7 7-1 Chapter 7Removing and Replacing Cardiograph Components This chapter contains information on removing and replacing the PageWriter Trim cardiograph battery, fuses, and paper tray. If you require further technical assistance, contact the nearest Philips Response Center. See “Contacting a Philips Response Center” on page 1-32. About the Cardiograph Components The cardiograph includes the following customer-installable components. „Battery compartment „AC fuses „Paper tray

Removing and Replacing Cardiograph ComponentsRemoving and Replacing the Battery 7-2PageWriter Trim Cardiograph Service Manual Removing and Replacing the Battery All of the PageWriter Trim cardiograph models use the same single battery (part number 989803130051). Removal and replacement procedures are the same for all models. Removing the Battery To remove the battery: 1 Unplug the AC power cord. Make sure the green AC power indicator light is not illuminated. Figure 7-1 AC power light For an overview of cardiograph components, see “Tour of PageWriter Trim Cardiographs” on page 1-12. 2Pull up on the battery door tab and remove the door. Figure 7-2 Removing the battery door 3 Pull open the battery tab and pull out the battery. Figure 7-3 Removing the battery AC power light Battery door tab

Removing and Replacing Cardiograph ComponentsRemoving and Replacing the AC Fuses PageWriter Trim Cardiograph Service Manual7-3 WARNINGProperly dispose or recycle any depleted batteries according to local regulations. Do not disassemble, puncture, or incinerate the batteries. Replacing the Battery To replace the battery: 1 Insert the new battery into the battery slot, and push it in until it clicks into place. 2Re-attach the battery door. Removing and Replacing the AC Fuses The AC fuse needs to be replaced when the cardiograph is plugged into AC power, but the green AC power indicator light on the front of the cardiograph does not illuminate. Only use replacement AC fuses with Philips part number 453564000371 or use a 1.5 amp (250V) time-delay fuse the same size and configuration as the original fuse. Figure 7-4 Replacing the AC Fuse To replace the AC fuse: 1 Unplug the cardiograph from AC power. Pull out the AC power cord from the AC power connector on the rear of the cardiograph. 2Locate the AC fuse, which is directly below the AC power connector. 3Push down on the center tab on the fuse and pull out the fuse from the fuse holder slot. 4Insert the new fuse using the same orientation, with the center tab facing up. 5Push the fuse all the way into the fuse holder slot. The fuse snaps into place.

Removing and Replacing Cardiograph ComponentsRemoving and Replacing the Paper Tray 7-4PageWriter Trim Cardiograph Service Manual Removing and Replacing the Paper Tray The paper tray is the same for all PageWriter Trim models. Removing the Paper Tray To remove the paper tray: 1 Pull out the paper tray until it stops. 2Remove any unused paper. 3Carefully lift tray, and pull it out completely. Replacing the Paper Tray To replace the paper tray: 1 Insert the paper tray into the opening. 2Replace the paper (see page 3-3). 3Close the paper tray.

8 8-1 Chapter 1Parts and Accessories This chapter provides information on ordering replacement parts that are intended for installation by the customer, and other supplies, and accessories for the PageWriter Trim cardiographs. It shows illustrated, detailed views of each of the PageWriter Trim subassemblies, together with each element’s part number to facilitate part ordering. Ordering Replacement Parts To order replacement parts, use prefix number 45. In the US, call 877-447-7278 Outside the US, contact your local Philips Medical Systems Response Center. See “Contacting a Philips Response Center” on page 1-32. Ordering Supplies and Accessories To order medical supplies for the cardiograph, see “Supplies and Ordering Information” on page 1-20. In the US, call 1-800-225-0230. Outside the US, contact your local Philips Medical Systems Sales Office, or your authorized Philips Medical Systems Dealer or Distributor. You can also visit our web site at: http://shop.medical.philips.com.

