    							Wireless LAN Installation InstructionsInstalling the Wireless LAN Card
    PageWriter Trim Cardiograph Service ManualC-9
    13Click the OK button (upper right corner of screen) to close the Aironet Client Utility. The 
    PageWriter Trim Wireless LAN Installation screen appears. Click the 
    Reset Cardiograph 
    button. The cardiograph automatically restarts. 
    14
    When the R/T (real-time) ECG screen appears, press the Ta b key ( ) to highlight the 
    Config button on the Command Toolbar. Press the space bar to select the button. The 
    Configuration screens appear.
    15Press and hold the Alt key and then press the N key to select the Network tab. On the 
    Network Configuration screen, press the Alt key and then press the E key to select the 
    Wireless Network tab. 
    16Select Obtain IP Address Automatically or Specify IP Address and enter all relevant 
    information. Consult your network administrator for more information on configuring IP 
    address settings.
    17Press the Enter key when done. The Configuration Management window appears. Press 
    the Enter key to select the 
    Yes button and to save the settings. Select No to close the 
    window and return to the Wireless LAN Network screen. 
    18The R/T ECG screen appears. Press the On/Standby button for three seconds, then release 
    it to shut down the cardiograph. Press the On/Standby button again to restart the 
    cardiograph. The new network setting are applied.
    To configure a TraceMasterVue Remote Site for the purpose of transmitting ECGs, see 
    “Configuring a TraceMasterVue Remote Site” on page C-10.
    For information on the Status indicators on the wireless adapter, see Table C-3 below.
    Table C-3 Wireless Adapter LED Indicators 
    Green LED Status Amber LED Status What LEDs indicate
    Blinking quickly 
    (every 0.5 seconds)Blinking quickly Wireless adapter is not associated to an 
    access point.
    Off Blinking quickly Wireless adapter is associated to an 
    access point (with Power Save Mode set 
    to default Fast PSP)
    Off Off Wireless adapter is not receiving power, 
    or an error has occurred.
    Blinking slowly 
    (every 1.5 seconds)Blinking quickly Wireless adapter is associated to an 
    access point (with Power Save Mode set 
    to CAM; not recommended due to 
    increased power consumption).
    CONFIG 
    						
    							Wireless LAN Installation InstructionsEnabling LEAP Credentials
    C-10PageWriter Trim Cardiograph Service Manual
    Enabling LEAP Credentials
    Follow this procedure after entering the user name, user password, and user domain settings 
    used with LEAP in the Aironet Client Utility (ACU). For information on entering settings in 
    the ACU, see “Installing the Wireless LAN Card” on page C-3. The following procedure must 
    be completed in order to enable LEAP credentials used with LEAP authentication. This 
    procedure must be completed when:
    „transmitting ECGs from the Archive for the first time
    „the wireless LAN card is inserted or removed from the cardiograph while the cardiograph 
    is in active use (not in Standby)
    „the cardiograph is returned to active use after a full power shut down, and the wireless 
    LAN card is fully inserted into the PC card slot on the rear of the cardiograph when the 
    cardiograph is restarted 
    To enable LEAP credentials:
    1
    After specifying LEAP settings in the Aironet Client Utility (ACU), restart the 
    cardiograph.
    2When the R/T (real-time) ECG screen appears, press the Ta b key ( ) to highlight the 
    Archive button on the Command Toolbar. Press the space bar to select the button. The 
    Archive screen appears.
    3Press the Ta b key to highlight the Login WLAN button (bottom left corner of screen). 
    Press the space bar to select the button. 
    NOTENo information is entered. LEAP credentials are enabled by merely selecting the Login WLAN 
    button.
    4
    Press the Esc key ( ) to exit the Archive screen. 
    5Press the Ta b key ( ) to highlight the Config button on the Command Toolbar. Press 
    the space bar to select the button. The Configuration screens appear.
    6Press and hold the Alt key and then press the S key to select the System tab. On the 
    System screen, under 
    Diagnostics (right side of screen) press the Ta b key to highlight the 
    Network button. Press the space bar to select the button. The Network connection 
    window appears. The networking information for the cardiograph appears. 
    To configure a TraceMasterVue Remote Site for the purpose of transmitting ECGs, see 
    “Configuring a TraceMasterVue Remote Site” on page C-10.
    Configuring a TraceMasterVue Remote Site
    The Remote Site settings on the cardiograph are used to define specific TraceMasterVue ECG 
    Management Systems that receive ECGs from the PageWriter Trim cardiograph. 
    For more comprehensive information on configuring the PageWriter Trim cardiograph with a 
    TraceMasterVue ECG Management System, see the Installing TraceMasterVue and 
    Configuring Communication Guide available for download from the Philips InCenter web site 
    (
    incenter.medical.philips.com). Registration is required to access materials on the site. For 
    ARCHIVE 
    						
