Home > Philips > Cardiograph > Philips Trim 1 Service Manual

Philips Trim 1 Service Manual

    Download as PDF Print this page Share this page

    Have a look at the manual Philips Trim 1 Service Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 619 Philips manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    							Installing PageWriter Trim I SoftwareObtaining Software
    PageWriter Trim Cardiograph Service ManualA-3
    To load the software upgrade files to the PC card:
    1
    Insert the memory card into the PC card adapter as shown. 
    2Insert the PC card into a compatible slot on a computer or other device.
    3Go to the InCenter site (incenter.medical.philips.com) and enter your login information.
    NOTEFor information on registering for the InCenter site, see “Using the Philips InCenter Site” on 
    page 1-26.
    4
    On the top menu bar (top of screen), click on Service. From the drop-down menu, select 
    Software, Software Downloads, and then Diagnostic ECG. The Diagnostic ECG 
    Software Downloads screen appears. 
    Figure A-3 InCenter Service Menu
    5
    Select PageWriter from the side menu (left side of screen). Then select PageWriter 
    Trim I
    . The PageWriter Trim I software download screen appears. 
    6Click the PageWriter Trim I A.01.01 or A.01.03 link. 
    NOTEIf upgrading a cardiograph from legacy software (A.00.03 or lower) to software version A.01.03, 
    the cardiograph must first be upgraded to software version A.01.01. See Upgrading Legacy 
    Software on page A-1 for more information.
    7
    On the following software download screen, click on the PageWriter Trim A.01.03 or 
    A.01.01 
    link. 
    NOTEThe Documentation link contains all user documentation (instructions for use, quick reference 
    cards) for the selected software version.  A
    PC card adapterBMemory card
    AB 
    						
    							Installing PageWriter Trim I SoftwareObtaining Software
    A-4PageWriter Trim Cardiograph Service Manual
    8The InCenter software download license agreement appears. If you agree to the terms of 
    the agreement, click 
    I agree. 
    NOTEA message may appear that the Download Manager application needs to be installed. Click as 
    indicated to install the ActiveX control in order to continue with the software download. See the 
    InCenter main page for more information on installing and using ActiveX controls with the 
    InCenter site. 
    9
    Select a file destination and then click Save. The Philips Download Manager dialog 
    appears. The software files are saved to the specified destination. After the files are saved, 
    click 
    Launch. The WinZip Self Extractor dialog appears.
    10Select a destination for the files and then click Unzip. The files are saved to the specified 
    destination. If the save operation is successful, the top-level directory structure should 
    appear as shown in Figure A-4.
    NOTEThe top-level directory structure is identical for software version A.01.01 and version A.01.03, 
    and appears for both software versions as shown in Figure A-4.
    Figure A-4 Trim I PC Card Directory
    11
    Remove the PC card from the computer or other device and proceed to the next section. 
    						
    							Installing PageWriter Trim I SoftwareInstalling the Software Upgrade
    PageWriter Trim Cardiograph Service ManualA-5
    Installing the Software Upgrade 
    To install the software upgrade:
    1
    Peel off the protective covering from the rear panel of the cardiograph to expose the PC 
    card slot.
    Figure A-5 Removing the protective cover 
    2
    To protect against static discharge, touch the PC card (with memory card inserted into the 
    adapter) to the grounding post on the rear of the cardiograph. 
    Figure A-6 Grounding post on rear of cardiograph
    AGrounding post
    A 
    						
