    							Installing PageWriter Trim II, III, and Rx SoftwareInstalling the Software Upgrade
    PageWriter Trim Cardiograph Service ManualB-7
    5Connect the cardiograph to AC power. Ensure that the green AC power indicator light is 
    lit. 
      WARNINGAlways connect the cardiograph to AC power when performing the software upgrade. If 
    the cardiograph is not connected to AC power when performing the software upgrade it 
    may be damaged.
    6Ensure that the cardiograph is shut down (screen is black) and that the Standby light is not 
    lit. If necessary, press and hold the On/Standby button to exit Standby. Then, press and 
    hold the On/Standby button to shut down the cardiograph. 
    7Press the On/Standby button. After about twenty seconds, the cardiograph beeps and the 
    PageWriter Trim software identification screen appears.
    Figure B-18 The Software Identification Screen
    8
    As soon as the software identification screen appears, press the Copy and Page buttons 
    (right side of cardiograph) at the same time.
    NOTEThe software identification screen only appears for five seconds. If the software identification 
    screen disappears before the buttons are pressed, press and hold the On/Standby button to shut 
    down the cardiograph. Press the On/Standby button again and repeat the procedure.  
    						
    							Installing PageWriter Trim II, III, and Rx SoftwareInstalling the Software Upgrade
    B-8PageWriter Trim Cardiograph Service Manual
    9The Software Installation Utility screen appears. 
    Figure B-19 Software Installation Utility screen
    10
    Ensure that the correct localization option is selected next to the Option field on the 
    Software Installation Utility screen. See Table B-7, “Software Localization Options,” on 
    page B-8 for a complete listing of all available localization options. 
    Table B-7 Software Localization Options 
    Localization
    Option CodeDescription Language Lead Standard Paper Size
    AB0 Taiwan Traditional 
    ChineseAAMI A
    AB2 China Simplified 
    ChineseIEC A4
    AB4 Singapore & 
    Hong KongEnglish AAMI A4
    AB9 Portugal Portuguese 
    (Brazilian)IEC A4
    ABA USA/Canada
    (English)English AAMI A
    ABB European 
    EnglishEnglish IEC A4
    ABB Romania 
    EnglishEnglish IEC A4
    ABC Canada (French) French AAMI A
    ABD Germany German IEC A4
    ABE Spain Spanish IEC A4
    ABF France French IEC A4 
    						
    							Installing PageWriter Trim II, III, and Rx SoftwareInstalling the Software Upgrade
    PageWriter Trim Cardiograph Service ManualB-9
    11Press the Ta b key to highlight the Install Operation drop-down menu or click on the 
    menu with the mouse. 
    ABG Australia English AAMI A4
    ABH Netherlands Dutch IEC A4
    ABM Latin American Spanish AAMI A
    ABN Norway Norwegian IEC A4
    ABS Sweden Swedish IEC A4
    ABU UK English IEC A4
    ABX Finland Finnish IEC A4
    ABZ Italy Italian IEC A4
    AC4 Brazil Portuguese AAMI A
    AC6 South Korea English AAMI A4
    AC8 Argentina Spanish AAMI A
    ACB Russia English IEC A4
    ACE Denmark & 
    Faroe Islands Danish IEC A4
    ACJ India English IEC A4
    AKB Czech Republic English IEC A4
    AKD Poland English IEC A4
    AKJ Israel & Gaza 
    StripEnglish IEC A4
    AKR Slovak Republic English IEC A4
    AKV Chile & others Spanish AAMI A
    AR0
    *Japan Japanese IEC A4
    AR2 Greece English IEC A4
    ARF Hungary English IEC A4
    *. The Japanese AR0 option is no longer available. The supported software revision level for this localization 
    option is A.00.03. There is no software upgrade path for this localization option.
    Table B-7 Software Localization Options (continued)
    Localization
    Option CodeDescription Language Lead Standard Paper Size 
    						
    							Installing PageWriter Trim II, III, and Rx SoftwareVerifying the Software Upgrade
    B-10PageWriter Trim Cardiograph Service Manual
    12Press the Ta b key to highlight the Install Operations menu, or click on the menu with the 
    mouse. 
    13For software version A.01.01 only: press the down arrow key to select Kernel. 
    CAUTIONDo not select Kernel if installing software version A.01.03. Go to step 16.
    14Click on the Start Installation button or press Ta b to highlight the button and then press 
    the Enter key ( ). The kernel installation takes several minutes to complete, and the 
    cardiograph automatically restarts.
    15After about forty seconds, the cardiograph beeps and the PageWriter Trim software 
    identification screen appears.As soon as the software identification screen appears, press 
    the 
    Copy and Page buttons at the same time.
    NOTEThe software identification screen only appears for five seconds. If the software identification 
    screen disappears before the buttons are pressed, press and hold the On/Standby button to shut 
    down the cardiograph. Press the On/Standby button again and repeat the procedure. 
    The Software Installation Utility screen is displayed.
    16For all software versions: press the down arrow key to select All Except Kernel. 
    17
    Click on the Start Installation button or press the Ta b key to highlight the button then 
    press Enter. It takes several minutes to upgrade the application, and then the cardiograph 
    automatically shuts down. Remove the PC card from the cardiograph.
    Go to the next procedure to verify that the software installation is complete. 
    Verifying the Software Upgrade 
    Complete the following procedure to verify that the software installation procedure on the 
    PageWriter Trim II, III, or Rx model cardiographs is complete. 
    To verify that the software installation is complete:
    1
    Ensure that the installation PC card is removed from the cardiograph.
    2Press the On/Standby button. After about forty seconds, the cardiograph beeps and the 
    PageWriter Trim software identification screen appears. 
    						