    							Wireless LAN Installation InstructionsConfiguring a TraceMasterVue Remote Site
    PageWriter Trim Cardiograph Service ManualC-11
    information on registering to use the InCenter site, see “Using the Philips InCenter Site” on 
    page 1-26. The guide is updated on a regular basis, check for updates at the Philips InCenter 
    web site. 
    To configure a Remote Site with the wireless network connection:
    1
    Press the Ta b key ( ) on the keyboard to highlight the Config button. Press the space 
    bar to select the button. The Configuration screens appear.
    2Press and hold down the Alt key on the keyboard and then press the R key. The Remote 
    Sites screen appears. 
    3Press the Ta b key to highlight the Network button (left side of screen).
    4Press the space bar select the setting. 
    5Press the Ta b key to highlight the settings Receive (Query) or Copy/Transfer (left side of 
    screen). Press the space bar to select the check box.
    NOTESelecting the Receive (Query) check box enables the user to search the Remote Site (in the 
    Archive) for ECG files to print on the cardiograph. Selecting the Copy/Transfer check box 
    enables the user to transfer ECGs from the Archive to a TraceMasterVue ECG Management 
    System.
    6
    Under the Network tab (right of screen), press the Ta b key to select the Server URL field. 
    The cursor appears in the field. 
    7Type in the URL address of the Remote Site server. The Server URL is the 
    TraceMasterVue file server URL. The Server URL must be entered in the following 
    format: 
    http://computername/EMSCOMM/ 
    Computername is the TraceMasterVue file server computer name or IP address. Examples 
    of valid URLs include: 
    http://tracemaster/EMSCOMM/ or http://10.101.2.42/EMSCOMM/.
    8Press the Ta b key to highlight the User Name field. The cursor appears in the field. Type 
    in the user name. 
    NOTEThe specified user name must be a member of a TraceMaster group that has View privileges to 
    TraceMasterVue. 
    9
    Press the Ta b key to highlight the Password field. The cursor appears in the field. Type in 
    the password assigned to the specified user. 
    10Press and hold down the Alt key on keyboard.Press the Z key. The Net Connect screen 
    appears.
    11Press the Ta b key to highlight Always Connected connection. 
    12Press the space bar to select the setting. 
    13At the top of the screen a check mark and the word Modified appears. Press the Ta b key to 
    highlight the 
    Add New button. 
    CONFIG 
    						
    							Wireless LAN Installation InstructionsConfiguring a TraceMasterVue Remote Site
    C-12PageWriter Trim Cardiograph Service Manual
    14Press the space bar. The Add New Remote Site window appears.Type in a name for this 
    Remote Site. Press the Ta b key to highlight the 
    OK button. Press the space bar to select the 
    button. 
    The Remote Site appears in the 
    Remote Sites Archive drop-down list (top of screen) and 
    may be selected in the Archive. Press the Enter key (on keyboard) to save the changes and 
    to exit the Configuration screen. The new changes are applied.
    For information on using the Archive to transfer ECGs to a TraceMasterVue ECG 
    Management System, see Chapter 4 “Orders and Archive” of the PageWriter Trim 
    Cardiograph Instructions for Use available on the User Documentation CD, or download 
    the file from the Philips InCenter web site (
    incenter.medical.philips.com).
    For information on troubleshooting wireless connectivity, see “Wireless Troubleshooting” 
    on page 5-24. 
    						
    							D
    D-1
    Appendix AAssembling the Cardiograph Cart and
    Patient Cable Arm
    This Appendix includes information on how to assemble the optional cardiograph cart, and 
    optional patient cable arm for the cardiograph cart. For information on ordering the 
    cardiograph cart or patient cable arm, see “Upgrades and Accessories” on page 1-18. 
    Figure D-1 The PageWriter Trim Cardiograph Cart
    1 
    						
    							Assembling the Cardiograph Cart and Patient Cable ArmAssembling the Cart
    D-2PageWriter Trim Cardiograph Service Manual
    Assembling the Cart 
    Follow this procedure to assemble the cardiograph cart.
    To assemble the cart:
    1
    Insert the beam into the cart base. 
    2Hold the beam steadily. Turn the cart onto the side to expose the bottom of the cart.
    3Place the ground strap onto the screw end. 
    						
    							Assembling the Cardiograph Cart and Patient Cable ArmAssembling the Cart
    PageWriter Trim Cardiograph Service ManualD-3
    4Attach the bolt screws and tighten using the provided wrench.
    5Turn the cart upright.
    6Attach the bin to the cart beam.
    7Insert the bin dividers. 
    						
    							Assembling the Cardiograph Cart and Patient Cable ArmAttaching the Cardiograph to the Cart
    D-4PageWriter Trim Cardiograph Service Manual
    8Pull the tray out from the top shelf. Attach the top shelf to the beam using the provided 
    bolts. Tighten the bolts.
    Attaching the Cardiograph to the Cart 
    CAUTIONFollow the procedure below to ensure that the cardiograph is securely fastened to the cart before use.
    To attach the cardiograph to the cart:
    1
    Align the front feet of the cardiograph with the front locking holes on the cart. Align the 
    rear feet of the cardiograph with the rear screw holes on the cart. Lower the cardiograph 
    onto the cart and snap into place.  
    						
    							Assembling the Cardiograph Cart and Patient Cable ArmUsing the Cart Wheel Positioners and Brake
    PageWriter Trim Cardiograph Service ManualD-5
    2Insert the screws through the bottom of the cart and through the screw holes. Tighten the 
    screws with a Phillips head screwdriver.
    Using the Cart Wheel Positioners and Brake
    The cart includes one wheel brake and two wheel positioners. Lock the wheel positioners at all 
    times when using the cart. The wheel positioners keep the cart straight when moving forward 
    or backward, or when turning corners. The wheel positioners also help the cart maneuver in 
    tight spaces. 
    To use the cart wheel positioners and brake:
    1
    Align the front wheels so that they are straight. Step on both wheel positioners. Move the 
    cart forward until the wheels lock into position. The cart will move forward or backward 
    in a straight line.  
    						
    							Assembling the Cardiograph Cart and Patient Cable ArmConnecting the Patient Data Cable Bracket
    D-6PageWriter Trim Cardiograph Service Manual
    2Step on the gray rear wheel brake to lock the cart wheels. The cart will not move. Step on 
    the wheel brake again to unlock the wheels.
    Connecting the Patient Data Cable Bracket 
    To connect the patient data cable bracket:
    1
    Insert patient data cable into the bracket.
    2Attach the patient data cable bracket to the mounting hole located on the rear of the 
    cardiograph as shown. Tighten the screw with a Philips head screwdriver. 
    						