    7-1
    Chapter 7Removing and Replacing
    Cardiograph Components
    This chapter contains information on removing and replacing the PageWriter Trim 
    cardiograph battery, fuses, and paper tray. If you require further technical assistance, contact 
    the nearest Philips Response Center. See “Contacting a Philips Response Center” on 
    page 1-32.
    About the Cardiograph Components
    The cardiograph includes the following customer-installable components. 
    „Battery compartment
    „AC fuses
    „Paper tray 
    						
    							Removing and Replacing Cardiograph ComponentsRemoving and Replacing the Battery
    7-2PageWriter Trim Cardiograph Service Manual 
    Removing and Replacing the Battery
    All of the PageWriter Trim cardiograph models use the same single battery (part number 
    989803130051). Removal and replacement procedures are the same for all models.
    Removing the Battery 
    To remove the battery:
    1
    Unplug the AC power cord. 
    Make sure the green AC power indicator light is not illuminated. 
    Figure 7-1 AC power light
    For an overview of cardiograph components, see “Tour of PageWriter Trim Cardiographs” 
    on page 1-12.
    2Pull up on the battery door tab and remove the door.
    Figure 7-2 Removing the battery door
    3
    Pull open the battery tab and pull out the battery.
    Figure 7-3 Removing the battery
    AC power 
    light
    Battery door tab 
    						
    							Removing and Replacing Cardiograph ComponentsRemoving and Replacing the AC Fuses
    PageWriter Trim Cardiograph Service Manual7-3
    WARNINGProperly dispose or recycle any depleted batteries according to local regulations. 
    Do not disassemble, puncture, or incinerate the batteries.
    Replacing the Battery
    To replace the battery:
    1
    Insert the new battery into the battery slot, and push it in until it clicks into place. 
    2Re-attach the battery door.
    Removing and Replacing the AC Fuses 
    The AC fuse needs to be replaced when the cardiograph is plugged into AC power, but the 
    green AC power indicator light on the front of the cardiograph does not illuminate. 
    Only use replacement AC fuses with Philips part number 453564000371 or use a 1.5 amp 
    (250V) time-delay fuse the same size and configuration as the original fuse.
    Figure 7-4 Replacing the AC Fuse
    To replace the AC fuse:
    1
    Unplug the cardiograph from AC power. Pull out the AC power cord from the AC power 
    connector on the rear of the cardiograph. 
    2Locate the AC fuse, which is directly below the AC power connector.
    3Push down on the center tab on the fuse and pull out the fuse from the fuse holder slot. 
    4Insert the new fuse using the same orientation, with the center tab facing up. 
    5Push the fuse all the way into the fuse holder slot. The fuse snaps into place.  
    						
    							Removing and Replacing Cardiograph ComponentsRemoving and Replacing the Paper Tray
    7-4PageWriter Trim Cardiograph Service Manual 
    Removing and Replacing the Paper Tray
    The paper tray is the same for all PageWriter Trim models. 
    Removing the Paper Tray
    To remove the paper tray:
    1
    Pull out the paper tray until it stops. 
    2Remove any unused paper.
    3Carefully lift tray, and pull it out completely.
    Replacing the Paper Tray
    To replace the paper tray:
    1
    Insert the paper tray into the opening. 
    2Replace the paper (see page 3-3).
    3Close the paper tray. 
    						
    							8
    8-1
    Chapter 1Parts and Accessories
    This chapter provides information on ordering replacement parts that are intended for 
    installation by the customer, and other supplies, and accessories for the PageWriter Trim 
    cardiographs. It shows illustrated, detailed views of each of the PageWriter Trim 
    subassemblies, together with each element’s part number to facilitate part ordering.
    Ordering Replacement Parts
    To order replacement parts, use prefix number 45.
    In the US, call 877-447-7278 
    Outside the US, contact your local Philips Medical Systems Response Center. See 
    “Contacting a Philips Response Center” on page 1-32.
    Ordering Supplies and Accessories
    To order medical supplies for the cardiograph, see “Supplies and Ordering Information” on 
    page 1-20. 
    In the US, call 1-800-225-0230.
    Outside the US, contact your local Philips Medical Systems Sales Office, or your authorized 
    Philips Medical Systems Dealer or Distributor.
    You can also visit our web site at: 
    http://shop.medical.philips.com. 
    						
