the TraceMasterVue server from the cardiograph. Follow the procedure in Table 5-
    Table 5-16 Pinging the TraceMasterVue server from the cardiograph
    To ping the TraceMasterVue server from the PageWriter Trim 
    cardiograph:
    1
    If the cardiograph is in active use, press the On/Standby button for 
    three seconds, then release it to shut down the cardiograph. 
    2Connect a USB mouse to the USB port ( ) on the rear cardiograph 
    (if not previously connected). 
    3Press the On/Standby button to restart the cardiograph. When the 
    PageWriter Trim software identification screen appears, press Ctrl 
    () + Shift ( ) + the Trim Knob at the same time.
    Note: The software identification screen disappears after five seconds. If the 
    key sequence is not entered in time, restart the cardiograph.
    Figure 5-20 PageWriter Trim Software Identification Screen
    4
    An Access Code window appears. If the Access Code window does not 
    appear, try the left CTRL+SHIFT+Trim Knob sequence again.
    5Type in 1573. The Service Utility screen appears. 
    6Click on the pull-down menu under Diagnostic Tests. Press the up or 
    down arrow key (on keyboard) to select 
    Network Ping. 
    7Click the Start button. The Ping Test window appears. 
    8Type in the IP address of the TraceMasterVue server. Click the Ping 
    button.
    If the test succeeds, ensure that you can ping the cardiograph from the 
    server, as described next. Then retest ECG transmission. 
    If the test fails, ping the access point. Consult your network 
    administrator for information on pinging the access point.  
    						
    							TroubleshootingWireless Troubleshooting
    5-30PageWriter Trim Cardiograph Service Manual
    16.
    10If the access point ping test fails, most likely there is an IP configuration issue. Consult 
    your network administrator for further assistance. It may be helpful to connect the 
    cardiograph to a hardwired Ethernet connection for further IP configuration 
    troubleshooting.
    Resolving an Unexplained Reply Received from the Remote 
    Site
    Use the following procedure if the error message Unexpected Reply received from Remote 
    Site
     appears when transmitting ECGs to a TraceMasterVue ECG Management System.
    To resolve an unexplained reply received from the remote site:
    1
    On the Configuration screen, under the Remote Sites tab, check that the Remote Site 
    server URL is in the correct format. Ensure that the Remote Site Server URL includes a 
    forward slash at the end of the URL. An example of a valid URL is: 
    http://192.168.0.1/EMSCOMM/
    2
    If the URL is correct, consult the TraceMasterVue system administrator for more 
    information. The TraceMasterVue server may need to be reset.
    To ping the cardiograph from the server:
    Before proceeding, have available the IP address of the target cardiograph. 
    1On the TraceMasterVue server, select Start > Run, and type cmd in the 
    Run dialog box. 
    2In the Command shell, type the following:
    ping 
    For example, ping 161.99.23.55 Table 5-16 Pinging the TraceMasterVue server from the cardiograph 
    						
    							TroubleshootingWireless Troubleshooting
    PageWriter Trim Cardiograph Service Manual5-31
    Checking the Wireless Adapter Association to an Access 
    Point
    Use the following procedure when a wireless adapter is not associating with an access point. 
    1Check the LEDs on the wireless adapter. See Table 5-17 for further steps.
    2Press the On/Standby button for three seconds, then release it to shut down the 
    cardiograph.
    3Connect a USB mouse to the USB port ( ) on the rear of the cardiograph.
    4Press the On/Standby button. 
    5When the software identification screen appears, press the Ctrl () + Shift () 
    + 6 ( ) keys at the same time.
    Table 5-17 LED Indicators
    LED Indicators Recommended Solution
    Red LED: blinking quickly
    Green LED: blinking quickly1Move the cardiograph to an area with a 
    strong wireless signal. The green LED 
    should then blink slowly or turn off 
    completely indicating that the wireless 
    adapter is associated with an access point.
    2If the green LED continues to blink 
    quickly indicating that the wireless adapter 
    is still not associated to an access point, go 
    to step 2 in this procedure.
    3If LEAP authentication is used, ensure that 
    the LEAP credentials are enabled. For 
    more information, See Enabling LEAP 
    Credentials on page C-10..
    Red LED: off
    Green LED: off
    1Ensure that the wireless adapter is securely 
    inserted into the PC card slot on the rear of 
    the cardiograph. 
    2Press the On/Standby button to put the 
    cardiograph into Standby.
    3Press the On/Standby button again.
    4Check the status LEDs on the wireless 
    adapter. If the LEDs are still not 
    illuminated, contact the Philips Response 
    Center for further assistance. 
    Red LED: blinking quickly
    Green LED: off or blinking slowlyThe wireless adapter is associated to an access 
    point. Go to Step 2 in this procedure. 
    						
    							TroubleshootingWireless Troubleshooting
    5-32PageWriter Trim Cardiograph Service Manual
    Figure 5-21 PageWriter Trim Software Identification Screen
    6
    The PageWriter Trim Wireless LAN Installation screen appears. Click on the Configure 
    Wireless LAN Card
     button. 
    Figure 5-22 PageWriter Trim Wireless LAN Installation screen
    7
    The Aironet Client Utility (ACU) window appears. Ensure that the correct profile is 
    selected in the 
    Select Active Profile drop-down list. 
    8Ensure that the message 350 Series Radio is Associated appears on the ACU window. If it 
    does not, proceed to step 9.
    9Ensure that the access point is turned on and is operating.
    10If the wireless adapter still cannot associate with the access point, contact your network 
    administrator to ensure that all settings are correct for both the wireless adapter and for the 
    access point. 
    – If MAC or IP filtering is enabled at the access point, ensure that the wireless adapter 
    MAC address or IP address is added to the list. 
    						
    All Philips manuals Comments (0)