    							Installing PageWriter Trim II, III, and Rx SoftwareVerifying the Software Upgrade
    PageWriter Trim Cardiograph Service ManualB-11
    Figure B-20 The Software Identification Screen
    3
    As soon as the software identification screen appears, press the left Ctrl + Shift+ Trim 
    Knob at the same time. 
    4The Access Code window appears. If it does not appear, press the left Ctrl + Shift+ Trim 
    Knob sequence again.
    5Type in 1573 and then press the Enter key.
    6The Service Utility screen appears. Under Diagnostic Tests (center of screen) select Self 
    Test
     from the menu. Ensure that the Print Results When Applicable check box is 
    selected. Enter text in the 
    Test Info: field to identify the specific test date or test operator.
    Figure B-21 The Service Utility main screen
    7
    Click Start. A progress indicator appears at the lower right corner of the screen. Follow 
    the instructions on the screen to complete each portion of the Self Test. 
    8For the Trim Knob test, turn the Trim Knob in a clockwise direction for four full rotations, 
    and then turn the Trim Knob in a counterclockwise direction for four full rotations. Press 
    the Trim Knob. Verify that the Trim Knob graphic on the screen represents the actual 
    position of the Trim Knob when turning. Click the 
    Pass button if the graphic on the screen 
    and the Trim Knob are consistent. 
    						
    							Installing PageWriter Trim II, III, and Rx SoftwareVerifying the Software Upgrade
    B-12PageWriter Trim Cardiograph Service Manual
    9For the Screen Test, a blank screen appears with the buttons Color and Pattern at the 
    bottom of the screen. Click both buttons to display the test image.
    NOTEThe color gradations in the image on the cardiograph display will be more gradual and smoother 
    than those shown in Figure B-22.
    Figure B-22 Screen Test Image
    10
    Look for the following details in the test image appearing on the screen:
    – The progression of shading (from light to dark) in the red, green, and blue bars should 
    be smooth and without breaks.
    – The gray lines (on top of color bars) should be straight and intersect the cross hairs at 
    five points on the screen.
    If the screen does not look similar to the test image in Figure B-22, the screen display 
    failed the Screen Test. 
    11After examining the image, click the Close button.
    The Test Result Confirmation window appears. 
    12Select Yes if the image on the screen is displayed correctly. 
    Select 
    No if the image on the screen did not display correctly. 
    A message appears confirming that the Screen Test failed or passed. 
    13Next, the printer performs a self test and the cardiograph prints out the printer test page. 
    Inspect the printer test page using the test points explained in Table B-4. If the printer test 
    page matches the printer test points as described in Table B-4 the cardiograph has passed  
    						
    							Installing PageWriter Trim II, III, and Rx SoftwareVerifying the Software Upgrade
    PageWriter Trim Cardiograph Service ManualB-13
    the printer test. If the printer test page does not match the printer test points, the 
    cardiograph has failed the print test. 
    Table B-4 Printer Test Page Description
    Test Point Description
    A The stepped bars are sharp edged and printed cleanly without distortion 
    or missing segments 
    B The spacing between the vertical lines is 25 mm with a discrepancy of no 
    more or less than 2%
    C The diagonal lines should be straight and printed cleanly without 
    distortion or breaks in the lines
    D The character set is printed cleanly without distortion or missing 
    characters, and all characters are clearly legible 
    						
    							Installing PageWriter Trim II, III, and Rx SoftwareVerifying the Software Upgrade
    B-14PageWriter Trim Cardiograph Service Manual
    Figure B-23 Printer Test Page
    14
    Click the OK button if the cardiograph passed the printer test. 
    15When the Self Test is complete, the cardiograph prints out a Self Test Report. On the 
    printed test report, check the software revision level next to 
    Release Version (lower right 
    side of report). The revision level should read 
    A.01.01 (Figure B-24 on page B-15) or 
    A.01.03 (Figure B-25 on page B-16). Check that the correct localization option appears 
    next to 
    Country Options, see Table B-7 on page B-8.
    A
    B
    C
    D 
    						
    							Installing PageWriter Trim II, III, and Rx SoftwareVerifying the Software Upgrade
    PageWriter Trim Cardiograph Service ManualB-15
    Figure B-24 Software Version A.01.01 Self Test Report
    A
    ASoftware Revision Level and Localization Option Information 
    						
    							Installing PageWriter Trim II, III, and Rx SoftwareVerifying the Software Upgrade
    B-16PageWriter Trim Cardiograph Service Manual
    Figure B-25 Software Version A.01.03 Self Test Report
    If an incorrect software revision level or incorrect localization option appears on the Self Test 
    Report, reinstall the software following the same procedure. If an incorrect revision level or 
    localization option appears again on the report, or if the cardiograph fails any portion of the 
    Self Test, contact the nearest Philips Response Center for further assistance. See Contacting a 
    Philips Response Center on page 1-32.
    Software Version A.01.03 PIM Kernel Revision 
    For software version A.01.03, the PIM Kernel Revision Level may appear on the Self Test 
    Report as either 
    D.008 (as shown above in Figure B-25) or as version G.003. Both PIM Kernel 
    versions are valid for software version A.01.03.
    B
    A
    APIM Kernel Revision Level Information
    BSoftware Revision Level and Localization Option Information 
    						