    							Parts and AccessoriesPatient Interface Module (PIM) Assembly and Parts
    8-2PageWriter Trim Cardiograph Service Manual
    Patient Interface Module (PIM) Assembly and Parts 
    Figure 8-1 Patient Interface Module (PIM) Assembly Parts
    Table 8-1 Patient Interface Module (PIM) Assembly Parts
    Reference Number Description Part Number
    1 Patient Data Cable Replacement Kit 453564034571
    2 Patient Interface Module (PIM) PCA Set 453564044081 
    						
    							Parts and AccessoriesCart Assembly and Parts
    PageWriter Trim Cardiograph Service Manual8-3
    Cart Assembly and Parts
    Figure 8-2 Cart Assembly
    Table 8-2 Cart Assembly Parts List
    Reference Number Description Part Number
    1 Rear Handle Kit 453564054341
    2 Swivel Caster Kit 453564050721
    3 Locking Caster Kit 453564054301
    4 Directional Locking Caster Kit 453564054291
    5 Cart Base Kit (includes all four casters shown) 453564054311
    6 Front Handle Kit 453564054331
    7 Cart Upper Housing Kit 453564054321 
    Not shown Complete Unassembled Replacement Cart 453564050691 
    						
    							Parts and AccessoriesCart Assembly and Parts
    8-4PageWriter Trim Cardiograph Service Manual 
    						
    							A
    A-1
    Appendix BInstalling PageWriter Trim I Software
    A software installation procedure takes approximately fifteen minutes to complete.
    Software Upgrades
    The upgrade path for a specific cardiograph is dependent upon the software version that is 
    currently installed on the unit. Philips recommends that all cardiographs be upgraded to 
    software version A.01.03 for optimal performance. See Table A-1 for more information on the 
    appropriate upgrade path. 
    Upgrading Legacy Software
    Legacy software is defined as any PageWriter Trim I software version that is lower than 
    A.01.01. Upgrading legacy software to the Philips recommended software version A.01.03 
    requires performing two sequential software upgrades. First, upgrade the cardiograph software 
    from A.00.02 to A.01.01, and then upgrade from software version A.01.01 to version A.01.03. 
    Special Note about Japanese Localization Option
    The Japanese AR0 option for the PageWriter Trim I cardiograph is no longer available. The 
    supported software revision level for all Japanese option cardiographs is legacy software 
    version A.00.03. There is no software upgrade path for the Japanese localization option.
    Table A-1 PageWriter Trim I Software Upgrade Path 
    Cardiograph Model Available Software RevisionLowest Software Version 
    Required for Upgrade
    Trim I A.01.03 (recommended)A.01.01
    Trim I A.01.01 A.00.02
    *
    *. See “Upgrading Legacy Software” below for important information on upgrading from legacy software version A.00.02 to
    software version A.01.01 or higher. Philips recommends upgrading all PageWriter Trim I cardiographs to software version
    A.01.03 for optimal performance.  
    						
    							Installing PageWriter Trim I SoftwareObtaining Software
    A-2PageWriter Trim Cardiograph Service Manual
    Software Revision Information 
    Table  A-2 provides detailed software revision information. 
    Follow the same procedure to upgrade the software in an existing working unit or to install 
    software on a new system. 
    Obtaining Software 
    The software installation procedure requires the use of a computer with a compatible PCMCIA 
    card slot, and internet access. The procedure also requires the use of a blank CompactFlash 
    (CF) memory card and a PC card adapter. The memory card and adapter, referred to as a PC 
    card, may be purchased from Philips using part number 453563479701. The following 
    procedure describes how to download the cardiograph software files to the PC card. If the PC 
    card with the applicable software is available, proceed to page A-5.
    Downloading Software Files from Philips InCenter
    All software upgrade files may be downloaded from the Philips InCenter web site found at: 
    www.incenter.medical.philips.com. The InCenter site requires an active login and password. 
    Instructions for obtaining a login and password are located on the right side of the main 
    InCenter home page. 
    Table A-2 Software Version Number Information
    Component A.00.02 A.00.03 A.01.00 A.01.01 A.01.03
    Loader L.03.12 L.03.12 L.03.12 L.03.12 L.03.12
    Kernel K.02.20 K.02.20 K.03.00 K.03.01 K.03.01
    Application A.07.02.01 A.07.02.02 A.08.00.02 A.08.01.01 A.08.03.01
    PIM Loader 1.009 1.009 1.009 1.009 1.009
    PIM Kernel D.008 D.008 D.008 D.008 D.008 or 
    G. 0 0 3
    *
    *. See “Software Version A.01.03 PIM Kernel Revision” on page A-13 for more information. 
    						