    							Installing PageWriter Trim I SoftwareVerifying the Software Installation
    PageWriter Trim Cardiograph Service ManualA-13
    Figure A-11 Software Version A.01.03 Self Test Report
    If an incorrect software revision level or incorrect localization option appears on the Self Test 
    Report, reinstall the software following the same procedure. If an incorrect revision level or 
    localization option appears again on the report, or if the cardiograph fails any portion of the 
    Self Test, contact the nearest Philips Response Center for further assistance. See Contacting a 
    Philips Response Center on page 1-32.
    Software Version A.01.03 PIM Kernel Revision 
    For software version A.01.03, the PIM Kernel Revision Level may appear on the Self Test 
    Report as either 
    D.008 (as shown above in Figure A-11) or as version G.003. Both PIM Kernel 
    versions are valid for software version A.01.03. ARelease Version and Localization Option Information
    A 
    						
    							Installing PageWriter Trim I SoftwareSpecial Note for Software Version A.01.01
    A-14PageWriter Trim Cardiograph Service Manual
    Special Note for Software Version A.01.01
    The following section provides additional information about software version A.01.01.
    Special Note about Software Version A.01.01 Self Test 
    Report
    Figure A-12 Software Version A.01.01 Self Test Report with Failed Version Test
    If the cardiograph fails the Self Test, check the Version Test field under Self-Test Results 
    (upper right side of report). If this is the only field that fails the Self Test, and the Self Test 
    reports the failure as 
    Software Version (shown as B in Figure A-12), confirm that software 
    revision information is exactly the same as the information shown in Table A-4, “PageWriter 
    Trim I Cardiograph A.01.01 Valid Software Revision Information,” on page A-14. If the 
    information on the Self Test and the information in Table A-4 are identical, the A.01.01 
    software installation is complete. If the information is not identical, reinstall the software and 
    run the test again. For further assistance, contact the Philips Response Center, see page 1-32. AFailed Version Test Field
    BFailed Self Test Information
    Table A-4 PageWriter Trim I Cardiograph A.01.01 Valid Software Revision 
    Information
    Software Element Loader Kernel Application
    PageWriter Trim I A.01.01 Software
    L.03.12 K.03.01 A.08.01.01
    PIM Software1.009 D.008 Not applicable
    A
    B 
    						
    							Installing PageWriter Trim I SoftwareSpecial Note for Software Version A.01.03
    PageWriter Trim Cardiograph Service ManualA-15
    Special Note for Software Version A.01.03 
    NOTEAlways ensure that the Patient Interface Module (PIM) is connected to the cardiograph before pressing 
    the On/Off button and returning the cardiograph to active use. 
    Figure A-13 Connecting the PIM to the cardiograph
    Please note that the following sequence of events may occur if the PIM is not connected to the 
    cardiograph, or a defective PIM is connected to the cardiograph when it is returned to active 
    use with installed software version A.01.03.
    To return the cardiograph to active use if the PIM is not connected or a defective PIM is 
    connected:
    1
    If the Patient Interface Module (PIM) is not connected to the cardiograph or a defective 
    PIM is connected, press the On/Off button. 
    2Five squares ( ) will appear at the upper left corner of the display.
    3The display will then go blank for approximately 50 seconds.
    4The error code 0207 will appear on the display.
    5Press any key other than the On/Off key to return the cardiograph to active ECG mode 
    use.
    6Connect the PIM or replace the defective PIM.
    7The cardiograph will operate normally and requires no further intervention. 
    Special Note about PIM Repairs 
    If a Patient Interface Module (PIM) used with a PageWriter Trim I cardiograph with installed 
    software version A.01.03 requires repair or replacement, ensure that following the repair or 
    replacement that the cardiograph has installed PIM software version 
    G.003. For further 
    assistance, contact the nearest Philips Response Center. See Contacting a Philips Response 
    Center on page 1-32. 
    						
    							Installing PageWriter Trim I SoftwareSpecial Note about PIM Repairs
    A-16PageWriter Trim Cardiograph Service Manual 
    						
