Home > Philips > Cardiograph > Philips Trim 1 Service Manual

Philips Trim 1 Service Manual

    Download as PDF Print this page Share this page

    Have a look at the manual Philips Trim 1 Service Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 619 Philips manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    							Assembling the Cardiograph Cart and Patient Cable ArmConnecting the PIM the Cardiograph
    PageWriter Trim Cardiograph Service ManualD-7
    Connecting the PIM the Cardiograph
    Connect the patient data cable on the PIM to the PIM connector port ( ) on the rear panel 
    of the cardiograph. 
    To connect the PIM to the cardiograph:
    XConnect the patient data cable to the PIM connector port on the rear of the cardiograph. 
    WARNINGTo ensure safety and prevent damage to the system, only connect the patient data cable 
    to the PIM connector port on the rear of the cardiograph. 
    Placing the PIM in the Holder
    Place the PIM in the holder when not in use. Ensure that the patient data cable rests in the 
    recessed section of the PIM holder.
    Figure D-2 Inserting the PIM into the holder
    CAUTIONEnsure that the PIM patient data cable and lead wires do not drag on the ground or become tangled in 
    the cart wheels.  
    						
    							Assembling the Cardiograph Cart and Patient Cable ArmAssembling the Patient Cable Arm
    D-8PageWriter Trim Cardiograph Service Manual
    Assembling the Patient Cable Arm 
    Follow this procedure to assemble the optional patient cable arm that attaches to the 
    cardiograph cart. For information on ordering the optional patient cable arm, see “Upgrades 
    and Accessories” on page 1-18.
    Figure D-3 The Patient Cable Arm Assembly Kit 
    To assemble the cart:
    1
    Remove the bin dividers from the storage bin.  
    						
    							Assembling the Cardiograph Cart and Patient Cable ArmAssembling the Patient Cable Arm
    PageWriter Trim Cardiograph Service ManualD-9
    2Place a protective cloth on the floor and turn over the cart. 
    3Remove the bolt screws, grounding strap, and counterweight as shown using the tools 
    provided in the kit.  
    						
    							Assembling the Cardiograph Cart and Patient Cable ArmAssembling the Patient Cable Arm
    D-10PageWriter Trim Cardiograph Service Manual
    4Remove the cart base.
    5Remove the caster base from the bottom base as shown using a Philips head screwdriver. 
    						
    							Assembling the Cardiograph Cart and Patient Cable ArmAssembling the Patient Cable Arm
    PageWriter Trim Cardiograph Service ManualD-11
    6Remove the front casters from the base as shown.
    7Insert the replacement front casters from the kit as shown. Tighten the screws as shown. 
    						
    							Assembling the Cardiograph Cart and Patient Cable ArmAssembling the Patient Cable Arm
    D-12PageWriter Trim Cardiograph Service Manual
    8Attach the caster base to the cart base as shown. Tighten the screws with a Philips head 
    screw driver.
    9Attach the cart base counterweight and grounding strap as shown. 
    						
    							Assembling the Cardiograph Cart and Patient Cable ArmAssembling the Patient Cable Arm
    PageWriter Trim Cardiograph Service ManualD-13
    10Reattach the bottom base to the cart beam as shown.
    11Flip the cart upright. 
    						
    							Assembling the Cardiograph Cart and Patient Cable ArmAssembling the Patient Cable Arm
    D-14PageWriter Trim Cardiograph Service Manual
    12Connect the patient cable arm holder onto the rear of the beam as shown. Ensure that the 
    bolt holes are properly aligned. Tighten the screws.
    13Insert the patient cable arm into the holder and insert the mounting screw. Tighten the 
    screw. 
    						
    							Assembling the Cardiograph Cart and Patient Cable ArmAssembling the Patient Cable Arm
    PageWriter Trim Cardiograph Service ManualD-15
    14Remove the screws from the Patient Interface Module (PIM) holder. 
    15Insert the PIM into the holder and connect the bottom housing.  
    						
    							Assembling the Cardiograph Cart and Patient Cable ArmAssembling the Patient Cable Arm
    D-16PageWriter Trim Cardiograph Service Manual
    16Secure the PIM holder housing as shown. Tighten the screws.
    17Attach the cable holders to the arm as shown. 
    						