Philips Trim 1 Service Manual
Have a look at the manual Philips Trim 1 Service Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 619 Philips manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.
Installing PageWriter Trim II, III, and Rx SoftwareInstalling the Software Upgrade PageWriter Trim Cardiograph Service ManualB-7 5Connect the cardiograph to AC power. Ensure that the green AC power indicator light is lit. WARNINGAlways connect the cardiograph to AC power when performing the software upgrade. If the cardiograph is not connected to AC power when performing the software upgrade it may be damaged. 6Ensure that the cardiograph is shut down (screen is black) and that the Standby light is not lit. If necessary, press and hold the On/Standby button to exit Standby. Then, press and hold the On/Standby button to shut down the cardiograph. 7Press the On/Standby button. After about twenty seconds, the cardiograph beeps and the PageWriter Trim software identification screen appears. Figure B-18 The Software Identification Screen 8 As soon as the software identification screen appears, press the Copy and Page buttons (right side of cardiograph) at the same time. NOTEThe software identification screen only appears for five seconds. If the software identification screen disappears before the buttons are pressed, press and hold the On/Standby button to shut down the cardiograph. Press the On/Standby button again and repeat the procedure.
Installing PageWriter Trim II, III, and Rx SoftwareInstalling the Software Upgrade B-8PageWriter Trim Cardiograph Service Manual 9The Software Installation Utility screen appears. Figure B-19 Software Installation Utility screen 10 Ensure that the correct localization option is selected next to the Option field on the Software Installation Utility screen. See Table B-7, “Software Localization Options,” on page B-8 for a complete listing of all available localization options. Table B-7 Software Localization Options Localization Option CodeDescription Language Lead Standard Paper Size AB0 Taiwan Traditional ChineseAAMI A AB2 China Simplified ChineseIEC A4 AB4 Singapore & Hong KongEnglish AAMI A4 AB9 Portugal Portuguese (Brazilian)IEC A4 ABA USA/Canada (English)English AAMI A ABB European EnglishEnglish IEC A4 ABB Romania EnglishEnglish IEC A4 ABC Canada (French) French AAMI A ABD Germany German IEC A4 ABE Spain Spanish IEC A4 ABF France French IEC A4
Installing PageWriter Trim II, III, and Rx SoftwareInstalling the Software Upgrade PageWriter Trim Cardiograph Service ManualB-9 11Press the Ta b key to highlight the Install Operation drop-down menu or click on the menu with the mouse. ABG Australia English AAMI A4 ABH Netherlands Dutch IEC A4 ABM Latin American Spanish AAMI A ABN Norway Norwegian IEC A4 ABS Sweden Swedish IEC A4 ABU UK English IEC A4 ABX Finland Finnish IEC A4 ABZ Italy Italian IEC A4 AC4 Brazil Portuguese AAMI A AC6 South Korea English AAMI A4 AC8 Argentina Spanish AAMI A ACB Russia English IEC A4 ACE Denmark & Faroe Islands Danish IEC A4 ACJ India English IEC A4 AKB Czech Republic English IEC A4 AKD Poland English IEC A4 AKJ Israel & Gaza StripEnglish IEC A4 AKR Slovak Republic English IEC A4 AKV Chile & others Spanish AAMI A AR0 *Japan Japanese IEC A4 AR2 Greece English IEC A4 ARF Hungary English IEC A4 *. The Japanese AR0 option is no longer available. The supported software revision level for this localization option is A.00.03. There is no software upgrade path for this localization option. Table B-7 Software Localization Options (continued) Localization Option CodeDescription Language Lead Standard Paper Size
Installing PageWriter Trim II, III, and Rx SoftwareVerifying the Software Upgrade B-10PageWriter Trim Cardiograph Service Manual 12Press the Ta b key to highlight the Install Operations menu, or click on the menu with the mouse. 13For software version A.01.01 only: press the down arrow key to select Kernel. CAUTIONDo not select Kernel if installing software version A.01.03. Go to step 16. 14Click on the Start Installation button or press Ta b to highlight the button and then press the Enter key ( ). The kernel installation takes several minutes to complete, and the cardiograph automatically restarts. 15After about forty seconds, the cardiograph beeps and the PageWriter Trim software identification screen appears.As soon as the software identification screen appears, press the Copy and Page buttons at the same time. NOTEThe software identification screen only appears for five seconds. If the software identification screen disappears before the buttons are pressed, press and hold the On/Standby button to shut down the cardiograph. Press the On/Standby button again and repeat the procedure. The Software Installation Utility screen is displayed. 16For all software versions: press the down arrow key to select All Except Kernel. 17 Click on the Start Installation button or press the Ta b key to highlight the button then press Enter. It takes several minutes to upgrade the application, and then the cardiograph automatically shuts down. Remove the PC card from the cardiograph. Go to the next procedure to verify that the software installation is complete. Verifying the Software Upgrade Complete the following procedure to verify that the software installation procedure on the PageWriter Trim II, III, or Rx model cardiographs is complete. To verify that the software installation is complete: 1 Ensure that the installation PC card is removed from the cardiograph. 2Press the On/Standby button. After about forty seconds, the cardiograph beeps and the PageWriter Trim software identification screen appears.