    							B
    B-1
    Appendix CInstalling PageWriter Trim II, III, and
    Rx Software
    A software installation procedure takes approximately twenty minutes to complete.
    Software Upgrades
    The upgrade path for a specific cardiograph is dependent upon the software version that is 
    currently installed on the unit. Philips recommends that all cardiographs be upgraded to 
    software version A.01.03 for optimal performance. See Table B-5 for more information for 
    the appropriate upgrade path. 
    Upgrading Legacy Software
    Legacy software is defined as any PageWriter Trim II/III/Rx software version that is lower 
    than A.01.01. Upgrading legacy software to the Philips recommended software version 
    A.01.03 requires performing two sequential software upgrades. First, upgrade the cardiograph 
    software from A.00.02 to A.01.01, and then upgrade from software version A.01.01 to version 
    A.01.03. 
    A token number and token label are required if upgrading from legacy software version 
    A.00.03 to A.01.01 or higher. A token number is an identification number that is assigned to a 
    specific cardiograph, and that is cross-referenced to the serial number of the same 
    cardiograph. A token number is required in order to complete any legacy software upgrade. 
    For information on obtaining a token number and token label in order to complete a legacy 
    software upgrade procedure, contact the nearest Philips Response Center. See “Contacting a 
    Philips Response Center” on page 1-32 for a listing of contact telephone numbers. For 
    information on applying the token label to the cardiograph, or managing token numbers, see 
    “PageWriter Trim II, III and Rx Token Label” on page 1-19.
    Table B-5 PageWriter Trim II/III/Rx Software Upgrade Path 
    Cardiograph Model Available Software RevisionLowest Software Version 
    Required for Upgrade
    Trim II/III/Rx A.01.03 (recommended)A.01.01
    Trim II/III/Rx A.01.01 A.00.02
    *
    *. See “Upgrading Legacy Software” below for important information on upgrading from legacy software version A.00.02 to
    software version A.01.01 or higher. Philips recommends upgrading all PageWriter Trim II/III/Rx cardiographs to software
    version A.01.03 for optimal performance.  
    						
    							Installing PageWriter Trim II, III, and Rx SoftwareSaving Custom Settings (Software version A.00.03 and lower only)
    B-2PageWriter Trim Cardiograph Service Manual
    Special Note about Japanese Localization Option
    The Japanese AR0 option for the PageWriter Trim II and III model cardiographs is no longer 
    available. The supported software revision level for all Japanese option cardiographs is legacy 
    software version A.00.03. There is no software upgrade path for the Japanese localization 
    option.
    Table  B-6 provides detailed software revision information. 
    Follow the same procedure to upgrade the software in an existing working unit or to install 
    software on a new system. 
    Saving Custom Settings (Software version A.00.03 and 
    lower only)
    NOTESoftware version A.01.01 and higher automatically saves custom settings. Only follow this procedure 
    when upgrading from legacy software version A.00.03 and lower to software version A.01.01.
    Custom settings (excluding all network settings) must be saved to a PC card before installing 
    the software upgrade. All network settings must be recorded on these instructions in the spaces 
    indicated before installing the software upgrade. Network settings must be manually reentered 
    on the cardiograph after the software installation procedure is complete. 
    NOTEEnsure that all custom settings specified on the configuration screens are accurate and complete 
    before saving the custom settings file. 
    To save custom settings:
    1
    Insert a PC card into the PC card slot on the rear of the cardiograph. 
    Ensure that the PC card is fully inserted into the slot.
    2Press the Ta b key ( ) or turn the Trim Knob to highlight the Config button on the 
    Command Toolbar.
    3Press the space bar or the Trim Knob to select the button. The Configuration screens 
    appear.
    Table B-6 Software Version Number Information
    Component A.00.03 A.01.00 A.01.01 A.01.03
    Loader L.03.12 L.03.12 L.03.12 L.03.12
    Kernel K.02.22 K.03.00 K.03.01 K.03.01
    Application A.07.02.02 A.08.00.02 A.08.01.01 A.08.03.02
    PIM Loader 1.009 1.009 1.009
    PIM Kernel D.008 D.008 D.008 G.003 or 
    D.008
    *
    *. See Software Version A.01.03 PIM Kernel Revision on page B-16 for more information.
    CONFIG 
    						
    							Installing PageWriter Trim II, III, and Rx SoftwareSaving Custom Settings (Software version A.00.03 and lower only)
    PageWriter Trim Cardiograph Service ManualB-3
    4Press and hold the Alt key on the keyboard, and then press the N key to select the 
    Network tab.
    All configured cardiograph Network settings appear on the screen. Record all network 
    settings as indicated in the following steps. 
    5Under Network ID (right side of screen) write down the Computer Name exactly as it 
    appears in the field. 
    NOTEThe Computer Name is case-sensitive (upper and lower case letters) and is limited to 15 
    characters.
    6
    Write down all of the networking information listed on the screen in the table below, if 
    necessary. 
    Keep this information in a safe place. This information will be manually entered on the 
    same screen after the software installation is complete.
    NOTEAll cardiograph networking configuration information is deleted when the software upgrade is 
    installed. This information is not saved with the custom settings file.
    7
    Press and hold the Alt key on the keyboard and then press the S key to select the System 
    tab. 
    8Turn the Trim Knob to highlight the Save Configuration File button under Configuration 
    Management
    , then push the Trim Knob to select the button. 
    The 
    Input Configuration file name window appears. 
    9Type in a name for the custom settings file. 
    Record the file name in the space below.
    10Press the Enter key on the keyboard. 
    The Custom Settings file is saved to the PC card. A message appears that the file has been 
    successfully saved. 
    11Remove the PC card from the PC card slot.
    Proceed to install the software upgrade.
    Computer Name
    IP Address
    Subnet Mask
    Default Gateway
    Secondary WINS Server
    Custom Settings File Name 
    						