    							Installing PageWriter Trim I SoftwareInstalling the Software Upgrade
    A-6PageWriter Trim Cardiograph Service Manual
    3Ensure that the Patient Interface Module (PIM) is securely attached to the connector on the 
    rear of the cardiograph.
    Figure A-7 Connecting the PIM to the cardiograph
    4
    Insert the PC card into the PC card slot. 
    Figure A-8 Inserting the PC Card
    5
    Connect the cardiograph to AC power. Verify that the green LED on the front of the 
    cardiograph is lit.
      WARNINGAlways connect the cardiograph to AC power when performing the software upgrade. If 
    the cardiograph is not connected to AC power when performing the software upgrade it 
    may be damaged.
    6Press the On/Off button to turn on the cardiograph. 
    7When the five triangles (\\\\\) appear on the screen, press the Patient ID and 
    Auto buttons at the same time. Press the buttons within two seconds of the triangles  APC card eject button
    A 
    						
    							Installing PageWriter Trim I SoftwareInstalling the Software Upgrade
    PageWriter Trim Cardiograph Service ManualA-7
    appearing on the screen. If the buttons are not pressed in time, press the On/Off button to 
    turn off the cardiograph. Press the On/Off button again and repeat the procedure.
    8A prompt appears to enter a password. Press the following keys in sequence: Patient ID, 
    Auto, Rhythm, Copy, Page. Press the Stop key to delete your last input if it is wrong. 
    The Biomed Service Program screen appears. 
    9Turn the Trim Knob to highlight Settings, then press the Trim Knob to select it.
    10Ensure that the correct localization option is selected, see Table A-3, “Software 
    Localization Options,” on page A-7 for a complete listing of all available localization 
    options. 
    Table A-3 Software Localization Options 
    Localization 
    Option Code Description Language Lead StandardPaper
    Size
    AB0 Chinese, 
    Traditional Traditional Chinese AAMI A 
    AB2 Chinese, 
    SimplifiedSimplified Chinese IEC A4
    AB4 Singapore & Hong 
    Kong EnglishEnglish AAMI A
    AB9 Portugal, 
    PortuguesePortuguese IEC A4
    ABA U.S.A./Canada 
    EnglishEnglish AAMI A
    ABB European English English IEC A4
    ABC Canadian French French AAMI A
    ABD European German German IEC A4
    ABE European Spanish Spanish IEC A4
    ABF European French French IEC A4
    ABG Australian English English AAMI A4
    ABH European Dutch Dutch IEC A4
    ABM Latin American 
    SpanishSpanish AAMI A
    ABN Norway
    EnglishEnglish IEC A4
    ABS Sweden
    EnglishEnglish IEC A4 
    						
    							Installing PageWriter Trim I SoftwareInstalling the Software Upgrade
    A-8PageWriter Trim Cardiograph Service Manual
    11Turn the Trim Knob to highlight the arrow icon, then press the Trim Knob to return to the 
    main menu.
    12Turn the Trim Knob to highlight Upgrade, then press the Trim Knob to select it. ABU United Kingdom 
    EnglishEnglish IEC A4
    ABX Finland
    EnglishEnglish IEC A4
    ABZ European Italian Italian IEC A4
    AC4 Brazil Portuguese Portuguese AAMI A
    AC6 South Korea
    EnglishEnglish AAMI A4
    AC8 Argentina Spanish Spanish AAMI A
    ACB Russia
    EnglishEnglish IEC A4
    ACE European 
    Denmark & Faroe 
    IslandsEnglish IEC A4
    ACJ India English English IEC A4
    AKB English
    Czech RepublicEnglish IEC A4
    AKD English
    PolandEnglish IEC A4
    AKJ Israel & Gaza 
    Strip EnglishEnglish IEC A4
    AKR English
    Slovak RepublicEnglish IEC A4
    AKV Chile & Others 
    SpanishSpanish AAMI A
    AR0 Japan Japanese 
    KanjiJapanese IEC A4
    AR2 English
    GreeceEnglish IEC A4
    ARF English
    HungaryEnglish IEC A4
    Table A-3 Software Localization Options (continued)
    Localization 
    Option Code Description Language Lead StandardPaper
    Size 
    						