Parts and AccessoriesPatient Interface Module (PIM) Assembly and Parts 8-2PageWriter Trim Cardiograph Service Manual Patient Interface Module (PIM) Assembly and Parts Figure 8-1 Patient Interface Module (PIM) Assembly Parts Table 8-1 Patient Interface Module (PIM) Assembly Parts Reference Number Description Part Number 1 Patient Data Cable Replacement Kit 453564034571 2 Patient Interface Module (PIM) PCA Set 453564044081

Parts and AccessoriesCart Assembly and Parts PageWriter Trim Cardiograph Service Manual8-3 Cart Assembly and Parts Figure 8-2 Cart Assembly Table 8-2 Cart Assembly Parts List Reference Number Description Part Number 1 Rear Handle Kit 453564054341 2 Swivel Caster Kit 453564050721 3 Locking Caster Kit 453564054301 4 Directional Locking Caster Kit 453564054291 5 Cart Base Kit (includes all four casters shown) 453564054311 6 Front Handle Kit 453564054331 7 Cart Upper Housing Kit 453564054321 Not shown Complete Unassembled Replacement Cart 453564050691

Parts and AccessoriesCart Assembly and Parts 8-4PageWriter Trim Cardiograph Service Manual

A A-1 Appendix BInstalling PageWriter Trim I Software A software installation procedure takes approximately fifteen minutes to complete. Software Upgrades The upgrade path for a specific cardiograph is dependent upon the software version that is currently installed on the unit. Philips recommends that all cardiographs be upgraded to software version A.01.03 for optimal performance. See Table A-1 for more information on the appropriate upgrade path. Upgrading Legacy Software Legacy software is defined as any PageWriter Trim I software version that is lower than A.01.01. Upgrading legacy software to the Philips recommended software version A.01.03 requires performing two sequential software upgrades. First, upgrade the cardiograph software from A.00.02 to A.01.01, and then upgrade from software version A.01.01 to version A.01.03. Special Note about Japanese Localization Option The Japanese AR0 option for the PageWriter Trim I cardiograph is no longer available. The supported software revision level for all Japanese option cardiographs is legacy software version A.00.03. There is no software upgrade path for the Japanese localization option. Table A-1 PageWriter Trim I Software Upgrade Path Cardiograph Model Available Software RevisionLowest Software Version Required for Upgrade Trim I A.01.03 (recommended)A.01.01 Trim I A.01.01 A.00.02 * *. See “Upgrading Legacy Software” below for important information on upgrading from legacy software version A.00.02 to software version A.01.01 or higher. Philips recommends upgrading all PageWriter Trim I cardiographs to software version A.01.03 for optimal performance.

Installing PageWriter Trim I SoftwareObtaining Software A-2PageWriter Trim Cardiograph Service Manual Software Revision Information Table A-2 provides detailed software revision information. Follow the same procedure to upgrade the software in an existing working unit or to install software on a new system. Obtaining Software The software installation procedure requires the use of a computer with a compatible PCMCIA card slot, and internet access. The procedure also requires the use of a blank CompactFlash (CF) memory card and a PC card adapter. The memory card and adapter, referred to as a PC card, may be purchased from Philips using part number 453563479701. The following procedure describes how to download the cardiograph software files to the PC card. If the PC card with the applicable software is available, proceed to page A-5. Downloading Software Files from Philips InCenter All software upgrade files may be downloaded from the Philips InCenter web site found at: www.incenter.medical.philips.com. The InCenter site requires an active login and password. Instructions for obtaining a login and password are located on the right side of the main InCenter home page. Table A-2 Software Version Number Information Component A.00.02 A.00.03 A.01.00 A.01.01 A.01.03 Loader L.03.12 L.03.12 L.03.12 L.03.12 L.03.12 Kernel K.02.20 K.02.20 K.03.00 K.03.01 K.03.01 Application A.07.02.01 A.07.02.02 A.08.00.02 A.08.01.01 A.08.03.01 PIM Loader 1.009 1.009 1.009 1.009 1.009 PIM Kernel D.008 D.008 D.008 D.008 D.008 or G. 0 0 3 * *. See “Software Version A.01.03 PIM Kernel Revision” on page A-13 for more information.