    							Installing PageWriter Trim II, III, and Rx SoftwareObtaining Software
    B-4PageWriter Trim Cardiograph Service Manual
    Obtaining Software 
    The software installation procedure requires the use of a computer with a compatible PCMCIA 
    card slot, and internet access. The procedure also requires the use of a blank CompactFlash 
    (CF) memory card and a PC card adapter. The memory card and adapter, referred to as a PC 
    card, may be purchased from Philips using part number 453563479701. The following 
    procedure describes how to download the cardiograph software files to the PC card. If the PC 
    card with the applicable software is available, proceed to page B-6.
    Downloading Software Files from Philips InCenter Site
    All software upgrade files may be downloaded from the Philips InCenter web site found at: 
    www.incenter.medical.philips.com. The InCenter site requires an active login and password. 
    Instructions for obtaining a login and password are located on the right side of the main 
    InCenter home page. 
    NOTEFor more information on registering for the InCenter site, see “Using the Philips InCenter Site” on 
    page 1-26.
    To load the software upgrade files to the PC card:
    1
    Insert the CompactFlash (CF) card into the PC card adapter.
    2Insert the PC card into the card reader on a computer or other device.
    3Go to the InCenter site (incenter.medical.philips.com) and enter your login information.
    4On the top menu bar (top of screen), click on Service. From the drop-down menu, select 
    Software, Software Downloads, and then Diagnostic ECG. The Diagnostic ECG 
    Software Downloads screen appears. 
    Figure B-14 InCenter Service Menu
    5
    Select PageWriter from the side menu (left side of screen). Then select the applicable 
    cardiograph model (
    II, III, or Rx). The PageWriter Trim software download screen 
    appears. 
    6Click the PageWriter Trim A.01.01 or A.01.03 link. 
    NOTEIf upgrading a cardiograph from legacy software (A.00.03 or lower) to software version A.01.03, 
    the cardiograph must first be upgraded to software version A.01.01. See Upgrading Legacy 
    Software on page B-1 for more information.
    7
    On the following software download screen, click on the applicable software version for 
    your language.  
    						
    							Installing PageWriter Trim II, III, and Rx SoftwareObtaining Software
    PageWriter Trim Cardiograph Service ManualB-5
    NOTESThe US English and European software link includes all supported European languages.
    The Documentation link contains all user documentation (instructions for use, quick reference 
    cards) for the selected software version. 
    8
    The InCenter software download license agreement appears. If you agree to the terms of 
    the agreement, click 
    I agree. 
    NOTEA message may appear that the Download Manager application needs to be installed. Click as 
    indicated to install the ActiveX control in order to continue with the software download. See the 
    InCenter main page for more information on installing and using ActiveX controls with the 
    InCenter site. 
    9
    Select a file destination and then click Save. The Philips Download Manager dialog 
    appears. The software files are saved to the specified destination. After the files are saved, 
    click 
    Launch. The WinZip Self Extractor dialog appears.
    10Select a destination for the files and then click Unzip. The files are saved to the specified 
    destination. If the save operation is successful, the top-level directory structure should 
    appear as shown in B-15.
    NOTESThe top-level directory structure is identical for software version A.01.01 and version A.01.03, 
    and appears for both software versions as shown in Figure B-15.
    The ENU folder will appear as SCN for Simplified Chinese, or TCN for Traditional Chinese.
    Figure B-15 Trim II/III/Rx PC Card Directory
    11
    Ensure that the software files are saved to the PC card. Remove the PC card from the 
    computer or other device and proceed to the next section. 
    						
    							Installing PageWriter Trim II, III, and Rx SoftwareInstalling the Software Upgrade
    B-6PageWriter Trim Cardiograph Service Manual
    Installing the Software Upgrade 
    For ease of navigation on the installation screens, connect a USB mouse to the USB connector 
    on the rear of the cardiograph. If a USB mouse is not available, use the Tab key and the Enter 
    key on the keyboard to navigate on the software installation screens.
    To install the software upgrade:
    1
    Connect a USB mouse to the USB connector ( ) on the rear of the cardiograph.
    2To protect against static discharge, touch the PC card (with memory card inserted into the 
    adapter) to the grounding post on the rear of the cardiograph. 
    Figure B-16 Grounding post on rear of cardiograph
    3
    Insert the PC card into the PC card slot on the rear of the cardiograph. Ensure that the PC 
    card is fully inserted into the slot.
    4Ensure that the Patient Interface Module (PIM) is securely attached to the connector on the 
    rear of the cardiograph.
    Figure B-17 Connecting the PIM to the cardiograph
    AGrounding post