Installing PageWriter Trim II, III, and Rx SoftwareVerifying the Software Upgrade PageWriter Trim Cardiograph Service ManualB-11 Figure B-20 The Software Identification Screen 3 As soon as the software identification screen appears, press the left Ctrl + Shift+ Trim Knob at the same time. 4The Access Code window appears. If it does not appear, press the left Ctrl + Shift+ Trim Knob sequence again. 5Type in 1573 and then press the Enter key. 6The Service Utility screen appears. Under Diagnostic Tests (center of screen) select Self Test from the menu. Ensure that the Print Results When Applicable check box is selected. Enter text in the Test Info: field to identify the specific test date or test operator. Figure B-21 The Service Utility main screen 7 Click Start. A progress indicator appears at the lower right corner of the screen. Follow the instructions on the screen to complete each portion of the Self Test. 8For the Trim Knob test, turn the Trim Knob in a clockwise direction for four full rotations, and then turn the Trim Knob in a counterclockwise direction for four full rotations. Press the Trim Knob. Verify that the Trim Knob graphic on the screen represents the actual position of the Trim Knob when turning. Click the Pass button if the graphic on the screen and the Trim Knob are consistent.
Installing PageWriter Trim II, III, and Rx SoftwareVerifying the Software Upgrade B-12PageWriter Trim Cardiograph Service Manual 9For the Screen Test, a blank screen appears with the buttons Color and Pattern at the bottom of the screen. Click both buttons to display the test image. NOTEThe color gradations in the image on the cardiograph display will be more gradual and smoother than those shown in Figure B-22. Figure B-22 Screen Test Image 10 Look for the following details in the test image appearing on the screen: – The progression of shading (from light to dark) in the red, green, and blue bars should be smooth and without breaks. – The gray lines (on top of color bars) should be straight and intersect the cross hairs at five points on the screen. If the screen does not look similar to the test image in Figure B-22, the screen display failed the Screen Test. 11After examining the image, click the Close button. The Test Result Confirmation window appears. 12Select Yes if the image on the screen is displayed correctly. Select No if the image on the screen did not display correctly. A message appears confirming that the Screen Test failed or passed. 13Next, the printer performs a self test and the cardiograph prints out the printer test page. Inspect the printer test page using the test points explained in Table B-4. If the printer test page matches the printer test points as described in Table B-4 the cardiograph has passed
Installing PageWriter Trim II, III, and Rx SoftwareVerifying the Software Upgrade PageWriter Trim Cardiograph Service ManualB-13 the printer test. If the printer test page does not match the printer test points, the cardiograph has failed the print test. Table B-4 Printer Test Page Description Test Point Description A The stepped bars are sharp edged and printed cleanly without distortion or missing segments B The spacing between the vertical lines is 25 mm with a discrepancy of no more or less than 2% C The diagonal lines should be straight and printed cleanly without distortion or breaks in the lines D The character set is printed cleanly without distortion or missing characters, and all characters are clearly legible
Installing PageWriter Trim II, III, and Rx SoftwareVerifying the Software Upgrade B-14PageWriter Trim Cardiograph Service Manual Figure B-23 Printer Test Page 14 Click the OK button if the cardiograph passed the printer test. 15When the Self Test is complete, the cardiograph prints out a Self Test Report. On the printed test report, check the software revision level next to Release Version (lower right side of report). The revision level should read A.01.01 (Figure B-24 on page B-15) or A.01.03 (Figure B-25 on page B-16). Check that the correct localization option appears next to Country Options, see Table B-7 on page B-8. A B C D
Installing PageWriter Trim II, III, and Rx SoftwareVerifying the Software Upgrade PageWriter Trim Cardiograph Service ManualB-15 Figure B-24 Software Version A.01.01 Self Test Report A ASoftware Revision Level and Localization Option Information
Installing PageWriter Trim II, III, and Rx SoftwareVerifying the Software Upgrade B-16PageWriter Trim Cardiograph Service Manual Figure B-25 Software Version A.01.03 Self Test Report If an incorrect software revision level or incorrect localization option appears on the Self Test Report, reinstall the software following the same procedure. If an incorrect revision level or localization option appears again on the report, or if the cardiograph fails any portion of the Self Test, contact the nearest Philips Response Center for further assistance. See Contacting a Philips Response Center on page 1-32. Software Version A.01.03 PIM Kernel Revision For software version A.01.03, the PIM Kernel Revision Level may appear on the Self Test Report as either D.008 (as shown above in Figure B-25) or as version G.003. Both PIM Kernel versions are valid for software version A.01.03. B A APIM Kernel Revision Level Information BSoftware Revision Level and Localization Option Information