    							Installing PageWriter Trim I SoftwareVerifying the Software Installation
    PageWriter Trim Cardiograph Service ManualA-9
    13Turn the Trim Knob to highlight Upgrade All and then press the Trim Knob to select this 
    installation option.
    14Turn the Trim Knob to highlight Start, then press the Trim Knob to select it. Information 
    on the status of the upgrade installation appears on the screen. 
    15When the installation is complete, the cardiograph automatically shuts down. 
    Remove the PC card from the cardiograph. Go to the next procedure to verify that the new 
    software revision has been successfully installed.
    Verifying the Software Installation
    Complete the following procedure to verify that the software application has been successfully 
    installed. 
    To verify the software installation is complete:
    1
    Ensure that the PC card is removed from the cardiograph.
    2Press the On/Off button to turn on the cardiograph. 
    3When the five triangles (\\\\\) appear on the screen press the Patient ID and 
    Auto buttons at the same time.
    4Press the following buttons in sequence: Patient ID, Auto, Rhythm, Copy, Page. Press 
    the 
    Stop button to delete your last input if it is wrong. The Biomed Service Program 
    screen appears. 
    5On the Main Menu screen, turn the Trim Knob to highlight Test, then press the Trim 
    Knob to select it. 
    6Press the Trim Knob and then turn the Trim Knob to highlight Self Test. Press the Trim 
    Knob to select it.
    7Turn the Trim Knob to highlight Start, then press the Trim Knob to select it. 
    8Directions to complete the Self Test appear on the screen. When indicated, touch each 
    button on the cardiograph except the On/Off button. The numbers 0-8 begin to appear on 
    the screen. Continue the test by pushing the Trim Knob, then turn it clockwise once, and 
    counterclockwise once to complete the test. The numbers 0-6 should appear on the screen 
    when the test is complete. Press the 
    Patient ID button if all 6 numbers appear on the 
    screen. 
    9Next, a series of characters appear on the screen. Verify that the characters appear 
    correctly on the screen. Press the 
    Patient ID button if the characters appear correctly on 
    the screen. 
    10Next, the printer performs a self test and the cardiograph prints out the printer test page. 
    Inspect the printer test page using the test points explained in Table A-3. If the printer test 
    page matches the printer test points as described in Figure A-9 on page A-11 the  
    						
    							Installing PageWriter Trim I SoftwareVerifying the Software Installation
    A-10PageWriter Trim Cardiograph Service Manual
    cardiograph has passed the printer test. If the printer test page does not match the printer 
    test points, the cardiograph has failed the print test. 
    Table A-3 Printer Test Page Description 
    Test Point Description
    A The stepped bars are sharp edged and printed cleanly without distortion 
    or missing segments 
    B The spacing between the vertical lines is 25 mm with a discrepancy of 
    no more or less than 2%
    C The diagonal lines should be straight and printed cleanly without 
    distortion or breaks in the lines
    D The character set is printed cleanly without distortion or missing 
    characters, and all characters are clearly legible 
    						
    							Installing PageWriter Trim I SoftwareVerifying the Software Installation
    PageWriter Trim Cardiograph Service ManualA-11
    Figure A-9 Printer Test Page
    11
    Press the Patient ID button if the cardiograph passed the printer test. 
    12When the Self Test is complete, press the Patient ID button on the cardiograph to print out 
    a report of the test results.
    A
    B
    C
    D 
    						
    							Installing PageWriter Trim I SoftwareVerifying the Software Installation
    A-12PageWriter Trim Cardiograph Service Manual
    13On the printed test report, check the software revision level next to Release Version 
    (lower right side of report). The revision level should read 
    A.01.01(Figure A-10 on 
    page A-12) or 
    A.01.03 (Figure A-11 on page A-13). Check that the correct localization 
    option appears next to 
    Country Options, see Table A-3 on page A-7.
    Figure A-10 Software Version A.01.01 Self Test Report 
    ARelease Version and Localization Option Information
    A 
    						
    All Philips manuals Comments (0)